Ramblegard RG43 User manual

Display Monitor
(RG43)
User Instructions


The Ramblegard Central Monitor Display is an
affordable falls management and nurse-call
solution for Care Homes & Hospital wards.
Caregivers can conveniently read a large LED
display from a distance, and the panel can also
forward signals to a Caregiver Alert Pager that
will display call type and device ID.
Alarms can be reset using a wireless reset
button, cordless monitor, pager, or at the panel.
Straightforward programming allows users to
quickly set up the system, and easily add/remove
devices based on changes in daily requirements.

1.1 Open battery case and insert 6 x AA alkaline batteries or
remove the pull tab. Batteries are intended as a back up only,
in case of power loss and will give approximately 4-8 hours of
operation. The product must be connected to a power supply for
normal operation.
1. 2 Attach the antenna to improve the range of wireless devices.
1.3 Your panel should be pre-configured to operate with your kit.
If not, refer to programming section below for details.
1.4 Adjust volume to desired level by using the wheel on the
right side of the monitor. Tracking should only be ON if mobile
devices (wheelchairs or pendant call buttons) are in use.
Otherwise, set this switch to OFF. Switch on the button lock to
prevent accidental activation of the buttons on the front of the
device.
1. 5 Place display monitor on a desk or mount on the wall in a
clearly visible location with the power cable safely and conveniently
routed.
1. Set-Up

2. Programming
2.1
Panel Setting
To access the panel settings menu, press and hold the PROGRAM
+ SELECT1 buttons for ~5s. The display will beep once. Use the
SELECT1 button to cycle between the settings, and SELECT2 to
change.
Call Forward
:
To configure the call-forward feature, tap SELECT1
until the tens digits displays “t”. Then use SELECT2 to adjust the
setting:
Set P = Pause Setting
0 OFF
1 Not Used
2 Not Used
3 Signals to RG41 (Pager Unit)
4 Panel receives from master, does not transmit
Site Code: Setting a site code can be useful if several panels are
in use, and will ensure pagers only receive alarms from their area.
• Tap SELECT1 until the tens digit displays “A”.
• Use SELECT2 to set 0-7
• This setting must match whatever has been selected on the
wireless devices in the area
Pause Setting: This allows a caregiver to pause the panel for a
programmable interval, allowing patients to be removed from
their beds without triggering an active alarm.
• Tap SELECT1 until the tens digit displays “P”.
• Use SELECT2 to change the setting:
Set P = Pause Setting
0 Panel cannot be paused
1-9 Pause 1-9min
A Pause for up to 20min
B Pause for up to 40min
C Pause for up to 60min
D Pause for up to 80min
E Pause for up to 100min
F Pause for up to 120min

Alarm Tone: To change this setting, tap SELECT1 until the tens
digit displays “b”. Use SELECT2 to choose between b1 (ascending
harmonic scale) or b2 (standard beep).
Delay Setting: At times, a caregiver may not require an
immediate alarm when a patient exits the bed (e.g. using the toilet
at night) but would like to know if they have not returned to bed
after a certain time. Use the delay setting to choose the desired
time before the panel alarms. The delay setting only applies to
Pad-type alarms (bed, chair, and floor). To change this setting, tap
SELECT1 until the tens digit displays “d”. Use SELECT2 to change
the setting.
2. Programming (cont.)
Set d = Delay Setting on Panel
0 Immediate alarm for bed exit
1-9 Alarm delays 1-9min after bed exit
A Alarm delays 10min after bed exit
B Alarm delays 15min after bed exit
C Alarm delays 20min after bed exit
D Alarm delays 25min after bed exit
E Alarm delays 30min after bed exit
F Alarm delays 35min after bed exit
Pager Code: Setting a pager code can be useful if several panels
are in use, and will ensure pagers only receive alarms from their
area.
• Tap SELECT1 until the tens digit displays “C”
• Use SELECT2 to set 0-9
• This setting must match whatever you have selected on the
pagers in the area.
2.2 Add/Remove Devices
To enter individual device pairing mode, press and hold PROGRAM
+ SELECT2 for ~5s. The display will beep once and briefly display
[PP] when programming has been entered, followed by [01], which
is the default slot.
• Use the SELECT1 button to adjust the tens-digit
• SELECT2 to adjust the units to choose a memory slot.
• The PAD, CALL, and SENSOR LEDs on the left side will indicate
whether any devices occupy the memory slot.

2. Programming (cont.)
To add a device to RG43 memory, just trigger an alarm from the
device you want to add. The indication LED for the device type will
turn on to indicate the memory slot now contains this device type
(Call, Pad, or Sensor). Multiple devices can be added to the same
memory slot. Adding a wireless reset button to a memory slot will
allow the button to reset of any active alarm in this slot.
To delete a device from RG43 memory, select the memory slot
and hold RESET for ~5s. You should see the indication LED turn off
to indicate the device is no longer present in memory.
2.3 Auto-Programming
The RG43 panel now supports auto-programming when used
together with an RG2 Companion Wireless Monitor.
RG2 Companion Wireless Monitor must be set to t=2 or t=3 for
this to work.
On the RG2:
• Press PROGRAM + PAUSE for ~2s
• For t=2, use PAUSE to adjust tens digit and RESET to adjust the
units if the monitor will be moved to a new room
• For t=3, the display settings cannot be adjusted, and the
monitor will display [AP] to indicate it is ready for auto-pairing.
• Take the monitor next to the RG43 panel to synchronise the
displays.
On RG43:
• Hold PROGRAM + REVIEW for ~5s to enter pairing mode
• The RG43 will briefly display [AP], and all devices paired with the
RG2 monitor will be uploaded to the panel.
• The panel will indicate the number of devices that are received
from the RG2 monitor, followed by the new display ID or IDs.
Example 1. If an RG2 monitor with a bed and chair sensor programmed in and are set to t=2, then is uploaded with
new display [34], the RG43 panel will rst show [d2] to indicate that two devices are paired into this display, followed
by [34] to conrm what will now display when a device is triggered.
Example 2. If an RG2 monitor set to t=3 with two bed sensors programmed to displays [34] and [35], the
RG43 will rst show [d2] to indicate that two devices are being paired, followed by [34], [35] to conrm
what will now display when these devices are triggered.

