Schick CDR Wireless User manual

CDR Positioning System
User Guide
Schick Technologies, Inc.
30-00 47th Avenue
Long Island City, NY 11101
(718) 937-5765
(718) 937-5962 (fax)
PART NUMBER B1051044REV. B

Copyright 2003 by Schick Technologies, Inc.
All Rights Reserved
August 6, 2003
!
Printed in the United States of America
This document was originally prepared in English
CDR and CDR Wireless are trademarks of Schick Technologies, Inc.
CDR is a registered trademark and is covered by US Patent Numbers
5,912,942 and 6,134,298. Additional patents are pending.
Many of the designations used by manufacturers and sellers to
distinguish their products are claimed as trademarks. Where those
designations appear in this document, and Schick Technologies, Inc.
was aware of a trademark claim, the designations have been printed i
n
caps or initial caps.

CDR Positioning B1051044 Rev. B
i
Contents
1. Introduction...................................................................................1
1.1. Welcome ...................................................................................................................................1
1.2. Your CDR Positioning System .................................................................................................1
1.3. Using Sensor Sheaths................................................................................................................1
1.4. Using Aiming Rings and Arms.................................................................................................1
2. Procedures....................................................................................2
2.1. Sheath Placement (Wired Sensors)...........................................................................................2
2.2. Sheath Placement (Wireless Sensors) .......................................................................................3
2.3. Universal Bite Tab Holder ........................................................................................................4
2.4. Universal Endodontic Tab Holder.............................................................................................5
2.5. Universal Anterior Periapical Tab Holder.................................................................................6
2.6. Universal Bitewing Tab Holder................................................................................................7
2.7. Universal Posterior Periapical Tab Holder................................................................................8
3. Service and Reference................................................................9
3.1. Part Numbers.............................................................................................................................9
3.2. Product Support and Service.....................................................................................................9
3.3. Recommended Date of Use.......................................................................................................9
3.4. Recommended Storage Conditions...........................................................................................9

B1051044 Rev. B CDR Positioning
ii
Safety Notices
Choking Hazard
Proper precautions must be taken when using holders in the CDR Positioning System. To
prevent accidental choking, ensure the holders are correctly positioned and stabilized when
placed inside the patient's mouth. Athough the edges and corners of the Sensor holders and
sheaths have been rounded to increase patient comfort, be aware of any condition that may
cause the patient to gag or choke while the holder is in position.
Do Not Use Holder if Damaged
Sensor holders are carefully inspected before they are shipped to you for normal use. In the
unlikely event of mechanical damage either prior or during shipment, discard the holder
and do not use it.
Always Use Sheaths with Holders
When used correctly, sheaths provide an effective barrier against exposure to bacteria and
infection. Use new Sensor sheaths and holders for each new patient. The sheath is an
integral part of the CDR system and must always be placed over the Sensor before
positioning the holder.

CDR Positioning B1051044 Rev. B
1
1.Introduction
1.1. Welcome
Your new CDR Positioning System provides maximum flexibility in helping you place
your CDR Sensors precisely where you want them for the best digital images. Rounded
edges and corners around the tabs significantly increase comfort for your patients during
positioning, so you can perform your work efficiently and with greater satisfaction.
1.2. Your CDR Positioning System
Each tab type in the CDR Positioning System is color-coded and area-specific, so you can
always find the right tab for the intraoral areas you intend to X-ray. The CDR Positioning
System contains the following tabs:
• Universal Bite Tab Holders (red)
• Universal Endodontic Tab Holders (green)
• Universal Anterior Periapical Tab Holders (white)
• Universal Bitewing Tab Holders (blue)
• Universal Posterior Periapical Tab Holders (yellow)
The CDR Positioning System is designed to accommodate all CDR Sensors, wired and
wireless. Holders are disposable, single-use items. They can be repositioned several times,
if necessary, for the same patient, but they must be disposed of at the end of that patient's
exam.
1.3. Using Sensor Sheaths
Holders must always be used with Sensor sheaths, which provide an effective deterrent
against cross-contamination.
Sheaths are available for Size 1 and Size 2 wired and wireless Sensors. When placing the
sheath over the Sensor, make sure the Sensor is inserted completely into the sheath. The
sheath should fit snugly over the Sensor with the seams of the sheath lined up along the
edge of the Sensor -- not across the flat imaging area.
If needed, the holder can be lifted from the sheath and repositioned. Removing saliva from
the sheath may be helpful if the holder is used repeatedly during the same exam.
1.4. Using Aiming Rings and Arms
Your CDR Positioning System is compatible with other positioning instruments (aiming
rings and arms) that you have purchased previously from Schick Technologies or from any
of our authorized dealers.

