Prius Healthcare USA Rhythm Turn User manual

Rhythm Turn
Low Air Loss
and
Dynamic Turning
Mattress Replacement System
User Manual
Prius Healthcare USA
4027 Tampa Road, #3200
Oldsmar, FL 34677, USA
TEL: (813)854-5464
FAX: (813)854-5442
1707

Content
1. The Purpose of this Manual .................................................................1
2. Product Description.................................................................................1
Master Control Unit Features
Mattress Features
3. Technical Data............................................................................................2
Master Control Unit
Rhythm Turn Mattress Replacement
Symbol Definition
4. Instruction for Proper Use....................................................................3
5. Cleaning.........................................................................................................6
The Mattress
The Master Control Unit
Replace Air Filter
Waste Disposal
6. Storage and Care .......................................................................................7
7. Maintenance and Troubleshooting...................................................8
8. EMC Related Notification ......................................................................9
9. Warranty.................................................................................................... 12

Warning
Connect the Master Control unit to a proper power source.
Don’t use the system in the presence of any flammable gases.
( such as Anesthetic Agents)
Keep the pump and mattress away from source of liquid of open flame.
Keep the mattress away from sharp object
The device is not AP/APG protected.
Don’t place the heating device close to the mattress system.
Caution
Use the mattress under the physician’s instruction.
Re-positioning the patient in certain period of time is still necessary
when using this system.
The Control unit can only be repaired by authorized distributor.
(The circuit diagram, repairable component parts list, and service manual
are released only to authorized distributor).
Don’t drop the control unit.
Do not store the system in direct sunlight or extreme cold conditions

1
1. The Purpose of this Manual
This operation manual is mainly focused on the set up, cleaning and routine
maintenance of the Rhythm Turn-Function Air Therapy Support System. We
recommend you keeping this manual handy to answer most of the question related to
the system.
2. Product Description
The Rhythm Turn system, operated blower unit, is a very unique innovation of a
specialized mattress replacement. The system is primarily designed for at risk patients
or step-down intensive care units. It features continuous lateral rotation therapy in two
different degree (20 degree and 40 degree), which gently turns the patient from side to
side to significantly lower the risk of infection, pneumonia and other pulmonary
complications – illnesses that significantly ad to patient care costs and length of stay.
Master Control Unit Features
The Master Control Unit is user friendly designed and most of the functions are
self explained.
Rotation angle can be independently selected for 20 degree of 40 degree.
Rotation time can be adjusted in 5 min increments to 99 min. Or the caregiver
can even select the Static Function that will seize the Rotation Function and
provide only the True Low Air Loss Therapy.
Auto Firm Function provides a uniformly firmness for nursing procedure.
Power failures produce an audio alarm for added safety.
10 digital scales of Soft/Firm Comfort Control.
Double insulation to provide max silent operation.
Foot board mounting rack provides the convenience of placement.
Mattress Features
Individual air cushion design for maximum pressure distribution.
Each air cushion has orifices to provide true Low Air Loss therapy.
LAL Turning Mattress Replacement, eliminating the compromising effects of an
existing mattress.
Permanent inflated bed rails for added safety.

2
3. Technical Data
Master Control Unit
M o de l Name
Rhythm Turn Control Unit
M o de l N o.
FC-PHR0026
Size
17.7”(L) x 6.8”(W) x 10.8”(H)
Weight
13.5 lbs
Dwell Time
3~95
Max Operating Pressure
50 mmHg
Rated Voltage
AC 110~120V
Rated Frequency
60 Hz
Fuse Rating
T5AH / 250V
Max Current
5A
Classification
Class I, Type BF
Not AP or AGP type
Operation Temperature
15°C ~ 35°C
Operation Humidity
30% ~75%
Mode of Operation
Continuous
Standard
IEC 60601-1,
CAN/CSA C22.2 No. 601.1,
IEC 60601-1-2
Rhythm turn Mattress Replacement
Model Name
Rhythm Turn Air Mattress
M o de l N o.
FM-PHR0013
Size (cm)
91(W) x 203(L) x 20(H)
Weight (Kg)
20 Kg
Cells Material
TPU
Cover Material
AD with Quilting
Base Material
Nylon laminated PVC
Symbol Definition
Refer To Instruction Manual /
Booklet
NOTE Background colour: blue
Symbol : white
Waste Disposal
Type BF Applied Part
Alternating Current
Warning
Keep Dry

