- 3 -
1. FIRST STEPS BEFORE START
Remove plastic protection from the screen.
Attach three-axis Hall probe to the instrument.
Turn on teslameter. Use push-button located at the back of teslameter.
Instrument may be supplied with internal
battery or external AC adapter with
stabilized 9 V output - see POWER
socket. Adapter plug-in automatically
disconnects the battery. Do not use
rechargeable batteries.
Note: connecting of external power supply
with voltage higher than 9,5 V
automatically switch off teslameter.
When teslameter is initially turned on -
logo „RX-25” will be illuminated for few
seconds. Immediately after this the
instrument automatically starts to do
zeroing relative to the magnetic field
existing around of probe.
For more precise and absolute
measurements, after at least one minute of
instrument being on, it is recommended to
repeat probe zeroing by placing probe into
attached Zero Field Chamber and then
choosing zeroing function from Menu.
2. FRONT PANEL OPERATION
Access to Menu ( see page 5) and a choice of desired modes and functions is
showed on the Menu diagram. All operations of the instrument are controlled by
four buttons only: Esc (menu/measure change), (up cursor), (down cursor),
-ENTER (circular horizontal cursor ). From any position at Menu it is
instantaneously possible to enter to measure mode and inversely simply by
using the Esc button!.
Black label position on the display means active function at the time.
Note that menu on the display of horizontal functional branches is subdivided
vertically on the screen. Therefore, horizontal cursor initially moves vertically
and then horizontally. This corresponds exactly to horizontal positions of Menu.