Profi-pumpe DIESEL User manual

Technical changes, misprints and mistakes reserved! Newest information about our products can be found online
Technische Änderungen, Druckfehler und Irrtümer vorbehalten! Aktuelle Informationen zu unseren Produkten nden Sie auf:
http://www.pro-pumpe.de
1. ADBLUE®* & DIESEL
DIGITAL-DURCHFLUSSMESSER
(DPY01105-4)
FLOWMETER
DURCHFLUSSMESSER
Version 16.02-1.0
OPERATING INSTRUCTIONS (Translation)
BEDIENUNGSANLEITUNG (Original)
de
*AdBlue, a registered trademark / eingetragenes Warenzeichen of / vom Verband der Automobilindustrie e. V.

Contents
1. Introduction ......................................................................................................................................................... 2
2. In general ............................................................................................................................................................ 2
3. Main components ................................................................................................................................................ 2
4. Operating modes ................................................................................................................................................. 3
5. LCD display .......................................................................................................................................................... 3
6. User buttons ........................................................................................................................................................ 3
7. Battery housing ................................................................................................................................................... 3
8. Installation ........................................................................................................................................................... 3
9. Daily use ............................................................................................................................................................. 4
10. Callibration .......................................................................................................................................................... 5
11. Meters conguration ............................................................................................................................................ 9
12. Maintenance ...................................................................................................................................................... 10
13. Recognising and repairing of mistakes ............................................................................................................. 10
14. Guarantee regulations ....................................................................................................................................... 11
15. Notes on Product Liability ................................................................................................................................. 11
16. Notes on Disposal ............................................................................................................................................. 11
17. EU Declaration of Conformity ............................................................................................................................ 11
12. Technical Data ................................................................................................................................................... 12
English 32 English
1. INTRODUCTION
We would like to congratulate you on the purchase of our K24 Electronic digital meter. We appreciate your trust. That‘s
why funtional security and operational safety stands by us on rst place.
To prevent damage to persons or property, you should read this user manual carefully. Please
observe all safety precautions and instructions for proper use of the K24 meter. Failure to follow
the instructions and safety precautions can result in injury or property damage.
Please keep this manual with the instructions and safety instructions carefully in order to at any
time you can restore them.
2. IN GENERAL
K24 Electronic digital meter featuring a turbine measurement system, designed for precise measuring of low viscosity uids.
It is divided into two using macrogroups:
