Sacher Lasertechnik TEC-500 Guide

Technical Documentation
Page 1
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
Alignment Procedure for Littman/Metcalf (TEC-500)
Laser Cavities
1 Security Precautions................................................................................................................ 2
1.1 Radiation Protection............................................................................................................ 2
1.2 ESD Protection.................................................................................................................... 2
1.3 Protection of Optical Elements............................................................................................ 2
1.4 Identification of Each Element within the Laser Head........................................................3
2 Collimation of the Laser Diode ............................................................................................... 5
2.1 Laser Diode Beam Propagation........................................................................................... 5
2.2 Determining the Status of the Laser Collimation................................................................ 6
2.3 Adjustment of the Laser Collimation .................................................................................. 7
3 Cavity Alignment....................................................................................................................... 8
3.1 Pre-Adjustment of the Reflection Prism.............................................................................. 8
3.2 Minimizing the Laser Threshold....................................................................................... 10
4 Mode-Hop Free Tuning.......................................................................................................... 12
4.1 Principles of Mode-Hop Free Tuning................................................................................ 12
4.2 Monitoring the Tuning Performance via Fabry Perot Interferometer ............................... 12
4.3 Monitoring the Tuning Performance via Power Monitoring............................................. 12
4.4 Monitoring the Tuning Performance via Wavemeter........................................................ 14
4.5 Optimizing the Tuning Performance................................................................................. 14
5 Summary.................................................................................................................................. 16
Thank you for ordering a laser system with Sacher Lasertechnik. The provided alignment
procedure will enable you to achieve the best performance with the laser system.

Technical Documentation
Page 2
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
1 Security Precautions
Before starting to work on the alignment procedure of the Littman/Metcalf laser system,
there will be two types of security precautions required, protection against laser radiation
and protection of the laser diode against electro-static discharge (ESD).
1.1 Radiation Protection
Make sure to provide adequate protection against laser radiation. E.g. protective eyewear
is absolutely required. Further precautions are required for higher power laser systems.
Consult your laser security officer for detailed instructions before removing the protective
cover of the laser head and before the starting the alignment procedure.
1.2 ESD Protection
Laser diodes are electro-static discharge sensitive (ESD) devices. Most premature laser
diode damages are caused by ESD damages. Make sure to take adequate precautions
against ESD. This can be done e.g. via a wristband which connects you to the laser
ground. Consult your security officer for detailed instructions before starting the alignment
procedure.
1.3 Protection of Optical Elements
The optical elements within the laser head are sensitive components. A simple finger print
will cause a irreversible damage. Do not touch an optical surface of any of the optical
components.

Technical Documentation
Page 3
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
1.4 Identification of Each Element within the Laser Head
Identify each element inside of the laser head and compare its position and function with
the description within the operation manual of the laser system.
Fig. 1: CAD drawing of the Littman/Metcalf laser head. The red arrow indicates
laser radiation (A). Laser beam (A) is the zero order diffraction of the diffraction
grating. Green arrows indicate fine adjustments screw. Black arrows indicate
optical elements.
(2) Cavity
Alignment Screw
(1) Wavelength
Alignment Screw
(A) output beam
(3) Collimation
Adjustment Screw
Reflection PrismDiffraction
Grating
Optical Isolator
(optional)
Fiber Coupler
(optional)

Technical Documentation
Page 4
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
Fig. 2: CAD drawing of the Littman/Metcalf laser head. The green arrow indicates
the fine adjustment screw for the wavelength adjustment.
Fig. 1 and Fig. 2 show a CAD drawing of our Littman/Metcalf type of laser head. The red
arrow indicates laser radiation. Laser beam (A) is the laser output beam. It is the zeroth
order diffraction of the diffraction grating. Green arrows indicate fine adjustment screws.
Fig. 3: Photo of the Littman/Metcalf laser head. The green arrows indicates the fine
adjustment screws for the collimation and the cavity alignment. Black arrows
indicate optical elements.
Fig. 3 shows a photo our Littman/Metcalf type of laser head. The green arrows indicate
fine adjustment screws. Contact us in case you feel uncertain on any of the described
elements.
(1) Wavelength
Alignment Screw
(2) Cavity
Alignment Screw
(3) Collimation
Adjustment Screw
Optical Isolator
(optional)
Fiber Coupler
(optional)

