Seamaty SMT-120V User manual

成 都 斯 马 特 科 技 有 限 公 司
Please read this manual carefully before using analyzer.
SMT-120V
User Manual
Fully Automated Veterinary
Chemistry Analyzer

P/N: E00SMT120VCO Version:A/0
1
Contents
Good to Know 3
Section 1: Introduction of Analyzer 7
1.1 Analyzer overview 7
7
8
8
9
9
10
11
11
11
11
12
12
13
14
14
14
16
17
18
18
1.2 Brief introduction
1.3 Principle
1.4 Structure and composition
1.5 Function
1.6 Application
1.7 Specification
2.1 Unpacking
2.2 Installation procedure
2.2.1 Set up analyzer
2.2.2 Set up printing paper
2.2.3 Set up external printer
4.1:Frequently-used buttons
4.1.1: Setting
4.1.2: Sample data
4.1.3: QC data
4.1.4: Preparing analysis
4.2: Soft keyboard
Section 2: Installation of Analyzer
Section 3: Turning the Analyzer On and Off
Section 4: Operation
19
19
20
20
21
22
25
26
26
26
28
28
28
29
30
5.1 Sample requirements
5.2 Preparing the reagent disc
5.2.1 Storage and preparing
5.2.2 Dispense the sample
5.3 Sample test
5.4 Test process summary
5.5 Calibration & Q.C.
5.5.1 Calibration
5.5.2 Quality control
5.6 Printing out
5.6.1 Build-in printer
5.6.1.1 Sample test report
5.6.1.2 QC test report
5.6.2 External printer
Section 5: Test and Result

P/N: E00SMT120VCO Version:A/0
2
Contents
30
31
32
32
32
32
32
35
35
36
37
40
41
42
43
6.1 Cleaning the analyzer
6.1.1 Cleaning the exterior case
6.1.2 Cleaning the display
6.1.3 Cleaning the inside of drawer
6.1.4 Comparison of before and after cleaning
6.2 Updating the analyzer software
6.2.2 Updating by WIFI
6.3 Troubleshooting
6.4 Consumables replacement
Section 7: Packing, Storage andTransportation
Section 8: Contact Information
5.6.2.2 QC test report
5.6.2.1 Sample test report
Section 6: Maintenance & Service
6.2.1 Updating by USB flash drive

P/N: E00SMT120VCO Version:A/0
3
Good to Know
Dear customers:
Thank you for your trust and using SMT-120V fully automated veterinary chemistry analyzer. In
order to let you have a comprehensive understanding of this analyzer, we provide this manual
for you with contents including introduction, usage, maintenance, packing, storage,
transportation, etc. To ensure the good operation of analyzer and the reliability and accuracy of
test results, please pay a lot attention on the "WARNING", "CAUTION" and "Note" in this
manual.
Note: Use this manual appropriately
Scope of application:
User who received the operation training from Seamaty1)
User who received the operation training from authorized distributors of Seamaty2)
We do our best to provide correct and reliable information in this manual, but it is inevitable to
have mistakes or omission. If you have any doubts, please kindly contact us to specify them.
Contact information has been provided in the last page of this manual.
In order to improve the performance and reliability of the analyzer, we will make some real-time
changes to the hardware or software, which may lead to the difference from this manual.
Seamaty reserves the right to revise this manual and software version without notice!
No person or organization may reproduce, modify or translate the contents of the instructions
without the written consent of Seamaty.
Seamaty has the final interpretation of the contents of this manual.
The illustrations used in this manual are showing as samples, which may differ from the actual.
Be sure to use the analyzer under the conditions specified in this manual. Failure to do so may
result in failure of analyzer or unreliability of test results.
Instructions

