SAPHYMO MiniTRACE CSDF User manual

www.saphymo.de
Manual
Type CSDF

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 2 08/2012
0 Contents
0Contents...............................................................................................................2
1Foreword ..............................................................................................................4
2Description ...........................................................................................................5
2.1 Display ..........................................................................................................5
2.2 Alarm function ...............................................................................................6
2.3 Operation Modes of the MiniTRACE CSDF ..................................................6
2.4 Measurements...............................................................................................9
2.4.1 Information for your own radiation protection.........................................9
2.4.2 Examples of natural radionuclides in food and limits for artificial
radionuclides in food.............................................................................................9
2.4.3 Measurement tasks for the MiniTRACE CSDF ....................................10
2.5 Accessories for measuring..........................................................................21
2.5.1 MiniTRACE function test using a small test source..............................25
2.6 The design of the instrument.......................................................................27
2.7 Protective rubber casing..............................................................................28
3Standard calibration for Co-60 and Cs-137 ........................................................29
4Maintenance and care........................................................................................30
4.1 Changing the batteries ................................................................................30
4.2 Cleaning (Decontaminating) MiniTRACE CSDF .........................................30
4.2.1 Removing the grid................................................................................30
4.2.2 Cleaning the case ................................................................................31
4.3 Technical properties and functions..............................................................31
4.3.1 Calibration factors and efficiency .........................................................33
5MiniTRACE user software ..................................................................................34
6Warranty information ..........................................................................................35
7Appendix ............................................................................................................36
7.1 DKD certificate for the Co-60 radiation emitter............................................36

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
08/2012 Page 3
Figures
Figure 1: Information shown on the display .............................................................5
Figure 2: MiniTRACE CSDF display – alarm is activated ........................................6
Figure 3: Using the mode button to select an operation mode ................................8
Figure 4: Opened protective lid..............................................................................10
Figure 5: Display in dose rate mean mode ............................................................11
Figure 6: Display in cps mean mode......................................................................12
Figure 7: Measuring device for wipe tests .............................................................15
Figure 8: Wipe test measurement..........................................................................16
Figure 9: Counting up measurement .....................................................................17
Figure 10: Food measurement flow chart ................................................................18
Figure 11: Food measurement ................................................................................18
Figure 12: Display for background (control) measurement in the “Food” menu.......19
Figure 13: Result display for background (control) measurement
in the “Food” menu .................................................................................19
Figure 14: Display in the “Food” menu.....................................................................20
Figure 15: Result display in the “Food” menu..........................................................20
Figure 16: Food measurement kit (See Chapter 2.4.3.7).........................................22
Figure 17: Accessories for the wipe test (See Chapter 2.4.3.5.4)............................22
Figure 18: Belt case.................................................................................................22
Figure 19: Transparent plastic pouch ......................................................................23
Figure 20: Transportation box..................................................................................23
Figure 21: Peli Box ..................................................................................................24
Figure 22: Emergency case.....................................................................................24
Figure 23: Test source (see Chapter 2.5.1).............................................................25
Figure 24: Communication kit with MiniTRACE .......................................................25
Figure 25: Dose rate of the test source....................................................................26
Figure 26: “CMean” menu........................................................................................26
Figure 27: Side view ................................................................................................27
Figure 28: Front view ...............................................................................................27
Figure 29: Protective rubber casing.........................................................................28
Figure 30: Changing the batteries ...........................................................................30
Figure 31: Removing the grid (part 1)......................................................................30
Figure 32: Removing the grid (part 2)......................................................................31
Figure 33: Energy response of the MiniTRACE CSDF (based on Cs-137,
with lid closed and GM tube energy compensated)................................32
Figure 34: Energy response of the MiniTRACE CSDF
(with open lid, NOT energy compensated) ............................................32
Tables
Table 1: Status reports in the 5-digit alphanumeric portion of the LCD display ......5
Table 2: Overview of measurements....................................................................10
Table 3: 1-Sigma statistical uncertainty with different background dose rates .....18
Table 4: List of accessories..................................................................................21
Table 5: Technical data ........................................................................................31
Table 6: 4 π-Efficiency [Bq] MiniTRACE CSDF for different radionuclides..........33
Table 7: Calibration factors and efficiency [Bq/cm
2
] MiniTRACE CSDF
based on ISO 7503-1 .............................................................................33