Technical Documentation
Page 5
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
2 Collimation of the Laser Diode
Proper beam collimation of the laser is required before the cavity alignment can be
performed. The collimation procedure consists of several steps. Before discussing details
of the collimation procedure, the physical basics of laser diode emission will be briefly
reviewed.
2.1 Laser Diode Beam Propagation
Laser diodes are epitactic grown multi-layer semiconductor crystals. The epitactic layers
define multiple electric p-n-junctions. Only a small part of the laser diode crystal generates
the laser radiation. This active area of the laser diode crystal has a typical size of 1µm x
3µm x 1500µm (height x width x length). The width is measured parallel to the p-n-junction.
The height is measured perpendicular to the p-n-junction.
Due to the large difference between height and width of most laser diodes, the emission
point for the laser radiation measured parallel and perpendicular and to the p-n-junction is
different. This effect is commonly referred to as astigmatism of laser diodes.
The emission point of the laser radiation measured in the direction parallel to the p-n-
junction is located almost at the facet of the laser diode. The emission point measured
perpendicular to the p-n-junction is located behind the laser facet, inside of the laser diode.
A typical distance between these two emission points is between 5µm and 10µm for single
mode laser diodes.
Emitter
Width
Fast
Axis
Slow
Axis
Fig. 4: Schematic view of a laser diode together with the laser emission. The blue
indicated area shows the p-n-junction of the laser diode. The yellow indicated area
shows the active region of the laser diode. The laser radiation is indicated on the
right hand side of the laser diode. The schematic indicates the beam profile of the
astigmatic beam propagation.

Technical Documentation
Page 6
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
Due to the astigmatism of laser diodes, the divergence of the laser beam is different for the
direction parallel and perpendicular to the n-p-junction. The collimated laser beam
measured in the direction parallel to the p-n-junction shows a larger divergence than
measured perpendicular to the p-n-junction. Fig. 4 shows a schematic of a laser diode with
the astigmatic beam propagation.
Commonly, the direction of the laser beam measured parallel to the p-n-junction is referred
to as slow axis of the laser beam. The direction of the laser beam measured perpendicular
to the p-n-junction is referred to as fast axis of the laser beam.
2.2 Determining the Status of the Laser Collimation
When collimating laser diodes with a spherical lens, it is not possible to perform a
simultaneous collimation for fast and for slow axis. The laser diode is either collimated for
the fast axis with accepting a beam divergence of the slow axis, or the laser diode is
collimated for the slow axis with accepting a beam divergence of the fast axis.
Fast
Axis
Slow
Axis
Emitter
Width Collimation
Lens
Fig. 5: Schematic view of the beam propagation of a collimated laser diode. Due to
the astigmatism of the laser diode, the slow axis shows a focus behind the
collimation lens. After the focus, the slow axis diverges and becomes larger than
the fast axis.
Commonly, a proper collimation of the fast axis is chosen with accepting a slight beam
convergence for the slow axis.
For analyzing the beam propagation, a laser camera is required. In case a laser camera is
not available, a screen together with a beam viewer may give some indication of the beam
propagation.
Make sure to wear eye protective glasses and a ESD protective wrist band before starting
any work at the collimator.

Technical Documentation
Page 7
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
2.3 Adjustment of the Laser Collimation
The collimator offers a fine adjustment screw indicated as (3) in Fig. 1 for adjusting the
beam collimation. Clockwise rotation of the fine adjustment screw will increase the
distance between the emission facet of the laser diode and the collimation lens.
Fig. 6: Photo of the laser collimator together with the inside-hex adjustment tool.
Rotation of the adjustment tool will result into an adjustment of the lens relative to
the laser diode. Do not rotate the adjustment screw more then +/- ¼ of a rotation.
Verify the beam propagation according to the schematic given in Fig. 5. The laser diode is
well prepared for using it within the external cavity if the beam propagation shows the point
of lowest diameter at a distance of approximately 1.5m distance from the collimator. The
beam propagation be adjusted via rotation of the adjustment tool shown in Fig. 6. Do not
rotate the adjustment screw more than +/- ¼ of a rotation.