l With body made of inconductive plastic material of light colour, designed to be used with water / urea solution
l With body made of conductive plastic material of dark colour (assessed resistance: 50 ohm), designed to be used
with DIESEL FUEL, WATER and windscreen uids. The card can be rotated with respect to its housing, thus allow-
ing easy display readings in any position. The card housing, easily accessible, is closed by a plastic cover sealed
through a rubber protection acting as a gasket as well. The whole unit can be easily removed by unscrewing the 4
screws xing thecard and the cover.
Turbine measurement system. The turbine is placed inside a hole through the body of K24, tted with threaded inlet
and outlet. The body of K24 is made of a plastic material that allows several types of threads with relevant combina-
tions. K24 has 2 rubber protections, designed to act as gaskets, too, and thus reducing the number of its components.
The liquids compatible with K24 must be at low viscosity, namely: Diesel fuel, Water, Water/urea solution, Kerosene,
Windscreen - Petrol
3. MAIN COMPONENTS
Please read the user manual before using the K24 Electronic digital meter Warning sign
SAFETY INSTRUCTION AND WARNINGS
WhilexingtheK24
card, make sure the bat-
tery contact cable is not
placed above the circular
housing of the bulb.
3.1. DISPLAY POSITIONING
The square shape of the k24 body
allows the card to be rotated in
its housing, thus ensuring great
versatility in positioning.
4. OPERATING MODES
The user can choose between two different operating modes:
l
Normal Mode: Mode with display of Partial and Total dispensed quantities.
l
Flow Rate Mode: Mode with display of Flow Rate, as well as Partial dispensed quantity.
The meter features a non-volatile memory for storing the dispensing data, even in the event of a complete power break
for long periods. The measurement electronics and the LCD display are tted in the top part of the K24 which remains
isolated from the uid-bath measurement chamber and sealed from the outside by means of a cover.
6. USER BUTTONS
The k24 features two buttons (reset and cal) which individually perform two main functions and, together, other
secondary functions. The main functions performed are:
lFor the reset key, resetting the partial register and resettable total (reset total)
lFor the cal key, entering instrument calibration mode.
Used together, the two keys permit entering conguration mode, useful for changing the units of measurements and
calibration factor.
7. BATTERY HOUSING
The k24 is powered by two standard type 1.5 V batteries (size AAA). The battery housing, easily accessible, is closed
by a metal cover sealed through a rubber protection acting as a gasket as well. The whole unit can be easily removed
by unscrewing the 4 screwsxingthe cover and the protection to the body.
8. INSTALLATION
Theinstallationmustbeperformedbyaqualied
professional.
K24 features a threaded, perpendicular inlet and
outlet (1 and female that can be combined together).
It has been designed to be easily installed in any
position: xed in-line or mobile on a dispensing
nozzle. In order to improve the life of the turbine, it is
recommended to t a strainer before the meter itself.
At the female inlets, tighten the coup-
lingsatamax.torqueof55N/m.
ITH THE GAS-FEMALE INLETS, DO
NOT USE CONICAL THREADED COU-
PLINGS.
RESET button
LCD Display
CAL button
5.LCDDISPLAY
The „LCD“ of the METER features two numerical registers and various indi-
cations displayed to the user only when the applicable function so requires.
Key:
1. Partial register (5 gures with moving comma FROM 0.1 to 99999) in-
dicating the volume dispensed since the reset button was last pressed
2. Indication of battery charge
3. Indication of calibration mode
4. Totals register (6 gures with moving comma FROM 0.1 to 999999),
that can indicate two types of Total:
4.1. General Total that cannot be reset (TOTAL)
4.2. Resettable total (Reset TOTAL)
5. Indication of total multiplication factor (x10 / x100 )
6. Indication of type of total, (TOTAL / Reset TOTAL)
7. Indication of unit of measurement of Totals: L=Litres Gal=Gallons
8. Indication of Flow Rate mode
9. Indication of unit of measurement of Partial: Qts=Quarts, Pts=Pints, L=Litres, Gal=Gallons
1
2
3
4 5 6
7
8
9
88,8,8,8
888888,8,8,8
*AdBlue, a registered trademark of Verband der Automobilindustrie e. V. *AdBlue, a registered trademark of Verband der Automobilindustrie e. V.

