Instruction Manual
HI 93737
Silver ISM
http://www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI93737iswarrantedfor two years against defects in workman-
shipand materialswhenusedforits intendedpurposeandmaintained
accordingtoinstructions.
Thiswarranty islimitedto repairorreplacement freeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required,contactthedealerfromwhom youpurchasedthe
instrument.Ifunderwarranty, report the model number, dateofpur-
chase,serialnumberandthe nature of the failure. Iftherepairisnot
coveredby the warranty,youwillbe notified ofthechargesincurred. If
theinstrumentistobereturnedtoHannaInstruments,firstobtaina
ReturnedGoodsAuthorizationNumber from the Customer Service de-
partmentand thensenditwith shipmentcostsprepaid.Whenshipping
anyinstrument, makesureitisproperly packagedforcompleteprotec-
tion.
Tovalidate yourwarranty,filloutandreturntheenclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrightsare reserved. Reproduction inwholeor in part isprohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket, Rhode Island, 02895,USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93737 meter measures the silver content in water
and wastewater in the 0.000 to 1.000 mg/L (ppm) range.
Themeter usesan exclusivepositive-locking systemto ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid form and are supplied in bottles.
The amount of reagent is precisely dosed to ensure maxi-
mum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0.000 to 1.000 mg/L (ppm)
Resolution 0.001 mg/L
Accuracy ±0.005 mg/L ±10% of reading
TypicalEMC ± 0.001 mg/L
Deviation
Light Source Light Emitting Diode @ 555 nm
Method Adaptation of the 1-(2-pyridylazo)-2-
naphtolPANmethod.Thereaction between
silver and reagents causes an orange tint
in the sample
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI93737A-0 BufferReagentA 1 mL
HI93737B-0 Buffer ReagentB 1 mL
HI93737C-0 Indicator Reagent C 2mL
HI93737D-0 Fixing Reagent D 2 mL
HI93703-51 DispersingAgent 4-6drops
(onlywhen necessary, see note)
REAGENT SETS
HI93737-01 Reagents for 50 tests
HI93737-03 Reagents for 150 tests
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximumconcentrationreadable(see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
Note: for best results perform your tests
between20-24°C.
• When the LCD displays "- - -", it is
ready.
• Fill two graduated beakers with 25 mL of sample.
• Add 1.0 mL of HI 93737A reagent
to one beaker (the blank) and swirl
gently to mix.
• Add 1.0 mL of HI 93737B reagent
to the other beaker (the sample).
Swirl gently to mix and wait for 2
minutes.
• Add 1.0 mL of HI 93737C reagent
to each beaker, swirl and wait for 2
minutes.
• Add 1.0 mL of HI 93737D reagent
to each beaker, swirl and wait for 2
minutes.
• Fill one cuvet up to 1.5 cm (¾") below
the rim with 10 mL of the blank.
• Place the cuvet into the holder
and ensure that the notch on
the cap is positioned securely
into the groove.
• Press ZERO and SIP will appear on the LCD.
ON
OFF
A
B
ZERO
1.5 cm
• Wait for a few seconds and the
display will show "-0.0-". Now the
meter is zeroed and ready for mea-
surement.
• Fill a second cuvet up to 1.5 cm (¾")
below the rim with 10 mL of the re-
acted sample.
• Insert the second cuvet into the
instrument.
• Press READ and "SIP" will appear during measurement.
• Theinstrument directly displays concentration in mg/L of
silver on the LCD.
Note: a temperature above 30°C may cause turbidity. In
this case, before zeroing and taking readings, add
2-3 drops of Dispersing agent (HI93703-51) to each
cuvetandswirl until complete dissolution of turbidity.
INTERFERENCES
Interference may be caused by:
Al3+ above 30 mg/L
Ca2+ above 1000 mg/L as CaCo3
Cd2+ above 20 mg/L
Cl-above 8000 mg/L
CO2+ above 1.5 mg/L
Cr3+ above 20 mg/L
Cr6+ above 40 mg/L
Cu2+ above 15 mg/L
F-above 20 mg/L
Fe2+ above 1.5 mg/L
Fe3+ above 10 mg/L
K+above 500 mg/L
Mn2+ above 25 mg/L
Mg2+ above 1000 mg/L as CaCo3
Na+above 5000 mg/L
Ni2+ above 1.5 mg/L
Pb2+ above 20 mg/L
Zn2+ above 30 mg/L
1.5 cm
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C(86°F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLID
E
OFF
RecommendationsforUsers
Before using these products, make sure that they are entirely suitable for the environ-
mentinwhichtheyareused.
Operationoftheseinstrumentsinresidentialareacouldcauseunacceptableinterferences
toradioandTV equipments, requiring the operatortotakeallnecessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments'EMCperformance.
Toavoiddamagesorburns,donotperformanymeasurementinmicrowaveovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
REAGENT SETS
HI93737-01 Reagents for 50 tests
HI93737-02 Reagents for 150 tests
OTHER ACCESSORIES
HI710009 Blue rubber boot
HI710010 Orange rubber boot
HI721310 9V battery (10 pcs)
HI731318 Tissue for wiping cuvets (4 pcs)
HI731321 Glass cuvets (4 pcs)
HI731325 Caps for cuvets (4 pcs)
HI93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93737R1 11/98
PRINTED IN PORTUGAL

