Schleich GLP2-i/e User manual

SCHLEICH
TECHNOLOGY IN ADVANCE
GLP2-i/e
Operating manual

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 2
File name Q:\WINWORD\USER\ML\BAGLP2i\BAGLP2i_e.doc
Revision 1.3
GLP2 software version 3.107 and over
Editor J. Strauch / M. Danne / G. Schellnack
Date of print-out 19.03.2009 17:41

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 3
Index
1.0 Introduction................................................................................................ 7
1.1 Technical data of the tester....................................................................................................7
1.2 Technical data of the test methods..........................................................................................7
1.2.1 PE test ..........................................................................................................................7
1.2.2 Continuity test (2 Wire) ...................................................................................................7
1.2.3 Insulation test ................................................................................................................7
1.2.4 Function test ..................................................................................................................8
1.2.5 Leakage current test .......................................................................................................9
1.2.6 Leakage current test at medical devices (corresponds to figure 16 of EN60601) .....................9
1.2.7 High voltage test .......................................................................................................... 11
1.2.8 Residual votlage test ..................................................................................................... 11
1.2.9 Resistance test ............................................................................................................. 11
1.2.10 FI-Schutzschalter Prüfung ............................................................................................ 12
1.3 Range of application............................................................................................................ 13
1.4 Liability ............................................................................................................................. 14
2.0 Safety ....................................................................................................... 15
2.1 Safety symbol .................................................................................................................... 15
2.2 Attention ........................................................................................................................... 15
2.3 Regulations for safety at work .............................................................................................. 15
2.4 Important information for users of high-voltage testing equipment ........................................... 17
2.4.1 Usage.......................................................................................................................... 17
2.4.2 Working with high-voltage test pistols ............................................................................. 17
3.0 General information .................................................................................. 19
3.1 Introduction ....................................................................................................................... 19
3.2 Range of application............................................................................................................ 19
3.3 Copyright........................................................................................................................... 19
4.0 Introduction.............................................................................................. 20
4.1 General explanations........................................................................................................... 20
4.2 Basic test process ............................................................................................................... 20
4.3 Operating elements / GLP2-i ................................................................................................ 21
4.4 Operating elements GLP2-e.................................................................................................. 22
4.5 External operating elements................................................................................................. 23
5.0 General programming of the tester ........................................................... 24
5.1 Operation of the touch panel ................................................................................................ 24
5.1.1 Main menu................................................................................................................... 24
5.1.2 Input menu.................................................................................................................. 24
5.1.3 Testing menu ............................................................................................................... 25
5.1.4 Menu inquiry / acknowledgement.................................................................................... 26
5.1.5 Programming menu ...................................................................................................... 27
5.2 General information / after switching on ................................................................................ 28
5.3 Load test program aus der Datenbank ................................................................................... 31
5.4 “go/no go counter”.............................................................................................................. 32

