Select Medical OLA 8 User manual

User Manual
Dynamic Pressure Redistribution Mattress
& Mattress Overlay
(UM20180417 – Review V1.1)

Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

USER MANUAL - OLA 8 & 4
TABLE OF CONTENTS
Statements and Symbols.......................................
Important Noce...................................................
Introducon...........................................................
Contact Informaon..............................................
Product Overview..................................................
Safety.....................................................................
Symbols & Denions............................................
Control Unit & Maress Parts................................
Installaon.............................................................
Operaon...............................................................
Cleaning & Decontaminaon................................
Storage..................................................................
Troubleshoong.....................................................
Maintenance..........................................................
Specicaons.........................................................
Electromagnec Compability..............................
Warranty and Service............................................
2
2
2
2
3
4
6
8
10
12
16
18
20
21
23
24
29
1
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

Thank you for choosing the OLA 8 or OLA 4 pressure redistribuon
system. This manual should be read carefully before using the maress
as it contains important safety and maintenance informaon to ensure
long lasng and reliable service.
INTRODUCTION
IMPORTANT NOTICE
STATEMENTS AND SYMBOLS
Before operang this medical equipment, it is important to read this
manual and understand the operang instrucons and safety
precauons. If you have any quesons regarding the use of this
equipment please contact your supplier.
Refer to manual
NB: Tips or informaon users should be aware of
Warning to highlight potenal hazards that, if disregarded, could lead to
injury or death.
CONTACT INFORMATION
For any service, warranty, sales or customer service informaon on this
product please contact your supplier or if in doubt contact Select Medical Ltd.
at the following address:
Select Medical Ltd, Unit 10 Philips Rd, Whitebirk Ind Estate, Blackburn,
BB1 5NA.
Customer Service: +44 (0)1254 685538 Sales: +44 (0)1254 668899
Email: inf[email protected]
www.selectmedical.co.uk
Cauon to highlight potenal hazards that if disregarded could lead to
equipment damage or failure.
2Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

PRODUCT OVERVIEW
Environment
Your dynamic maress system is intended for use in the following
environments:
• A care environment where medical supervision and monitoring are
provided (e.g. nursing homes, care home, rehabilitaon facilies etc).
• A domesc environment where the maress is used to alleviate or
compensate for an injury or disability.
Intended Use
OLA 8 is an 8” replacement, dynamic pressure redistribuon system
suitable for individuals up to very high risk of developing a pressure ulcer or
for those with exisng ssue damage.
OLA 4 is a 4” pressure relieving overlay suitable for individuals up to high risk
of developing a pressure ulcer or for those with exisng ssue damage.
OLA 8 & 4 provide regular periods of pressure reducon to vulnerable ssue
areas, aiding blood and lymphac ow which is vital to maintaining healthy
ssue. The maress system is designed to be used on standard or proling
beds.
For assistance in seng up, using or maintaining your dynamic maress
system, or to report unexpected operaon refer to the contact details found
on page 2.
Features
OLA 8 & 4 Maress:
• One in two cell-cycle design giving opmum therapy
• Mul-stretch, waterproof and vapour permeable cover
• CPR pull cord for rapid deaon
• Machine washable up to 95°C
• 1 year warranty
3
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

OLA 8 Control Unit (Q2-02):
• Audible low pressure alert
• Pressure adjustment for opmum therapy
• External, easy replacement pump lters
OLA 4 Control Unit (D30):
• Low pressure alert
• Pressure adjustment for opmum therapy
• External, easy replacement pump lters
• The maress system & control unit must be installed and used
in accordance with the informaon provided in this manual.
• The maress system is typically not suitable for children. If it is
to be used by a child ensure a risk assessment has been
undertaken.
• Before using the system ensure that the mains lead is free from
damage and is posioned so as not to cause an obstrucon or
trip hazard.
• Exposure of the control unit to any liquid while it is plugged in
could cause a severe electrical hazard.
• Use care when handling or transporng the control unit.
Dropping or other sudden impacts may result in damage to the
unit.
• Do not open the control unit or aempt to repair or service
the unit. Repairs and servicing should always be undertaken by
suitably trained personnel.
• If the control unit is not funconing properly, or has been
damaged, unplug the unit and take it out of service
immediately.
• Do not use the system near a heat source or naked ame.
• Do not use with hot water boles or electric blankets.
• Do not use liquids near the control unit if plugged in.
• Do not place any objects, such as blankets, on or over the
control unit.
SAFETY
General Safety
4Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