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 4 08/2012
1 Foreword
Thank you for purchasing the MiniTRACE CSDF contamination and survey meter. In order to
gain a better understanding of its functions, please take a little time to read through this
instruction manual.
MiniTRACE CSDF is a multifunctional contamination and survey meter that has been
developed by Saphymo GmbH in Frankfurt/Main.
MiniTRACE CSDF is especially suited for use in nuclear plants, reprocessing sites, research
centers, hospitals, and for occupational groups such as the armed forces, police or fire
department.
MiniTRACE CSDF offers ease of use. It is the ideal instrument for identifying possible
contamination simply and quickly (values are displayed in Bq, Bq/cm
2
or cps, depending on
the mode). You can also use it as a multifunctional survey meter. With its energy compen-
sated Geiger-Mueller pancake detector you can determine the ambient dose rate equivalent
(H*(10)), displayed in µSv/h. The device is also suitable for food measurements and wipe
tests.
The MiniTRACE CSDF enables the user to select different measurement cycles for the count
up mode directly on the instrument or to choose between different radionuclides.
Additionally, the optional MiniTRACE software enables you to set the alarm thresholds or
calibrate the instrument (see additional software manual).

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
08/2012 Page 5
2 Description
2.1 Display
The display is composed of a 6-digit 7-segment display for the measurement results as well
as a 5-digit alphanumeric display. The MiniTRACE CSDF shows the unit in “cps”, “Bq”,
“Bq/cm
2
”, “µSv/h”, “counts” and Bq/l.
Note: The dose rate displayed is energy compensated to the ambient dose rate
equivalent (H*(10)). To measure the dose rate correctly the lid must be closed.
The following status reports can be displayed on the MiniTRACE CSDF display:
Bq/cm2
se c
µSv/h
Figure 1: Information shown on the display
Display Acoustic
Alarm
Note
Unit: Bq,
Bq/cm
2
, cps,
µSv/h or Bq/l
-
Available measurement units
CAPT Yes
Detector error:
If the detector is defective, the message “CAPT” will appear in the display.
The warning is triggered when the detector fails to send an impulse for a
period of 40 seconds.
SAT Yes
Saturation:
If the measurement range is saturated (value out of range), the message
“SAT” will appear in the display. This message will remain until the
measurement area is no longer saturated.
No
Battery warning:
If the battery symbol appears in the display, the battery should be
changed. The instrument can still be operated for 10 hours after the battery
symbol appears.
E < % No
Statistical uncertainty in the mean value mode:
The mean value is determined for a user-defined period. The default is set
to 3 Sigma and adjustable by software.
Table 1: Status reports in the 5-digit alphanumeric portion of the LCD display

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 6 08/2012
2.2 Alarm function
One alarm threshold for the MiniTRACE CSDF can be set via software for the following units:
•Alarm 1: Dose rate mode in µSv/h (default: 20 µSv/h)
•Alarm 2: Counts in cps (default: 0, meaning deactivated)
•Alarm 3: Activity in Bq (default: 60 Bq)
•Alarm 4: Surface contamination in Bq/cm
2
(default: 4 Bq/cm
2
)
The alarm can be activated or deactivated using the mode button (See Figure 3). If you
would like to activate the alarm, go to the “AL oN” menu and wait three seconds. If you would
like to deactivate the alarm, go to the “AL oFF” menu and wait three seconds. The value of
the alarm threshold is shown on the display.
Figure 2: MiniTRACE CSDF display – alarm is activated
Note: After one of the operation modes has been selected, the status (here alarm
activated) and the value (here 20.00 µSv/h) of the alarm setting are displayed.
2.3 Operation Modes of the MiniTRACE CSDF
The following measurement modes can be set:
•Rate mode (RATE): Dose rate in “µSv/h”. To use this function properly the lid must
be closed. The results will then be energy compensated to the ambient dose
equivalent (H*(10)).
•Rate mean mode (RMEAN): Dose rate mean value in “µSv/h”. To use this function
properly the lid must be closed. The results will then be energy compensated to the
ambient dose equivalent (H*(10)). To reset the calculated mean value, press the
mode button. This function is used to gain measurement results of a higher precision
in a stable radiation environment. The longer the measurement, the higher the
accuracy.
•CPS mode (CPS): Count rate in “cps”.
•CPS mean mode (CMEAN): Count rate mean value, no unit is displayed. To reset
the calculated mean value, press the mode button. This function is used to gain
measurement results of a higher precision in a stable radiation environment. The
longer the measurement, the higher the accuracy.
•Bq mode (BQ): Activity in “Bq”. You can use the mode button to choose between
different radionuclides (along with their appropriate calibration factors; 4π-calibration).
See Table 6.
•Bq/cm^2 mode (BQCM2): Surface contamination in “Bq/cm
2
”. You can use the mode
button to choose between different radionuclides (along with their appropriate cali-
bration factors). See Table 7. Calibration based on ISO 7503-1.
Indicates
whether the
alarm is
activated