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 4
5.5 Programming ..................................................................................................................... 33
5.5.1 Test programs.............................................................................................................. 34
5.5.1.1 General information ................................................................................................ 34
5.5.1.2 Name of test programs............................................................................................ 35
5.5.1.3 Storing test programs ............................................................................................. 36
5.5.1.4 Deleting test programs............................................................................................ 37
5.5.1.5 Loading test programs............................................................................................. 38
5.5.2 Set values.................................................................................................................... 38
5.5.3 Test program ............................................................................................................... 39
5.5.3.1 General information on test programs ....................................................................... 40
5.5.3.2 PE / ISO special test ............................................................................................... 40
5.5.3.3 Visual test ............................................................................................................. 41
5.5.3.4 Add. process .......................................................................................................... 43
5.5.4 General data ................................................................................................................ 46
5.5.5 Interfaces .................................................................................................................... 53
5.5.6 Printer......................................................................................................................... 55
5.5.6.1 Example of a print-out without definitions (short) ....................................................... 58
5.5.6.2 Example of a print-out with definitions (long)............................................................. 58
5.5.7 Password..................................................................................................................... 59
5.6 Single test ......................................................................................................................... 60
5.6.1 Special features of the single test high voltage AC ............................................................ 61
5.7 Information........................................................................................................................ 62
5.8 Time and date (optional) ..................................................................................................... 63
5.9 Connection diagrams........................................................................................................... 63
5.10 Status messages, error messages ....................................................................................... 64
6.0 Test methods ............................................................................................ 69
6.1 PE resistance test ............................................................................................................... 69
6.1.1 Explanations on the PE resistance test............................................................................. 69
6.1.2 Programming of the PE resistance test -GLP2-i/e ............................................................. 77
6.1.3 Examples for application of PE resistance test .................................................................. 80
6.1.3.1 Manual test process of the PE resistance test ............................................................. 80
6.1.3.2 Automatic test process of the PE resistance test ......................................................... 81
6.1.4 Schematics of the PE resistance test ............................................................................... 82
6.2 Insulation resistance test ..................................................................................................... 83
6.2.1 Explanation of the insulation resistance test ..................................................................... 83
6.2.2 Programming of the insulation resistance test -GLP2-i/e ................................................... 89
6.2.3 Examples for application of insulation resistance test ........................................................ 91
6.2.3.1 Manual test process of the insulation resistance test ................................................... 91
6.2.3.2 Automatic test process of the insulation resistance test ............................................... 91
6.2.4 Schematics of the insulation resistance test ..................................................................... 92
6.3 Function test ...................................................................................................................... 94
6.3.1 Explanation of the function test ...................................................................................... 94
6.3.2 Programming of the function test -GLP2-i/e .................................................................... 95
6.3.3 Programming of the power test -GLP2-i .......................................................................... 98
6.3.4 Programming of the apparent power test -GLP2-i/e.......................................................... 99
6.3.5 Programming of the COSφ test -GLP2-i/e ..................................................................... 100
6.3.6 Programming of the visual test -GLP2-i/e...................................................................... 101
6.3.7 Examples for application of function test........................................................................ 103
6.3.8 Schematics of the function test..................................................................................... 104
6.4 Leakage-current test ......................................................................................................... 111
6.4.1 Explanation of the leakage-current test ......................................................................... 111
6.4.2 Programming of the leakage-current test according to EN60990 ....................................... 113
6.4.3 Schematics of the leakage-current test according to EN60990 .......................................... 116
6.4.4 Programming of the leakage-current test according to EN60601 ....................................... 118

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 5
6.4.5 Schematics of the leakage-current test according to EN60601 .......................................... 123
6.5 Substitute leakage current test........................................................................................... 127
6.5.1 Explanation of the substitute leakage current test........................................................... 127
6.5.2 Programming of the substitute leakage-current test -GLP2-i/e......................................... 128
6.6 High-voltage test .............................................................................................................. 130
6.6.1 Explanation of the high-voltage test .............................................................................. 130
6.6.2 Programming of the high-voltage test -GLP2-i/e ............................................................ 141
6.6.3 Adjustment of the high voltage..................................................................................... 145
6.6.3.1 Manual adjustment of the high voltage.................................................................... 145
6.6.3.2 Automatic adjustment of the high voltage................................................................ 145
6.6.4 Examples of application of the high-voltage test ............................................................. 146
6.6.4.1 Manual test process of the high-voltage test ............................................................ 146
6.6.4.2 Automatic test process of the high-voltage test ........................................................ 146
6.6.5 Schematics of the high-voltage test .............................................................................. 147
6.6.6 Option ARC detector for high voltage test AC .................................................................159
6.6.6.1 Programming of the ARC detector test..................................................................... 160
6.7 Continuity test.................................................................................................................. 161
6.7.1 Explanation of the continuity test.................................................................................. 161
6.7.2 Programming of the continuity test -GLP2-i/e................................................................ 162
6.8 Resistance test ................................................................................................................. 164
6.8.1 Explanation of the resistance test ................................................................................. 164
6.8.2 Programming of the resistance test -GLP2-i/e................................................................ 165
6.9 Mathematical test ............................................................................................................. 167
6.9.1 Explanation of the mathematical test ............................................................................ 167
6.9.2 Programming of the mathematical test .......................................................................... 168
6.10 Turn-to-turn tests ........................................................................................................... 170
6.10.1 Turn-to-turn tests at armatures .................................................................................. 170
6.10.2 Programming of turn-to-turn tests .............................................................................. 171
6.11 Rotor test....................................................................................................................... 173
6.11.1 Explanation of the rotor test....................................................................................... 173
6.11.2 Programming of the rotor test .................................................................................... 174
6.12 FI-Schutzschalter Prüfung ................................................................................................ 176
6.11.1 Erklärung der Prüfung eines Fehlerstrom-Schutzschalters .............................................. 176
6.11.2 Erklärung der Programmierung der FI-Schutzschalterprüfung ......................................... 177
7.0 Connections at the GLP2i/e..................................................................... 179
7.1 Connection of the test object at rear panel........................................................................... 179
7.2 Connection of the test object at front panel.......................................................................... 180
7.3 Control plug ..................................................................................................................... 181
7.4 Connections at the rear panel, partly optional....................................................................... 182
7.5 PE test probe at the front panel .......................................................................................... 183
7.6 Serial connection cable between tester GLP1-e and GLP2-i/e and PC ....................................... 184
7.7 Connection of the RS485 network (optional)......................................................................... 185
7.8 Parallel printer connection LPT1 (option)............................................................................. 186
8.0 Interface protocol ................................................................................... 187
8.1 General information .......................................................................................................... 187
8.2 Simplified interface protocol ............................................................................................... 188
8.3 Protocol for reading in the actual values .............................................................................. 188
8.4 Helps for special characters................................................................................................ 189
9.0 Annex.............................................................................................................................. 190