B1051044 Rev. B CDR Positioning
2
2.Procedures
2.1. Sheath Placement (Wired Sensors)
We have supplied specific use instructions for each holder in this section, but the first steps
are common to all holders and sheaths used with wired Sensors, as shown below.
Table 1. Overview of Sheath and Holder Placement with Wired Sensors
Step Instruction Illustration
1. Select the appropriate size Sheath
for the Wired Sensor
2. Slide the Sensor into the Sheath
3.
Position the Holder on the Sheath
(Actual position will vary with
holder and area to be X-rayed.
Continue with the following
paragraphs for specific placement
information)

CDR Positioning B1051044 Rev. B
3
2.2. Sheath Placement (Wireless Sensors)
We have supplied specific use instructions for each holder in this section, but the first steps
are common to all holders and sheaths used with wireless Sensors, as shown below.
Table 2. Overview of Sheath and Holder Placement with Wireless Sensors
Step Instruction Illustration
1. Select the appropriate size Sheath
for the Wireless Sensor
2.
A. Slide the Sensor into the
Sheath with the adhesive-side
of the white strip and the back
of the Sensor facing you.
B. Remove the protective layer
from the adhesive strip. Fold
the sheath around the Sensor,
making sure the adhesive side
faces the back of the Sensor,
holding it snugly.
C. Tear the excess length from the
Sheath along the perforated
seam.
3.
Position the Holder on the Sheath
(Actual position will vary with
holder and area to be X-rayed.
Continue with the following
paragraphs for specific placement
information)

B1051044 Rev. B CDR Positioning
4
2.3. Universal Bite Tab Holder
REFERENCE
This tab is color-coded RED. As an example, tab
placement on a CDR wired Sensor sheath is shown below.
Placement on wireless Sensor sheaths is identical.
ILLUSTRATION
Placing Bite Tab on Sheath
(Horizontal) (Step 3) Placing Bite Tab on Sheath
(Vertical) (Step 3)
USAGE
1. Place the sheath on the Sensor, following the steps for
either wired (Table 1) or wireless (Table 2) Sensors.
2. Select bite tab and remove protective strip from
adhesive.
3. For horizontal bitewings, place bite tab horizontally
on the sheath in the center of the Sensor. For vertical
bitewings, place bite tab vertically on the sheath in the
center of the Sensor.
4. Position Sensor in patient's mouth. Make other X-ray
head or positioning instrument adjustments as needed.
When ready, take X-ray.
5. Reposition tab if required.
6. Dispose of tab and sheath at the end of this patient's
exam.

CDR Positioning B1051044 Rev. B
5
2.4. Universal Endodontic Tab Holder
REFERENCE
This tab is color-coded GREEN. As an example, tab
placement on a CDR wired Sensor sheath is shown below.
Placement on wireless Sensor sheaths is identical.
ILLUSTRATION
Placing Endodontic Tab on
Sheath (Anterior) (Step 3) Placing Endodontic Tab on
Sheath (Posterior) (Step 3)
USAGE
1. Place the sheath on the Sensor, following the steps for
either wired (Table 1) or wireless (Table 2) Sensors.
2. Select endodontic tab and remove protective strip
from adhesive.
3. For posterior images, place endodontic tab at the
center of the Sensor. For anterior images, place
endodontic tab at the edge of the Sensor near the
cable.
4. Position Sensor in patient's mouth. Make other X-ray
head or positioning instrument adjustments as needed.
When ready, take X-ray.
5. Reposition tab if required.
6. Dispose of tab and sheath at the end of this patient's
exam.

B1051044 Rev. B CDR Positioning
6
2.5. Universal Anterior Periapical Tab Holder
REFERENCE
This tab is color-coded WHITE. As an example, tab
placement on a CDR wired Sensor sheath is shown below.
Placement on wireless Sensor sheaths is identical.
ILLUSTRATION
Placing Anterior Tab on
Sheath (Maxillary) Step 3 Placing Anterior Tab on
Sheath (Mandibular) Step 3
USAGE
1. Place the sheath on the Sensor, following the steps for
either wired (Table 1) or wireless (Table 2) Sensors.
2. Select anterior tab and remove protective strip from
adhesive.
3. For maxillary and mandibular images, place anterior
tab at the center of the Sensor.
4. Position Sensor in patient's mouth. Make other X-ray
head or positioning instrument adjustments as needed.
When ready, take X-ray.
5. Reposition tab if required.
6. Dispose of tab and sheath at the end of this patient's
exam.

