Protimeter BaleMaster GRN6165 Installation instructions

Protimeter BaleMaster® GRN6165
User Instructions
AAS-INS6165, Rev. A
August 2017
A
mphenol
Advanced Sensors
PROTIMETER
BaleMaster (GRN6165)
Protimeter Moisture Meter
Page 2 Protimeter BaleMaster GRN6165
1 SafetyConsiderations
Caution note for the Bale Probe- The Bale Probe point is extremely
sharp and should be handled with due care. The point should be covered with
the cap provided with the probe when not in use
Calibration of unit - The accuracy specifications of the product are valid
for one year from the date of calibration.
Periodic calibration checking, as outlined in section 4 , is advised.
Only operate the measuring instrument properly, for its intended
purpose and within the parameters specified in
the technical data. Readings from moisture meters are not definitive but are
used to help a professional make informed judgement to the material’s
moisture condition. Conductive material such as salts, carbon and metal can
give false positive readings.
2Introduction
The Protimeter BaleMaster is a conductivity moisture meter designed for
measuring moisture level of baled products such as straw and hay. The
instrument is calibrated for wheat straw and may be used to take relative
measurements in other baled products.
Protimeter BaleMaster GRN6165 Page 3
3 Moisture Measurement inBales
Measure moisture in bales as follows:
Note: Before using the stainless steel probe, be sure it is thoroughly clean
and dry. The probe must not be wet when measuring moisture.
a.Connect the Bale Probe to the instrument and switch ON by pressing and
holding button.
b.Push the Bale Probe in to the bale to the required depth. Note the reading
on thedisplay.
Note: When assessing the moisture content (mc) of whole bales, it is
recommended that the average value is obtained from numerous readings
taken from different sides and at different depths.
c.To obtain the correct average value, use the following moisture
measurement ranges as aguideline:
- Less than 15% mc: bales are in a safe condition for storage.
- Greater than 15% mc: bales are not in a safe condition for storage.
Additional drying is recommended.
d.After use, switch the instrument OFF by pressing and holding the
button, or leave the instrument to switch OFFautomatically.
4 CalibrationCheck
Calibration can be checked using the Protimeter Calcheck device (Part #
BLD5086) and a 2-pin moisture probe (Part # BLD5079), connected via the
Bale probe socket. Correctly calibrated BaleMaster instruments register a
value in the range 14.0 – 14.4 at 20°C across the Calcheck.
5 Care andMaintenance
When the BaleMaster is not in use, keep it in its pouch together with its
accessories. Clean and dry the probe, and wrap it in the bubble wrap. Then
store the probe for later use. Replace the batteries when the low battery power
symbol appears on the display, or remove the batteries if the instrument is not
going to be used for extended periods.
Page 4 Protimeter BaleMaster GRN6165
6 ReferenceMode
Measure the material until the meter reading is stable. Then, press for 2
seconds. This will store the reading until the mode changes or the meter turns
off. Now, all readings taken thereafter will be displayed as normal, and below
you will see a second reading that shows if the material is measured above or
below the original reading. Reference mode can be useful when trying to
establish what materials are above or below a point of reference or dry standard.
7 OperatingBaleMaster
Switch On:
Press the ON/OFF button.
The unit turns on, with the LCD displaying all the segments and sweeping the
LED bar graph.
Protimeter BaleMaster GRN6165 Page 5
Measurements:
The numeric measurement and color LED will be shown, as well as “DRY”
(green) or “AT RISK” (yellow) or “WET” (red), based on the measurement
shown.
8.5-15 DRY (Green), 15.1-19.5 AT RISK (Yellow), 19.6-36.8 WET (Red)
Page 6 Protimeter BaleMaster GRN6165
Reference Mode of Measurement:
While the first reading is displayed on the screen, press and hold the
button for 2 seconds to enter the Reference Mode. The display will be similar to
the one shown on the next page.
To return to the normal measurement mode, press again.
Settings:
Press and hold the button and switch on the unit by pressing . Hold
both the buttons until the BaleMaster displays theversion.
Release buttons while it shows the version number. The unit enters
Settings Mode.
Protimeter BaleMaster GRN6165 Page 7
DRY, AT RISK and WET Settings:
