Shinko CUX II Series User manual

3194M01
CUXⅡSERIES
ELECTRONIC COUNTING SCALE
OPERATION MANUAL
HIJK
Shinko Denshi Co., Ltd. 01-12-1998 NS

INDEX
1. SPECIFICATIONS....................................................................................3
1-1. GENERAL SPECIFICATIONS..........................................................................................6
1-2. STANDARD MODELS ...................................................................................................6
2. UNPACKING & INSTALLATION ..................................................................7
2-1. UNPACKING...............................................................................................................7
2-2. INSTALLING THE WEIGHINGPAN.................................................................................7
2-3 TO LEVEL THE SCALE..................................................................................................7
2-4 PERFORMANCE TEST...................................................................................................8
3. OPERATION KEY FUNCTIONS...................................................................9
4. BEEP SIGNALS AND CONDITION SYMBOLS............................................... 10
4-1 MESSAGES BY BEEP...................................................................................................10
4-2 MESSAGES BY SYMBOLIC LETTERS..............................................................................10
5. OPERATION ..........................................................................................11
5-1. TARE.......................................................................................................................12
5-2. UNIT WEIGHT SET (SAMPLING)..................................................................................13
1.) QUANTITY SET SAMPLING ...................................................................................13
2.) DIRECT ENTERING OF UNIT WEIGHT.....................................................................14
3.) SCS (AUTOMATIC RE-SAMPLING SYSTEM)............................................................15
5-3. RE-SAMPLING..........................................................................................................16
5-4. COUNTING...............................................................................................................16
1.) NORMAL COUNTING...........................................................................................16
2.) ACCUMULATE COUNTING....................................................................................16
3.) FIXED UNIT WEIGHT METHOD..............................................................................17
5-5. MEMORY LOOK-UP....................................................................................................17
5-6. MEMORY STORE.......................................................................................................17
5-7. MEMORY DATA CALL UP IN WORKING AREA.................................................................19
5-8. LIMIT SET AND COMPARING.......................................................................................19

5-9. FUNCTION SET.........................................................................................................21
A.) FUNCTION SET..................................................................................................21
B.) FUNCTION TABLES.............................................................................................22
C.) CONTENTS OF INTERFACE..................................................................................23
6. CALIBRATION ...................................................................................... 24
7. TROUBLE SHOOTINGS........................................................................... 26

1. SPECIFICATIONS
1-1. General Specifications
Measuring System : Tuning-fork system.
Sampling System : Semi-automatic counting and storing system(SCS sampling)
Storing by known sample quantity entered via ten-key
Storing by known Unit Weight entered via ten-key
Memory Bank : Memorable various Unit Weight and Tare Weight up to 300
units. ( ID 1 to 30 Unit Weight with TARE W.)
Accumulation function : Accumulate to measured count data, and adds an optional
number(quantity) inputted by the ten-key.
Limit Function : Judges if the object has quantity in a specific range by setting by
setting upper/ lower limits.
Display : Quantity ; VFD 7 digits/12.5 mm H
Total Weight ; LCD 7 digits/12.5 mm H
Average Piece Weight ; LCD 7 digits/12.5 mm H
Power Supply : 9VDC/400mA with AC Adapter
Options : Output for Shinkoprinters. / Bidirectional RS232C output.
Relay contact output.
Rechargeable Battery option (Operable for approx. 5 hours).
1-2. Standard Models
MODEL CUXII-600 CUXII-1500
CUXII-3000
CUXII-6000
CUXII-12K
CAPACITY 600g 1500g 3000g 6000g 12000g
READABILTY 0.01g 0.05g 0.05g 0.1g 1g
HIGH RES’TION
READABILITY 0.002g 0.01g 0.01g 0.02g 0.1g
MINIMUM UNIT
WEIGHT FOR SCS
10 mg 25 mg 50 mg 100 mg 200 mg
RESOLUTION
IN COUNTING
1 mg 2.5 mg 5 mg 10 mg 20 mg
COUNTABLE
UNIT WEIGHT 1 mg 2.5 mg 5 mg 10 mg 20 mg
PAN SIZE 140mm dia
234 x 204 mm
WEIGHT 3.2 kg 3.9 kg
1

