Resonance Technology Serene Sound User manual

Installation / Operation Manual
S e r e n e S o u n d
A u d i o S y s t e m
Resonance Technology, Inc.
18121 Parthenia Street
Northridge, California 91325, USA
+1 (818) 882-1997
+1 (8 ) 428-MRTV
www.mrivideo.com
Innovative Audio Entertainment /
2 Way Communication Systems

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Notice:
This document contains proprietary information protected by copyright. Rx only. All rights
reserved. No part of the document may be photocopied or reproduced without the prior
written consent of Resonance Technology, Inc. Information contained in this document is
covered by one or more of the following U.S. patents 5627902, 5877732, 5432544,
5412419.

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
TABLE OF CONTENTS:
Page
Introduction 1
Safety Instructions 2-3
Block Diagram 4
Materials Checklist 5-6
Room Layout and Installation Overview 7
How to Connect
Transducer
Power cables
8
-
9
Running MR Laser Link Cable 1
Components Interconnection 11-13
Auto Voice Setup 14-15
Control Room Setup 16
Operation 17-2
Patient Setup
2
1
Cleaning 22
Troubleshooting 23

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Congratulations on your purchase of the MRI-Compatible Serene Sound Audio system. This device
represents more than 25 years of development and state-of-the-art engineering. We are confident this
product will give you the tools you need for comforting the patient while undergoing MRI scans. This
installation/operation manual outlines how to properly care for and operate this system.
Thank you for choosing to purchase this system from Resonance Technology, Inc., the leader in fMRI and
MRI patient comfort systems.
Suggestions on how to improve this system are always welcome.
Page 1
Introduction

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
At Resonance Technology, Inc., patient safety is our top priority. Please review this section completely as its contents are vital to
the safety of the installer, the clinician/operator, and the patient.
USE RESTRICTIONS
Warranty and Shelf life
The Serene Sound System comes with a one-year original manufacturer warranty and a shelf life of two years from
the date of installation. Optional extended warranty may be purchased for this system. With patient safety in
mind, Resonance Technology, Inc. recommends daily to weekly visual inspection on heavily used parts of the
system like the Headset. There is no set maintenance requirements for the Serene Sound System, but Resonance
Technology offers its customers product servicing for this Serene Sound System components that are no longer
functioning properly.
Restrictions on Using Non-Resonance Technology, Inc. Components with the Serene Sound system
The original manufacturer’s warranty will be voided if any non-Resonance Technology, Inc. Power Supply is used to
provide power the Serene Sound Transducer.
In addition, Resonance Technology, Inc. cannot be held responsible or liable for any unauthorized use of this
equipment. If you have any questions about how to operate this system, please read this user manual or call
Resonance Technology, Inc. customer service at (818) 882-1997 or email support@mrivideo.com.
HEALTH CONCERNS
Caution: Sound Volume.
Do not use the audio headset on patients at high volume. It is advised against continuous loud and extended audio
play. If your patients experience ringing in their ears, reduce the audio volume or discontinue use the system use
until the patient feels comfortable.
MRI ENVIRONMENT HAZARDS
Installation of materials inside the MRI suite must be done with extreme caution and only by authorized personnel.
Care must be taken to keep ferromagnetic materials such as tools, filter plates, screws, etc. at least three meters
(approximately 1 feet) away from the energized magnet. Absolutely no work should be done near the filter panel
when a scan is in progress.
All cabling inside the MRI environment should either be connected or terminated properly. Failure to do so may
result in skin burns related to RF energy. All cables should be run straight and never looped, as this may also
cause serious skin burns inside the MRI room.
In addition, no persons with ferromagnetic prosthetic devices such as pacemakers or joint replacements should
enter the MRI suite at any time. Extremely high magnetic forces have the potential to dislodge ferrous items at
high velocities that can result in serious injury or death.
Page 2
WARNING
Prior to every use, inspect all system components that come in contact with the patient. Discontinue product
usage immediately if any damage is evident or presents other potential hazards. Use of damaged
components may cause injury to the clinician or the patient. Examples of hazardous damage include but is not
limited to the following:
Microphone boom separated from headset housing (exposing wires)
Cable padding torn, exposing cable inside
Any damage potentially exposing wires to the patient
If the headset becomes damaged, stop using the headset immediately and notify Resonance Technology, Inc. Customer
Service for assistance at (818) 882-1997 or (8 ) 428-6788. Use of broken components can cause injury to the clinician or
the patient.

