Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
At Resonance Technology, Inc., patient safety is our top priority. Please review this section completely as its contents are vital to
the safety of the installer, the clinician/operator, and the patient.
USE RESTRICTIONS
Warranty and Shelf life
The Serene Sound System comes with a one-year original manufacturer warranty and a shelf life of two years from
the date of installation. Optional extended warranty may be purchased for this system. With patient safety in
mind, Resonance Technology, Inc. recommends daily to weekly visual inspection on heavily used parts of the
system like the Headset. There is no set maintenance requirements for the Serene Sound System, but Resonance
Technology offers its customers product servicing for this Serene Sound System components that are no longer
functioning properly.
Restrictions on Using Non-Resonance Technology, Inc. Components with the Serene Sound system
The original manufacturer’s warranty will be voided if any non-Resonance Technology, Inc. Power Supply is used to
provide power the Serene Sound Transducer.
In addition, Resonance Technology, Inc. cannot be held responsible or liable for any unauthorized use of this
equipment. If you have any questions about how to operate this system, please read this user manual or call
Resonance Technology, Inc. customer service at (818) 882-1997 or email support@mrivideo.com.
HEALTH CONCERNS
Caution: Sound Volume.
Do not use the audio headset on patients at high volume. It is advised against continuous loud and extended audio
play. If your patients experience ringing in their ears, reduce the audio volume or discontinue use the system use
until the patient feels comfortable.
MRI ENVIRONMENT HAZARDS
Installation of materials inside the MRI suite must be done with extreme caution and only by authorized personnel.
Care must be taken to keep ferromagnetic materials such as tools, filter plates, screws, etc. at least three meters
(approximately 1 feet) away from the energized magnet. Absolutely no work should be done near the filter panel
when a scan is in progress.
All cabling inside the MRI environment should either be connected or terminated properly. Failure to do so may
result in skin burns related to RF energy. All cables should be run straight and never looped, as this may also
cause serious skin burns inside the MRI room.
In addition, no persons with ferromagnetic prosthetic devices such as pacemakers or joint replacements should
enter the MRI suite at any time. Extremely high magnetic forces have the potential to dislodge ferrous items at
high velocities that can result in serious injury or death.
Page 2
WARNING
Prior to every use, inspect all system components that come in contact with the patient. Discontinue product
usage immediately if any damage is evident or presents other potential hazards. Use of damaged
components may cause injury to the clinician or the patient. Examples of hazardous damage include but is not
limited to the following:
Microphone boom separated from headset housing (exposing wires)
Cable padding torn, exposing cable inside
Any damage potentially exposing wires to the patient
If the headset becomes damaged, stop using the headset immediately and notify Resonance Technology, Inc. Customer
Service for assistance at (818) 882-1997 or (8 ) 428-6788. Use of broken components can cause injury to the clinician or
the patient.