CDR Positioning B1051044 Rev. B
7
2.6. Universal Bitewing Tab Holder
REFERENCE
This tab is color-coded BLUE. As an example, tab
placement on a CDR wired Sensor sheath is shown below.
Placement on wireless Sensor sheaths is identical.
ILLUSTRATION
Placing Bitewing Tab on
Sheath (Horizontal) (Step 3) Placing Bitewing Tab on
Sheath (Vertical) (Step 3)
USAGE
1. Place the sheath on the Sensor, following the steps for
either wired (Table 1) or wireless (Table 2) Sensors.
2. Select bitewing tab and remove protective strip from
adhesive.
3. For vertical bitewings, place bitewing tab vertically on
the sheath in the center of the Sensor. For horizontal
bitewings, place bitewing tab horizontally on the
sheath in the center of the Sensor.
4. Position Sensor in patient's mouth. Make other X-ray
head or positioning instrument adjustments as needed.
When ready, take X-ray.
5. Reposition tab if required.
6. Dispose of tab and sheath at the end of this patient's
exam.

B1051044 Rev. B CDR Positioning
8
2.7. Universal Posterior Periapical Tab Holder
REFERENCE
This tab is color-coded YELLOW. As an example, tab
placement on a CDR wired Sensor sheath is shown below.
Placement on wireless Sensor sheaths is identical.
ILLUSTRATION
Placing Periapical Tab on
Sheath (Mandibular) (Step 3) Placing Periapical Tab on
Sheath (Maxillary) (Step 3)
USAGE
1. Place the sheath on the Sensor, following the steps for
either wired (Table 1) or wireless (Table 2) Sensors.
2. Select periapical tab and remove protective strip from
adhesive.
3. For maxillary images of posterior teeth, place
periapical tab at the bottom edge of the Sensor. For
mandibular images of posterior teeth, place periapical
tab at the top edge of the Sensor.
4. Position Sensor in patient's mouth. Make other X-ray
head or positioning instrument adjustments as needed.
When ready, take X-ray.
5. Reposition tab if required.
6. Dispose of tab and sheath at the end of this patient's
exam.

CDR Positioning B1051044 Rev. B
9
3.Service and Reference
3.1. Part Numbers
When ordering replacement holders, please refer to the following table.
Table 3. Sensor Holder Pack, Kit, and Starter Set Numbers
Part Number Description
B1073051 Universal Bite Tab Holder (50/pkg) red
B1073052 Universal Bitewing Holder (50/pkg) blue
B1073053 Universal Endodontic Holders (50/pkg) green
B1073054 Universal Anterior Holders (50/pkg) white
B1073055 Universal Periapical Holders (50/pkg) yellow
B1070009 Universal Sheaths Size 2 (300/pkg)
B1070008 Universal Sheaths Size 1 (300/pkg)
B1210030 CDR Wireless Size 2 Sensor Sheaths (300/pk)
B1110030 CDR Wireless Size 1 Sensor Sheaths (300/pk)
B1073061 Universal Starter Kit – Size 2
B1073060 Universal Starter Kit – Size 1
B1073062 Universal Holder Starter Set
In the United States, Schick Technologies products are available exclusively through
Patterson Dental Supply, Inc. Call your Patterson representative, local Patterson branch, or
1-800-873-7683 for more information. Outside the United States, contact the authorized
international dealer for Schick Technologies products in your country or region
You may have noticed part numbers on the Sensor holders themselves. These are Schick
Technologies' own internal part numbers and are likely to be different from the order
numbers used by your dealer or representative for Schick Technologies products.
3.2. Product Support and Service
In the event you have questions about your CDR Positioning System or require assistance
from us, please contact your local distributor for our products in the United States, or the
authorized international dealer for Schick Technologies products in your country or region.
3.3. Recommended Date of Use
For best results, check the label on the holder bag and use the holders by the date specified.
3.4. Recommended Storage Conditions
For best results, check the label on the holder bag and store the holders within the specified
temperature range.

B1051044 Rev. B CDR Positioning
10
NOTES
Other manuals for CDR Wireless
2
Table of contents
Other Schick Accessories manuals