ScienceCube KDS-1037 User manual

KDS-1037
- 1 -
High Concentration
CO2Gas Sensor
Figure 1. High Concentration CO2Gas Sesor
Description
The ScienceCube CO2gas sensor is used to monitor gas concentration carbon dioxide in variety of
biology and chemistry experiments. It measures in range of 0 to 10 volume % (100,000 ppm) by
recording the amount of infrared radiation absorbed by carbon dioxide molecules.
The sensor uses a hot metal filament as an infrared source (IR), which is located at one end of the
sensor tube.At the other end of the sensor tube is an infrared sensor that measures how much
radiation gets through the sample without being absorbed by the carbon dioxide molecules. The
detector measures infrared radiation absorbed in the narrow band centered at 4.26㎛. The greater
concentration of the absorbing gas in the sampling tube, the less radiation will make it from the source
through the sensor tube to the IR detector.
The temperature increase in the infrared sensor produces a voltage that is amplified and read by an
interface.
CO2gas moves in and out of the sensor tube by diffusion through the six vent holes in the sensor
tube. When the sensor is collecting data, the IR source blinking on and off – it takes a new reading
about once every 3 second.
Suggested experiments
Measuring CO2levels (respiration) from small animals and insects.
Monitoring CO2changes in a plant terrarium during photorespiration and photosynthesis in
light/dark (combination with the light sensor).
Measuring CO2levels during cellular respiration of peas or beans.
Monitoring production of CO2during chemical reactions
Na2CO3+ HCl CO2+ H2O + NaCl
Monitoring CO2 consumption during chemical reactions:
2NaOH + CO2Na2CO3+ H2O
Important : The sensor tube can never be placed into any liquids. The sensor is
intended only for measuring gaseous, not aqueous, CO2concentration.

KDS-1037
- 2 -
Measuring CO2levels in classrooms. The air in a classroom during the lesson can reach as
high as 1000 to 1200 ppm CO2.
Determining the rate at which CO2diffuses through a gas diffusion tube.
Monitoring production of CO2during fermentation or respiration of sugars.
How it Works
The NDIR (Non Dispersive Infrared) type utilizes the phenomena that an asymmetry gas molecule,
consisting of 3 atoms or more, absorbs the energy from its inherence vibration selectively.
For example, H2O gives 1.4 absorbance at 1.9㎛wavelength, while SO2, CO2, CO, NO give a
maximum absorbance at 4.0㎛, 4.3㎛, 4.7㎛, 5.3㎛wavelengths, respectively.
CO2gas concentration is calculated from the electric signal that was transformed from the
absorbance that changes with the gas concentration. The absorbance is measured at the course of a
light with 4.3㎛wavelength, emitted from Infrared lamp, reaches gas sensor position.
This is the type that overcomes the shortcomings of a Single type and generally called as Dual type.
In this type, two sensors are packed in one case, so that one terminal is used for measuring standard
while the other is used for measuring infrared absorbance change that was affected by CO2
concentration.
Standard value is the absorbance that was measured at the wavelength, in which the absorbance
does not affected by CO2and other gases.
Calibration
For many measurements it will not be necessary
to calibrate the CO2gas sensor. The sensor has
measurement range between 0 to 100,000 ppm1.
The output voltage is between 0 and 4 volt.
Most of the time this calibration is sufficient.
For more accurate measurements, however, the
sensor can be calibrated at one known CO2level
using the calibration button on the sensor box.
Default calibration graph of CO2gas sensor:
value (ppm) = 25,000 * Vout (V)
Coefficients of the calibration function:
a = 25,000; b = 0
Figure 2. Calibration of the CO2gas sensor.
1The CO2sensor measures in units, parts per million, or ppm. In gaseous mixtures, 1 part per million
refers to 1 part by volume in 1 million volume units of the whole. This unit can be recalculated to
percent by dividing a value in ppm by 10,000. 100,000 ppm is equal to 10 %.
The level of CO2in the earth’s troposphere has gradually increased from 317 ppm in 1960 to current
levels of nearly 370 ppm. Exhaled human breath has CO2concentration of about 50,000 ppm.
CO2 Concenturation (ppm)
Sensor Output (V)
100,000 ppm
0 ppm 4.0V
0V

