PowerMetrix PowerMaster MTA15Z User manual

Rev: 1.0 Page 1 of 13
MTA15Z
PowerMaster
®
Accessory
Product Manual
Revision 1.1

Rev: 1.0 Page 2 of 13
Copyright Notice:
Copyright © by TEC, 2013. All rights reserved.
Disclaimer
This documentation is provided for information purposes. TEC makes no warranty of any kind
with regard to this material, including, but not limited to, the implied warranties of merchantability
and fitness for a particular purpose. TEC shall not be liable for errors, omissions, or
inconsistencies which may be contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of the material. Information in this document
is subject to change without notice and does not represent a commitment on the part of TEC.

Rev: 1.1 Page 3 of 13
Table of Contents
General Information ........................................................................................4
1
Setting Up.................................................................................................6
2
PowerMaster Procedure............................................................................9
2.1
With Site Selected.............................................................................................................9
2.2
With No Site Selected (Component Test)........................................................................11
3
Error Messages.......................................................................................12

Rev: 1.1 Page 4 of 13
General Information
Product Introduction
The MTA15Z is a self-contained accessory to the PowerMaster
®
analyzer. This accessory
accommodates testing of any ANSI socket-based meter using a load provided by the
PowerMaster
®
. The MTA15Z includes a zero-insertion force (ZIF) socket that is automatically
configured by the PowerMaster
®
and requires no jumpers for testing. For mechanical meters, the
user can pull the supplied pins for the lid, rotate the case upright, and test appropriately.
Currents are injected via the PowerMaster
®
by the 6mm terminated lead set. Voltages are
injected using the standard voltage lead set, and are fused for each phase on the MTA15Z.
Once the service type is selected in the PowerMaster
®
, the user will select a Phantom Load
Setup to provide the test. Results are saved as normal in the PowerMaster
®
and will be
downloaded using the Meter Site Manager 2 software.
Included Accessories
Qty.
Part #
Description
1 10-340-3309 RS232 Communication Cable
1 10-340-0066 Power Cable
2 22-900-2022 Pull pin for lid
1 10-340-0068 AC power cable (N. America)
1 10-340-0069 AC power supply (100-240V)
1 10-340-4096G Green safety ground lead
1 50-950-0006 Green alligator clip
Spare Parts
Qty.
Part #
Description
3 34-100-0144 KLKR Fuse 0.5A, 600V, slow acting
2 22-900-2022 Pull pin for lid
Safety
Operation of the PowerMaster
®
and the supplied accessories and adapters can present the user
to potentially hazardous conditions. Please follow all required safety procedures set forth by the
user’s safety organization within the company. If no safety organization exists, please follow all
applicable OSHA rules and standards for PPE (Person Protective Equipment) when working in
high voltage and low voltage environments. This equipment should be used by trained and
qualified personnel ONLY.

Rev: 1.1 Page 5 of 13
Tips for Testing:
1. Connect the green safety ground to a true earth ground before testing begins. Remove
the green safety ground last after testing is complete.
2. Never disconnect the CURRENT or VOLTAGE leads from the PowerMaster
®
while driving
a load into the MTA15Z.
3. Pressing the “ON” key on the PowerMaster
®
while driving a load can be used as an
emergency shutoff. The auxiliary power switch on the front panel of the PowerMaster
®
will provide the same function as well.
Specifications
Supported PowerMaster
®
models 7332 and 7335
Supported meter forms 1S, 2S, 3S, 4S, 5S, 6S, 8S, 9S, 10S,
11S, 12S, 13S, 14S, 15S, 16S, 17S,
25S, 26S, 29S, 35S, 36S, 39S, 45S,
46S, 56S, 66S, 76S
Fuse (x3) KLKR 0.5A, 600V; slow acting
Case Pelican 1450
Dimensions 14.6” x 10” x 6”
Weight 14.8 lbs (6.7 kg)
Warranty 1 year (does not include fuses)
Technical Support
For technical support, please contact the factory at (865) 218-5838 (877-966-5851 toll free) and
ask for “Powermetrix Technical Support.” You may alternatively email Powermetrix at
help@powermetrix.com. The support staff will answer questions about the operation and care of
your equipment, assist you in troubleshooting a problem, and help you overcome common
application difficulties whenever possible. If it becomes necessary for your equipment to be
returned to us for any reason, you will be issued an RMA number during the technical support
contact.
Feedback
Powermetrix depends on information from our customers to continue the attributes of quality,
dependability, and simplicity associated with our products. We invite you to contact our Technical
Support office.