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 6
Declaration of Conformity
The undersigned declare on behalf of
Schleich Antriebs-und Steuerungstechnik GmbH
An der Schleuse 11
D-58675 Hemer
That the tester
GLP2-i/e
to which this declaration relates
is in compliance with the below mentioned
EEC Directives (including all amendments)
73/23/EEC Low Voltage Directive
89/336/EEC Electromagnetic compatibility
98/37/EG | EN60204 Machinery Directive
and that following standards have been applied
EN61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use
EN61326-1 EMC requirements / Electrical equipment for measurement, control and laboratory use
Hemer, 6.6.2001
Dipl.Ing. M. Lahrmann Karl-Dieter Schleich
Geschäftsleitung Geschäftsleitung

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 7
1.0 Introduction
1.1 Technical data of the tester
Supply voltage : 230V AC (option: 400V AC)
Protection by fuses, GLP2-i : 10Amp (option: 16A, 32A, 64A)
Password upon delivery : 123
1.2 Technical data of the test methods
The test methods listed below are not necessarily integrated into the testing device. Es ist Möglich, dass
eine Variante einer Prüfmethode eine Andere ausschließt.
1.2.1 PE test
Display: Ohm or Volt
Measuring range: 0 ... 1 Ohm / 0 ... 10V
Resistance test: e.g. according to EN60335-1
Voltage drop test: e.g. according to EN60204-1
Resolution: 0.001 Ohm / 0.01V
Accuracy: 0.5% of measuring range
Limit values: 0.01 ... 1 Ohm -freely programmable
Test current: 10A; (option: 30A short-time operation, 30A continuous operation)
Test current -special feature: 1...30A / the test current can be programmed in steps of 1A
Test current regulation: electronic constant-current source and monitored
Test voltage: <6VAC or <12VAC
Frequency of test voltage: 50Hz (depending on the mains of the tester)
Test time: 0.1 ... 60 sec. -freely programmable
(0.1 ... 1 sec. -freely programmable / short-time operation)
1.2.2 Continuity test (2 Wire)
Measuring range: 1 Ohm ... 10K Ohm / accuracy related to the measuring value = 1.5% +- 1 Ohm
Measuring range: 10K Ohm ... 100K Ohm / accuracy related to the measuring value = 2.5% +- 100 Ohm
Test points: L-N / L-PE / N-PE / PE1-probe
1.2.3 Insulation test
Display: in MOhm
Measuring range: 0 ... 1000 MOhm
Resolution: 0.1 Mohm (value < 100 Mohm)
Resolution: 1 Mohm (value > 100 Mohm)
Accuracy: 0..99Mohm 1% of measuring value (U-min 500V)
Accuracy: 100..499Mohm 1,5% of measuring value (U-min 500V)
Accuracy: 500..1000Mohm 2.5% of measuring value (U-min 500V)
Limit values: 0.1 ... 1000 MOhm -freely programmable
Test current: max. 3mA
Test voltage: 30V ... 1000V / programmable in steps of 10V
Test voltage regulation: electronically regulated and monitored
Test time: 0.1 ... 60 sec. -freely programmable
Tests: L+N <->PE, L<->PE, N<->PE, L<->N, probe<->PE, probe<->L+N, probe<->L,
probe <->N, (L+N means that, for the test, L and N are automatically connected inside the tester.)

