Restek ProFLOW 6000 Series User manual

www.restek.com
Restek ProFLOW 6000
Electronic Flowmeter
Version 5.6
(Cat.# 22656)
Instruction
Manual

2 www.restek.com
Table of Contents
1.0 Introduction ............................................................................................................................................. 3
2.0 Specifications........................................................................................................................................... 4
3.0 Installing the batteries ......................................................................................................................... 4
4.0 Battery power consumption.............................................................................................................. 5
4.1 Battery lifetime.............................................................................................................................. 5
4.2 Battery charge indicator ............................................................................................................ 5
5.0 Operating instructions ......................................................................................................................... 5
6.0 Interpreting results................................................................................................................................ 6
6.1 Flow range display....................................................................................................................... 6
7.0 Data collection on the PC ................................................................................................................... 7
8.0 ProFLOW 6000 menu structure......................................................................................................... 9
8.1 Unit power up/power down..................................................................................................... 9
8.2 USB activation................................................................................................................................ 9
8.3 Adjust LCD image backlight...................................................................................................10
8.4 Show battery charge indicator..............................................................................................10
8.5 Firmware version information ...............................................................................................11
8.6 Adjust auto shutoff duration..................................................................................................11
8.7 Display calibration date ...........................................................................................................12
8.8 Display unit serial number......................................................................................................12
9.0 Troubleshooting ...................................................................................................................................12
10.0 Product back label legend................................................................................................................13
11.0 Volumetric vs. mass flow measurements....................................................................................14
12.0 Bubble flowmeters ..............................................................................................................................14
13.0 Product case specifications..............................................................................................................14
14.0 Calibration and service ......................................................................................................................14
Declaration of Conformity ................................................................................................................31

www.restek.com 3
Restek ProFLOW 6000
Electronic Flowmeter
1.0 Introduction
Restek’s ProFLOW 6000 Flowmeter is specifically designed for use with gas chromatography (GC) systems. The
probe is applied directly to the gas flow stream, and the measured flow rate is presented on the LCD screen.
Units of flow are measured in mL/min.
This unit provides continuous real-time measurements of gas streams ranging from 0.50 mL/min to 500 mL/min.
Because the technology uses volumetric flow measurement, the unit is compatible with all laboratory gases.
CAUTION: Do NOT exceed maximum operating flow rates. Recalibration may be required if the unit
has been subjected to extreme flow rates.
Always use appropriate laboratory safety practices when operating this device. Wear laboratory
safety goggles when operating this unit.
NOTE: There are no serviceable parts in this unit. Opening the device—other than to change the batteries—or
tampering with the internal parts will void the factory warranty.
NOTE: To ensure accurate measurements and effective clearance of the flow gas from the unit, DO NOT
obstruct the vent holes on the back of the unit.
NOTE: Modifying the length of tubing can cause a shift in calibration.
Figure 1: Product diagrams
BACK
FRONT

4 www.restek.com
2.0 Specifications
Table I:
3.0 Installing the batteries
This unit uses two AA alkaline batteries.
To install batteries, extend the wire stand. Open the cover. Insert the batteries with the
polarity (⊕and ⊖) correctly aligned. Close the cover. (Figures 2 and 3.)
Precautions for battery replacement:
•Load the new batteries with their polarity (⊕ and ⊖) aligned correctly.
• Do not use rechargeable batteries.
Figure 2: Extend the wire stand before
opening battery door.
Figure 3
Insert the batteries
as marked.
+
–+
–
Type of measurement Volumetric flow
Accuracy of measurements ± 2% of flow reading or ± 0.2 mL/min, whichever is greater
Power requirements 2 AA alkaline batteries 1.5 V DC each/3 V DC 200 ma
Operating flow range 0.50 to 500 mL/min
Operating temperature range 32 °-120 °F (0 °- 48 °C)
Available communication USB type B data port
Warranty One year (excludes recalibration)
Calibration NIST traceable
Yearly recalibration is recommended
Certifications CE, Ex (see section 10.0)
Compliance WEEE, RoHS 2, China RoHS 2 (see section 10.0)