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
08/2012 Page 7
•Count up mode (CNTUP): Pulses are counted during the selected time interval. The
interval can be set via the mode button. Possible intervals are: 60, 120, 300, 600,
1,800 and 3,600 seconds.
•Food mode: Contamination of food in “Bq/L”.
Note: If you turn on the instrument the last selected operation mode is still activated.
The default calibration factors are set for a distance of 3 mm between grid and
surface. For more details see Chapter 3.
The MiniTRACE CSDF is UNABLE to identify radionuclides!

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 8 08/2012
press I-buttonpress mod e button
activate
RATEmode
turn on
pressI-button
Rate mode
Rate mean mode
(E < XX % in the displa y)
CPS mode
CPS mean mode
(E < XX % in the display)
Bq mode (BQ)
Bq/cm^ 2 mode (BQCM2)
Count up mode (CNTUP)
Food
Speaker on (SP ON)
Speaker off (SP OFF)
Ala rm (ALON)
Ala rm off (ALOFF)
OFF
press
mode button
press
mode button
press
mode button
press
mode button
press
mode button
press
mode button
press
mode button
press
mode button
press
mode button
press
mode button
press
mode button
wait 3 sec.
unit µSv/h
alarm on/off
alarm threshold
in the display
activate
RATEm ean
mode
wa it 3 sec.
reset mean value
press
mode button
activate
CPSm ode
wa it 3 sec.
unit CPS, a la rm on/off
alarm threshold
in the d isplay
activate
CPSmea n
mode
wait 3 sec.
Reset mean value
press
mode-button
activate
Bq mode
wa it 3 sec.
unit Bq, used rad ionuc lude
in the display
press
mode button
select
radionuc lide
wa it 3 sec.
activate
Bq/cm ^ 2
m ode
wa it 3 sec.
unit Bq/c m^ 2, used ra dionuclude
in the d isplay
press
mode button
select
radionuc lide
wa it 3 sec.
activate
count up
m ode
wa it 3 sec.
measurement time
in the display
press
mode button
selec t measurement cycle
(60, 120, 300, 600, 1800, 3600 sec .
wait 3 sec.
Reset
measurement
cyc le
press
mode-button
Reset mean value
press
mode button
See flowchart in c hap ter “Food m easurement”
wa it 3 sec.
activate
sp eaker
wa it 3 sec.
deac tivate
speaker
wa it 3 sec.
activate
alarm
wait 3 sec.
Dea c tivate
alarm
wait 3 sec.
Note:
a la rm threshold can be set via MiniTRACE Software)
pressI-button
(2 times short)
pressI-button
(2 times short)
pressI-button
(2 timesshort)
pressI-button
(2 times short)
press I-button
(2 times short)
turn off
instrument
pressI-button
IR
press
mode button
activate
IR-c omm unic ation
wa it 3 sec.
pressI-button
(2 times short)
Figure 3: Using the mode button to select an operation mode