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 8
1.2.4 Function test
The function test is either 1-phase or 3-phase. The operating manual deals with the standard variants. For
the accuracy of function tests that are not indicated here, please refer to our offer.
Function test, 1-phase:
Test voltage: 230V 1-phase AC potential-free
Test voltage fluctuation: depending on mains supply
Voltage measuring range: 260 Veff AC, DC
Resolution: 0.1 V
Measuring: TRMS
Test voltage for leakage current: 230V +10%
Current measuring range: 0 ... 2(5, 10, 16) A AC
Resolution: 0.01 A
Accuracy: 0.25% of measuring range
Measuring: TRMS
Set values: 0.01 ... 2A - freely programmable
Correction: with mathematical correction of the current consumption
in case of mains voltage fluctuations, connectable
Apparent -active power measuring range: 0... 1150VA (W)
Resolution: 1 VA (W)
Accuracy: 1% of measuring range
Correction: with mathematical correction of the power in case of mains
voltage fluctuations, connectable
cosPhi measuring range: 0...1
Resolution: 0.01
Test time: 0.1 seconds ... 24 hours -freely programmable
Starting time: freely programmable
Short-circuit test: before and during the function test
Display: continuous display of voltage and current during the test
Tolerances: in +% and -% or as +limit value and -limit value
Function test, 3-phase:
Test object : 1-phase or 3-phase
Test voltage: 400V AC 3-phase current (mains voltage)
Voltage measuring range: 250Veff AC against neutral lead
Resolution: 1V
Measuring: TRMS / measured per phase
Current measuring range: 0 ... 16A AC
Resolution: 0.01A
Accuracy: 0.5% of measuring range
Measuring: TRMS / measured per phase
Set values: 0.1 ... 16A freely programmable
Test methods: current consumption, apparent power (between two phases), active power (between two
phases), cosPhi
cosPhi test: 0...1 / measured per phase
Resolution: 0.01
Test time: 0.1 seconds ... 24 hours -freely programmable
Short-circuit test: before and during the function test

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 9
1.2.5 Leakage current test
Leakage current: max. 30mA
Resolution: 0.01mA
Accuracy: 0.2% of measuring range
Testing according to EN60990 requires 3 measuring circuits for the evaluation.
1: unweighted leakage current (EN60990 (5) Measuring Equipment Fig.3)
2: weighted according to perception and reaction (EN60990 (5) Measuring Equipment Fig.4)
3: weighted after release (EN60990 (5) Measuring Equipment Fig.5)
Touch-current frequency: max. 500Hz
Evaluation: r.m.s value (unweighted) or peak value
Warm-up time: 0 seconds ... 24 hours
Max. current at "warm leakage current test": like function test
IMPORTANT
In case the test object needs more than 5A, it is necessary to order the increased current additionally.
TESTING STANDARDS
EN60990 / VDE 0106
EN60335-1 / VDE 0700-1 Measuring at household devices
EN60950 / VDE 0805 Measuring at office devices (communication devices)
EN60598-1 / VDE 0711-1 Measuring at lamp
EN60990 (5) Measuring Equipment Fig.3
EN60990 (5) Measuring Equipment Fig.4
EN60990 (
5) Measuring Equipment Fig.5