• Do not use the control unit near ammable gas or in oxygen rich
environments as this poses a re risk or risk of explosion.
• Always assess the risk of intenonal or unintenonal tampering
of the control unit.
Risk Assessment
It is the responsibility of the carer/care provider to carry out the necessary
risk assessment to ensure the safety of the paent. This should be carried out
before using the maress system.
A risk assessment should include, but is not limited to:
• Product combinaons (bed frame, maress, side rails etc.)
• Extent of ssue damage (if any)
• Entrapment
• Paent falls
• Small adults (and children)
• Paents with learning dicules
• Unauthorised people with access to the controls
Contraindicators
Paent condions for which the applicaon of pressure relief on an
alternang maress system is a contraindicaon are as follows:
• Cervical or skeletal tracon
• Unstable spinal fractures
Other contraindicaons may be relevant which are specic to the paent or
care environment.
Maress Load
5
OLA 8 OLA 4
Minimum Weight Limit 32kg (5 stone) 32kg (5 stone)
Maximum Weight Limit 190kg (30 stone) 114kg (18 stone)
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

SYMBOL DEFINITIONS: CONTROL UNIT & MATTRESS
Control Unit
Warning: beware of potenal hazard
Refer to manual: failure to do so could introduce a hazard
Type BF Applied Part
Applied Part: The parts of the device that come into physical
contact with the user/occupant in order for it to carry out its
intended funcon.
Type BF: Applied parts which are electrically isolated from earth
and other parts of the medical equipment - Complying with
specic requirements for protecon against electric shock to
IEC 60601-1
W.E.E.E Label
(Waste Electrical and Electronic Equipment)
Class II electrical device
The user/occupant is protected by at least two layers of insulaon
between the current carrying parts (e.g. mains cable) – If damage
is noced to the control unit or mains cable assembly turn o at
the mains supply and contact your provider or Select Medical Ltd.
immediately.
IP21 Protected from touch by ngers and objects greater than 12
millimetres. Protected from condensaon.
The following symbols are found on the control unit:
6Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

7
The following symbols are found on the maress:
Disinfect by wiping the surface using a hypochlorite soluon diluted
1000ppm
95 Machine wash up to 95°C
Tumble dry on a low seng
Do not use harsh abrasives or Phenol cleaners
Do not iron
Ensure system is dry before storing
Do not place heavy objects on surface of cover other than the
paent
Do not use when damp, ensure surface is dry before use
Do not fold. Roll pack the system
Do not use sharp objects
Only use in conjuncon with appropriate medical advice
Maress
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

CONTROL UNIT/MATTRESS PARTS
Control Unit
8
OLA 8 - Q2-02 Control Unit OLA 4 - D30 Control Unit
1. Control Panel
2. On/O switch
3. Mains Power Cable
4. Female Air Connector Port
5. Air Filter
6. Fuse Holders
7. Cushion Bar
8. Hooks
1. Control Panel
2. On/O switch
3. Mains Power Cable
4. Female Air Connector Port
5. Air Filter
6. Cushion Bar
7. Hooks
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

Maress
1. Top Cover
2. Air Cells
3. Male Air Connector
4. Base Cover
5. Securing Straps (elascated corner straps only on OLA 4)
6. CPR Pull Cord
9
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

10
INSTALLATION
Before installing the maress system please read the warning and
cauon notes carefully. These highlight risk areas to ensure paent
safety.
• Ensure the maress is only used with compable equipment/
accessories.
• Ensure the maress is of the correct type for the paent.
• Ensure the CPR pull-cord is easily accessible at all mes.
• Ensure the plug is accessible at all mes so the maress can be
disconnected from the mains supply quickly, if required.
• Ensure the mains cable is plugged into an appropriate power
source at all mes.
• Ensure the mains cable is not taut, parcularly if being used on
a proling bed that moves up and down (check all posions).
• Ensure that the mains cable does not become compressed,
trapped or damaged by the bed frame or other equipment.
• Replace any damaged cable immediately as these cables can
create a risk of electrocuon and/or re.
• A CE marked extension cable must only be used when it is not
possible to reach a wall socket with the equipment mains cable.
• If an extension cable is used never overload it by plugging in
appliances that together will exceed the maximum current
rang stated for the extension cable.
• Do not use block adaptors.
• Ensure extension cables or sockets are not placed under the
bed frame as liquids could leak onto them posing an electrical/
re risk.
• Ensure the mains supply is compable with the control unit
(see page 23 for electrical specicaon)
• Avoid placing the maress system in direct sunlight as this
could damage the maress cover.
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