3
4. Instruction for Proper Use
1. Remove the existing mattress from the bed frame.
2. Replace the standard mattress with LAL Turning Mattress Replacement and make
sure orient mattress so that the air tube is placed at the foot of the bed.
3. Secure the straps beneath the mattress to the bed frame.
4. Hang the Master Control Unit on the foot-board of the bed frame. Attach the air
tubes connectors to socket on the left panel of the Master Control Unit. Be careful
on the color matching between the connectors and socket. (black connectors to
black socket, red connector to red socket)
5. Ensure the air hoses are not kinked under the mattress. (Could be verified by
simple visual check) For detail Air hose connection please refer Explode Diagram.
6. Zip the low shear top cover to the mattress. The top cover should be loosely fit to
the mattress.
7. Carefully plug power cord into a properly grounded power source. Turn on the
master mechanical power switch on the right side panel. The STANDBY LED
should illuminates.
8. Push the STANDBY/OPERATE button of the front panel. The OPERATE LED should
now be lighted up and the Master Control Unit should now start to spin.
9. Push the AUTO FIRM button for fast inflation. Allow 4 ~ 7 minutes for full inflation.
After the mattress is fully inflated, the caregiver can now transfer the patient on to
the mattress. Push the AUTO FIRM again to release the fast inflation mode. (Note:
The mattress can be inflated with patient lying on top).
10. Static Function: Push the static button and adjust the Contort Control by pressing
the SOFT/FIRM button to achieve the maximum patient comfort. On this mode the
system provides True Low Air Loss therapy. Perform a hand check by placing
hand under the patient buttocks between cells and foam. The patient should have
at least 4 cm of clearance between the buttocks and the bottom of the mattress.

4
11. Turning time can be adjusted by the CYCLE button. The time can be adjusted from
3 minutes to 95 minutes. (When Static Function is selected, the timing window
would not show any digits).
12. The Master Control Unit is equipped with power failure alarm. With this function
enabled, the Control unit generates a horn sound to signal to caregiver that main
power failure. The alarm can be disabled by pushing the Alarm Reset Button on
the front panel.
Caution: Immediate response by the operator is required with power
f ailure alarm.
13. LOCK-OUT: The Master Control Unit is also equipped with a manual locking-out
function. All function keys will be automatically disabled if the LOCK –OUT button
has not been activated. When lock-out has been engaged, the “LOCK OUT” button
will illuminate.
UNLOCKING
Unlocking the control panel is easy. Simply press the “LOCK OUT” button on the
control panel for 3-5 seconds or recycle the power by turning off and on the main
power switch.
14. 20 degrees turning function can be activated by pressing the “TURN” and select
desire turning therapy. “LEFT TURN” a l lo w s the mattress to turn to left and back
to horizontal. “RIGHT TURN” would have the same effect but turning to the right.
The “FULL TURN” al lo w s for full function of turning to left and right and s h o u ld
always activate with timer setting. The timer can be set by “CYCLE” button.

5
15. 40 degrees turning function can be activated by pressing the “BT” button and
follow the operation instruction on step 14.
Mattress Turning Illustration

6
5. Cleaning
The Mattress
The mattress should be cleaned on the bed weekly using a damp soft cloth and mild
detergent.
If top sheet (Top cover) or base (Bottom cover) becomes overly soiled, put on clean
gloves, plastic gown and eye protection before removing top sheet or base and dispose
of according to standard in function control procedures. Replace with clean covers.
Covers can be washed and thermally disinfected in a washing machine following below
procedure: (Never use phenol based cleaning solutions).
Industrial cleaning
Break washes c o ld 10 minutes
Main washes 60∘C 6 minutes
Main washes 72∘C 10 minutes
Extraction 2 minutes
3 Cold Rinses
Extraction 5 minutes
Domestic cleaning
Pre-wash c o ld
Main Wash 72∘C 10 minutes
Extraction 2 minutes
Cold Rinses
Extraction 5 minutes
Tumble Drying or Tunnel Drying is not recommended.
Mattress Cells can be wiped over with a solution of sodium hypochlorite 1000ppm or
any other non-phenolic germicidal solution.
The Master Control Unit
CAUTION
SWITCH OFF THE ELECTRICAL SUPPLY TO THE PUMP AND DISCONNECT THE
POWER CORD FROM THE MAIN SUPPLY BEFORE CLEANING AND INSPECTION
The master control unit should also be cleaned weekly using a damp soft cloth and m i ld
detergent.
The pump casing is manufactured from ABS plastic and if the case is soiled the pump
can be wiped down with a sodium hypochlorite solution to dilution of 1000ppm or any
EPA- approved hospital grade disinfectant. (Don’t use phenol base cleaning
solution).
The air filter should also be cleaned and checked as often as possible at a minimum of
every six months. Air Filter can be removed by pinching center of the filter and
pulling outward from the back of the Therapy control unit.