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 10
1.2.6 Leakage current test at medical devices (corresponds to figure 16 of EN60601)
Leakage current: max. 30mA
Resolution: 0.01mA ... 100nA (depending on measuring value)
Accuracy: 0.1% of measuring range
Touch-current frequency: max. 500Hz
Test points: 8 (as patient connections or enclosure parts)
Test probe connections: 2
FE point: 1 (ground)
Max. test current at the connections for the current supply: like function test (however max. 16A)
TESTING STANDARDS
EN60601-1 / VDE 0750-1 Measuring at medical devices
With this article number, it is possible to test the following figures of the standard:
figure 16, figure 20, figure 23, figure 26, figure 27
For automatic testing according to the following figures (S10, S7 or MD2), it is necessary to order 2
measuring point extensions (article number 4000271): figure 17, figure 19
For automatic testing according to the following figures (interruption of the current supply with S2 and S3),
the option: cut-off relay (article number 4000323 ) needs to be ordered additionally: figure 17, figure 19
For automatic testing according to the following figures (S9 and additional test voltage), the 2nd test
voltage (article number 4000324) needs to be ordered additionally: figure 18, figure 19, figure 21, figure
22, figure 24, figure 25
Please note that the 2nd voltage source can only be connected to the 1st measuring point.
It is not possible to connect the source automatically in the form of a matrix to further points.

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 11
1.2.7 High voltage test
High voltage AC
High voltage: 0 ... 2750V (5500V)AC - supply-sinusoidal AC voltage (true r.m.s.)
Resolution: 1Veff
Voltage monitoring: both minimum and maximum voltage
Test current: >200mA (according to VDE / IEC regulations)
beginning at approx. 1300V
Current measuring: apparent current or active current
Current display: apparent current or active current
Resolution: 0.1mA
Limit current evaluation: apparent current or active current
Limit current pre-setting: 0.1 ... 100mA -freely programmable, as soon as the limit current is exceeded,
there will be a switch-off with optical and acoustical indication
Response time: < 50ms
Burning function: burning without limit current pre-setting, protection
against overload by thermo switch
Minimum current: 0 ... 100mA -freely programmable
Nominal power: >500VA (according to VDE / IEC regulations)
beginning at approx. 500V
Test time: 0.1 seconds ... 24 hours and without time allowance
Switching-on: at zero crossing of mains voltage
Safety: key switch, external safety contact and Emergency Stop
In case of voltages over 5500V, please refer to the offer for technical details.
High voltage DC
For the data of the high-voltage test DC, please refer to the technical details of the offer.
1.2.8 Residual votlage test
automatic residual voltage test
test voltage: max. 260V
test: 1-phase
measuring delay: 5 sec. (freely programmable 1...10sec.)
test repeat: 10 (freely programmable 1...100)
time of test: random, no relation to phase position
graphical indication of the voltage: no
manual residual voltage test
test voltage: max. 260VAC
measuring leads: 4 pcs. incl. 4 pcs. alligator clips
length of measuring lead: 2m
measuring delay: 5 sec. (freely programmable 1...10sec.)
time of test: random, no relation to phase position
graphical indication of the voltage: no
1.2.9 Resistance test
Measuring range: 4ranges (0...10Ohm / 0...100Ohm / 0...10Kohm / 0...100KOhm)
accuracy: 0.5 % of range
4-wire resistance measurement: yes

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 12
1.2.10 FI-Schutzschalter Prüfung
Messbereich Strom: 0 –500mA (AC RMS)
Messfehler: +/- 1mA
Messbereich Zeitmessung: 10ms – 2000ms
Messfehler: + 20ms
Da es sich bei dieser Prüfung um eine AC Prüfung handelt und das aufschalten des Relais nicht mit der
Netzspannung synchronisiert ist, entsteht dabei ein Messfehler von 0-20ms.