1. Carefully open the packaging.
2. Although unlikely, please check the product for any signs of damage. Do
not use if damaged and contact your provider or Select Medical Ltd (see
page 2).
3. Place the maress on top of the bed frame (OLA 8), or on top of the
exisng maress (OLA 4), with the top cover facing upwards and the air
hose at the foot end of the bed.
4. Aach the maress to the bed frame by securing with the adjustable
straps. Addionally, OLA 4 has elascated corner straps to tuck under the
exisng maress.
5. Check the CPR pull-cord is securely in posion.
6. Using the hooks on the back of the control unit, hang the unit over the
frame/board at the foot end of the bed. If there is no foot frame/board
lay the unit on the oor, under the bed with the front control panel facing
upwards.
7. Aach the male air connector on the maress to the female air
connector port on the control unit/pump, ensuring the air hose is not
kinked or trapped between parts of the bed frame/other equipment.
8. Plug the mains cable into a suitable mains supply and switch on the
control unit. At this stage both the mains power and low pressure
indicators will illuminate.
9. The maress will start to inate and will be completely inated within
30 - 40 minutes.
10. Once fully inated, adjust the straps that aach the maress to the bed
frame, ensuring the maress is held in place securely.
11. Cover the maress loosely with a sheet, ensuring it does not interfere
with cell alternaon. 11
• On proling beds it is essenal that adjustable straps are
secured around the movable secons of the bed frame,
otherwise the maress may be damaged.
• If you are placing the control unit on the oor it is advisable to
place the unit on a rm surface.
• Ensure the mains cable is posioned so as not to cause a trip
hazard.
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

12
OPERATION
Control Panel
OLA 8 - Q2-02 Control Unit
1. Pressure Adjustment Dial
2. Alert Mute/Reset
3. Low Pressure Indicator
4. Normal Pressure Indicator
1. Pressure Adjustment Dial
Turn the dial to set the system for opmum performance.
2. Alert Mute/Reset
The audible/visual alert idenes when the pressure is low. To mute the
audible alert press the buon. The visible alert indicator will now ash.
Once the fault has been reced, re-press the buon to reset the alert.
3. Low Pressure Indicator
A visible indicator (orange) warns that the pressure is below an acceptable
level.
4. Normal Pressure Indicator
A visible indicator (green) idenes that the pressure has reached the
preset level.
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

13
Maress Operaon
1. Turn on the power on the control unit. The pump starts to inate the
maress to the pressure selected on the dial.
2. The low pressure indicator (orange) will illuminate as inaon commences.
3. OLA 8 /Q2-02 only - the audible alert is acvated, press the ‘alert mute’
buon to mute the alarm and its indicator will ash.
OLA 4 - D30 Control Unit
1. Pressure Adjustment Dial
2. Low Pressure Indicator
3.Normal Pressure Indicator
1. Pressure Adjustment Dial
Turn the dial to set the system for opmum performance.
2. Low Pressure Indicator
A visible indicator (orange) warns that the pressure is below an acceptable
level.
3. Normal Pressure Indicator
A visible indicator (green) idenes that the pressure has reached the
preset level.
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

14
4. Once opmum pressure is reached (about 30-40 minutes) the ‘normal
pressure’ indicator will come on and the ‘low pressure’ indicator (and audible
alarm on the OLA 8/Q2-02 ) will turn o.
NB: If the ‘low pressure’ indicator (audible alarm) will not go o, refer to
troubleshoong on page 20.
5. Adjust the ‘pressure/comfort control’ dial to provide a comfortable
pressure level for the paent. Pressure range:
• OLA 4 / D30: 30-80mmHg
• OLA 8 / Q2-02: 20-60mmHg
6. Using clinical judgement and with connuous monitoring of the paent
for up to 72 hours, increase or decrease the pressure levels using the dial to
suit the paents comfort levels. If possible, having regular dialogue with the
paent is key.
NB: The maress can be used in an upright posion, however the pressure
seng may need to be increased. Use clinical judgement to ensure paent
comfort and eecve pressure relief is maintained.
CPR Funcon
In an emergency rapid deaon of the maress may be required. The CPR
pull cord is located at the head end of the maress.
To re-inate push the CPR cord back into the closed posion. The maress
will start to inate. Wait for opmal pressure to be reached before using the
maress.
• Carers/care providers should always familiarise themselves with
the posion of the CPR valve.
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