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Only system components explicitly designated for use in the MRI suite should be placed inside the MRI suite.
Components not designated for use inside the MRI suite may present a projectile hazard and can become airborne,
causing property damage, serious bodily injury, or death. Please refer to the installation block diagram to
determine which components belong inside the MRI suite.
Resonance Technology, Inc. will not be held liable for any injuries or property damage which may occur as the
result of improper use or installation of this product. By agreeing to this notice, users certify that they are familiar
with basic safety procedures in an MRI room environment and that they have read and understand these safety
precautions.
For questions regarding installation procedures or this manual, Resonance Technology, Inc. technical support staff
may be reached Monday through Friday 8 a.m. to 5 p.m., Pacific Standard Time at (818) 882-1997, or by email at
support@mrivideo.com.
Caution: Electric Shock.
Failure to observe all operating and maintenance instructions may cause damage to your product, property
damage, or injury or death from electric shock, fire or other causes.
Caution: Do Not Disassemble.
Only trained, authorized personnel should perform any required service on this product. Failure to comply with this
warning may result in property damage, injury or death from electric shock, fire or other causes.
Caution: Avoid Extreme Heat, Wet, Humid, Dusty and Smoky Environments.
This product may be deformed by high temperature, direct sunlight and dropping or other mechanical shock. Do
not expose the unit to rain or moisture.
Caution: Unplug the Product When Not in Use.
Always unplug this product when not in use for extended periods, such as during vacations and MRI maintenance.
Page 3

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Block Diagram
Page 4
Patient Headset
Serene Sound
Transducer
Serene Sound
Controller
Tech Remote
Control
Transducer Power Supply
MR Laser Link Cable (15 feet)
Penetration
Panel
Waveguide
Transducer Power
Cable (3m / 1 ’)
Transducer Power
Cable (15m / 5 ’)
Penetration Panel
10 meter long
power cable
connected inside
the magnet room
and into the
Transducer unit.
2 meter long power cable connected
outside the magnet room into the DB-9
filter and Transducer power supply [Inside
Equipment oom]
DB
-
9 Filter installed on the
penetration panel [Body inside the
M I SUITE]
In order to connect the Transducer power cable to the power supply outside the magnet room, you need to use a DB-9 filter
provided with the system. Make sure to connect the two-meter long cable to both the power supply and penetration panel and
connect the 1 meter long cable to both the penetration panel and the Transducer.
Control Room
Speakers
Tablet
Hospital Grade
power cord

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Materials Checklist
The Serene Sound system comes complete with all the necessary components for full installation. The following
checklist is provided for verification:
Main System Components
Part Number Quantity Photo Description MR Safe Installation
Location
RTC-66 -17 -37 - 1
Serene Sound Controller
Control Room
RTC-66 - 5 -535- 1
Serene Sound Controller Power
Supply
Control Room
RTC-65 - 4 - - 2 1
Small Speakers
Control Room
RTC-65 - 67- - 1
Slim Softshell Headset with
Headband
Magnet Room
RTC-66 - 5 -54 - 1 1
Serene Sound Transducer
Magnet Room
RTC-65 - 5 - 5- 1
Transducer Switching [black]
Power Supply
Equipment
Room
1
Touch Screen Tablet
Control Room
RTC-65 - 2 - - 1
Technologist Remote Control
Control Room
Penetration panel accessories (Magnet Type Confirmed with purchase)
Part Number Qty Photo Description MR Safe Install Location
RTC-65 - 82-1 77-A 1
Filter Plate for Philips Penetration Panel
Pen panel installation
RTC-ALS-FPP 1
Filter Plate for Siemens Penetration Panel (large version)
Pen panel installation
RTC-65 - 81-1 8 - 1 Filter Plate for Siemens Penetration Panel (small version)
Pen panel installation
RTC-65 -3 -6 5- 1 9-pin D-Sub Filter
Pen panel installation
RTC-1 1-21 - 1- 1 1
Transducer DC Power cable – 3 meters (~1 ’)
From Pen panel to
Transducer Power Supply
(equipment room)
RTC-1 1-245- 1- 1 1
Transducer DC Power cable – 15 meters (~5 ’)
From Pen panel to
Transducer (magnet
room)
Page 5