P/N: E00SMT120VCO Version:A/0
4
Symbols used in analyzer
Symbols
How to use analyzer correctly
Meaning Explanation
Warning
Caution
Warning
Pinch Point
Note
Biohazard
The information should be learned on how to avoid the operator
from potential harm. Please refer to any marks in this manual.
It is the important information you should know to avoid any
potential damage of analyzer.
Keep hands and fingers clear during operation
The important information shows in the screen during operation
of analyzer.
Consider all material from veterinary sources to be potentially
infectious, and please handle them with precautions.
Before operation, please do as follows at first
Check that the actual configuration of the analyzer is exactly the same as the packing list. 1)
If you have any questions, please contact our salesperson or distributor.
Please be sure to fill in the guarantee card, and send product warranty receipt back to 2)
us as soon as possible. If you email us, please send it to sales@seamaty.com
Please read all documents carefully and save them for future need.3)
Warning
The analyzer should be protected from working in humid and corrosive atmospheres.
If peculiar smell or smoke generated during the operation, please cut off electricity in time
and notify the manufacturer or distributor for maintaining.
The operator should avoid touching the inner circuit and only the certificated personnel can
repair the analyzer.Operator should wear protective gloves, masks, and overalls.
User has obligation to provide an electromagnetic compatibility conformed working
environment to ensure the analyzer working well.
Instructions

P/N: E00SMT120VCO Version:A/0
5
Caution
Biohazard
Guarantee
This analyzer complies with the requirements of equipment emission and immunity in EN
61326-1:2013 Class B and EN 61326-2-6:2013.
The design and inspection of the analyzer is according to standard of IEC 61010-1:2012
and EN 61010-1:2010.
Do not use this analyzer near strong radiation sources, such as unshielded RF sources, as
this may interfere with the operation of the analyzer.
Keep the working environment aeration; avoid strong electromagnetic interference, dust,
sunshine and other strong direct light. Make sure the analyzer is placed on the worktable
stably.
Do not damage the electrical wire, hold plug tightly when pull the plug out and do not drag
the wire.Do not put container with water or small metal object on top of the analyzer, to avoid
water or metal object falling in inside, which may result in short-circuit and damage of the
analyzer.
The instrument should use power wire with three cores and make sure grounding well.
Do not place this analyzer in a position where it is difficult to disconnect the switch.
The production date of the analyzer is shown on the label at the back of the analyzer.
The reagent discs in this manual are Pulitai reagent discs which match with SMT-120V
chemistry analyzer.
In accordance with the standard laboratory practice, consider the potential biological risk of
animal blood samples, and handle them based on the corresponding level of biosecurity on
all samples, quality control, related containers and used reagent discs.
Considering the impact of environmental and unexpected factors, the recommended service
life of SMT-120V veterinary chemical analyzer is 5 years on the condition of appropriate use.
Seamaty undertakes to provide 24 months of free warranty service (excluding consumables)
for the analyzer from the date of receipt by user, except in the following cases:
Not strictly in accordance with the instructions in the use of analyzer or not use of standard
matched consumables;Man-made damage;Dismantle the analyzer without the permission of
Seamaty;The warranty is only for the first user, resale will cause the warranty invalid.
Instructions

P/N: E00SMT120VCO Version:A/0
6
Compliance with Safety Measures
For using analyzer safely and efficiently, please observe the following precautions:
Avoiding failure of analyzer1)
The working environment of the analyzer should meet the requirements in this manual.
Preventing electric shock2)
Do not open the analyzer without the authorization of Seamaty, and prevent the liquid
splashing into the analyzer. In order to prevent the occurrence of electric shock and other
safety incidents, if the liquid was into the analyzer inadvertently, please contact our after-
sales service team in time before power on.
Preventing pollution3)
It is a must to wear protective gloves during operation, or else there is a potential risk of
biological infection. There is also a potential biosafety risk in the absence of biosafety-
treated of used reagent disc. Once the skin is in direct contact, immediately wash and
disinfect the contact area and consult a doctor.
Operating reagent disc4)
Reagent beads may contain corrosive substances, please follow the instructions in
accordance with the manual. The operator will not come into contact with the reagent
beads sealed in the reagent disc rotor during normal use unless the rotor breaks. In case
of this, try to avoid direct contact with the reagent disc and prevent the reagent beads
entering into the respiratory tract.
Disposing reagent disc5)
For further guidelines on handling and disposing of hazardous laboratory wastes, please
refer to your local health and safety regulations.
Instructions