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
08/2012 Page 9
2.4 Measurements
2.4.1 Information for your own radiation protection
•Protect yourself from an unnecessary dose. The easiest way is to reduce the length
of exposure.
•
Increasing the distance reduces the dose due to the inverse square law.
•Radiation doses should all be kept as low as reasonably achievable.
•Protect yourself against contamination by using the rubber gloves and the tweezers.
•Do not touch the surface with your hand or the MiniTRACE to prevent the spread of
contamination.
•Avoid incorporation and inhalation. Incorporation or inhalation of radioactivity is much
more dangerous than contamination of the skin.
For more information see:
Euratom: http://www.euratom.org
IAEA: http://www.IAEA.org
WHO: http://www.who.int
2.4.2 Examples of natural radionuclides in food and limits for artificial
radionuclides in food
In all foods you will find some natural radionuclides, mainly the long-lived K-40 and the long-
lived radionuclides of the uranium-radium and thorium decay series. Alongside the nutrients
necessary for growth, plants and animals also absorb radionuclides, due to their similar
chemical properties. The level of radionuclide concentration in food is determined by the
radioactivity of the source media used (soil, water), the availability of nutrients and other
substances from soil and water and other conditions at the location of the plant or animal
production.
Higher radionuclide concentrations are found in the livers and kidneys of animals (in
comparison to the meat) due to their function as "pollution filters" as part of the metabolism of
various elements, above all radionuclides.
Some interesting numbers (literature source BfS/Germany):
•Natural K-40 in:
ofruits: approx. 50 Bq/kg
omilk: approx. 50 Bq/l
omeat: approx. 100 Bq/kg
opeas/beans: approx. 380 Bq/kg
oman: approx. 40-60 Bq/kg
The national threshold for food differs from country to country.
In Germany the following thresholds apply: Regulation 1707/86 dated 30.5.86
(2) for:
•Total Cesium 370 Bq/kg for milk and baby food
•600 Bq/kg for other food.

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 10 08/2012
More information can be found at:
Japan:
http://www.mhlw.go.jp/english/topics/foodsafety/index.html
http://www.mhlw.go.jp/english/topics/foodsafety/dl/110318-1.pdf
Germany:
http://www.bmu.de/atomenergie_sicherheit/doc/47216.php
2.4.3 Measurement tasks for the MiniTRACE CSDF
With the MiniTRACE CSDF you choose between the following operation modes:
Mode Unit Lid open Lid closed
Dose rate H*(10) (Menu RATE)
Preferred direction observed, see Figure 27
µSv/h - X
Dose rate mean value H*(10) (Menu RMEAN)
Preferred direction observed, see Figure 27
µSv/h - X
Count rate (Menu CPS) cps X (α, β, γ) X (γ)
Count rate mean value (Menu CMEAN) cps X (α, β, γ) X (γ)
Activity (Menu BQ) Bq X (α, β, γ) Not
recommended
Surface contamination (Menu BQCM2) Bq/cm
2
X (α, β, γ) Not
recommended
Count up (Menu COUNTUP) cps X (α, β, γ) X (γ)
Food (Menu FOOD) Bq/l - X (γ)
Table 2: Overview of measurements
Note: Measurement results in the “Surface contamination” (Bq/cm
2
) mode are based
on ISO 7503-1. For the activity mode (Bq) we have calibrated the instrument in
a 4 πgeometry.
The internal calibration factors are default values and not individually calibrated.
Figure 4: Opened protective lid
Protective
grid
Pancake Geiger-
Mueller detector
behind the grid
Lid open,
includes serial
number, infor-
mation and
energy filter