Rev: 1.1 Page 6 of 13
1 Setting Up
Grounding
For the PowerMaster
®
to properly stabilize and achieve the highest possible accuracy, the
MTA15Z must be properly grounded. Using the greed safety ground lead, insert the cable into
the green “VOLTAGE DRIVE” (BOTTOM) female receptacle. If using an alligator clip, connect to
a known earth ground. When using in a laboratory application, Powermetrix recommends the AC
Plug to Safety Banana Plug Adapter (part# 10-340-0045). This accessory allows the Auxiliary
power, Auxiliary neutral, and ground to be easily terminated for a standard US three-prong AC
wall outlet.
Voltage Lead Connections
1. Connect the Voltage Lead Set + Safety Ground + Aux Power cable to the PowerMaster
®
.
2. Remove all alligator clips from the ends.
3. Separate the “MEASURE” and “RETURN” from the stackable banana jacks.
4. Insert the cables labeled “MEASURE” into the “VOLTAGE MEASURE” receptacles
according to their color code.
5. Insert the cables labeled “DRIVE” into the “VOLTAGE DRIVE” receptacles according to
their color code.
6. Insert the green lead with alligator clip into the “VOLTAGE DRIVE” receptacle. Connect
the alligator clip to a known earth ground.
7. Connect the AUX PWR and AUX NEU to a voltage source between 100-530VAC.
Voltage cables from PowerMaster
®
Auxiliary power connected to voltage source
(100-530VAC)
WARNING: Input voltage must be supplied from the Model 7332 or 7335 only. Any other source
voltage from the service or another device will cause damage to the MTA15Z that is
not covered under the 1-year warranty.

Rev: 1.1 Page 7 of 13
Current Lead Connections
1. Connect the 3-Phase Current Direct Probes (terminated with 6mm plugs) to the
PowerMaster
®
.
2. Insert the probes into the “TEST CURRENT IN”. According to the color code, plug the
phase DRIVE into the blackfemale receptacle (DRIVE, BLACK) and the RTN into the
white female receptacle (RETURN, WHITE) for each phase.
Forms 1,3 Forms 2, 4, 5, 12, 13, 25, 26,
35, 45, 56, 66, Forms 6, 8, 9, 10, 11, 14, 15,
16, 17, 29, 36, 39, 46, 76
Removal
To remove the probes, push into the receptacle gently and now the probe will unlock for
removal.
Communication and Power Connections
1. Connect the DB9 cable to “RS232” on the PowerMaster
®
and “CONTROL IN” on the
MTA15Z.
2. Connect the gray power cable to “AUX ANALOG” on the PowerMaster
®
and “DC IN” on
the MTA15Z.
Cables from PowerMaster
®
Cables to MTA15Z
Alternate Power Connection (using the AC Power Supply and Cable)
1. Connect the AC Power Supply and AC Power Cable together.
2. Plug the cable into the “DC IN” on the MTA15Z.
3. Plug the AC Power Supply into an AC source between 100-240VAC.

Rev: 1.1 Page 8 of 13
Testing Mechanical Meters
1. Open the case and remove the pull pins located at the hinge of the lid.
2. Remove the lid and place aside.
3. Insert the meter into the appropriate socket.
4. Rotate the MTA15Z in a vertical position so the case stands upright.
5. For pulse detection, use a Manual Pushbutton Switch (10-100-3311) or a Photo Disk
Detector with Flexible Arm Mount (10-100-3326).
6. Test as normal.