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 13
1.3 Range of application
The safety and function tester GLP2-i/e is designed for testing voltage-free 2-poles only (unless it is a 3-
phase model or a model made to specification).
Any usage other than this is regarded as not appropriate.
For any damage resulting from inappropriate usage, the manufacturer/supplier can not be hold liable; the
user bears all risks.
Before the tester is switched on for the first time or after it has been transported, it needs to adapt to the
ambient temperature first. Please unpack the device and put it on a table. You should wait at least 30
minutes before switching it on.
The device must not be exposed to the condensation of water.
Attention!
We are pointing out that for the daily check of the testing device regarding its correct
functioning and accuracy, it is necessary to use a calibrated black box (test dummy). This black
box is connected to the test leads of the testing device and it is tested, whether the data of the
black box are the same as the measured values. In case the testing device has various
measuring ranges, it is necessary that for each measuring range the respective black box is
used.
In order to make sure that the comparison data are correct, the black box must be checked
annually.
For testing with the black-box, the testing device has a suitable program. Right after switching
on the testing device, this program asks the person who is doing the test to do the self-test with
the black-box. Only after the self-test has been completed correctly, the device enables itself for
further tests.
Please note: In the SetUp of the testing device it is possible for the user to deactivate the self-test.
However, doing this is not in accordance with the intended purpose of the device. If the self-test is
deactivated, there will also be a message on the display.
Attention!
We should like to point out: Every year and after every modification (repair), which could influence the
measuring equipment, the testing device must be calibrated and, if necessary, adjusted. The
supplier/manufacturer will test the device regarding its accuracy and confirm the calibration in a certificate.
Attention!
For test methods, their measuring ranges and accuracies, please refer to the enclosed calibration certificate.

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 14
1.4 Liability
The testing device has been designed according to the latest developments in technology, checked and is
therefore operationally reliable. In case the tester is maintained or repaired improperly by persons not
authorized by Schleich GmbH, the liability will in any case be passed on to the owner. The same applies if
the testing device is operated in any other way than described in this manual. (See chapter Operative
range and appropriate usage).
The user is obliged to operate the tester only in perfect condition.
Messrs. Schleich GmbH is liable for mistakes and omissions, with the exception of further claims, only
within the framework of the guarantee commitments in the order confirmation (respectively the
commitments in the offer).
In view of the permanent development and improvement of our products, we reserve the right for technical
modifications at any time. Such modifications, mistakes and misprints do not imply a claim for damages.
Furthermore, only original Schleich GmbH spare parts and accessories may be used. For any damage
arising from non-observance of the instructions in this operating manual Messrs Schleich GmbH do not
assume liability. Only the German original of this operating manual is binding. The above information does
not extend the liability-and guarantee agreements of the conditions of sale and delivery of Messrs.
Schleich GmbH.

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 15
2.0 Safety
2.1 Safety symbol
You will find this symbol with all safety precautions and warnings in this operating manual where danger to
life and limb exists. Please follow these warnings and act particularly careful in these cases. All safety
precautions and warnings need to be passed on to other users. Apart from the information in this operating
manual, the generally valid safety and accident prevention regulations must be observed.
2.2 Attention
Attention!
You will find this Attention ! sign at places in this operating manual, which must be observed very carefully,
so that the guidelines, regulations and indications and the correct operating process, are observed. This is
also necessary to prevent that the testing device and/or other parts of the equipment are damaged or
destroyed.
2.3 Regulations for safety at work
The following regulations for safety at work are to be strictly adhered to:
The safety and function tester GLP2-i/e has been designed according to the latest developments in
technology and is operationally reliable. However, should unqualified personnel use this testing device, in
an inappropriate way or for improper use, this testing device can be dangerous.
Each person in the factory of the user that deals with the installation, the operation and the maintenance of
the testing device GLP2-i/e must have read and understood the complete operating manual and particularly
the information on safety. We recommend that the user has this confirmed in writing (see annex).
The safety and function tester GLP2-i/e serves for testing voltage-free test objects, only. Any use other
than that is not in accordance with the application of this device. Any damage resulting from improper use
will not be covered by the supplier/manufacturer. In such cases the user bears all risks.
It is also necessary to follow the installation-, operating-and maintenance/repair conditions specified by
the manufacturer/supplier.
The safety and function tester GLP2-i/e may only be operated, maintained and repaired by authorized,
trained and instructed personnel, which must have participated in a special training covering all possible
risks.
Upon any kind of work concerning commissioning, operation, repairs and maintenance, the switching-off
and switching-on procedures indicated in the operating manual must be considered.
Any working manner that impairs the safety of the testing device must be refrained from.