Technical Documentation
Page 8
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
3 Cavity Alignment
This section describes the preparation of the prism arm for the alignment procedure as
well as the alignment procedure itself. Optional optical components as optical isolator or
fiber coupler may need to be removed.
3.1 Pre-Adjustment of the Reflection Prism
The following steps need to be taken for performing the cavity alignment. As tools, a beam
viewer and/or a NIR detection card are required.
A) Rotate the wavelength adjustment screw until angle between the prism arm and the
grating is close to the expected value for the operating the laser system at the desired
wavelength.
B) Preset the laser current with a value between the expected threshold current with
optical feedback and the threshold current without optical feedback.
C) Make sure to wear eye protective glasses and a ESD protective wrist band.
D) Enable the laser injection current.
(1)
(2)
(3)
(4)
Fig. 7: Schematic view of the laser spot visible on a screen placed into the zeroth order
beam of the laser head. The steps 1 to 4 describe the steps of the alignment
procedure.

Technical Documentation
Page 9
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
E) As soon as the injection current is enabled, the zeroth order diffraction of the
diffraction grating will be visible. A good distance to observe the zero order diffraction
is approximately 0.5m from the grating. In case you cannot see the spot, you need to
darken the laboratory.
F) When rotating the prism arm, the zeroth order diffraction beam will not move due to
the Littman/Metcalf design.
G) Rotate the prism arm until the first order diffraction directs into the collimation lens of
laser collimator after passing the grating a second time.
H) As soon as the first order diffraction of the grating is directed into the lens of the
collimator, a second weak spot will be visible close to the zero order diffraction at
approximately 0.5m distance from the diffraction grating. Fig. 7 (1) shows an
approximate view of the two spots.
I) Rotate the prism arm in such a way that the two spots are close to each other, and
both spots visible as indicated in Fig. 7 (2).
J) Adjust the cavity alignment screw in such a way that the two spots are on the same
horizontal line as shown in Fig. 7 (3). As soon as the strong moving spot and the weak
spot are close enough to each other, the weak spot will ‘jump’ into the strong spot and
the laser will start to operate under external cavity conditions.
K) Keep the position of the weak spot in mind and rotate the prism arm in such a way that
the strong spot is in an identical position as the weak spot. The spot should become
much brighter as soon as the laser is above threshold as indicated in Fig. 7 (4).
Laser Power
Cavity Alignment
Fig. 8: Schematic of the dependence of the laser power as a function of the cavity
alignment.

Technical Documentation
Page 10
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
L) The optimization of the cavity will be performed via the cavity alignment screw. Once
laser action is found, no more than +/- 1 rotation of the cavity alignment screw is
required.
M) A typical dependence of the laser power on the cavity alignment is shown in Fig. 8.
Usually, the best adjustment is achieved as soon as the laser power reaches its
maximum value. Make sure that the laser is not operating at a side maximum by
verifying the cavity alignment according to Fig. 8. Otherwise you will not be able to
achieve the specified values.
N) After optimization of the cavity alignment screw, you may try to optimize the laser
collimation for improving the laser power. No more than 1/8 of a turn of the collimator
adjustment screw will be required.
O) Repeat step L) and M) until the maximum power is achieved.
Please note that it may be difficult to determine which cavity alignment maximum will
provide the provides the highest power and the best laser performance.
Furthermore, there are types of laser diodes where the alignment for the highest power
does not coincide with the alignment for the best laser performance.
Therefore after optimizing the laser for the highest power, it is recommended to check the
minimization of the threshold current due to the feedback of the external cavity.
3.2 Minimizing the Laser Threshold
When optimizing the cavity adjustment via the cavity alignment screw, you may find
several maximum within the laser power as a function of the cavity alignment screw as
indicated in Fig. 8. Only one of these power maximum will provide you the best laser
performance.
Please note that the maximum with the highest power does not necessarily provide the
best over all performance of the laser system. A procedure for choosing the best
alignment is to check which power maximum offers the lowest threshold current.
A simple method for monitoring the threshold current as follows. Required items are an
oscilloscope, a power meter with a monitor exit and a signal generator.
A) Connect the laser current modulation input of the laser controller to signal generator
which provides a triangular voltage signal.
B) Alternatively, you may use the internal Ramp Generator function of the laser controller
together with the Current Coupling function of our laser controller for generating the
laser current modulation.