English 54 English
9. DAILY USE
The only operations that need to be done for daily use are partial and/or resettable total register resetting.
The user should use only the dispensing system of k24. Occasionally the meter may need to be congured or calibra-
ted. To do so, please refer to the relevant chapters. Below are the two typical normal operation displays.
One display page shows the partial and reset total registers. The other shows the partial and general total. Switcho-
ver from resettable total to general total display is automatic and tied to phases and times that are in factory set and
cannot be changed.
10. CALIBRATION
10.1. DEFINITIONS
Calibration factor or „K factor“. Multiplication factor applied by the system to the electrical pulses received, to trans-
form these into measured uid units.
FACTORY K FACTOR:
Factory-set default factor. It is equal to 1.000. This calibration factor ensures utmost precision in the following opera-
ting conditions:
uidTemperature: Dieselfuel20°C
Flowrate: 10-120litres/min
Even after any changes have been made by the user, the factory k factor can be restored by means of a simple
procedure.
10.2. WHY CALIBRATE
When operating close to extreme conditions, such as for instance with uids close to acceptable range extremes (like
diesel fuel at low temperatures) or in extreme ow rate conditions (close to minimum or maximum acceptable values),
an on-site calibration may be required to suit the real conditions in which the k24 is required to operate.
10.3. CALIBRATION PROCEDURE
K24 permits making quick and precise electronic calibration by changing the calibration factor (k factor).
There are 2 different ways of calibration:
l On-site calibration, performed by means of a dispensing operation.
l Direct calibration, performed by directly changing the k factor.
To enter the calibration phases it is necessary to press and hold down the „CAL“ button.
Why enter the calibration phases?
l Display the currently used calibration factor
l Return to factory K factor after a previous calibration with user k factor
l Change the calibration factor using one of the two previously indicated procedures.
In calibration mode, the partial and total dispensed quantities indicated on the display screen take on different
meanings according to the calibration procedure phase. During the calibration, the k24 cannot perform any normal
dispensing operations. In calibration mode, the totals are not increased.
WARNING
The k24 features a non-volatile memory.
It keeps the calibration and dispensing data stored even after replacing new batteries or long peri-
ods of inactivity.
IMPORTANT
Theowrateismeasuredwithreferencetotheunitofmeasurementofthe
Partial. For this reason, in case of the unit of measurement of the Partial
andTotalbeingdifferent,asintheexampleshownbelow,itshouldbe
rememberedthattheindicatedowraterelatestotheunitofmeasurement
ofthepartial.Intheexampleshown,theowrateisexpressedinQts/min.
Theword„GAL“remainingalongsidetheowratereferstotheregisterof
theTotals(ResetorNONReset)whichareagaindisplayedwhenexiting
fromtheowratereadingmode.
To return to „Normal“ mode, press the CAL key again. If one of the two keys RESET or CAL is accidentally pressed
during the count, this will have no effect.
IMPORTANT
Even though in this mode they are not displayed, both the Reset Total and the General Total (Total)
increase. Their value can be checked after dispensing has terminated, returning to „Normal“mode,
by quickly pressing CAL.
PARTIAL RESET
To reset the Partial Register,nish dispensing and wait for the Remote Display to show a
Flow Rate of 0.0 as indicated in the illustration then quickly press RESET
9.4. DISPENSING WITH FLOW RATE MODE DISPLAY
Itispossibletodispenseuids,displayingatthesametime:
lthe dispensed partial
l the Flow Rate in [Partial Unit / minute] as shown on the following display page:
Procedure for entering this mode:
l wait for the Remote Display to go to Standby, meaning the display screen shows Total only
l quickly press the CAL key.
l Start dispensing
The ow rate is updated every 0.7 seconds. Consequently, the display could be relatively unstable at lower ow rates.
The higher the ow rate, the more stable the displayed value.
9.1. DISPENSING IN NORMAL MODE
Normal mode is the standard dispensing. While the count is
made, the partial and resettable total are displayed at the same
time (reset total).
Should one of the keys be accidentally pressed during dispen-
sing, this will have no effect.
A few seconds after dispensing has ended, on the lower register,
the display switches from resettable total to general total: the
word reset above the word total disappears, and the reset total is replaced by the general total. This situation is
called standby and remains stable until the user operates the k24 again.
9.2. PARTIAL RESET
The partial register can be reset by pressing the reset key when
the meter is in standby, meaning when the display screen shows
the word „TOTAL“
After pressing the reset key, during reset, the display screen rst
of all shows all the lit-up digits and then all the digits that are not
lit up.
At the end of the process, a display page is rst of all shown
with the reset partial and the reset total and, after a few mo-
ments, the reset total is replaced by the non resettableTotal.
9.3. RESETTING THE RESET TOTAL
The reset total resetting operation can only be performed after
resetting the partial register. The reset total can in fact be reset by
pressing the reset key at length while the display screen shows
reset total as on the following display page:
Schematically, the steps to be taken are:
l Wait for the display to show normal standby display page (with
total only displayed),
l Press the reset key quickly
lThe meter starts to reset the partial
l While the display page showing the reset total is displayed
Press the reset key again for at least 1 second
l The display screen again shows all the segments of the display
followed by all the switched-off segments and nally shows the
display page where the reset Reset Total is shown.
NOTE: 6digitsareavailableforTotals,plustwoiconsx10/x100.Theincrementsequenceisthefollowing:
0.0>99999.9>999999>100000x10>999999x10>100000x100999999x100
*AdBlue, a registered trademark of Verband der Automobilindustrie e. V.