7
Replace Air Filter
1. Remove Air filter and Replace a new Filter.
2. Use a soft bristle brush to remove dust and difficult dried-on soil.
Waste Disposal
This Product has been supplied from an environmentally aware
manufacturer that complies with the WEEE.
This product may contain substances that could be harmful to the
environment if disposed of in places (landfills) that are not appropriate
according the legislation. Please be environmentally responsible and
recycle this product through your recycling facility at its end of life.
6. Storage and Care
Master Control Unit:
Check the power cord and plug for abrasions or excessive wear.
Plug in the unit and verify air flows from the units hose connection ports
Place in plastic bag for storage.
Mattress Replacement:
Check the air manifold for kinks or breaks. Replace if necessary.
Pull out the CPR plug at the head of the mattress and disconnect the air feed
tubes. All the air will now be expelled. Starting at the head end, the mattress
can now be rolled. Use the base mounted straps for containment.
Place in plastic bag of storage.
It is recommended the following guidelines are used whenever this system is being
stored or transported another location:
Temperature limitations: 5°C ~ 60°C
Relative Humidity 30% ~ 75%
Filter

8
7. Maintenance & Troubleshooting
No daily maintenance is required. It is intended this equipment should only be
serviced by properly qualified, authorized technical personnel. In case of minor
tr o u b le please refer as following Troubleshooting.
Symptom Inspection Procedure Possible Solution
Air is pumping out from the
control unit but mattress is
not inflating.
1.Is the power source
correct? Improper voltage
may cause the pump to
function abnormally and
damage the control unit.
1.Use power regulator.
2.
Is there any kinking tube?
2.
Adjust the air tubes to enable
smooth air flow.
3.Is there any air leakage
from the air cells?
3.Replace with new air cells
4.Is there any air leakage
from air tube between
mattress and control unit?
4.Replace with new air tubes
5.Has the air tube connector
been connected properly?
5.Re-connect the air tubes.
The Control Unit is not
functioning.
1.
Check the power cord and
the power voltage.
2. Check the fuse
1.Use a power regula tor
2. Replace with a new fuse
Some of the air cells are not
properly inflated.
1.Is the connection between
air cells and the manifold
kinked?
2.
Is there any air leakage
from the air cells?
1.Check for any kinking
between air cells and manifold.
2.Replace new air cell if faulty.

9
8. EMC Related Notifications
Guidance and manufacturer’s declaration – electromagnetic emissions
The air pump is intended for use in the electromagnetic environment specified below. The customer or the
user of the air pump should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The air pump uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The air pump is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public
lo w-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Recommended separation distances between
portable and mobile RF communications equipment and the air pump
The air pump is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the air pump can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the air pump as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d =
1,2
80 MHz to 800 MHz
d =
1,2
800 MHz to 2,5 GHz
d =
2,3
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

10
Guidance and manufacturer’s declaration – e le c tr o magnetic immunity
The air pump is intended for use in the electromagnetic environment specified below. The
customer or the user of the air pump should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV for power
supply lines
±1 kV for
input/output
lines
±
2 kV for power
supply lines
±1 kV for
input/output
lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±
1 kV line(s) to
line(s)
±
2 kV line(s) to
earth
±
1 kV line(s) to
line(s)
±
2 kV line(s) to
earth
Mains power quality should be that of
a typical commercial or hospital
environment.
interruptions
and
voltage
variations
on power
supply
input lines
IEC
61000-4-11
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in
U
T)
for 5 sec
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in
U
T)
for 5 sec
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
air pump ] requires continued
operation during power
mains interruptions, it is
recommended that the air pump
be powered from an uninterruptible
power supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.

11
Guidance and manufacturer’s declaration – electromagnetic immunity
The air pump is intended for use in the electromagnetic environment specified below. The customer or the
user of the air pump should assure that it is used in such an environment.
Immunity
test
IEC 60601 test
le ve l
Compliance
le ve l
Electromagnetic environment
–
gui dance
Conducted
RF
IEC
61000-4-6
Radiated
RF
IEC
61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the air pump,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1,2
d = 1,2 80 MHz to 800 MHz
d = 2,3 800 MHz to 2,5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a should
be less than the compliance level in each frequency
range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the air pump is used exceeds the applicable RF compliance level above, the air pump
should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the air pump.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

12
9. Warranty
Prius Healthcare Products guarantees that this equipment is free from defects in
material and workmanship. Our obligation under this warranty is limited to the
repair of equipment returned to the service address given below, transportation
charges prepaid, within 12 months after delivery to the original purchaser for all
equipment.
We agree to service and/or adjust any equipment returned for that purpose and to
replace or repair any part, which is proven to be defective at no charge.
This warranty excludes equipment damage through shipping, tampering, improper
maintenance, careless, accident, negligence or misuse, or products which have been
altered, repaired or dismantled other than with the manufacture’s written
authorization and by its approved procedures and by properly qualified
technicians.
In no event shall Prius Healthcare Products be liable for any direct, indirect of
consequential damages or losses resulting from the use of equipment.

Prius Healthcare USA
4027 Tampa Road, #3200
Oldsmar, FL 34677, USA
TEL: (813)854-5464
FAX: (813)854-5442
AL300104 V1.02
Other manuals for Rhythm Turn
1