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
08/2012 Page 11
2.4.3.1 Ambient dose rate equivalent measurement – H*(10) in
µSv/h
Select the “RATE” menu by using the mode button. Once “RATE” is displayed, wait 3
seconds; the menu will start automatically. For further details see Figure 3.
This mode reacts very quickly to changes in the radiation level.
To determine the ambient dose rate equivalent (H*(10)) correctly, the lid must be closed!
When the lid is open, the ambient dose rate equivalent is NOT calculated correctly.
The value is given in µSv/h.
Note: The ambient dose rate equivalent gives you important information for radiation
protection.
Dose limits for radiation protection in Germany are (in 2011):
•Population 1 mSv/a
•Professionally exposed persons: 20 mSv/a
2.4.3.2 Mean value of the ambient dose rate equivalent – H*(10) in
µSv/h
Select the “RMEAN” menu by using the mode button. Once RMEAN is displayed, wait 3
seconds; the menu will start automatically. For further details see Figure 3.
We recommend this menu if the ambient dose rate equivalent is “stable” in your environment
and you would like to determine the dose rate more precisely. The longer the measurement,
the higher the accuracy. This mode does NOT react to rapid changes in the radiation level.
To determine the mean value of the ambient dose rate equivalent (H*(10)) correctly, the lid
must be closed! When the lid is open, the mean value of the ambient dose rate equivalent is
not calculated correctly. The value is given in “µSv/h”. In the top left corner of the display,
statistical uncertainty is given. By default a statistical uncertainty of 3 sigma is calculated.
You can change the sigma value using the MiniTRACE software. For more details see the
software manual.
Note: The ambient dose rate equivalent gives you important information for radiation
protection.
Figure 5: Display in dose rate mean mode
2.4.3.3 Count rate measurement in cps
Select the “CPS” menu by using the mode button. Once “CPS” is displayed, wait 3 seconds;
the menu will start automatically. For further details see Figure 3.
This mode reacts very quickly to changes in the radiation level.
With this menu you can check the contamination of a surface, for example. The value is
given in counts per second (cps). With the lid closed only γ-radiation is measured. If you
open the protective lid (material: aluminum, thickness: 1 mm) you will see the Geiger-Mueller

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 12 08/2012
pancake detector. With the lid open you can measure α-, β- and γ-radiation. This detector is
very sensitive and protected by a grid.
Note: All given measurement values are gross values. This means that background
values are NOT subtracted.
Basically you should do contamination measurements (i.e. on surfaces) in 2 steps:
1. Measurement with lid closed to determine the γ-contamination
2. Measurement with lid open to determine the α-, β- and γ-contamination
To determine the β-component another step is necessary:
3. Measurement with lid open and a sheet of paper on the detector. Subtract the value
from measurement 3 from measurement 1; the result contains the β-contamination.
To determine the α-component another step is necessary:
Measurement with lid open and a sheet of paper on the detector. Subtract the value from
the result without and with paper and you will get the α-contamination.
To determine the α- and β-components another step is necessary:
Subtract the value from measurement 2 from measurement 1; result contains the α- and
β–contamination.
Note: For a higher precision of the measurement result the radiation background has
to be measured first and then subtracted from the final result.
α–radiation can be shielded with a piece of paper.
β-radiation is almost completely shielded by closing the lid.
2.4.3.4 Mean value of the count rate in cps
Select the “CMEAN menu” by using the mode button. Once “CMEAN” is displayed, wait 3
seconds; the menu will start automatically. For further details see Figure 3.
With this menu you can check the contamination of a surface, for example. If the count rate
is “stable” in your environment and you would like to determine the count rate more precisely
you can use this menu. The longer the measurement, the higher the accuracy. This mode
does NOT react to rapid changes in the radiation level.
The value is given in counts per second (cps). In the top left corner of the display, the
statistical uncertainty is given. By default a statistical uncertainty of 3 sigma is calculated.
You can change the sigma value using the MiniTRACE software. For more details see the
additional software manual.
With the lid closed only γ-radiation is measured. If you open the protective lid (material:
aluminum, thickness: 1 mm) you will see the Geiger-Mueller pancake detector. With the lid
open, you can measure α-, β-, and γ-radiation. This detector is very sensitive and protected
by a grid. See also 2.4.3.3.
Figure 6: Display in cps mean mode