2. UNPACKING & INSTALLATION
2-1. Unpacking
Unpack the container carefully. Examine the packaging and the device for any
damage, and report it to the shipper if there is any. Try to keep the scale
upright.
Check for enclosures as follows:
①Scale main unit
②Weighing pan with pan base
③AC Adapter
④Operation manual
Note : 1. If any items are missing or any damage is discovered immediately notify
your place of purchase or a dealer.
2. Keep box and shock absorber pad after installation. They are required for
transportation.
2-2. Installing the Weighing Pan
Place the pan base packed with the weighing pan on the scale. Fix it by driving
the nut on the pan base to the screw in the center of the scale. Place the
weighing
pan properly on the pan base.
* The receiving element of the pan base is directly linked with the delicate weighing
mechanism. A tension or a side force may damage the unit.
2-3 To Level the Scale
The scale must be level for proper operation. Use the 4 level adjusters located at
the 4 corners of the scale main unit to adjust the horizontal position of the scale
until the bubble of the bubble level lies within the red circle. Ensure that the
scale sits solidly and does not wobble due to uneven adjustment.
2

2-4 Performance Test
1.Connect the AC adapter at the rear of the scale, then plug the cord in line
outlet.
2.Press the ON/OFF key. All display, segments and characters will blink for
about 4 seconds as a self test.
3. Verify that the display changes by touching the pan slightly, and that it
returns immediately to the original by releasing it.
** INSTALLATION PRECAUTIONS
The CXII is a precision instrument and requires careful handling. Try to select a good,
clean environment for installing the scale.The following factors may cause the scale to
return inaccurate measurements.
1. Installation on a soft surface that may flex when objects are put on it or put on the
scale.
2. Environments that have greatly varying temperature and humidity.
3. Environments subject to vibration or an unstable surface for installation of the scale.
4. Environments subject to air flow from heaters or other air conditioning units.
5. Environments subject to corrosive gases or large amounts of dust.
6. Environments subject to direct sunlight.
3

3. OPERATION KEY FUNCTIONS
ON/OFF : ON/OFF key for the balance.
RE : Improving unit weight by increasing samples.
SCS : Used to command automatic calculation and storage of
average unit weight of sample materials in the unit.
T: Tare and parameter change
R.M. : Used to recall memorized Unit weight (and Tare) from memory-bank,
inputting code number by the ten-key.
No. : Used to set code numbers for various Unit weight (and Tare)
banking, and memory look up for Unit weight (and Tare).
TARE PRESET : Taring a known tare weight inputting by the ten-keys.
When needing to confirm tare weight, look up current tare value shown
ondisplay for about 2 seconds after press TARE PRESET.
UNT.W. SET : Used to unit weight setting.
Storing of a known unit weight inputting by the ten-keys.
LMT : Used to limit function( comparator ) setting.
Storing of limit value for setting points.
MODE :Function/mode change. Function modes such as bar graph indication,
Auto-Zero, Auto-Power off, stable judgment, resolution select,
Interface, unit selection and the calibration mode could be called by
Pressing the MODE key.
ADD : A key performs like an adding key of a calculating function, adds a
counted result to the total.
SMPL SET : Used to sample quantity setting.
Sampling by inputting known sample quantity from the ten-key.
0to 9key : Ten numeric keys, called commonly as ten-keys.
.key : Decimal point key, effective only in the setting of the unit weight and
the tare preset.
C/AC : Used to clear the numbers inputted by the ten-key, unit weight data
and tare preset data clear.
4

4. BEEP SIGNALS AND CONDITION SYMBOLS
CUXII signals you various messages by four different beep signals and four
symbolic letter as shown as under. These signals and symbol tell you;
(a) instruction to the next operation, (b) warning for faulty operation,
(c) indication of unsuitable data, (d) indication of measuring
condition.
4-1 Messages by beep
Beep Signal Conditions
Short beep (one time) : Confirmation for pressing the key.
Long beep (one time) : Completion for set data storing.
Short beep (two times)
: Warning for faulty key operation. (Try to re-input)
Short beep (three times)
: Warning for miss key operation. (Stop the
operation)
4-2 Messages by symbolic letters
Symbol Conditions
LIGHT
The UNIT W. is too light for the balance resolution. Impossible to
count, use a balance of smaller capacity.
While flickering, it is possible to count, but a counting error may be
produced due to the rough uniformity of the samples or depending
on the accuracy of the balance.
ADD Please add samples. -- while SCS operation --
NON UNI
The samples are not uniformed or contain some irregular material.
Unload samples until the symbol goes off or pick the irregular
material off from sampling material.
-- while SCS operation --
EXCESS Too many samples have been added. -- in SCS operation –
Unload samples until the symbol goes off.
5