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI96704C User manual

Hanna Instruments

Hanna Instruments HI96704C User manual

Hanna Instruments HI 991000 User manual

Hanna Instruments

Hanna Instruments HI 991000 User manual

Hanna Instruments HI 98180 User manual

Hanna Instruments

Hanna Instruments HI 98180 User manual

Hanna Instruments HI 755 User manual

Hanna Instruments

Hanna Instruments HI 755 User manual

Hanna Instruments HI6321 User manual

Hanna Instruments

Hanna Instruments HI6321 User manual

Hanna Instruments HI 931101 User manual

Hanna Instruments

Hanna Instruments HI 931101 User manual

Hanna Instruments HI 739 User manual

Hanna Instruments

Hanna Instruments HI 739 User manual

Hanna Instruments HI 746 User manual

Hanna Instruments

Hanna Instruments HI 746 User manual

Hanna Instruments HI96716C User manual

Hanna Instruments

Hanna Instruments HI96716C User manual

Hanna Instruments HI 99181 User manual

Hanna Instruments

Hanna Instruments HI 99181 User manual

Hanna Instruments HI 98186 User manual

Hanna Instruments

Hanna Instruments HI 98186 User manual

Hanna Instruments BL 7916 User manual

Hanna Instruments

Hanna Instruments BL 7916 User manual

Hanna Instruments HI8417 User manual

Hanna Instruments

Hanna Instruments HI8417 User manual

Hanna Instruments HI97723 User manual

Hanna Instruments

Hanna Instruments HI97723 User manual

Hanna Instruments HI 8730 User manual

Hanna Instruments

Hanna Instruments HI 8730 User manual

Hanna Instruments Total Chlorine HI 711 User manual

Hanna Instruments

Hanna Instruments Total Chlorine HI 711 User manual

Hanna Instruments HI 9210N User manual

Hanna Instruments

Hanna Instruments HI 9210N User manual

Hanna Instruments HI 22091 User manual

Hanna Instruments

Hanna Instruments HI 22091 User manual

Hanna Instruments HI97708 User manual

Hanna Instruments

Hanna Instruments HI97708 User manual

Hanna Instruments HI96713C User manual

Hanna Instruments

Hanna Instruments HI96713C User manual

Hanna Instruments HI727 User manual

Hanna Instruments

Hanna Instruments HI727 User manual

Hanna Instruments HI 2209 User manual

Hanna Instruments

Hanna Instruments HI 2209 User manual

Hanna Instruments HI 720 User manual

Hanna Instruments

Hanna Instruments HI 720 User manual

Hanna Instruments HI98193 User manual

Hanna Instruments

Hanna Instruments HI98193 User manual

Popular Measuring Instrument manuals by other brands

TSI Instruments HM685 owner's manual

TSI Instruments

TSI Instruments HM685 owner's manual

VONROC LL503DC Original instructions

VONROC

VONROC LL503DC Original instructions

Fostex XR-7 owner's manual

Fostex

Fostex XR-7 owner's manual

OTT HydroMet MF pro Basic user manual

OTT HydroMet

OTT HydroMet MF pro Basic user manual

Extech Instruments MO230 user guide

Extech Instruments

Extech Instruments MO230 user guide

GF MSA 2 MULTI instruction manual

GF

GF MSA 2 MULTI instruction manual

Alicat L Series operating manual

Alicat

Alicat L Series operating manual

JensPrima innoCon 6800O Operating	 instruction

JensPrima

JensPrima innoCon 6800O Operating instruction

Geokon 4650 instruction manual

Geokon

Geokon 4650 instruction manual

Bendix CVS SmarTire TPMS user manual

Bendix CVS

Bendix CVS SmarTire TPMS user manual

FE PHW quick reference

FE

FE PHW quick reference

Solinst 101 Replacement manual

Solinst

Solinst 101 Replacement manual

Danfoss RAVK 10-30 operating guide

Danfoss

Danfoss RAVK 10-30 operating guide

Sharp DV-RW360B Operation manual

Sharp

Sharp DV-RW360B Operation manual

Siemens FIDAMAT 6 operating instructions

Siemens

Siemens FIDAMAT 6 operating instructions

Innovative instruments DMM B18 manual

Innovative instruments

Innovative instruments DMM B18 manual

VeEX FX86 user manual

VeEX

VeEX FX86 user manual

ZIROX XS22 manual

ZIROX

ZIROX XS22 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.