www.restek.com 5
NOTE: Store your ProFLOW 6000 in its protective storage case following use. Do not store this manual or any
other items on top of the ProFLOW 6000 or the unit may turn on when the storage case lid is closed.
4.0 Battery power consumption
4.1 Battery lifetime
The battery lifetime is dependent on the number of options the user has enabled.
The unit is shipped with the most energy-demanding options disabled (Table II).
The power saving functions can be changed.
èSee Section 8.2: USB Activation
èSee Section 8.3: Adjust LCD Image Backlight
èSee Section 8.6: Adjust Auto Shutoff Duration
Table II: Default settings for the ProFLOW 6000
Auto shutoff duration 6 minutes
LCD backlight 0 (off)
USB port disabled
4.2 Battery charge indicator
The unit includes a battery charge indicator. Replace batteries as needed.
èSee Section 8.4: Show Battery Charge Indicator
5.0 Operating instructions
CAUTION: Do not exceed maximum operating flow rates.
Recalibration may be required if the unit has been subjected to extreme flow rates.
Connect the white probe end tip to the output of the gas flow line to be measured. Be sure the probe tip
connection is completely sealed around the flow source outlet and is free of leaks (Figure 4). (White tubing ID
0.125"; OD 0.250")
Press and hold the (Power) button until the unit responds with a regular clicking sound. The Pro-
FLOW will immediately begin to provide flow measurements (Figure 5). Wait for the measured values to stabi-
lize. It takes a few seconds for the unit to reach a steady state with the gas flow line.
To power down the unit, press and hold the (Power) button until the unit stops clicking.
The unit is equipped with a timed auto shutoff option. (Default: 6 minutes.)
èSee Section 8.6: Adjust Auto Shutoff Duration
Figure 4: Probe connected to a GC gas outlet. Figure 5: LED displays the measured flow value.

6 www.restek.com
6.0 Interpreting results
The unit has an operating range of 0.50 mL/min to 500 mL/min (Figure
7). If the flow is less than 0.50 mL/min., the display will read“under range.”
If the flow exceeds 515 mL/min, the display will read“over range.” Exces-
sively high flow rates may damage this unit.
NOTE: units of mL/min are equivalent to ccm.
6.1 Flow range display
The unit automatically adjusts the resolution of the display depending
on the flow range being measured. Table III shows the resolution of the
flow ranges.
Figure 6: Example flow value.
Figure 7: Description of flow ranges.
Table III: Display resolution vs. flow range.
Flow Range
Display reads
“under range”
Display reads
flow values
Display reads
“over range”
0.50 mL/min 515 mL/min
Flow range Display resolution (mL/min)
0.50–9.99 0.01
10.0–99.9 0.1
100–500 1

www.restek.com 7
7.0 Data collection on the PC
WARNING: ONLY connect USB cable to USB port while unit is OFF.
The ProFLOW 6000 provides you with a data stream of real-time flow values via the USB port (Figure 1, p. 3). In
order to use this feature, you must first install the appropriate FTDI Virtual Com Port (VCP) Driver available at:
http://www.ftdichip.com/Drivers/VCP.htm
The VCP driver will cause the ProFLOW to appear as a standard RS-232 port. This will work on any operating
system for which there is an FTDI VCP driver. After installing the driver, connecting the device, and determining
which port it creates, you can access the data stream through any programmatic means, or by using any serial
terminal software.
For Windows systems:
To determine which port the ProFLOW is using, go to the Control Panel and open System. Go to the Hardware
tab and click the Device Manager button. Expand the Ports (COM & LPT) entry. Make sure the VCP driver is
installed and then connect a powered ProFLOW 6000 to the USB port. You will see the new COM port appear.
Open your serial terminal.
èSee Section 8.2: USB Activation
Figure 8: Screen capture of data collection.