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
08/2012 Page 13
2.4.3.5 Measurement of surface contamination
2.4.3.5.1 Measurement of the activity in Bq
With this menu you can check the contamination of a surface, for example, based on a 4 π-
calibration.
Select the “BQ” menu by using the mode button. Once “BQ” is displayed, wait three seconds.
You can select the radionuclide you would like to measure by pressing the mode button.
After additional three seconds, the display will indicate whether an alarm threshold has been
activated for the activity measurement. Then the measurement starts automatically. For
further details see Figure 3. The activity is given in “Bq”. See also 2.4.3.3, 2.4.3.5.32.4.3.5.4.
You can select the following radionuclides with the built-in calibration factors (4 π-
calibration):
•Am-241
•C-14
•Cl-36
•Co-60
•Cs-137
•Sr-90
•U-238
•User (see below)
If you would like to select another radionuclide, press the mode button once, wait three
seconds, and then select the radionuclide. The display will indicate whether an alarm
threshold has been activated for the activity measurement. The measurement will then start
automatically.
You can assign your own calibration factor for the radionuclide “User” using the software. For
further information see the software manual.
Note: MiniTRACE CSDF can not determine the radionuclide. Selection of the
radionuclide has to be done by the user.
2.4.3.5.2 Measurement of surface contamination in Bq/cm
2
With this menu you can check the contamination of a surface, based on ISO 7503-1
calibration. The measurement of the surface contamination can be done in two different
ways. For more information see Chapters 2.4.3.5.3 and 2.4.3.5.4.
Select the “BQCM2” menu using the mode button. Once “BQCM2” is displayed, wait three
seconds. You can then select the radionuclide you would like to measure by pressing the
mode button. After an additional three seconds, the display will indicate whether an alarm
threshold has been activated for the activity measurement. The measurement will then start
automatically. For further details see Figure 3. Activity is given in Bq/cm
2
.

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 14 08/2012
You can select the following radionuclides. Associated built-in calibration factors are based
on ISO 7503-1:
•Am-241
•C-14
•Cl-36
•Co-60
•Cs-137
•I131
•Sr-90
•U-238
•User (see below)
If you would like to select another radionuclide, press the mode button once, wait three
seconds, and then select the radionuclide. The display will indicate whether an alarm
threshold has been activated for the activity measurement. The measurement will then start
automatically.
You can assign your own calibration factor for the radionuclide “User” using the software. For
further information see the software manual.
Note: MiniTRACE CSDF cannot determine the radionuclide. Selection of the
radionuclide must be carried out by the user.
2.4.3.5.3 Surface contamination with a direct measurement
Select the “BQCM2” menu or “Bq” mode. For further details see Figure 3. To measure the
surface contamination with a direct measurement you will need to open the lid. With this
method you measure the nonadherent and adherent surface contamination.
With the lid open, move the detector slowly over the surface, without allowing the
MiniTRACE to touch the surface. We recommend activating the acoustic registration of
impulses (“SP oN” menu). If you find a contaminated area, position the MiniTRACE at this
point. The distance between the surface and the MiniTRACE grid should be as little as
possible; we recommend approx. 3 mm.
Before you start with this measurement you need to determine the background radiation and
subtract it from the later result. For further information see ISO 7503-1 or your national
guidelines.
Note: The closer you are to the surface, the better the results will be. Be sure not to
touch the surface with the MiniTRACE CSDF. In the event of contact, the
MiniTRACE CSDF could be contaminated and the contamination could be
spread.
2.4.3.5.4 Surface contamination with a wipe test
To work with this menu select “BQCM2” mode. For further details see Figure 3. The wipe test
is a common method for measuring surface contamination. It is an indirect measurement
technique; only nonadherent surface contamination can be determined. This test is useful if
external radiation levels are too high, if surfaces are inaccessible or if the radiation would not
be detectable. In these cases activity on the surface is sampled on wipe test smears. They
provide an efficient, convenient means of sampling contaminated areas with radioactivity on

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
08/2012 Page 15
either wet or dry surfaces. Afterwards the activity on wipe test smears is measured with
MiniTRACE CSDF.
The surface contamination is measured in “Bq/cm²“ and it is proportional to the net counting
rate and a calibration factor. The calibration factor is a function of the instrument's efficiency
for the specified radionuclide, the area wiped, the source efficiency and the removal factor.
Frequently the removal factor is unknown.
Note: Use only the supplied measuring cup and holder. Otherwise the result will be
inaccurate.
Procedure:
•Perform a background measurement and go to the “BQCM2” menu.
•Take one wipe test sample from the filter roll by tearing along the perforation. Do not
disconnect the filter from the foil.
•Protect yourself against contamination by using the rubber gloves and the tweezers.
•Wipe very carefully with the supplied filter over the averaging area.
•In Germany the averaging area should not be larger than 100 cm
2
. Please check the
guidelines in your country.
•Put the filter in the filter cup with the side that is possibly contaminated facing up.
•Put the filter cup in the measurement device.
•Position the CSDF on the measurement device. The lid must be open.
•Now you can start the measurement.
•After measurement is done, remove the wipe test sample from the cup carefully, if
you have found any contamination.
•For the exact procedure see ISO 7503-1 “Evaluation of surface contamination” or
your national guidelines.
Figure 7: Measuring device for wipe tests
Wipe test
sample in the
measuring
cup
Measurement
device for
exact positi-
oning of
MiniTRACE