KDS-1037
- 3 -
Sensor calibration
Place the 250-ml collection bottle in the air outside long enough to ensure that its contents
are replaced with fresh air. The calibration will be based on this sample having CO2
concentration of about 400 ppm. While still outdoors insert the rubber stopper into the bottle.
Do this by holding the stopper, not the probe box. You can now take the bottle and the sensor
to the location where the calibration is to be done.
Connect the CO2sensor to an interface.
Let the sensor warm up by collection data for at least 10 minutes.
When the readings are stabilized use a paper clip or the ballpoint to press down the
calibration button – release the button immediately after green LED blinks.
After about 3min., the reading should stabilize at a value of approximately 400 ppm
If the reading is significantly lower or higher than 400 ppm, simply press the button again to
repeat the process. .
Taking measurements with CO2sensor
Connect the sensor to an interface.
Allow the sensor to warm up for about 10 minutes. While the sensor is warming up, readings
increase to circa 5000 ppm and then slowly decrease back to the CO2level the sensor is
exposed to. Start the measurement.
Note: Since the sensor updates its reading every 3second, do not set the data rate faster that
1 reading per 3second.
Once you have started collecting data, you should see the green LED on the sensor box turn
on each time the sensor takes a new reading.
Important
The CO2sensor needs to warm up for 10 minutes anytime power is interrupted. The sensor
required a large current (250 mA).
For the Sciencecube Interface users is recommended to connect the USBAdapter to the
interface unit when using the CO2sensor.
The sensor can not take readings at a CO2concentration higher than 10 %. Once the CO2
concentration reaches this level the computer will continue to display a reading of 10 %, until
the actual level drops below 10 % again.
Even though the sensor responds rather quickly to change in CO2concentration, remember
that gas has to diffuse through the holes in the sensor tube before any changes in
concentration can be detected. Since diffusion of gases is a fairly slow process, there is
resulting delay in the readings. Wafting air toward the sensor tube speed up this process.
When you finished collecting data, simply remove the sensor from the gas-sampling bottle.
We recommend leaving the slit rubber stopper on the sensor tube. Store the sensor in the
box it was shipped in.
The sensor is designed to operate between 20°C and 30°C. The sensor can be used outside
of this temperature range; however there will be a loss in accuracy of readings, even if the 1-
point calibration at the lower level or higher temperature is done.Allow enough time for the
sensor to stabilise at the desired operating temperatures.
Warranty
The CO2sensor is warranted for a period of 12 months from the date of purchase provided the
electrode has been used in accordance with this instruction manual and under normal laboratory
conditions. The warranty does not apply when the CO2has been subjected to accident, alternate use,
misuse or abuse in any manner.
ScienceCube warrants this product to be free from defects in materials and workmanship for a period
of five years from the date of shipment to the customer.
This warranty does not cover damage to the product caused by abuse or improper use.

KDS-1037
- 4 -
Specifications
CO2range 0 – 100,000 ppm (0 – 10%)
Accuracy
(at standard 1 atm):
0 ppm to 10,000ppm : 1,000ppm or 10% of value, whichever is
greater
10,000ppm to 50,000ppm : 20% of value
Over 50,000ppm : qualitative only
Output voltage
Output impedance 0 – 4 V
1kOhm
Resolution using 12 bit AD 30.5 ppm CO2
Calibration information Slope: 25,000 ppm/V, offset 0V
(0V=0 ppm, 4V=100,000 ppm)
Warm up time 10 minutes (maximum)
(faster when air currents are provided)
Response time 90% of full-scale reading in 60 s
Normal operating temperature
range 25 5°C
Operating hum. range 5 - 95 % (non-condensing)
Revision : Dec. 2014
© 2014 KoreaDigital Co. Ltd.
#804 Ace Twin Tower 2, Guro, Seoul, 152-779 South Korea
• TEL (+82) 2-2109-8880 • FAX (+82) 2-2109-8881
Direct : (+82)2-2109-8839
www.sciencecube.com
This product is to be used for educational purposes only. It is not appropriate for industrial, medical,
research, or commercial applications.
Our products and the contents are subject to change without any notice.
In consequence we cannot assume responsibility for any consequential or other damage resulting from the
use of this instrument.
ScienceCube are registered trademarks or trademarks of Korea Digital Co., Ltd.
Other product and brand names may be trademarks or registered trademarks of their respective
organizations.
Note : This product is to be used for educational purposes only. ScienceCube CO2 Gas
Sensors are not designed nor recommended for any industrial, medical, or commercial process
such as life support patient diagnosis, control of a manufacturing process, or industrial testing of
an
y
kind.
Table of contents
Popular Accessories manuals by other brands

echoflex
echoflex Elaho E-VAC-SR installation guide

MTS Sensors
MTS Sensors Temposonics R Series datasheet

Exo Terra
Exo Terra PT2499 operating instructions

Manitowoc
Manitowoc K00067 installation instructions

Eaton
Eaton IntelliView Series Instruction leaflet

CHIEF
CHIEF Keyboard Tray Accessory KSA1021 installation instructions