The first screen to appear in Settings is the DRY, AT RISK and WET setting
screen.
This screen sets whether the indication on the display needs to be switched
ON or OFF. When it is ON, the moisture condition will be displayed on the
screen. When it is OFF, no indication is displayed on screen.
Use the button to turn ON/OFF thesettings.
If the unit detects no key press for 2 seconds in the Settings screen, it
moves to the next setting.
Pressing the button, after you have entered the desired setting, it will
save the setting and move the screen to the next setting.
Page 8 Protimeter BaleMaster GRN6165
Buzzer ON/OFF Settings:
Once the DRY, AT RISK and WET settings are entered, the next setting turns
the Buzzer ON or OFF.
To change the settings, press the button.
If the unit detects no key press for 2 seconds in the Settings screen, it
moves to the next setting.
Pressing the button, after you have entered the desired setting, will
save the setting and move the screen to the next setting.
Protimeter BaleMaster GRN6165 Page 9
Backlight Settings:
Once the Buzzer settings are entered, the next setting turns the Backlight
ON or OFF.
To change the settings, press the button.
If the unit detects no key press for 2 seconds in the Settings screen, it
moves to the next setting.
Pressing the button, after you have entered the desired setting, it will
save the setting and move the screen to the next setting.
Page 10 Protimeter BaleMaster GRN6165
Auto Off Time Settings:
When Auto Off is set, the unit will shut down automatically at a specified time
between 1 and 6 minutes, if there is no key press detected within the set time.
For example, if the Auto Off time is set as 1, then the unit will automatically
shut down after one minute when no key is pressed.
The duration will change based on the set time of 1- 6 minutes.
If the Auto Off time is set to be “0”, then the unit will not automatically turn
off. A user must manually turn it off by
pressing and holding the button for 5 sec.
Turn off time can be set by pressing the button, and can be saved by
pressing .
After you save this setting, the unit will exit to the measurement screen.
Protimeter BaleMaster GRN6165 Page 11
Battery Low Indication:
Whenever the battery is low, a low battery indication symbol will appear on the
screen (bottom right). This indicates that the battery is low and should be
replaced soon. The unit will continue to perform in battery condition within the
specified accuracy, and turns off when the battery reaches the limit.
Holding / Freezing the Reading:
While measuring, if the reading needs to be frozen for any
observation, press during measurement. A text“HOLD” will be
displayed on thescreen.
8 BatteryReplacement
A 550mAh battery will last continuously for more than 20 of operation. A Low
Battery indication on the screen indicates that the battery needs to be changed
in a short time.
Remove the screw on the back side of the unit.
Slide the battery lid downwards, applying a slight pressure with the thumb to
open the battery compartment.
Remove and replace the battery. Care must be taken to ensure that the
polarity is correct as below.
Page 12 Protimeter BaleMaster GRN6165
9 TechnicalSpecifications
Operating Conditions:
Operating temperature range: 0°C to 50°C Humidity: 0 to 90% RH,
non-condensing
Measurement Specifications:
Protimeter Balemaster (Part# GRN6165) consists of:
BaleMaster instrument (Part# 6165) and Bale probe (GRN6148)
The Balemaster instrument is calibrated for Wheat Straw Measurement
Range: 8.5 – 36.8% mc
10PhysicalSpecifications
Power:
9V- Alkaline 550mAh, Battery low indication on LCD
Size:
19cm x 6.5cm x 3.5cm (7.5" x 2.5" x 1.4")
Gross Weight (without battery):
~228g
Buzzer:
Audible buzzer for key tone and measurement indication, user configurable
Regulatory Compliance:
CE, RoHS, ETL
Protimeter BaleMaster GRN6165 Page 13
MODEL: GRN6165
Meter
POL6148
Bale Probe
Page 14 Protimeter BaleMaster GRN6165
Contact Information:
www.protimeter.com
www.amphenol-sensors.com
Copyright © 2017 Amphenol Thermometrics, Inc. All rights reserved.
Protimeter BaleMaster®is a registered trademark of Amphenol Thermometrics, Inc.
U.S.A.
Amphenol Thermometrics, Inc.
967 Windfall Road
St Marys, Pennsylvania 15857, USA
T: +1 814-834-9140
U.K.
Amphenol Thermometrics (U.K.) Limited
Crown Industrial Estate
Priorswood Road
Taunton, TA2 8QY, UK
T: +44 1823 335 200
Amphenol Thermometrics, Inc.
967 Windfall Road
St. Marys, Pennsylvania 15857, USA
Amphenol
Advanced Sensors