Product name: Fully automated veterinary chemistry analyzer
Model No.: SMT-120V
Size: 191*213*220mm
Net weight: 4.2Kg
Accessories: power adapter, power cord and pipettor.
Back view, seeing Figure 1-2
P/N: E00SMT120VCO Version:A/0
7
Section 1 Introduction of Analyzer
1.1 Analyzer Overview
Front and side view, seeing Figure 1-1
Figure 1-2Figure 1-1
1.2 Brief Introduction
SMT-120V veterinary chemistry analyzer is easy to use. The operator only need to collect blood
samples (heparin lithium anticoagulation of whole blood, plasma or serum) with 90-120 μl
added into the reagent disc rotor, and then insert the rotor into analyzer to accomplish the
testing automatically with immediately printing out result. In order to prepare for a second review,
it is recommended that the operator collects not less than 250 μl of sample at a time.
There are two USB ports of analyzer for connecting external devices, such as printer, mouse,
keyboard etc.
It will take 12 minutes to finish a test and the report will be printed out immediately and
automatically after testing.
Instructions

P/N: E00SMT120VCO Version:A/0
8
1.3 Principle
SMT-120V veterinary chemistry analyzer is based on embedded system, used with
recommended reagent disc. It is used to detect the concentration of ALT, TP, UREA and other
liquids which adapts test methods of end-point, rate, two-point and eight-wavelength
synchronous detection.
This analyzer utilizes Lambert - Beer law, and adapts the principle of absorption spectroscopy
or transmission turbidimetry method. Absorption spectroscopy method is mainly used in the
biochemical reaction test, and the operating principle works as follows: Preload multiple group
detection reagents in the reagent disc to form an independent reaction cuvette, insert collected
blood samples into reagent disc, and then insert reagent disc into analyzer. Through
instructions of embedded processor in the analyzer, the biochemical substances in the blood
will be distributed into the cuvettes to make chemical reaction with test reagents along with
color changes which are monitored photometrically by the analyzer. The microprocessor then
calculates the concentrations of the analytes to achieve the purpose of testing.
Transmission turbidimetry method is mainly used for immunochemical test, and the operating
principle is: It will produce immune complexes after the antigen-antibody binding occurred with
turbidity at a certain time. When the light passes through the solution, it will be absorbed by
immune complexes. The more the amount of immune complexes, the more light is absorbed.
The intensity of light being absorbed is proportional to the content of immune complexes within
a certain range. Measure the absorbance by measurement optics. The intensity of component
is proportional to the absorbance; at the same time, the absorbance is proportional to intensity
of antigen. The analyzer then calculates the concentrations of the immune complexes. After the
analysis is completed, the results print automatically.
1.4 Structure and Composition
The analyzer consists of aluminum exterior case, movement parts, temperature control
components, two-dimensional bar code scan module, printer components, optical components,
touch screen, operating software and power adapter.
The analyzer is compact, small with light weight and easy for transport. The features are:
Analuminumalloyexteriorcase
Ahighspeedmotorwhichdrivesthereagentdisktospin(movementparts)
Aphotometertomeasuretheconcentrationofsubstanceinliquid(opticalparts)
Twomicroprocessorsusedforanalyzercontrolandtestcalculation(temperature
controlcomponents,two-dimensionalbarcodescancomponents)
Instructions

P/N: E00SMT120VCO Version:A/0
9
Figure 1-3
Athermalprintercanprintouttestresults(printcomponents)
A6.5inchcolorcapacitivemulti-touchscreen(display)
Multipleselectionfunctionsofdealingwithaseriesoftestsandresults(operating
software),SeeFigure1-3
1.5 Function
6.5 inch capacitive multi-touch screen, Android system, multi-language support.
Independent channel testing, no carryover.
Advanced optical testing system, built-in 8 filters (340、405、450、505、600、630、850) nm.
Test method: end point, rate, fixed time etc.
Heparin lithium anticoagulation whole blood, serum and plasma samples.
The large capacity memory to meet customer needs.
Intelligent real-time quality control function to ensure accuracy.
External mouse and printer are available (based on USB).
15V vehicle power supported, it is strongly recommended to connect the analyzer after power
inverter
Build-in thermal printer.
1.6 Application
It is applied to lithium heparin anticoagulant whole blood, serum and plasma for veterinary
clinical biochemistry analysis, it is recommended to match with Pulitai reagent disc.
Instructions