8 www.restek.com
If you do not already have serial terminal software, free, open-source options are available online (i.e.,
RealTerm, etc.). You can download RealTerm* software from:
http://realterm.sourceforge.net/
After installation, click the Port tab and set the following:
Baud: 115200
Port: The appropriate VCP for your ProFLOW
Parity: None
Data Bits: 8 bits
Stop Bits: 1 bit
HardwareFlow Control: None
Go to the USB menu entry in the ProFLOW and turn the transmission on. Finally, click Open on the Port tab in
RealTerm and you will see the serial data stream begin in the terminal window (Figure 8).
If you would like to log the flow data, this can be done by clicking on the Capture tab. Set File to the name
and location of the log file that you would like to save and click either the Start: Overwrite or Start: Append
buttons appropriately (Figure 9).
*This software is not supplied or supported by Restek. User assumes all responsibility for the downloading and use of the program.
Figure 9: Screen capture of data collection.

www.restek.com 9
8.0 ProFLOW 6000 menu structure
8.1 Unit power up/power down
Press the (Power) button:
The LCD screen will display Followed by Followed by measured
the device intro screen: calibrated date: flow data:
01/10/2012
To power off, press and hold the (Power) button.
Other messages encountered at power up
If the unit is hooked up to a flow stream with a flow rate less than 0.50 mL/min,
the unit will report an“under range”status for the flow. This message will appear
until the flow rate exceeds 0.50 mL/min.
8.2 USB activation
To enable the USB, press the (Menu) button.
Use the (arrow) keys to select the USB menu:
Press the (menu) button again to enter the value select screen.
Use the (arrow) keys to toggle between USB “on”and“off.”
To return to the main menu screen, press the (menu) button.
To exit and return to measuring flow, press the (menu) button again.
èSee Section 7.0: Data Collection on the PC

10 www.restek.com
8.3 Adjust LCD image backlight
Press the (Menu) button.
Use the (arrow) keys to select the Backlight menu:
Press the (Menu) button again to enter the value select screen.
Use the (arrow) keys to select the backlight value.
Backlight Values: 0 (off) and 5 (maximum).
To return to the main menu screen, press the (Menu) button.
To exit and return to measuring flow, press the (Menu) button again.
8.4 Show battery charge indicator
Press the (Menu) button.
Use the (arrow) keys to select the Batteries menu:
Press the (Menu) button again.
The battery life is displayed.
To return to the main menu screen, press the (Menu) button.
To exit and return to measuring flow, press the (Menu) button again.

www.restek.com 11
8.5 Firmware version information
Press the (Menu) button.
Use the (arrow) keys to select the Firmware menu:
Press the (Menu) button again.
The most recent version of firmware is displayed.
To return to the main menu screen, press the (Menu) button.
To exit and return to measuring flow, press the (Menu) button again.
8.6 Adjust auto shutoff duration
To conserve battery life, the unit automatically turns off after 6 minutes. To customize the auto shutoff setting,
press the (Menu) button.
Use the (arrow) keys to select the Power Time menu:
Press the (Menu) button again to enter the value select screen.
Use the (arrow) keys to select the auto shutoff setting.
Values: 1–59 minutes or“constant on”(max.)
To return to the main menu screen, press the (Menu) button.
To exit and return to measuring flow, press the (Menu) button again.

12 www.restek.com
8.7 Calibration Date
To display the calibration date, press the (Menu) button.
Use the (arrow) keys to select the Cal Date menu:
Press the (Menu) button again to display the calibration date.
To return to the main menu screen, press the (Menu) button.
To exit and return to measuring flow, press the (Menu) button again.
cal date
calibrated:
12/12/2011
8.8 Serial Number
To display the flowmeter serial number, press the (Menu) button.
Use the (arrow) keys to select the Serial Num menu:
Press the (menu) button again to display the serial number.
To return to the main menu screen, press the (Menu) button.
To exit and return to measuring flow, press the (Menu) button again.
Serial num
s/n:
RE100001
9.0 Troubleshooting
*Contact [email protected] for return instructions for servicing a damaged unit. Additional charges may apply if the warranty has expired or
the unit is damaged due to misuse.
Call Restek Customer Service at 1-800-356-1688 or 1-814-353-1300, ext. 3 (or your Restek representative) if you have any questions about this
product or any other Restek product.
Problem Possible Cause(s) Suggested Solution(s)
Multiple readings are not
giving reproducible results
• Unit is out of calibration
• Value is being compared to a bubble flowmeter
• Return the unit to Restek for recalibration*
• See Section 12.0 for a discussion of the
weaknesses of bubble flowmeters
Unit does not power up • Dead batteries • Replace with 2 new AA alkaline batteries
Flow value display is
erratic/jumpy
• The ProFLOW 6000 is very sensitive to small
changes in flow
• Allow more time for flow to stabilize