Technical Documentation
Page 11
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
C) Connect the monitor output of the power meter to the Channel A of the oscilloscope.
D) Connect the SYNC signal of the signal generator to Channel B of the oscilloscope.
E) When activating the laser current modulation, the Laser-Current Laser-Power
Characteristic of the laser system will be visible on the oscilloscope screen.
F) Changes of the threshold can be in-situ monitored as a function of the cavity
alignment screw.
G) The optimization of the cavity will be performed via the cavity alignment screw. No
more than +/- 1 rotation of the cavity alignment screw is required.
H) After optimization of the cavity alignment screw, you may try to optimize the laser
collimation for improving the laser power. No more than 1/8 of a turn of the collimator
adjustment screw will be required.
I) Repeat step G) and H) until the maximum power is achieved.

Technical Documentation
Page 12
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
4 Mode-Hop Free Tuning
After optimizing the laser alignment via maximizing the laser power and minimizing the
laser threshold, the mode-hop free tuning may be optimized.
4.1 Principles of Mode-Hop Free Tuning
For achieving mode-hop free tuning, a synchronization between the grating determined
tuning and the cavity determined tuning need to be synchronized. This synchronization is
performed by adjusting the cavity length.
For this purpose, the laser diode mount offers the option of varying the cavity length. In
case the grating determined tuning is “faster” than the cavity determined tuning, the cavity
length needs to be increased. In case the cavity determined tuning is “faster” than the
grating determined tuning, the cavity length needs to be reduced.
After adjusting the cavity length, the mode-hop free tuning needs to be monitored. This
can be done via several possible methods. We recommend a combination of these
methods.
4.2 Monitoring the Tuning Performance via Fabry Perot Interferometer
A simple method for monitoring mode-hop free tuning is using a Fabry Perot interferometer
The laser system needs to be protected from optical feedback from the Fabry Perot
interferometer via an optical isolator. Optical feedback may cause multimode emission and
or in case of strong feedback, damages to the laser diode may occur.
The Fabry Perot interferometer allows to monitor the mode-hop free tuning behavior when
slowly scanning the laser wavelength via the piezo actuator. This method is very sensitive
to multimode emission behavior.
4.3 Monitoring the Tuning Performance via Power Monitoring
A second method for monitoring mode-hop free tuning behavior bases on monitoring the
laser power as a function of the voltage applied to the piezo actuator.
Required items are an oscilloscope, a power meter with a monitor exit and a piezo
amplifier with signal generator.
A) Connect the monitor output of the power with Channel A of the oscilloscope.

Technical Documentation
Page 13
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
B) Connect the SYNC signal of the signal generator powering the piezo amplifier with the
Channel B of the oscilloscope.
C) Enable piezo scanning at the laser controller. Choose a small to moderate scan
amplitude for the piezo voltage.
D) Monitor the laser power as a function of the piezo voltage via the oscilloscope trace.
Fig. 9: Schematic of the dependence of the laser power as a function of the piezo voltage
Fig. 10: Oscilloscope trace of the power modulation with and without mode-hop
E) Fig. 9 and 10 show typical laser power trace with applied modulation. A continuous
trace indicates mode-hop free wavelength tuning. A discontinuous power change
indicates a mode-hop.
This method is fast and effective way of monitoring mode-hop free tuning performance
during piezo scanning.
Mode-Hop