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Checking Materials (continued)
Common Initial Shipment System Components/Accessories
Part Number Qty Photo Description MR Safe Installation Location
RTC-CV 1
Installation / Operation Manual (This manual)
Stored/ Control Room
RTC-65 - 5 -27 - 1
1
Transducer Acrylic Stand
Magnet Room
RTC-65 - 5 -182-
Transducer Mounting Plate
(not shipped when Acrylic stand is shipped)
Magnet Room
RTC-ALS-HEC 5
Headset Earpiece Covers
Stored/ Control Room
RTC-65 - 6 - 52- 1
Headset Microphone Cushion cover
Stored/ Control Room
RTC-ALS-TWR 5
Tie Wraps
Control Room
RTC-ALS-TWH 2
Tie Wrap Holders
Control Room
RTC-ALS-HDH 1
Headset Hooks
Magnet Room
RTC-1 1-1 2- 1-1 1 1
2-Channel MR Laser Link Cable
Installed From control
room to magnet room
RTC-1 1-3 6- 3- 2
Hospital Grade AC Power Cord
Equipment Room/
Control Room
RTC-65 - 5 - 5-PF1 2
Remote Outlet Switch with Remote
Equipment Room/
Control Room
RTC-33 -17 -4 4- 1
Mini Stereo Male to Male Cable
Stored/ Control Room
Uncommonly Shipped System Component/Accessories unless requested/purchased by customer
Part Number QTY Photo Description MR Safe Install Location
RTC-65 - 64- - 2 1
Earbud Headset
Magnet Room
RTC-65 - 67-1 31- A 1
Earbud Headband
Magnet Room
2 Earbud HS – Yellow foam Round cushions
Magnet Room
RTC-65 - 69-1521-X 1 5
Earbud Ear tips and Foam
Stored/ Control Room
RTC-1 1-239- 1- 1 1
5-pin Round Female to 5-pin Round
Female Cable (12M/39ft)
Magnet room
RTC-651- -453- 1
Custom Headrest for 8-channel headcoil
Control Room
Page 6

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Room Layout and Installation Overview
Below is the standard setup for the Serene Sound system. Your individual installation may vary,
but will generally be distributed in three areas: 1) Control Room, 2) Computer/Equipment Room
and 3) Magnet Room.
The Control Room setup consists of placement and connection of the Serene Sound Controller,
Technologist Remote Control and Stereo system.
A single Laser Link Fiber Optics cable bundle will run from the Serene Sound Controller through the
penetration panel waveguide to the Serene Sound Transducer. With the exception of the headset,
all Magnet Room Serene Sound components must be installed to the side of the magnet shroud,
also the transducer power cable should never be in parallel with magnetic coil. Additionally, these
components should be placed in a light traffic area.
Page 7
Serene Sound Controller
Headset
9pin D
-
Sub
Filter
Connection
Transducer
power
Supply
Control Room Speakers
Transducer power cable
Tablet/Remote Control
Transducer
MR Laser link Cable

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
How To Connect The Transducer Power Cables
In order to connect the Transducer power cable to the power supply outside the magnet room, you
need to use a DB-9 filter provided with the system. Make sure to connect the two meter long cable
to both the power supply and penetration panel and connect the 1 meter long cable to both the
penetration panel and the Transducer.
GE Filter Panel Plate: Plug the Female 9 Pin “D” end of the cable that has been run through with
the fiber optics from the Control room to the Equipment room. Connect the cable’s Female end to
the J-51 plug found in the GE Filter Panel
Siemens Filter Panel Plate: Remove any of the blank plates on
the filter panel with a metric Allen
(hex) wrench and install the RTC supplied filter panel plate for the Siemens Magnet.
or
Page 8
10 meter long power
cable connected inside
the magnet room to the
Pen Panel.
2 meter long
power cable
connected outside
the magnet room
into the DB 9 filter
and power supply.
DB
9 Filter installed on the penetration panel used to connect the
2 meter power cable to the 10 meter power cable
Facility Penetration Panel

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
How to connect the Transducer Power Cables [Continued]
Phillips Filter Panel Plate: It is advised to drop the female end of the 2ft 9pin D cable through the filter panel into the
MRI room side and to attach it to the male end of the filter before putting back the removed plate.
Page 9
RTC recommends the Transducer Power Supply be installed outside the magnet room in the equipment room.
Remote
Outlet Switch
Hospital Grade
Power Cord
Transducer Power Cable
1 Feet [3 meters]
connected to the Pen
Panel 9pin Filter.
Transducer Switching [Black] Power Supply

