•Check the battery status on the display. At one remaining dot at the battery display, max
one injection can be completed. Charging time approx. 3 hours.
Operating time approx. 5 hours, when fully charged.
•Attach a needle on the cartridge barrel and insert an anaesthetic cartridge.
To avoid leakage at the membrane of the cartridge it is advised to screw on the needle
first and then insert the cartridge. The cartridge barrel fits standard 1.7/1.8 ml dental
cartridges and standard dental needles.
Make sure that the needle is completely mounted and that the needle fits the applied
cartridge barrel (standard metric thread (M6) or imperial inch thread (7/32”)).
•Mount the loaded cartridge barrel onto the handpiece. Before the barrel is screwed
onto the handpiece, the piston rod should be retracted to starting position. It will automati-
cally return to starting position, when CALAJECT™ is turned on. The piston rod can
also be returned to starting position by pressing ”R” on the display.
•Prime the needle, by applying a quick tap on the foot switch. The plunger will then move
forward for 2 sec. after which it will stop automatically. Now the piston is fully engaged
in the cartridge and CALAJECT™ is ready for injection. This priming mode is indicated by a
rotating light signal at the front panel.
Do not inject in prime-needle mode.
•Select program. When the foot switch is activated again, the chosen injection program
will be active.
• CALAJECT™ will stop automatically, when the cartridge is empty - a long beep will be heard.
Return the piston rod by pressing “R” at the display. Make sure to let the piston retract
completely, before you unscrew the barrel from the handpiece.
CALAJECT™ will stop automatically when it reaches the pre-programmed maximum pressure.
In such case, a long sound signal will be heard and the pressure bar-scale on the display will
turn off. Wait a moment or move the needle to a new position before you continue the injection.
Change the needle if you suspect a clogging of the cannula. Otherwise this may result in a high
back pressure and automatic stop. CALAJECT™ will also turn off automatically, if it has been
turned on for a continuous period of 6 hours (=battery protection). Can be restarted in the normal
way after such incident.
ASPIRATION
CALAJECT™ aspirates automatically whenever the foot switch is deactivated during injection
in Program I, II and III.
SAFE NEEDLE HANDLING
The cavity in the CALAJECT™ handpiece holder is designed for parking the needle sheath
during injection. It facilitates one-handed recapping after and between injections and allows
you to place the handpiece safely in the holder with reduced risk of injuring yourself on a used,
contaminated needle.