Particle Count Meter
3
The HandiLaz Mini is designed primarily for sampling in flow
cabinets, cleanrooms and other filtered air environments. Non
filtered air has a relatively high particle count so if the meter is
used in a normal indoor environment the meter will work correctly
but the residual particles must be flushed from the meter before it
will accurately measure in a low particle count area.
A zeroing filter is provided to assist with this.
Remove the filter from the sealed packet, remove the protective
cap from the sample tube, carefully fit the supplied clear tube onto
the sample tube and to the zeroing filter. Make sure the arrow on
the filter points towards the HandiLaz Mini.
Set the meter to continuous sampling mode (page 3-4 of the
manual) and run the meter until the display reads zero. It may
help to tap the sample tube gently with the fingers to dislodge any
remaining particles from inside.
Keep the Zeroing Filter in a sealed bag so it can safely be used
again. The filter needs to be replaced if the HandiLaz Mini gives
a low airflow warning, (Fault Code F), when using the Zeroing
Filter.
Filter shown in sealed pack as supplied