www.restek.com 13
10.0 Product back label legend
# Description
1 Serial number
2 Company name
3 This unit conforms to EU/EMC Directive 2004/108/EC; standards to which conformity is declared include 61326:1997 w/A3 Class A.
4 This Class A digital apparatus complies with Canadian ICES-003.
5 This complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
6 Country of origin
7 Product catalog number
8 Company address
9 This unit is WEEE compliant.
10 This unit is RoHS 2 and China RoHS 2 compliant.
10 China RoHS Logo 2
Number Font: Impact
China RoHS Logo 1
11
Ex nL EN60079-0: 2006; Electrical apparatus for explosive gas atmospheres—Part 0: General Requirements.
EN60079-15: 2005; Electrical apparatus for explosive gas atmospheres—Part 15: Construction, test, and marking of type of
protection “nL” energy limited apparatus.
IIC Group II applies to areas above ground environments.
Gas Group IIC relates to hydrogen and related gas types.
3G Category 3 relating to gas analysis; normal safety measure. Sufficient safety during normal operation. Normal operation described
as measuring flows of flammable or explosive gases in a nonflammable environment.
T6 During testing, neither internal nor external elements exceed 85 °C.
X Additional information:
Operating range: 32 °F ≤Tamb ≤120 °F
0 °C ≤Tamb ≤48 °C
Not intended for outdoor use or wet locations.
2010 Year of product design release.

14 www.restek.com
11.0 Volumetric vs. mass flow measurements
The Restek ProFLOW 6000 is a volumetric flow measurement device. Volumetric flow is the measurement of
the volume of gas through a conveyance per quantity of time. Standard units of measure for this parameter
are given in mL/min. The advantage of measuring volumetric flow is its independence to the composition of
the flow gas. It is not necessary to correct the flow values based on the gas composition, as is required for mass
flow devices.
Mass flow measures the weight of the gas flowing through the instrument per quantity of time. Mass flow
units of measure are commonly g/sec.
12.0 Bubble flowmeter measurements
If you employ bubble flowmeters in your laboratory, you may find that they give slightly different flow rate
values than the Restek ProFLOW 6000. This error is due to technology limitations inherent in the bubble flow-
meter device, error from variances in air humidity within the bubble chamber, and its direct contribution to the
measured flow rate. In the event a bubble flowmeter is used to measure flow gas where the gas is at elevated
temperatures, the error due to humidity contributions can be extreme. For the most accurate measurement of
laboratory gas flow rates, we recommend using the Restek ProFLOW 6000 over bubble flowmeters.
INJECTION MOLDED
NON-FILLED
CYCOLOY - PC/ABS C6200
MANUF.: SABIC
INJECTION OVERMOLD
VERSOLLAN
INJECTION MOLDED
NON-FILLED
CYCOLOY - PC/ABS C6200
MANUF.: SABIC
INJECTION MOLDED
WHITE SILICONE
BUTTONS
13.0 Product case specifications
14.0 Calibration and service
The Restek ProFLOW 6000 comes factory calibrated and carries a one-year warranty (excluding recalibration)
from time of purchase. All units are calibrated to NIST traceable standards.
Recommended schedule for recalibration is once every year from time of purchase. Customers will need to
return the unit to Restek for recalibration. At that time, preventative maintenance services can also be per-
formed. A fee will be charged for recalibration and servicing of the unit. Prolonged failure to recalibrate the
instrument may result in increased error.
Please have the serial number available when calling Restek with any concerns you may have.
Additional charges may apply if the warranty is expired or the damage is due to misuse. This
manual is also available in electronic format at www.restek.com.