Rev: 1.1 Page 9 of 13
2 PowerMaster Procedure
NOTE on selecting Service Types
Some service type files are specific to the MTA15Z. These apply to forms 1, 2, 3, 4, 5, 12, and
56. When selecting these meter forms, select the Service Type that ends with “M”. Otherwise,
the user will receive an error message when attempting to test.
2.1 With Site Selected
STEP 1
At the Main Menu, press “1” to
Select Site.
STEP 2
In the Site Editor, press F2 to
create a new site.
STEP 3
In the Site Editor screen, enter
the Site ID, meter form, service,
and service type according to the
Jumper Configuration Diagram
STEP 4
Press F2 to select a meter in the
database, or enter a Kt on Page
1. Press F6 to save and exit.
STEP 5
At the Site Editor, press F6 to
select the site to be tested.
STEP 6
Optionally enter the user’s
name(s) and any comments
regarding the installation. Press
F6 to continue.
STEP 7
At the Main Menu, press “3” to
enter Meter Testing.
STEP 8
At the Meter Testing menu, press
“3” to enter Phantom Load with
MTA15Z
STEP 9
Hook up pulse pickup to meter
(AUX DIG on PowerMaster).

Rev: 1.1 Page 10 of 13
STEP 10
Select the setup, then press F6 to
continue.
STEP 11
After stabilization is complete and
pulses are detected, press F6 to
continue.
STEP 12
After the test is complete, press
F6 to save and exit.

Rev: 1.1 Page 11 of 13
2.2 With No Site Selected (Component Test)
STEP 1
At the Main Menu, press “2” for
Component Test.
STEP 2
Press F2 to select “MTA15Z”.
STEP 3
In the Component Test screen,
select the Meter Model and
desired Service Type.
STEP 4
Enter the serial number (MFG
SN) and alternatively the Kt value.
Press F6 to continue.
STEP 5
Hook up pulse pickup to meter
(AUX DIG on PowerMaster).
STEP 6
Select the setup, then press F6 to
continue.
STEP 7
After stabilization is complete and
pulses are detected, press F6 to
continue
STEP 12
After the test is complete, press
F6 to save and exit.

Rev: 1.1 Page 12 of 13
3 Error Messages
In some instances, error messages may display if a problem occurs. If the user encounters one
or more of these error messages, please check the setup again.
Examples:
The service type selected is not compatible with
the MTA15Z. Go back to the Site Editor (Main
Menu, Select Site) and select a service type file
that ends with “M”.
The PowerMaster
®
cannot communicate with the
MTA15Z. Verify the following:
1. Verify cables are connected according to
Section 1 Setting Up, Communication and
Power Connections
2. Verify 24VDC output from gray power
cable connected to AUX ANALOG on the
PowerMaster
®
If the error continues, please contact Powermetrix
Technical Support.
The PowerMaster
®
cannot detect an auxiliary
power source between 100-530VAC. Check the
connections and/or verify the Auxiliary Power
switch is in the “on” position (- = on).

Rev: 1.1 Page 13 of 13
The PowerMaster
®
cannot communicate with the
WFG board. Attempt to power off/on to fix the
issue. If this is not successful, contact Technical
Support.
The PowerMaster
®
cannot stabilize the currents
and/or voltages. Attempt the following:
1. Verify the green safety ground lead is
connected to true earth
2. Verify the current leads are connected
according to Section 1 Setting Up, Current
Lead Connections
3. Attempt moving auxiliary leads to a
different AC source
4. Attempt to uncheck “Stabilize Load Box
Phases” in User Preferences (see
PowerMaster
®
user manual)
Table of contents
Other PowerMetrix Measuring Instrument manuals