Radcal AGLS User manual

Specifications –
Light Sensor
Luminance Properties
Range
Resolution
Aperture
Illuminance Properties
Range
Resolution
Calibration Uncertainty
Photopic Response
Cosine Response
Calibration Traceability
0.05 to 50,000
cd/m2
0.01
cd/m2
10-mm diameter
0.05 to 50,000 lux
0.01 lux
<3% or 0.05 (lux or
cd/m2
)
<3% Deviation from 1931 CIE
Standard Photopic Curve
<2.5% based on the CIE f2 Test
NIST and other International
standards
Overview –
The AGLS is a sensor accessory to the Accu-Gold family of products that
include the Accu-Gold digitizer and Accu-Gold Touch systems. The AGLS
is an excellent Quality Control/Quality Assurance solution for measuring
luminance and illuminance.
The base unit of the light sensor contains a silicon photo diode that is
exposed to ambient light through an optical filter which very accurately
simulates the spectral response of the human eye (the photopic
response). In this way, the light sensor measures illuminance of the
ambient in units of lux.
A luminance adapter can be attached to the base unit in front of the
photo‐diode for the measurement of luminance of a monitor in cd/m2units.
The aperture of the Luminance Adaptor is 10‐mm in diameter.
The Accu‐Gold digitizer and Touch systems automatically detect the
presence or absence of the luminance adapter and automatically select
the appropriate measurement and units.
Base Unit -
Illuminance
Luminance
Adapter
AGLS Light Sensor
Manual
Radcal Corporation
426 West Duarte Road
Monrovia, CA 91016-4591 USA
USA (626) 357-7921
Fax USA (626) 357-8863
email [email protected]
www.radcal.com
Radcal Part # MNL/AGLS
4094149-MANUAL AGLS Rev D
Printed: November 2020
Copyright © 2018, 2019, 2020 Radcal Corp.
Setting Up the AGLS with Accu-Gold 2.0
Software –
Getting Started
The Accu-Gold 2.0 software is a Windows application that remotely
operates an Accu-Gold stand-alone digitizer or the digitizer of the
Accu-Gold Touch Pro system through a USB or wireless connection. In
order to install the Accu-Gold 2.0 application and learn the basics of its
operation refer to the Accu-Gold 2.0 User’s Guide on the Flash Drive
provided with your instrument or the Radcal Support web page.
Making the Connections
Begin by plugging the AGLS into the Accu-Gold Touch AUX connector on
your Accu-Gold digitizer or Accu-Gold Touch Pro system.
If using a USB connection, ensure that the USB cable is connected to both
your computer/tablet and the instrument. For the Accu-Gold Touch Pro,
turn on the instrument and select USB Mode.
If using the wireless interface with an Accu-Gold digitizer, attach the
Nugget wireless adaptor to your digitizer and turn it on. For the Accu-Gold
Touch Pro, turn on the instrument and select Wi-Fi Mode.
Start the Accu-Gold 2.0 application and ensure that the software is properly
connecting to the instrument (see the Accu-Gold 2.0 User’s Guide for more
details). Select the desired connection method (USB or Wi-Fi):
Place the Accu-Gold 2.0 application in “Ready” mode. The display will
automatically adjust to show the AGLS values.
Setting Up the AGLS with Accu-Gold Touch
(stand-alone) –
Begin by plugging the AGLS into the Accu-Gold Touch AUX connector
and then turn on the Touch system. Click on Start Display to operate the
AG Touch in stand-alone mode.
Using the arrow keys, navigate to the appropriate screen for light sensor
measurement and place the Touch in “Ready” mode (See the AG Touch
Manual for more details).
Making Illuminance Measurements –
Remove the Luminance Adaptor and position the light sensor base as
desired to measure illuminance. This is accomplished by unscrewing the
adaptor in a counter-clockwise direction. A switch is automatically
released when the adaptor is removed instructing the system to make
illuminance measurements reported in lux. Press and hold the AGLS
activator button for several seconds. The lux value will appear in the
display; when it stabilizes release the button to capture the value.
Note: when switching between modes, press Pause, change the adapter,
then press Play.
Making Luminance Measurements –
Attach the Luminance Adaptor to the base by screwing it on in the
clockwise direction. Ensure that the adaptor is fully attached so that the
actuator switch is depressed instructing the system to make luminance
measurements in units of candela/meter2(cd/m2). Position the sensor as
shown below over the screen area to be measured. Press and hold the
AGLS activator button for several seconds. The luminance value in cd/m2
will appear in the display; when it stabilizes release the button to capture
the value.
Note: the adaptor must be used with the felt cap shown below and the cap
must be fully seated. Luminance calibration is specific to the cap,
luminance adaptor, and light sensor base. For this reason, care should be
taken to always use the same three components together.
Illuminance Base
Sensor
Activator
Button
On the start
screen click
"Start Display".
The measure
window will
automatically be
loaded.
Navigation arrowsReady mode button
“Light”
screen

Popular Accessories manuals by other brands

Goobay PowerBank 1000mAh LIPOL (XS-size) user manual

Goobay

Goobay PowerBank 1000mAh LIPOL (XS-size) user manual

Hearth and Home Technologies Heat-Zone-240V Air Duct Kit Installation and operation instructions

Hearth and Home Technologies

Hearth and Home Technologies Heat-Zone-240V Air Duct Kit Installation and operation instructions

Palram Applications BORDEAUX 2230 instructions

Palram Applications

Palram Applications BORDEAUX 2230 instructions

Wenglor P1NL Series operating instructions

Wenglor

Wenglor P1NL Series operating instructions

wtw FDO 70x IQ Series operating manual

wtw

wtw FDO 70x IQ Series operating manual

Hengst WGM-B operating instructions

Hengst

Hengst WGM-B operating instructions

Banner iVu PLUS TG quick start guide

Banner

Banner iVu PLUS TG quick start guide

Baumer O200 instruction manual

Baumer

Baumer O200 instruction manual

Lowrance EP-25 installation instructions

Lowrance

Lowrance EP-25 installation instructions

Edimax WS-2001P Quick installation guide

Edimax

Edimax WS-2001P Quick installation guide

Nordic DELTACO PB-Q1001 user manual

Nordic

Nordic DELTACO PB-Q1001 user manual

Torc Mojave Warranty & instructions

Torc

Torc Mojave Warranty & instructions

Ecolink H214101 installation instructions

Ecolink

Ecolink H214101 installation instructions

Varta 57959 manual

Varta

Varta 57959 manual

Ultra Products Indispensable Fastener Set ULT31578 Specification sheet

Ultra Products

Ultra Products Indispensable Fastener Set ULT31578 Specification sheet

Parkside PDSS B2 Operation and Safety instructions, Translation of the original instructions

Parkside

Parkside PDSS B2 Operation and Safety instructions, Translation of the original instructions

FIBARO FGWPA-111 operating manual

FIBARO

FIBARO FGWPA-111 operating manual

HIOTH TECHNOLOGY CT-HAWF1008-BL user manual

HIOTH TECHNOLOGY

HIOTH TECHNOLOGY CT-HAWF1008-BL user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.