5. OPERATION
Basic operation is divided into the four procedures
(A) Tare if necessary
(B) Unit weight set ( sampling)
(C) Re-Sampling if necessary
(D) Counting ( loading all the samples )
Procedure (A) has 2 ways
Procedure (B) has 4 ways
Besides normal counting method, we have 2 different methods
Other operations include
(E) Memory look-up
(F) Memory store
(G) Memory call-up
(H) Limit set and comparing
( I ) Function set
6

5-1. Tare
1.) Tare with actual tare load
Load tare.
Press T key.
2.) Tare with numerical keys.
Enter tare weight with numerical keys.
Press TARE PRESET key.
# You must pre-load the tare before sampling.
< Set data clear >
Press C/AC key until the display goes “0“.
< Look up tare >
Press TARE PRESET key.
Current tare value shown on display for about 2 sec.
7

5-2. Unit weight set (sampling)
1.) Quantity set sampling
Tare the balance by pressing Tkey.
Display : Exactly 0 .
Load samples precisely counted in advance.
Enter quantity* of the samples with the numerical keys.
Press SMPL SET key.
( * you cannot get accurate counting without correct quantity set.)
Features
-Since it is difficult to count many samples by manuals , we can usually load
few pieces.
-Re-sampling is strongly recommended.
-Sampling error can occur depending on how re-sampling is made. (depends
on operators )
< Set data clear >
Press C/AC until the display goes “0“.
8

2.) Direct entering of unit weight
Enter unit weight with numerical keys.
Press UNT.W. SET key.
Features
-Good when average weight is almost precisely known.
-Bad when unit weight set differs a lot from samples’average.
-Re-Sampling is recommended.
-Sampling error can occur depending on how re-sampling is made. (depends
on operators )
< Set data clear >
Press C/AC until the display goes “0“.
9

3.) SCS (Automatic re-sampling system)
1. Press SCS key.
Beep then flicker : on 5 * (on QUANTITY display)
(*Initial quantity can be changed by ten-key)
Appear : ADD (on WEIGHT display)
2. Load exactly five(5) samples.
Flicker to the RED LED in the SCS key and PuSh on WEIGHT display
3. Press SCS key.
4. Add an optional quantity but within number of acceptable addition
quantity (display on the UNIT W. position).
Adding samples are no needs count.( because acceptable number decrease at
adding.) If too many samples added, sampling cannot be proceeded. Scale tells
you by the sign Sub. Meaning subtract quantity.
Wait to stable after beep then flicker :ADD (on WEIGHT display)
Also change to the acceptable addition quantity.
5. Continue the previous operation until the scale automatically completes
sampling.
However, even before the conditioning, the reference unit weight may be fixed
on by pressing the SCS key at an optional time when the fixation is required.
(In this case, sampling accuracy is not guaranteed.)
Features
-Operators can just concentrate on counting a few pieces.
-Just following the instruction brings accurate sampling.
-Re-sampling is not necessary in most cases.
-No sampling error. (not depend on operators)
-Scale rejects samples in failure of unit weight set.
10

5-3. Re-Sampling
1. After unit weight is set, load or add (other) samples.*
*too many sampling will cause sampling error.
To avoid an error, bar graph is very useful.
2. Press RE key.
5-4. Counting
1.) Normal counting
Load all the ( remaining ) samples by degrees seeing bar graph.
*if the bar graph goes beyond max/min limits, Re-Sampling should be made
again.
2.) Accumulate counting
1. Unit weight has been set.
2. Load partial samples.
3. Press ADD key.
4. Remove samples.
5. Load other samples in the same way.
6. Press ADD key.
7. Remove samples.
8. If you press MODE key, quantity on thedisplay changes from quantity
for the loaded samples to the accumulated quantity and vice verse.
11

3.) Fixed unit weight method
# Enable the bar graph when using this counting method
1. Unit weight has been set.
2. Load or add (more) samples in a way the bar graph does not go beyond
min/max limits. In other words, limit the samples’quantity in this
condition.
3. Press RE key.
4. Repeat the same procedure until you can complete counting.
5-5. Memory look-up
1. Enter ID number with numerical keys
2. Press No. key.
5-6. Memory Store
#Memory store with currently set weight
Call up an ID to save data using Memory look-up.
Unit weight set
1. Enter unit weight with numerical keys.*
2. Press UNT.W. SET key.
*If you fail to enter weight and press numerical keys, the current unit
weight in the working area will be stored.
12