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 16 08/2012
Figure 8: Wipe test measurement
Calculation of the final result:
backgound
theoftotal
valuemeasuredvaluemeasuredcmBqvaluenet −=]/[
2
Note: To determine the background value put the CSDF in the measurement holder. It
is not necessary to use a new (uncontaminated) wipe test sample; put only the
measurement cup in the measurement holder.
Example:
Measurement result of the background: 5 Bq/cm
2
Measurement result of surface measurement: 105 Bq/cm
2
Net value = 105 Bq/cm
2
– 5 Bq/cm
2
= 100 Bq/cm
2
][
2
5.15]/[
2
2
cmareawiped
cmxcmBqvaluenet
resultFinal =
Example:
Net value = 100 Bq/cm
2
Wiped area = 100 cm
2
Area of the window of the GM-detector: 15.5 cm
2
2
2
2
/5.15
100
2
5.15/100 cmBq
cm
cmxcmBq
resultFinal ==

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
08/2012 Page 17
2.4.3.6 Count up mode
To work with this menu select “CNTUP” mode. For further details see Figure 3.
This mode
is the summation of pulses at a predetermined time interval. The following time
intervals are available: 60, 120, 300, 600 (10 min.), 1800 (30 min.) and 3600 (60
min.) seconds. When you select the count up mode, the display will indicate which
measurement interval has been selected. To keep this measurement interval, wait
three seconds. To change the measurement interval, press the mode button. The
end of the measurement is indicated by an acoustic signal (beep).
Figure 9: Counting up measurement
This menu is very helpful in conveying to students an insight into the statistical correlations
between the radioactivity measurements.
2.4.3.7Food measurement
The instrument is calibrated to measure Cs-137 contamination in liquid or solid food. The
results are only precise if the food is contaminated with Cs-137 and when determined using
the procedure described here.
For more information about radionuclides in food and limits for radiation in various products
see Chapter 2.4.2.
To check radioactivity in food, only two steps are necessary:
1. Determination of the background radiation (control measurement)
2. Food measurement
All measurements must be carried out with the lid closed.
Note: Due to limited sensitivity the instrument is not suitable for verifying official limits
for baby food. With any measurement errors, the manufacturer cannot be held
responsible for any damage.
Sum of the
pulses
Remaining
measurement
time

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 18 08/2012
food mode
(Food)
activate
food
mode
wait 3 sec.
end of food m easurem ent
activate
control
result of the
measurement
after approx. 2 h.
starts autom.
dose rate < 300 nSv/h
press mode button
press
mode button
saved c ontrol measurement
in the memory is OK
no control measurement in the memory
or last measurement is > 300 nSv/h
dose ra te > 300 nSv/h
end off food mea surement
ac tiva te
food
result of the
measurement
after approx. 1.5 h
Figure 10: Food measurement flow chart
Note: You can change the duration of the background (control) measurement using
the MiniTRACE software. But note: the shorter the measurement, the higher the
uncertainty. If you work with a 2-hour measurement time, the result will contain
a 1-sigma uncertainty of 30 % at a background level of 0.1 µSv/h. Be sure to
use an area with a low radiation background, as the lower the background level
the more certain the result. If you do need to measure with a high background
level, try to use a lead castle to shield the measurement system from back-
ground radiation. If your background level is higher than 0.3 µSv/h, food mea-
surement will be blocked and you will get an error message (C ERR) on the
display.
True value At 100 nSv/h At 200 nSv/h At 300 nSv/h
500 Bq/L 500 +/- 30 % 500 +/- 40 % 500 +/- 50 %
Table 3: 1-Sigma statistical uncertainty with different background dose rates
Figure 11: Food measurement