P/N: E00SMT120VCO Version:A/0
10
1.7 Specifications
Sample Type
Sample Volume
Bar Code
Testing Time
Testing Principle
Testing Method
Temperature
Absorbance
Resolution
Carryover
QC & Calibrate
Work Environment
Light Source
Storage
Printer
IT Connection
Weight
Optic System
Power Adapter
Display
Instructions

P/N: E00SMT120VCO Version:A/0
11
Section 2 Installation of Analyzer
2.1 Unpacking
Unpack the carton case and check with Packing List, and prepare to install the analyzer if
everything is right. If you have any questions, please contact with our after-sales service team.
Meanwhile, please write the guarantee and send it back, which is convenient for us to follow
and track your service. (If send email, please email to: sales@seamaty.com)
Instrument work environment : Temperature: 10-32 ℃,
Relative Humidity ≤85%
Note
Do not place the device in following environment :
Humidity, corrosive gas, dusty, strong electromagnetic interference; 1)
Crowd and non-aeration; 2)
Non-steady worktable;3)
Direct sunlight.4)
Requirements for power supply:
AC100-240V, 50-60Hz, , 105W Max
The analyzer can not use the same power supply socket with some high-power
equipment including centrifugal, refrigerator, and oven, etc.
The power supply should be connected with three-cores wire and and good grounding ;
the voltage between the neutral line and earth line<5V; or choose two core power cord
with power adapter provided by Seamaty.
2.2 Installation Procedure
2.2.1 Set up analyzer
1) Remove the analyzer from the shipping carton case and place in a stable worktable
Check if the exterior case of analyzer is intact2)
3) Plug the power jack into the analyzer, then plug the detachable power supply cord
into the power adapter and into a grounded electrical outlet.
Turn on the switch on the back of analyzer, indicator light on, the display will show4)
the home screen. Please keep safe of accessories have not been used.
Instructions

The analyzer is compatible with printer with HP PCL3 GUI printer job language, such as HP
Deskjet 1010. If you do not use an external printer, you could skip this section.
2.2.3 Set up external printer
Steps:
Set up printer according to the user manual;1)
Load A4 paper;2)
Connect the external printer and analyzer through USB cable and port;3)
Power on the printer.4)
P/N: E00SMT120VCO Version:A/0
12
The specification of thermal print paper is 50*57cm, the inner structure, seeing Figure 2-1
2.2.2 Set up printing paper
Figure 2-1
Steps:
1) Open the printer cover,
Remove the packing of printing paper, load the paper roll into the printer chamber2)
with correct direction, see Figure 2-2.
Figure 2-2
3) Pull out the roll paper to cross the printer slot
Close and lock printer cover.4)
Note: A printing paper roll has been preloaded in the analyzer, and please follow the
steps above to replace a new roll.
Instructions

P/N: E00SMT120VCO Version:A/0
13
Section 3 Turning the Analyzer On and Off
Turning on: Connect power cord with power adapter, and then connect with analyzer, turn on
the switch in the back as the indicator light (logo beneath the display) turns bright with the
screen showed as Figure 3-1
Figure 3-1
Loading and initializing, the screen shows as Figure 3-2 :
Figure 3-2
Turning off: Press down the switch button in the back of the analyzer when the drawer is closed,
disconnect the plug from analyzer, the indicator light off.
Note: To extend the lifespan of the analyzer and accessories, it required to follow the
procedures of powering off, and do not frequently switch on and off.
Instructions