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 16
The operator is obliged to report any change at the testing device, that impairs the safety, immediately to
his superior (e.g. to the person in charge for measuring devices) or to Messrs. Schleich GmbH.
Unauthorized modifications and changes, which impair the safety of the testing device, are not allowed. All
work at the testing device must be done when it is voltage-free and switched-off.
Before commissioning and after maintenance/repairs it must be checked, whether all protection devices are
installed properly and functional. Protection devices must never be removed.
For the operation of the tester GLP2-i/e, please take particular note of the regional safety regulations and
regulations for the prevention of accidents.

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 17
2.4 Important information for users of high-voltage testing equipment
Caution! High voltage!
Touching high voltage can result in loss of life! Equipment, measure leads and accessories must be in
perfect condition and need to be checked regularly. The equipment may only be used in dry state and in dry
rooms. Only suitable qualified personnel or trained personnel under supervision of specialists must work on
this equipment.
2.4.1 Usage
The relevant regulations, above all VDE 0104 (EN 50191), must be followed strictly. Every day before
starting work, equipment, accessories and supply line must be visually examined to ensure their perfect
condition. Every user must be familiar to the safety instructions in this manual and should be aware of all
possible risks. He is obliged to protect himself and others against possible accidents. Because of the
necessary personal responsibility, all users must be selected carefully and need to be trained in detail at
least once a year.
The training must be put on record. Maintenance and alterations of the tester are reserved to the
manufacturer. (Unauthorized alterations endanger the safety). For checking, the test pistols can be sent
(postage paid) to the manufacturer.
2.4.2 Working with high-voltage test pistols
High-voltage test pistols are always delivered with connected high voltage cable.
Variant 1, testing with 2 safety test pistols:
You will have full accident protection if you always use the test pistols in pairs, one in each hand. It is very
dangerous to connect the test object tightly at one pole, working with only one test pistol or to take two
test pistols in one hand!
Variant 2, testing with one safety test pistol:
If the high-voltage test is performed with one high-voltage test pistol and one firmly connected pole, it is
absolutely necessary that a suitable safety device like a hand-switch cover is available for the second hand.
The hand switch, which must be activated during the test, ensures that the operator cannot touch the high
voltage. “Putting your second hand in the pocket” is hazardous and prohibited!
Do not touch the test object while it is being supplied with high voltage by the test pistols! Its insulation
could break down during the test. Do not test on metal surfaces (the voltage could be carried over).
During non-working time the high voltage must be switched off!
The test pistols are well insulated, but as protection-insulated devices not grounded. Therefore, the user
may feel a tingling or a slight vibration if, after switching-on the voltage, he doesn't hold the test pistol
tightly, or has a very short distance from the test pistol or from the supply cable. You can feel the electric
field. A capacitive blind current of max. 0.08 mA can be flowing via the capacity between the voltage-
carrying lead and the hand. (According to VDE 0104 / EN 50191, there is no danger up to 3.0 mA AC und
12mA DC!).

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 18
Only use test pistols in perfect condition with new-value cables! Check test pistols and cable for obvious
damage every day before starting work! Faulty test pistols (even with the slightest fault at the device or at
the lead!) have to be put out of operation immediately and must be sent back to the manufacturer for
repair!
Attention!
For reasons of personal safety, a 3mA AC or 12mA DC test current limitation at the high-voltage test can
only be effected by means of hardware integrated in the testing device. An adjustment by software of the
maximum test current to 3mA or 12mA, at which the device is to switch off, does not serve for personal
safety, because the current may, for a short time, reach considerably higher values.
Only a limitation of the maximum possible current through hardware is allowed for personal protection.
This is done optionally inside the testing device by means of a resistor. Only in this case, the otherwise
necessary protection measures, as described in the VDE 0104 (EN 50191), can be reduced.
Test pistols with integrated push-button
If test pistols with integrated push-button are used, the test is started automatically by
activating the push-buttons of both test pistols. For this purpose, the lever at the test pistol needs to be
operated first, so that the test probe tip in the insulating tube is activated. Now the test probe tips must be
pressed on the test object. The push-buttons in the test pistols must be activated for the complete duration
of the test; otherwise, the test will be interrupted.
Fault analysis
If test pistols with integrated start switch or push-button are used and the test doesn’t start, for
finding out what the problem is, the test probe of the test pistol must be touched under no circumstances.
For checking the test pistols, they must always be disconnected from the tester. Only when they are
completely disconnected from the tester, they can be checked.