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
The MR Laser Link cable is fed through the waveguide in the Penetration Panel and then routed to
the Transducer unit inside the magnet room.
Caring for the Fiber Optics Cable:
Note: Take care to ensure the cable is not bent in radius of less than 15cm.
Page 10
Running the MR Laser Link Cable
r ≥ 90° or Right angle
r
r ≤ 90° or Right angle
r
MR Laser Link Cable
MR Laser Link Cable

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Components Interconnection
A. Magnet Room
Place the Serene Sound Transducer in a secure area away from traffic as illustrated below. Install
the audio headset firmly into the Transducer using the bracket provided. Be sure to connect the
headset and fiber before powering on the unit. Also make sure that the audio headset connector is
connected all the way into the Transducer audio socket for proper installation.
Page 11
Press down firmly on the audio headset
connector
into the Transducer and use a NON-FERROUS Flat
head screw driver to secure in place.
SERENE SOUND
TRANSDUCER
PREFERRED
LOCATION

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Components Interconnection
(continued)
Magnet Room (cont’d)
Attach the MR Laser Link connectors to the Transducer as illustrated below. Match the cable colors
to those on the unit’s labels.
Note: Remember not to bend the individual cables smaller than a 15 cm radius
.
Page 12
r≥ 90° or Right angle
r
r< 90° or Right angle
r
Individual Cable Individual Cable
Transducer
Power (DC)
Cable
Fiber Optic
Cable

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Components Interconnection
(continued)
B. Control
Room
Connect all appropriate cables to the Controller as illustrated below. RTC recommends connecting Power supply last
after all other connections are made.
Connect the 5-pin Round Cable from the Controller Power Supply to the Controller connector labeled
POWER.
Connect the Technologist Remote Control to the Controller connector labeled REMOTE.
Connect the Control Room Speakers: mini stereo audio cable to the “CONTROL ROOM AUDIO” [OUT]
and the Power to the USB Connection labeled “SPEAKERS”.
Connect the MR Laser Link cable to the Controller connector labeled Optic Link. Be sure the fiber optic
cable matches the color coded markings on the connection housing.
-Optional- Auto voice connection can be made in order to incorporate the MRI magnet Auto voice
command features.
Connect the mini stereo jack cable from record-out to computer audio input. This feature can be
used to record the patients’ headset microphone response into a computer microphone input. This
is an optional feature.
Apply power to the Controller first and then turn on the Transducer Power Supply. Wait until the
system is fully synchronized (logo lights rotating on the Transducer inside the Magnet Room).
Page 13
Audio input cable (mini stereo)
from the audio source
Fiber optics cable connects to
the Transducer in the Magnet
Room
Tech
Remote
Control
connection
Controller DC
power Supply
GE Magnet Auto
Voice Interface
Connection
Siemens & GE Auto Voice
Audio Cable
connection
Control Room Speaker
USB Power Connection
Control Room Speaker Mini Stereo
Cable Connection [OUT]

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Auto Voice Interface Setup
GE Auto Voice Setup
The GE Auto Voice has an interface box in order to interface the Serene Sound Controller to the MRI signal. Remove
the cable from GE OSG J7 and connect to the Interface box “J7-In”. And connect a cable from “J7-Out” back to the GE
OSG J7 location. The CV Controller receives the signal from interface box via interlink cable which connects to the
controller back panel.
Interface Box
Top View
Interface Box Outputs
Siemens Magnet Auto Voice Setup
Remove the (LINE OUT) mini stereo audio cable from the back of the Siemens computer and plug it into the back of the AV console
marked input using a mini stereo extension cable. From the back of the controller, plug a mini stereo cable (male-male) to output
signal to plug back onto the Siemens computer marked audio out.
If Auto voice command volume are low, have the Siemens engineer, adjust the sound settings on the computer.
Page 14
Remove Cable from GE OSG J7 and Connect to
Interface Box Connect from Interface Box to GE OSG J7
Connect Interface Box to back of Controller via Interlink cable
Connect Interface Box to back of Controller via
Interlink cable
Output connection from Interface Box to GE OSG J7