www.restek.com 15
Inhaltverzeichnis
1.0 Einleitung................................................................................................................................................16
2.0 Spezifikationen .....................................................................................................................................17
3.0 Die Batterien einbauen......................................................................................................................17
4.0 Stromverbrauch....................................................................................................................................18
4.1 Batterie Lebenserwartung ......................................................................................................18
4.2 Batterieladungsindikator.........................................................................................................18
5.0 Benutzungsanleitung .........................................................................................................................18
6.0 Interpretation der Ergebnisse..........................................................................................................19
6.1 Flussbereichsanzeige ................................................................................................................19
7.0 Datensammlung mit dem PC..........................................................................................................20
8.0 ProFLOW 6000 Menüstruktur.........................................................................................................22
8.1 Gerät Ein/Ausschalten...............................................................................................................22
8.2 USB Aktivierung ..........................................................................................................................22
8.3 Die LCD Hintergrundbeleuchtung einstellen ..................................................................23
8.4 Den Batterieladungsindikator anzeigen............................................................................23
8.5 Firmware Version Information ...............................................................................................24
8.6 Die Selbstausschaltdauer einstellen....................................................................................24
8.7 Kalibrier Datum anzeigen........................................................................................................25
8.8 Seriennummer anzeigen .........................................................................................................25
9.0 Troubleshooting ...................................................................................................................................25
10.0 Aufkleber auf der Rückseite .............................................................................................................26
11.0 Volumetrische gegen Masse Flussmessungen .........................................................................27
12.0 Messungen mit einem Bläschen Flowmeter..............................................................................27
13.0 Produkt Gehäuse Spezifikationen..................................................................................................27
14.0 Kalibrierung und Service...................................................................................................................27
Konformitätserklärung.......................................................................................................................31
Restek’s ProFLOW 6000
Elektronischer Flowmeter
(Art.-Nr. 22656)

16 www.restek.com
1.0 Einleitung
Restek’s ProFLOW 6000 Flow Meter wurde speziell für die Benutzung in Gaschromatographie (GC) Systemen
entwickelt. Die Prüfspitze wird direct an den Gasfluss gehalten und die gemessene Flussrate wird auf einem
LCD Bildschirm angezeigt. Der Fluss wird in mL/min gemessen.
Dieses Gerät bietet kontinuierliche real-time Messungen von Gasströmen im Bereich von 0.50 mL/min bis 500
mL/min. Da die verwendete Technologie auf volumetrischen Flussmessungen beruht ist das Gerät mit allen
Laborgasen kompatible.
VORSICHT: AUF KEINEN FALL DIE MAXIMAL ZULÄSSIGE FLUSSRATE ÜBERSCHREITEN.
REKALIBRIERUNG KANN ERFORDERLICH SEIN FALLS DAS GERÄT EXTREMEN FLUssRATEN
AUSGESETZT WURDE.
BEI BENUTZUNG DIESES GERÄTS IMMER AUSREICHENDE SICHERHEITSVORKEHRUNGEN TREFFEN.
TRAGEN SIE LABOR SCHUTZBRILLEN WENN SIE DIESES GERÄT BENUTZEN.
ANMERKUNG: Es gibt keine wartungsbedürftigen Teile in diesem Gerät. Bei Öffnen des Geräts, außer zum
Auswechseln der Batterien, oder Entfernen der internen Teile verfällt die Herstellergarantie.
ANMERKUNG: Um akkurate Messungen und effektive Entfernung des gemessenen Gases aus dem Gerät zu
gewährleisten, nicht die Ventilationsöffnungen auf der Rückseite des Geräts versperren.
ANMERKUNG: Eine Veränderung der Schlauchlänge kann die Kalibrierung verschieben.
Abbildung 1: Produkt Diagramme
Vorne
Prüfspitze
LCD Anzeige
Einschalttaste
Pfeiltaste zur
Werterhöhung
Menütaste
Pfeiltaste zur
Werterniedrigung
USB Port
Seriennummer/
Kalibrierungsnummer
Belüftungsöffnungen
Produktetikett
Batteriedeckel
Drahtbügelständer
Hinten