6 English English 7
OPERATION DISPLAY
1NONE
K24 IN STAND BY
2LONG CAL KEY KEYING
K24 enters calibration mode, shows „CAL“ and displays the calibration factor in use
instead of total. The words „Fact“and „USER“indicate which of the two factors is currently
in use
3LONG RESET KEY KEYING
K24 shows „CAL“and the partial at zero. K24 is ready to perform on-site calibration
4DISPENSING INTO SAMPLE CONTAINER
Without pressing any KEY, start dispensing into the sample
container.
Dispensing can be interrupted and started again at will. Continue dispensing until the
level of the uid in the sample container has reached the graduated area. There is no
need to reach a preset quantity.
5SHORT RESET KEY KEYING
24 is informed that the calibration dispensing operation is nished.
Make sure dispensing is correctly nished before performing this operation.
To calibrate the K24, the value indicated by the partial totaliser (example 9.800) must be
forced to the real value marked on the graduated sample container.
In the bottom left part of the display an arrow appears (upwards and downwards), THAT
SHOWS the direction (increase or decrease) of the USER K FACTOR value change
when the operations 6 or 7 are performed
6SHORT RESET KEY KEYING
Arrow direction changes. The operation can be repeated IF NECESSARY
7SHORT/LONGCALKEYKEYING
The indicated value changes in the direction indicate by the arrow - one unit for every
short CAL key keying
- continually if the CAL key is kept pressed.
(for the rst 5 units slowly and then quickly)
If the desired value is exceeded, repeat the operations from point (6)
10.6. IN- FIELD CALIBRATION PROCEDURE
10.5.IN-FIELDCALIBRATION
This procedure calls for the uid to be dispensed into a graduated sample container in real operating conditions (ow
rate, viscosity, etc.) requiring maximum precision.
WARNING
For correct K24 calibration, it is most important to:
lcompletely eliminate air from the system before calibrating
luseapreciseSampleContainerwithacapacityofnotlessthan5litres,featuringanaccurate
graduated indicator
lensurecalibrationdispensingisdoneataconstantowrateequivalenttothatofnormaluse,until
the container is full
lnotreducetheowratetoreachthegraduatedareaofthecontainerduringthenaldispensing
stage(thecorrectmethodduringthenalstagesofsamplecontainerllngconsistsinmaking
shorttop-upsatnormaloperationowrate)
lafter dispensing, wait a few minutes to make sure any air bubbles are eliminated from the sample
container; only read the Real value at the end of this stage, during which the level in the container
could drop
lif necessary, carefully follow the procedure indicated below
10.4. DISPLAY OF CURRENT „K FACTOR“ AND RESTO-
RING „FACTORY“ K FACTOR
By pressing the cal key while the appliance is in standby, the
display page appears showing the current calibration factor
used.
If you are using k24 with „factory k factor“ the display page
shown in the diagram will be displayed, with the word „FACT“.
If one „user k factor“has been set, the calibration factor set by the user (in our example 0.998) will be displayed. The
word „USER“ indicates a calibration factor set by the user is being used.
The ow chart alongside shows the
switchover logic from one display page
to another.
In this condition, the Reset key permits
switching from User factor to Factory
factor.
To conrm the choice of calibration fac-
tor, quickly press CAL while „USER“or
„FACT“are displayed.
After the restart cycle, the meter uses
the calibration factor that has just been
conrmed.
ATTENTION
When the Factory Factor
isconrmed,theoldUser
factor is deleted from the
memory
Indicated value
Real value
*AdBlue, a registered trademark of Verband der Automobilindustrie e. V.