Other Protimeter Measuring Instrument manuals

Protimeter Surveymaster BLD5365 User manual

Protimeter

Protimeter Surveymaster BLD5365 User manual

Protimeter MMS3 User manual

Protimeter

Protimeter MMS3 User manual

Protimeter MMS2 User manual

Protimeter

Protimeter MMS2 User manual

Protimeter Digital Mini BLD5702 User manual

Protimeter

Protimeter Digital Mini BLD5702 User manual

Protimeter TimberMaster BLD5605 User manual

Protimeter

Protimeter TimberMaster BLD5605 User manual

Protimeter HygroMaster2 User manual

Protimeter

Protimeter HygroMaster2 User manual

Protimeter BLD7705 User manual

Protimeter

Protimeter BLD7705 User manual

Protimeter MMS Plus User manual

Protimeter

Protimeter MMS Plus User manual

Protimeter BLD5765 User manual

Protimeter

Protimeter BLD5765 User manual

Protimeter Grainmaster i User manual

Protimeter

Protimeter Grainmaster i User manual

Protimeter BLD5765 User manual

Protimeter

Protimeter BLD5765 User manual

Protimeter Surveymaster User manual

Protimeter

Protimeter Surveymaster User manual

Protimeter TimberMaster BLD5609 User manual

Protimeter

Protimeter TimberMaster BLD5609 User manual

Popular Measuring Instrument manuals by other brands

Sigineer Power MH-M80 Installation and operation manual

Sigineer Power

Sigineer Power MH-M80 Installation and operation manual

GE Wiring Devices & Specialty Products GE5805WS6 user manual

GE Wiring Devices & Specialty Products

GE Wiring Devices & Specialty Products GE5805WS6 user manual

Hach HQ11d Basic user manual

Hach

Hach HQ11d Basic user manual

Mastech MS2128A Operation manual

Mastech

Mastech MS2128A Operation manual

Major MT 730 instruction manual

Major

Major MT 730 instruction manual

SolaX Power X1-Meter-WiFi Quick installation guide

SolaX Power

SolaX Power X1-Meter-WiFi Quick installation guide

Powell EcoVisor 01.41B.48080 Instruction bulletin

Powell

Powell EcoVisor 01.41B.48080 Instruction bulletin

IFM Electronic efector 300 SV Series operating instructions

IFM Electronic

IFM Electronic efector 300 SV Series operating instructions

koban KPA-01T manual

koban

koban KPA-01T manual

Ninglu AM706 Operation & installation manual

Ninglu

Ninglu AM706 Operation & installation manual

REMA DSD05 Series owner's manual

REMA

REMA DSD05 Series owner's manual

Doyon WM35 manual

Doyon

Doyon WM35 manual

EUTECH INSTRUMENTS DO 700 instruction manual

EUTECH INSTRUMENTS

EUTECH INSTRUMENTS DO 700 instruction manual

Knauer K-7400S user manual

Knauer

Knauer K-7400S user manual

IMV Easi-Scan quick guide

IMV

IMV Easi-Scan quick guide

ICM Controls ICM450 Installation, operation & application guide

ICM Controls

ICM Controls ICM450 Installation, operation & application guide

janitza UMG 508 Handbook

janitza

janitza UMG 508 Handbook

Hioki 3490 instruction manual

Hioki

Hioki 3490 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.