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 19
3.0 General information
This operating manual has been worked out with the utmost care. However, Messrs. Schleich GmbH cannot
be hold liable for possible mistakes in this operating manual and their consequences. We reserve the right
for technical changes. Messrs. Schleich GmbH shall not be liable for any mistakes in this manual.
3.1 Introduction
This operating manual has been written with the intention to be read and understood in all detail by those
responsible for the safety and function tester GLP2-i/e.
The complete documentation should always be kept within easy reach of the testing device. Details of great
importance for operating the tester are pointed out to in this operating manual.
Only the knowledge of this operating manual will ensure that mistakes with this testing device can be
avoided and guarantees trouble-free operation. It is therefore of great importance that this operating
manual is known to the relevant persons.
We recommend reading this operating manual carefully, because we will not accept liability for any damage
or operating trouble resulting from disregarding this operating manual. Should you nevertheless have
difficulty with the operation, please do not hesitate to contact our service department.
3.2 Range of application
The safety and function tester has been designed to be operated only within the range of application found
under point 1.1, technical data. This testing device may only be operated within the indicated range of
application; otherwise Messrs. Schleich GmbH accept no liability.
3.3 Copyright
Messrs. Schleich GmbH holds the copyright for this operating manual. This operating manual has been
written for the operating and supervising personnel of the tester. It contains instructions and technical
illustrations that may neither completely nor partly be copied, distributed or used for competition purposes
or given to a third party.

Safety and Functional Tester GLP2-i/e
During the test, test object and test leads must not be touched!
The safety measures required by law must be adhered to!
Schleich GmbH * D-58675 Hemer * 0049 / (0)2372 / 9498-0 * 0049 / (0)2372 / 9498-99 * http://www.schleich.com * [email protected] 20
4.0 Introduction
4.1 General explanations
The tester GLP2 i/e has been designed for 2-pole test objects. I.e., the test object has only two connections
for voltage and maybe one PE connection.
As an option, it is also possible to test multi-pole test objects. In this case the tester would, for example, be
equipped with a matrix.
Depending on the configuration of the device, the test object can be connected either at a PE socket at the
front and/or at an industrial connection at the rear panel.
At the industrial connection, the test object is connected via a connection box or via a direct lead.
If, despite the industrial connector at the rear panel, the socket at the front panel is to be used, it will be
necessary to connect a bridge plug at the back. At this bridge plug, the contacts 6, 7 and 8 need to be
bridged. Otherwise it is not possible to perform a PE test at the front socket.
The PE test probe can be connected at the front panel via 19-pole round plug. Alternatively, it is possible to
connect a probe or a PE to the connection box.
The measuring process is based on the fully-or half-automatic process of the individual test methods one
after the other. The order of the process is determined by the user.
Example of a process:
1. PE test
2. Insulation resistance test
3. High-voltage test (if the device has additional high voltage / option)
4. Current-consumption test
The measuring process is therefore based on an automatic sequence of individual test steps.
It is possible to program up to 200 test steps in up to 200 individual test processes. The test processes are
loaded out of the tester database or the remote-control software NetCom XI transmits them via the
interface.
4.2 Basic test process
A test process always follows the same basic pattern. The test process can be divided into the following
steps:
1. put test object into fixture or connect it
2. depending on the extent of danger, close protection cover or press 2-hand start
3. the individual test steps are executed successively (perhaps scanning PE points)
4. open protection cover (see point 2)
5. remove test object from the fixture or disconnect it
6. perhaps print out stickers or protocols
7. perhaps data transfer to the PC
After this, the above described test process is repeated.
Table of contents
Popular Test Equipment manuals by other brands

Indu-Sol
Indu-Sol PRolinetest user manual

Chauvin Arnoux
Chauvin Arnoux C.A 745 user manual

Seaward
Seaward Live Line Tester operating instructions

Rohde & Schwarz
Rohde & Schwarz CTS50 operating manual

Fluke Calibration
Fluke Calibration 4322-SYS-NAVAIR Service manual

Metronelec
Metronelec MENISCO ST88 NEO user manual