P/N: E00SMT120VCO Version:A/0
14
Section 4 Operation
Figure 4-1
There are four buttons on the touch screen showing as Figure 4-1, and the detailed function
explanations are as follows:
4.1 Frequently-used Buttons
When the loading is completed, the home screen will show interface as figure 4-1. Please check
the date and time in the status bar, and you could reset them according to section 4.1.1
Figure 4-2
This button is used to set system date, time, hospital name, printer mode, sample or quality
control range etc. Click “Setting” icon in the home screen, and enter into setting menu,
showed as Figure 4-2.
4.1.1 Setting
Instructions

P/N: E00SMT120VCO Version:A/0
15
Figure 4-3
Changing date and time: The date and time is factory preset to GMT+08:00 Beijing Standard
Time. To change date and time,
Hospital Name: Use soft keyboard to input hospital name and save it
Lis Settings: Upload the test data to the hospital or clinic management system.
Printer Mode: Select the build-in or external printer as your default printer and then press
“Save” icon.
Customizing Assay: Press “Assay” icon, and the screen shows as Figure 4-3, select
analyte in the list to modify the parameters and press “Save”. The analyzer includes a
number of factory-set analyte and reference parameters, but it is available for user to
modify thembased on the needs.
1) Press “System Time” icon in the setting screen
In the Date & Time Setting screen, set date, time, time zone etc. Click “Done” to save 2)
your time setting.
Find Drop-Down Menu in the right corner, click the icon, showing “Exit” button.3)
Press “Exit” to return back to Set screen4)
The analyzer includes two type units, one is System International units (SI) and the other is
Common units. The default is SI units, but user can modify it according to the needs,
seeing Figure 4-4
Figure 4-4
Instructions

P/N: E00SMT120VCO Version:A/0
16
Figure 4-5
Quality Control Setting: Press “QC” (Quality Control) icon, showing as figure 4-5. Select QC1
(low level) or QC2 (high level) to set value of each item and then press Save.
Figure 4-6
It is used to search and review the test results, to modify the sample information, and to print
out test results, seeing Figure 4-6
4.1.2 Sample Data
Press “Query” icon to show screen as Figure 4-7. User can search history results by Sample
No., Name or Date. Click “Sample Detail” icon to view or modify the sample information.
Click “Print” icon to print out the result.
Instructions

P/N: E00SMT120VCO Version:A/0
17
Figure 4-7
Figure 4-8
It is used to search and review the quality control test results, to modify the sample
information, and to print out result, seeing Figure 4-8
4.1.3 QC Data
Press “Query” icon to show screen as Figure 4-9. User can search history results by Quality
Control ID, Lot or Date. Click “QC Detail” icon to view or modify the sample information.
Click “Print” icon to print out the result.
Figure 4-9
Instructions

P/N: E00SMT120VCO Version:A/0
18
4.2 Soft Keyboard
Add sample or control into reagent disc and insert into drawer of analyzer, click “Prepare
Analysis” icon to start testing. If there are messages showed on screen, read it carefully
before operation or contact our after-sales service department if any problem occurred.
4.1.4 Prepare Analysis
The analyzer has a built-in soft keyboard. Click the space where you want to input text to
activate the soft keyboard.
Figure 4-11
Instructions

P/N: E00SMT120VCO Version:A/0
19
Section 5 Test and Result
5.1 Sample Requirements
Warning
Users should strictly follow the local clinical laboratory bio-safety regulations when handling
blood samples and operating the analyzer.
Note
The required sample volume to add in reagent disc is 100 μl, and 90-120 μl is acceptable.
It is recommended to collect no less than 250 μl blood.
Lithium heparin is the only anticoagulant recommended for use with the analyzer.
Caution
When collecting the sample in lithium heparin collection tubes, turn upside down 3-5
times slightly to ensure the sample and the anticoagulant mixed well.
Whole blood must be analyzed within 30 minutes of collection, or separated into plasma
or serum.
To prevent hemolysis, do not refrigerate or shake whole blood.
If not analyzed immediately, plasma or serum can be stored at 2-8° C for no longer than
24 hours after centrifugation; or store it at -20°C for up to 30 days with no self defrost
cycle. Under these conditions, there will be no clinically important changes in most analyte
concentrations.
In order to obtain the correct glucose test results, blood collection should only be taken
at least fasting 12 hours.
Instructions
Table of contents