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Auto Voice Setup (Continued)
Canon Auto Voice Setup
Remote the LINE OUT mini stereo audio cable from the back of the computer and plug it into the mini stereo extension
cable in order to plug it into the back of the Serene Sound Controller (green label marked “IN”)
From the Back of the Controller, plug a mini stereo cable (male-male) to the Green label “OUT” signal to plug back into
the computer marked audio out.
Be sure to test Auto Voice from Canon MRI System to make sure audio command signal is redirected to the RTC
headset.
Be sure to set the Auto Voice volume level on the computer to 1 . On most computers, use “ctrl” “esc” to pull up
window menu then go to control panel sound-speaker volume and raise it to 1 .
Philips Auto Voice Setup
Phillips DB9 male cable to X-16 Disconnected then plugged into the RTC Auto voice interface box. Then Connect
Interface box where the X-16 was removed from the Interface box to X16 using a short 9pin Female-Male Cable.
Page 15
Connect X16 to Phillips Auto
Voice Interface Box “INPUT”
Using a DB9 M-F Cable connect
Phillips Auto Voice Interface Box
“OUTPUT” to X16
Auto Voice Interface Box Mini stereo “OUTPUT” connects
to Auto voice “IN” on Controller back panel

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Control Room Setup
When operating normally, the logo lights on the Transducer are rotating and the sides of the Controller are blue.
However, after two to three hours of non-operation, the system goes into a SLEEP mode and the system appears to be
turned off. To wake up the system, just press down once on the TALK button of the Technologist Remote Control and
wait for the system to synchronize completely. Remember to always communicate with the patient to adjust the
headset audio volume. Always start at a low setting progressively to a higher setting until the patient can hear you
clearly. Below is an illustration how to connect the control room components:
(Typical Control Room Setup)
Page 16

Resonance Technology, Inc. Serene Sound Installation/Operation Manual Part Number: RTC-SS Rev 11
Operation
TECHNOLOGIST REMOTE CONTROL OPERATION:
!
Wheel #1: Patient Audio Volume (scroll up and down to adjust)
Scrolling the wheel up increases the patient’s headset volume.
Scrolling the wheel down decreases the patient’s headset volume.
Technologist Microphone – Press and hold the talk button. Then scroll the adjustment wheel
This adjustment is to change the volume that the patient can hear during communication.
Wheel #2: Patient Microphone Volume (scroll up and down to adjust)
Scrolling the wheel up increases the volume of the patient’s microphone volume heard in the control room.
Scrolling the wheel down decreases the volume of the patient’s microphone.
Wheel #3: Main Volume (Scroll up and down to adjust)
Scrolling the wheel up increases the control room volume (if speakers connected to the Controller Control room output connection).
Scrolling the wheel down decreases the control room volume.
Wheel #4: Main Menu (Scroll and press to select sub menu items)
Pressing down on this wheel activates the system menu. Scrolling the wheel up or down rotates through the system functions. To set the patient microphone in AUTO or
MANUAL mode use the menu options to select the desired mode. When in the MANUAL mode you must press the talk button to activate microphone. To adjust your
technologist microphone volume built into the technologist remote unit, simply press and hold the TALK button and at the same time scroll the wheel #1 to set the PTT
volume to about 6 to 7 . If your voice sounds too loud in the patient’s audio headset lower the PTT Make sure all cables have been connected securely into the
Transducer and that the components do not impede walkways. Ensure that the Transducer is not placed directly in front of the magnet bore or directly under the bed.
Before placing the headset on the patient, make sure that the technologist microphone volume has been set to a comfortable level for the patient. Place the audio headset
over the patient’s ears and adjust the headband. If necessary, position the headset microphone about one cm. away from the patient’s mouth.
Please Note: If the tech remote unit is not responding you can reset [Factory Reset] it to its original factory default settings by following these 5
simple steps:
1
st
: Set Patient Audio (Adjustment Wheel #1) to 1
2
nd
: Set Patient Microphone (Adjustment Wheel #2) to 2
3
rd
: Set the Main Volume (Adjustment Wheel #3) to 3
4
th
: Press and hold talk button for 15 seconds
5
th
: Release the talk button. All the settings have been reset to the original factory defaults. You now have to re-adjust all the wheels settings.
Page 17
Notice: Always adjust the audio input source to a level that will not cause distortion.
Otherwise a crackling noise will be heard on the patient audio headset!
Other manuals for Serene Sound
1
Table of contents