8 English English 9
OPERATION DISPLAY
8LONG RESET KEY KEYING
K24 is informed that the calibration
procedure is nished. Before doing this,
make sure the DISPLAYED factor is the
ACTUAL factor.
K24 calculates the new USER K FACTOR. This calculation could require a few seconds,
depending on the correction to be made. During this operation the arrow disappears but
the CAL indication remains.
If this operation is performed after operation (5), without changing the indicated value,
the USER K FACTOR would be the same as the FACTORY K FACTOR, thus it is
ignored
9NO OPERATION
At the end of the calculation, the new USER K FACTOR is shown for a few
seconds,afterwhichtherestartcycleisrepeatedtonallyachievestandby
condition.
ATTENTION: From now on, the indicated factor will become the calibration factor
used by the meter and will continue to remain such even after a battery change
10 LONG CAL KEY KEYING
K24 enters calibration mode, shows „CAL“ and displays the calibration factor in use
instead of total. The words „Fact“and „USER“indicate which of the two factors is currently
in use.
OPERATION DISPLAY
4LONG RESET KEY KEYING
We now go on to Direct change of the calibration factor: the word „Direct“ appears toge-
ther with the Currently Used calibration factor. In the bottom left part of the display, an
arrow appears (upwards or downwards) dening the direction (increase or decrease) of
change of the displayed value when subsequent operations 5 or 6 are performed
5SHORT RESET KEY KEYING
Arrow direction changes. The operation can be repeated to alternate the direction of the
arrowused by the meter and will continue to remain such even after a battery change
6SHORT/LONGCALKEYKEYING
The indicated value changes in the direction indicated by the arrow
l one unit for every short CAL key keying
l continually if the CAL key is kept pressed. The speed increase rises by keeping the
key pressed. If the desired value is exceeded, repeat the operations from point (5)
7LONG RESET KEY KEYING
K24 is informed that the calibration procedure is nished. Before performing this operati-
on, make sure the indicated value is that required
8NO OPERATION
At the end of the calculation, the new USER K FACTOR is shown for a few seconds,
after which the restart cycle is repeated to nally achieve standby condition.
ATTENTION: From now on, the indicated factor will become the calibration factor
used by the meter and will continue to remain such even after a battery change
9NO OPERATION
The K24 stores the new work calibration factor and is ready to begin dispensing, using
the USER K FACTOR that has just been calculated
Indicated value Real value
10.7. DIRECT MODIFICATION OF K FACTOR
This procedure is especially useful to correct a „mean error „obtainable on the basis of several performed dispensing
operations. If normal K24 operation shows a mean percentage error, this can be corrected by applying to the current-
ly used calibration factor a correction of the same percentage. In this case, the percentage correction of the USER K
FACTOR must be calculated by th(e operator)in the following way:
Example:
Error percentage found E% - 0.9 %
CURRENT calibration factor 1,000
NewUSERKFACTOR 1,000*[(100-(-0,9))/100]=
1,000*[(100+0,9)/100]=1.009
If the meter indicates less than the real dispensed value (negative error) the new calibration factor must be higher
than the old one as shown in the example. The opposite applies if the meter shows more than the real dispensed
value (positive error).
11. METERS CONFIGURATION
Some models of meter feature a menu with which the user can select the main measurement unit, Quarts (Qts), Pints
(Pts), Litres (Lit), Gallons (Gal). The combination of the unit of measurement of the Partial register and that of the
Totals is predened according to the following table:
To choose between the 4 available combinations:
l Wait for K24 to go to Standby,
l Press the CAL and RESET keys together. Keep these pressed until the word „UNIT“ appears on the screen
together with the unit of measurement set at that time (in this example Litres / Litres)
l Press the reset key to select the desired combination of unit of measurement, amongst those shown below.
l Save the new combination by pressing the cal key at length. K24 will pass through the start cycle and will then
be ready to dispense in the set units.
WARNING
The Resettable Total and Total registers will be automatically changed to the new unit of measure-
ment. NO new calibration is required after changing the Unit of Measurement.
Combination
no. Unit of Measurement
of the Partial Register Register Unit of Measurement
of the Totals Register
1 Litres (L) Litres (L)
2Gallons (Gal) Gallons (Gal)
3Quarts (Qts) Gallons (Gal)
4 Pints (Pts) Gallons (Gal)
OPERATION DISPLAY
1NONE
K24 IN STAND BY: not in counting mode
2LONG CAL KEY KEYING
K24 enters calibration mode, shows „CAL“and displays the calibration factor being used
instead of the partial. The words „Fact“and „USER“indicate which of the two factors
(factory or user) is currently being used
33 LONG RESET KEY KEYING
K24 shows „CAL“and the partial at zero.
K24 is ready to perform on-site calibration by dispensing
*AdBlue, a registered trademark of Verband der Automobilindustrie e. V. *AdBlue, a registered trademark of Verband der Automobilindustrie e. V.
This manual suits for next models
1
Table of contents
Languages:
Popular Measuring Instrument manuals by other brands

Powerfix Profi
Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot
Test Equipment Depot GVT-427B user manual

Fieldpiece
Fieldpiece ACH Operator's manual

FLYSURFER
FLYSURFER VIRON3 user manual

GMW
GMW TG uni 1 operating manual

Downeaster
Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments
Hanna Instruments HI96725C instruction manual

Nokeval
Nokeval KMR260 quick guide

HOKUYO AUTOMATIC
HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke
Fluke 96000 Series Operator's manual

Test Products International
Test Products International SP565 user manual

General Sleep
General Sleep Zmachine Insight+ DT-200 Service manual