Any devices that were previously programmed into these display
slots on the RG43 panel will be overwritten. If these devices were
previously programmed into a different displays slot on the RG43
panel, they will be automatically deleted.
The RG43 panel will only accept alarm signals from devices that
have been entered into memory. When the RG43 panel receives
an alarm signal it will display the device ID (01-99), sound an
audible alarm, and transmit the alarm details to a pager if the
panel has been configured to do so. An LED will flash on the left
side of the panel to indicate the type of device that made the call.
3. Operation
Device Type Delay Setting on Panel
Wireless Pad Devce ID + Pad LED
Call button Device ID + Pad LED
Door Sensor Device ID + Sensor/Door LED
Power LED: This will be steady ON when connected to a power
supply. It will flash if the display is using battery power, and a beep
will sound if AC is lost.
Panel Battery: This indicates the status of the panel only. It
will flash if there is low battery or batteries have not been
inserted. This LED can only be reset by changing the batteries.
The two LEDs next to the large digits indicate the status of
wireless devices:
Signal Lost: will turn on if tracking is active and the panel cannot
detect one of the devices in memory.
Device Battery: will turn on if a wireless device is low on
batteries.
• If one of these LEDs is flashing and no ID is displaying, press
REVIEW to check the relevant device.
Button Lock: The switch on the right side of the panel will
prevent any devices from accidentally being erased from memory.
The only buttons that will continue to function are REVIEW and
RESET when this is active.
Reset Button:When an alarm is active, a short press will silence
the alarm, and a long press will erase the alarm. If no alarms are
active, the panel can be put into Pause mode by holding RESET
for 3 seconds if this has been enabled (see Programming section).
While paused, the panel display will flash [PA].

Pause mode can also be entered using a RG41 Jill Pager by holding
CANCEL until the LCD displays PAUSE. Entering pause using a
pager does not have adjustable time, and the panel will exit after
2min or if a second alarm is received.
Review Button: This will cycle through events from the last 15
minutes, beginning with the most recent. If there have not been
any recent alarms there will be no display.
Select 1: When no alarms are present, pressing this will display
active pads in the area.
Frequently Used Devices
Cordless Pads: Up to 30 wireless pads can reliably be used
with the RG43 panel before signalling interference reduces the
reliability of alarms. Pads typically work within 25m of the panel in
indoor environments. Always test pads before putting them into
use with a patient. If tracking is set to ON and pad signals display
as lost, move these pads closer to the panel.
Wireless Reset Buttons: Wireless reset buttons can be
assigned to a memory slot to give a totally wireless falls man-
agement solution. If multiple devices are programmed into a
single memory slot, this will reset any alarm from the slot into
which it has been paired. When using wireless reset buttons,
Ramblegard recommends holding the button for 3s to ensure the
panel receives the signal.
Cordless Monitors: Cordless RG2 monitors can now be used
as a “kit” together with a RG43 panel. The monitor will act as
pause/reset for any devices paired with it, and can be conveniently
moved around a ward when a caregiver needs the monitor for a
different patient.
LEDs: The RG41 pager is a useful accessory to the RG43 panel,
which will display alarm type and location, can reset alarms, or
put the RG43 panel into PAUSE to remove a patient from bed
without triggering an active alarm. If pagers will be used with
the RG43, make sure the panel is configured to relay signals (t=3),
that pagers are configured with direct signalling OFF, and the site
code matches what has been assigned to the panel. If either of
these settings does not match, the pager will not receive the call
forward from the RG43 Panel.
3. Operation (cont.)

Paging: If a pager is not receiving signals, first check that
the RG43 panel is configured to repeat signals. The standard
Ramblegard solution includes a RG41 pager, so set Call Forward
to t = 3. If this does not solve the issue, check that the pager is
configured to turn Direct Signalling OFF.
• Check that the area code and paging code settings on the RG43
panel match what you have selected on your pager.
• Erase the pager memory and reprogram the settings.
• Contact your local distributor with details for further
assistance.
Signal Lost:If a device is displaying as signal lost on the RG43
panel, you can silence this indication by pressing RESET.
The SIGNAL LOST LED will continue to flash until the device has
been brought back into range or is removed from memory.
• If you would like to remove this device from the panel memory,
hold REVIEW + RESET for ~3s. The RG43 panel will flash [dE], then
the number to confirm it has been deleted.
LEDs: If you cannot reset an error LED, enter programming mode
and erase the memory slot. If multiple devices have been entered
into this slot you will need to reprogram them into memory.
4. Troubleshooting


T: 028 9448 3971
www.ramblegard.com
S&E CareTrade t/a Ramblegard
Elmfield House
Rathenraw Industrial Estate
Greystone Road
Antrim
BT41 2SJ
Other manuals for RG43