www.restek.com 17
2.0 Spezifikationen
Tabelle I:
Art der Messungen Volumetrischer Fluss
Genauigkeit der Messungen ± 2% Anzeige oder ± 0.2 mL/min, je nach Messbereich
Stromanspruch 2 AA Alkali Batterien 1,5 V DC je/3 V DC 200 ma
Arbeitsbereich des Gasflusses 0.50 bis 500 mL/min
Temperatur Arbeitsbereich 0 °- 48 °C
Vorhandene Kommunikation USB die art B data port
Garantie Ein Jahr (ohne Neukalibrierung)
Kalibrierung NIST nachweisbar.
Jährliche Rekalibrierung ist empfohlen
Zertifikate CE, Ex (unter Sektion 10.0)
Befolgungen WEEE, RoHS 2, China RoHS 2
3.0 Die Batterien einbauen
Dieses Gerät benutzt 2 AA Alkali Batterien.
Um die Batterie einzubauen muss der Drahtstehbügel auf der Rückseite aufgeklappt werden
Öffnen Sie die Klappe. Bauen Sie die Batterien unter Beachtung der Polarität (⊕und ⊖) ein.
Machen Sie die Klappe zu. (Abbildungen 2 und 3.)
Vorsicht beim Auswechseln der Batterien:
•Die neuen Batterien unter Beachtung der Polarität (⊕ und ⊖) einbauen.
• Keine aufladbaren Akkus benutzen.
Abbildung 2: Klappen Sie den Drahtstehbügel voll
auf bevor Sie die Batterienklappe öffnen.
Abbildung 3:
Bauen Sie die
Batterien wie
angezeigt ein.
+
–+
–

18 www.restek.com
ANMERKUNG: Bewahren Sie Ihren ProFLOW 6000 nach Gebrauch in seiner Originalbox auf. Bitte legen Sie
diese Anleitung oder andere Gegenstände nicht auf den ProFLOW 6000 da das Gerät sonst einschalten kann
wenn der Deckel des Aufbewahrungskoffers geschlossen wird.
4.0 Stromverbrauch
4.1 Batterie Lebenserwartung
Die Lebenserwartung der Batterie hängt von der Anzahl der von dem Benutzer ermöglichten Optionen ab.
Das Gerät wird so verschickt dass die Optionen die am meisten Energie verbrauchen auf„untauglich“ eingestellt
sind (Tabelle II).
Die Stromsparfunktionen können umgestellt werden.
èSektion 8.2: den USB aktivieren
èSektion 8.3: die LCD Hintergrundbeleuchtung einstellen
èSektion 8.6: die Selbstsauschaltungsdauer einstellen
Tabelle II: Standardeinstellungen für den ProFLOW 6000
Selbstsauschaltungsdauer 6 Minuten
LCD Hintergrundbeleuchtung 0 (aus)
USB port Ausgeschaltet
4.2 Batterieladungsindikator
Dieses Gerät ist mit einem Batterieladungsindikator ausgerüstet.Wechseln Sie die Batterie wenn angedeutet aus
èSektion 8.4: Batterieladungsindikator
5.0 Bedienungsanleitung
VORSICHT: AUF KEINEN FALL DIE MAXIMAL ZULÄSSIGE FLUSSRATE ÜBERSCHREITEN
REKALIBRIERUNG KANN ERFORDERLICH SEIN FALLS DAS GERÄT EXTREMEN FLUSSRATEN
AUSGESETZT WURDE.
Bringen Sie die weiße Prüfspitze an die zu messende Gasleitung an. Stellen Sie sicher dass die Verbindung der
Prüfspitze mit dem Outlet komplett dicht ist und es keine Lecks gibt. (Abbildung 4). (Weißer Prüfschlauch ⁄"
ID; ¼" AD)
Drücken und halten Sie die Einschalttaste bis das Gerät mit einem regelmäßigen Klickton anspricht.
Der ProFLOW fängt sofort an Flussmessungen anzugeben (Abbildung 5). Warten Sie bis sich die angezeigten
Messwerte stabilisiert haben. Es dauert ein paar Sekunden bis sich das Gerät mit der Gasleitung ausgeglichen hat.
Zum Ausschalten drücken sie die Einschalttaste bis das Klicken aufhört.
Das Gerät ist mit einer Selbstausschaltautomatikoption ausgerüstet. (Standardeinstellung: 6 Minuten.)
èSektion 8.6: Einstellen der Selbstausschaltdauer
Abbildung 4: Prüfspitze an einen GC Gas Outlet
angebracht
Abbildung 5: LED zeigt den gemessenen Flusswert an.