15
Using Inconnence Products with the Maress
Inconnence products, such as sheets or pads, can be used with the system,
however this may compromise the eecveness of the alternang pressure
distribuon.
Transporng the Maress & Power Cuts
If the maress is disconnected from the power supply so it can be moved, or
in the event of a mains power failure, carry out the following procedure to
maintain maress inaon:
1. Disconnect the male connector from the power unit by squeezing the two
tabs (A) and pulling away from the control unit (B).
2. Seal using the cap marked “Transport” which for safety is aached to the
male connector.
NB: Complete the acon quickly to limit air
loss.
3. Switch o the control unit.
4. Disconnect from the power supply.
5. The maress can now be moved.
• If inconnence products are being used it is important to carry
out a risk assessment and regular paent skin checks.
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

16
CLEANING & DECONTAMINATION
Cleaning
Cleaning is required regularly between paents to prevent cross infecon. It is
therefore important to clean and decontaminate the control unit and
maress following these procedures.
Control Unit
• The maress will remain inated for up to 24 hours - return the
system to the mains supply as soon as possible.
• Whilst unplugged alternang mode will not be operaonal and
pressure relief will not be provided.
• Do not remove the maress from the bed frame if the occupant
is sll on the maress.
• If it is essenal that the paent is moved whilst remaining on the
maress, the maress must be re-plugged in immediately once
the desired locaon has been reached to reduce the risk of ssue
damage.
• Never drag the maress, always carry it.
• Disconnect the mains cable from the power socket before
aempng to clean the control unit.
• Do not immerse or soak the pump.
• Do not spray any cleaning soluon directly on the surface of the
control unit.
• If any of the cleaning/washing instrucons are not followed the
product warranty will be invalidated.
• Do not use phenol based cleaning soluons, solvents, neat
bleach or abrasive products to clean the casing as this may
cause damage.
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

Maress
N.B: Before aempng to clean the maress the top cover should be checked
for physical signs of damage that may lead to strike-through (ingress of uid
through cover). Staining to the underside of the top cover is a sign of
strike-through.
17
1. Check for external damage – do not use if damage is found.
2. Place the pump on a work surface and using a clean cloth wipe the outside
of the case with a prepared sodium hypochlorite soluon (1000ppm).
3. The control unit should be cleaned by starng with the cleanest parts and
systemacally moving to the direst parts. Extra care should be taken around
areas where excess dirt or dust may gather.
4. Change the cloth if it becomes dirty.
5. Once clean, wipe down with a new clean cloth moistened with clean
water to remove detergent residue.
6. Dry o with a paper towel. Always allow the surfaces to dry thoroughly
before pung back into use.
• Frequent or prolonged exposure to higher concentraon
disinfectant soluons may prematurely age the fabric cover of
the maress.
• Do not use the cover if strike-through or damage is found – risk
of cross infecon. Replace with a new top cover.
• Do not use solvents or alcohol-based cleansers e.g. Phenicol,
Hibiscrub, Clearsol, Stericol or Hycoline as these will destroy the
maress materials.
• Do not autoclave.
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1

18
Decontaminaon
1. Unzip the top cover from the maress.
2. The top cover can be machine washed up to 95°C and tumble dried on a
cool seng.
3. Unsnap the air cells from the maress base on both sides.
4. Carefully clean with (1000ppm) prepared soluon of sodium hypochlorite
and allow to dry completely.
5. Make sure to disconnect all the air cells and spray the cleaning soluon on
all sides, including the connecng tubes and hoses.
6. Re-assemble the maress and lay it out at.
7. Ensure the maress is completely dry before either storing or using for
another paent.
STORAGE
General Cleaning:
1. Wipe down with a clean cloth moistened with a mild detergent and diluted
in warm water (40˚C).
2. Rinse with cold clean water and a clean cloth and allow to fully dry before
use.
Storage
1. Detach the control unit from the maress.
2. Pull the CPR cord unl it is open.
3. Ensure there is no air trapped in the cells.
4. Lay the maress out at and roll the maress from the foot end towards
the head end.
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1
Other manuals for OLA 8
1
This manual suits for next models
1
Table of contents
Other Select Medical Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Roche
Roche ACCU-CHEK TenderLink Instructions for use

Medela
Medela Invia Liberty Clinician Instructions for Use

Proactive
Proactive Protekt Aire 4000DX Operation manual

Integra
Integra Miltex Directions for use

Medical Solutions
Medical Solutions 6000 Series Operator's manual

ResMed
ResMed AirFit N30 user guide