Tare set
1. Enter tare weight with numerical keys. *
2. Press TARE PRESET key.
*If you fail to enter weight and press numerical keys, the current tare
weight in the working area will be stored.
[To clear data in MEMORY BANK]
1. Call up an ID to clear data from memory using memory look-up.
2. Press and hold down theC/AC key until the UNIT W. displays goes blank.
The ID memory data (both UNIT W. and TARE) are cleared.
Memory store with currently set weight
Unit weight and/or tare weight are set in the working area.
1. Call up an ID to save data using memory look-up.
2. Press UNT.W. SET key.
Unit weight is saved in the ID.
3. Press TARE PRESET key.
Tare weight is saved in the ID.
4. Return to the working area, press RE key.
13

5-7. Memory data call up in working area
1. Enter ID number with numerical keys.
2. Press No. key. (Memory data look-up.)
3. Press R.M.key. (The data set in the ID is called up to the working area.)
[To clear the working area data]
To clear data settings (both Unit W. and Tare),
press and hold down the C/AC key until the displays goes blank.
5-8. Limit set and Comparing
[Limit Set]
Limit function must be set enable ( see function table and set).
You can set either one point or two points limit.
Press LMT key. Li‑LO appears at Weighing display, and low limit value shows
that currently set at UNIT.W display.
Enter low limit value with numerical keys.
Press LMT key. Low limit set, and a beep will sound to sign. Then Li‑HI
appears at Weighing display, and high limit value shows that
currently set at UNIT.W display, then advance to the high
limit setting.
Enter high limit value with numerical keys.
Press LMT key. High limit set, and a beep will sound to sign. High limit set,
and go back to working area.
14

[Look up limit values]
Press LMT key. Li‑LO appears at Weighing display, and low limit value shows
that currently set at UNIT. W display.
Press LMT key. Li‑HI appears at Weighing display, and high limit value
shows that currently set at UNIT.W display.
Press LMT key. Go back to working area.
In case that only one point(13.Pn. 1) is selected to set, low limit value shows after
press LMT key at once go back to working area.
[Comparing ( how limit function work )]
Suppose two limits have been set;
When the quantity is under the low value, it is ranked as “LO “
When the quantity is between the low value and the high, it is ranked as “OK “
When the quantity is over the high value, it is ranked as “HI “
Notes
# Such settings as lower limit value >= Higher limit value will be error.
# Comparative judgment condition;
At one point(13.Pn. 1) :
LO : value < lower limit. OK: lower limit =< value.
At two point(13.Pn. 2) :
LO: value < lower limit
OK: lower limit =< value =< higher limit
HI : higher limit< value
# We get beep sound in response to comparing result.
The condition is selectable by function set.
15

5-9. Function set
A.) Function set
1.) Press MODE key until QUATITY display shows “Func “.
2.) The display then changes to “1. A.0 1 “
To change setting parameter, press Tkey
3.) To forward the next title, press MODE key.
To exit function setting mode and enter the working area, press any key
without pressing the MODE, Tkeys to return to the working area.
16

B.) Function tables
1. SEL. limit function *1: disable 2: enable
11.Co. enable when *1: anytime 2: in stable state
12.Li. enable where 0: except zero point or below(minus value).
*1: any area.
13.Pn. limit setting point(s) 1: one *2: two
14.bu. buzzer enable where *0: disable in any area
1: LO 2: OK 3: HI
4: LO+OK 5:OK+HI 6:LO+HI
2.S.S. SCS mode selection
*1: normal SCS mode
2: high grade SCS mode (precise sampling)
3.b.G. bar graph indication
0: disable
*1: count gap (fraction)
2: reminding for capacity
4.A.0 Auto Zero
0: not effective
*1: automatically adjusts to zero point
5.A.P. Auto power off
(effective with battery option) 0: not effective
*1: automatically turns off of power
6.S.d. stable judgement 1: rough …(* 2)…4: precise
7.d.i. weighing display resolution select
*1: normal mode 2: high resolution mode
8.IF. Interface
0: output is not effective
1: constant serial output (6-digit format)
*2: constant serial output (7-digit format)
3: RK type output format
A.un. unit selection *0: g 1: oz 2: lb
* mark shows the status set at the shipping time.
#other functions are not mentioned here because they are not related to operations.
17
This manual suits for next models
5
Table of contents
Other Shinko Scale manuals
Popular Scale manuals by other brands

Dini Argeo
Dini Argeo MCW Safety and maintenance manual

Soehnle
Soehnle STATIVWAAGE 7730 user manual

Intelligent Weighing Technology
Intelligent Weighing Technology AHS Series Operation manual

ISHIDA
ISHIDA Uni-5 Service manual

ADE
ADE Bella operating manual

American Weigh
American Weigh AMW-DISC-55 user manual