www.restek.com 19
Flussbereich
Anzeige
zeigt “under
range” an
Anzeige zeigt
(+) flow values an
Anzeige zeigt
“over range” an
0.50 mL/min. 515 mL/min.
6.0 Interpretation der Ergebnisse
Das Gerät hat einen Arbeitsbereich von 0.50 mL/min bis 500 mL/
min (Abbildung 7). Wenn der Fluss weniger als 0.50 mL/min. zeigt die
Anzeige“under range”an.
Wenn der Fluss 515 mL/min. übertrifft zeigt die Anzeige“over range” an.
Übermäßig hohe Flussraten können das Gerät beschädigen.
ANMERKUNG: Die Einheit mL/min ist gleichwertig mit ccm.
6.1 Flussbereichsanzeige
Das Gerät justiert automatisch die Auflösung der Anzeige, je nach
dem gemessenen Flussbereich. Tabelle III zeigt die Auflösung der
Flussbereiche.
Abbildung 6: Beispiel von
Flusswerten
Abbildung 7: Beschreibung der Flussbereiche.
Tabelle III: Anzeigenauflösung gegen Flussbereich.
Flussbereich Anzeigenauflösung (mL/min)
0.50–9.99 0.01
10.0–99.9 0.1
100–500 1

20 www.restek.com
7.0 Datensammlung mit dem PC
VORSICHT: Auf keinen Fall die maximale Flussrate überschreiten. Das Gerät könnte eine erneute
Kalibrierung benötigen wenn es extremen Flussraten ausgesetzt wurde.
Der ProFLOW 6000 bietet Ihnen einen Datenstrom von Real Time Flusswerten über den USB Port (Abbildung
1, Seite 3). Um dieses Feature zu benutzen müssen Sie zuerst den entsprechenden FTDI Virtual Com Port (VCP)
Driver installieren. Dieser ist erhältlich bei:
http://www.ftdichip.com/Drivers/VCP.htm
Durch den VCP Driver wird der ProFLOW wie ein Standard RS-232 Port erscheinen. Das funktioniert mit jedem
Operating System für den es einen FTDI VCP Driver gibt. Nachdem der Driver installiert ist, verbinden Sie das
Gerät mit dem PC und stellen Sie fest welchen Port es geschaffen hat. Jetzt haben Sie Zugang zum Datenstrom
durch jede Programmierungsart, oder durch jede Serial Terminal Software.
Für Windows Systeme:
Um festzustellen welchen Port der ProFLOW benutzt, gehen Sie zum „Control Panel“ und öffnen Sie„System“.
Gehen Sie zum “Hardware Tab” und klicken Sie auf „Device Manager“. Expandieren Sie die Ports (COM & LPT)
Eingabe. Stellen Sie sicher dass der VCP Driver installiert ist, und dann verbinden Sie einen eingeschaltenen
ProFLOW 6000 mit dem USB Port. Ein neuer COM Port wird erscheinen. Öffnen Sie Ihren Serial Terminal.
èSektion 8.2: USB Aktivierung
Abbildung 8: Schirmbild der Datensammlung.
This manual suits for next models
1
Table of contents
Languages:
Other Restek Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Aktakom
Aktakom ATK-2040 user manual

PeakTech
PeakTech 3270 Operation manual

Autometers Systems
Autometers Systems A-400 Series Programming and installation Brochure

Chroma
Chroma 16502 user manual

CARLO GAVAZZI
CARLO GAVAZZI LDI 35 - CONFIGURATION SOFTWARE operating instructions

Dewetron
Dewetron DEWE3-A4L Technical reference manual