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
08/2012 Page 19
If you measure something other than a liquid the sample will need to be prepared. Solid food
must be mashed and the measuring cup filled precisely to the 0.6 l mark.
Note: The calibration is only valid with the supplied measuring cup and holder.
Otherwise the result can be incorrect.
Measurement steps in detail:
Step 1: Determination of the background radiation (control measurement)
•Place the empty cup and the MiniTRACE CSDF in the measurement device.
•Go to the “Food” menu.
•MiniTRACE will start with the measurement of the background. The display will
indicate how long this measurement will take (in seconds). The result of the
background measurement will be given in the dose rate unit “µSv/h”. You can
interrupt this automatic measurement by pressing the mode button. On the display
the last measured value will be shown. If no value is available in the memory of the
MiniTRACE, you will get an error message (C ERR). In this case, you will need to
conduct a background measurement. We recommend not interrupting background
measurement if you have changed measurement location, or if the last background
measurement was conducted on the previous day.
•Background (Control) measurement takes about 2 hours (7,200 seconds) and is
important for the accuracy of the final result.
Figure 12: Display for background (control) measurement in the “Food” menu
•The end of the background (control) measurement is indicated by a 1 second beep.
On the display “C ok” is shown.
Figure 13: Result display for background (control) measurement in the “Food” menu
For further measurements the background (control) measurement is only required once at
the beginning. Do the background (control) measurement:
•If you change the location of the measurement.
•For measurements on another day.
C = Control
(background)
measurement
Value of the
background dose
rate in µSv/h
(here 0.13 µSv/h)
Remaining measure-
ment time in seconds
(2,700 sec = 45 min)
Result of the
background dose
rate in µSv/h
(here 0.06 µSv/h)

Instruction Manual for MiniTRACE CSDF
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 20 08/2012
Step 2: Food measurement
The measurement result is always given in “Bq/L”. For a result in “Bq/kg” you will need to
carry out two steps:
1. Weigh the food. To do so, subtract the weight of the measurement cup (0.148 kg)
from the total weight.
2. Convert the result from “Bq/L” into “Bq/kg”.
Formula:
[ ] [ ] [ ]
L0.6SaphymofromcuptmeasurementheofVolume*
[kg]foodsolidofNet weight
Bq/LfoodtheofresulttMeasuremen
Bq/kgActivity =
Example:
The net weight of the 0.6 L chopped food sample is 1.2 kg, and the result of the food
measurement was 800 Bq/L.
gBq/k400kg0.6*
L1.2
Bq/L800
Bq/kg ==
•Fill the measurement cup with the food, taking care to reach an exact capacity of
0.6 L by filling precisely to the measuring line in the cup!
•Start the measurement of the sample (food) by pressing the mode button. After
approx. 1 minute (“Wait - - - - -“ is displayed) measurement starts automatically. In the
display the remaining measurement time (in seconds) is given as well as the possible
measurement limit (e.g. L2500) that can be achieved.
Figure 14: Display in the “Food” menu
•The end of the measurement will be indicated by a beep and in the display the result
given in “Bq/L”.
Figure 15: Result display in the “Food” menu
•The mode button enables the user to cancel the measurement and go back to the
menu.
•After the end of measurement or canceling of the menu you can start another
measurement by pressing the mode button again. If the measurement result is higher
than 500 Bq/L, a result is given on the display. If the result measurement is lower
Remaining
measurement
time
Limit
(L = LOWER THAN)
that can be achieved
(e. g. 2,500 Bq/L)
F = Food
measurement
Measurement
unit
Limit
(L = LOWER THAN)
that can be achieved
(e. g. 300 Bq/L)
Table of contents
Other SAPHYMO Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Vertiv
Vertiv Liqui-tect LP3000 Installer/user guide

Bosch
Bosch LR 2 manual

Interacoustics
Interacoustics AT235 Instructions for use

Dolby Laboratories
Dolby Laboratories DM100 user manual

WATANABE ELECTRIC INDUSTRY CO., LTD
WATANABE ELECTRIC INDUSTRY CO., LTD AL-512 SERIES instruction manual

TASKING
TASKING iSYSTEM iC5700 CI BlueBox Hardware user manual