Technical Documentation
Page 14
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
4.4 Monitoring the Tuning Performance via Wavemeter
A third method for monitoring mode-hop free tuning behavior is to use a wavemeter.
In case the system is not equipped with an optical isolator, it needs to be installed prior to
the tuning optimization via Wavemeter. The laser system needs to be protected from
optical feedback from the Wavemeter via an optical isolator. Optical feedback may cause
multimode emission and or in case of strong feedback, damages to the laser diode may
occur.
The Wavemeter allows to monitor the mode-hop free tuning behavior when slowly
scanning the laser wavelength via piezo actuator or during motor operation.
4.5 Optimizing the Tuning Performance
Optimizing the mode-hop free tuning bases on three steps. Initially the mode-hop free
tuning behavior should be monitored via Fabry Perot Interferometer. After the initial
monitoring of the mode-hop free tuning behavior with a Fabry Perot interferometer, the
mode-hop free tuning may be optimized via monitoring the laser power as a function of the
voltage applied to the piezo actuator. As a first step, the cavity alignment needs to be
optimized.
A) Changing of the cavity alignment will change the number and position of the mode-
hops.
B) Optimize the mode-hop free tuning range via fine adjustment of the cavity adjustment
screw. No more than one rotation of the cavity alignment screw will be required for
optimization of the mode-hop free tuning of the external cavity laser system.
C) Once the initial cavity alignment is optimized, the cavity length needs to be optimized
for synchronizing the tuning “speed” of the grating determined tuning and the cavity
determined tuning. This is best monitored via wavemeter.
D) In case the laser shows mode-hops where the laser frequency jumps in direction of
tuning, the cavity length needs to be reduced.
E) In case the laser shows mode-hops where the laser frequency jumps against direction
of tuning, the cavity length needs to be enlarged.

Technical Documentation
Page 15
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
TEC-500 side view
Fig. 10: Photo of the laser cavity indicating Collimator base plate and Collimator base
plate screw for changing the cavity length
F) The cavity length may be adjusted via changing the position of the collimator base
plate where the collimator is mounted to. The collimator base plate allows a translation
of the collimator for changing the cavity length.
After performing the change of cavity length, steps A..F need to be repeated until the
tuning performance is fully mode-hop free.
Collimator
base plate
Collimator
base plate
screw

Technical Documentation
Page 16
Sacher Lasertechnik GmbH Tel.: +49 6421 305-0 Sacher Lasertechnik, LLC Tel.: 1-714-670-7605
Rudolf Breitscheid Str. 1-5 Fax: +49 6421 305-299 5765 Equador Way Fax: 1-714-670-7662
D-35037 Marburg, Germany Email: [email protected] Buena Park, CA 90620, USA Email: [email protected]
5 Summary
Following the instruction steps described in the section 1 to 4 will result into an excellent
laser performance.
In case there are questions or problems with following the given instructions, please
contact us.
When contacting us please provide a technical documentation of the results obtained
during the steps of the alignment procedure. This will allow us to give you an analysis of
possible issues.
Table of contents
Popular Measuring Instrument manuals by other brands

VOLTCRAFT
VOLTCRAFT FM-HP 40 92 37 operating instructions

Tempo Communications
Tempo Communications 930XC-20C instruction manual

Elcometer
Elcometer 138 user guide

Starrett
Starrett 799A Series instruction manual

Clas Ohlson
Clas Ohlson ET-933 manual

Vanguard Instruments
Vanguard Instruments ATRT-01 S3 user manual

Kofloc
Kofloc 3100 instruction manual

Keithley
Keithley 6517B Reference manual

NORTHROP GRUMMAN
NORTHROP GRUMMAN IR Patara Laser System Pre-installation guide

SCHOTT
SCHOTT handylab multi 12 operating manual

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS WATERPROOF TDSCAN 4 CONDUCTIVITY TESTER instructions

PCE Health and Fitness
PCE Health and Fitness PCE-RVI 6 user manual