Shinyei Technology FMM-MD User manual

Operation Manual
IM-Z-11002 Rev.3

Table of Contents
Table of Contents
1.Introduction......................................................................................................................1
1-1Supplied Components List .........................................................................................1
1-2Warning.......................................................................................................................2
1-3Cautions......................................................................................................................3
2.General Description...........................................................................................................4
2-1Specification ...............................................................................................................5
2-2Instrument Features....................................................................................................5
2-2-1Base Unit(FMM-MD).........................................................................................5
2-2-2Accessory.............................................................................................................6
2-3Formaldehyde Sensor Assy(FMM-SA)..................................................................6
2-3-1Description............................................................................................................6
2-3-2 Attach and Remove the sensor assy......................................................................7
2-4Default Settings...........................................................................................................7
2-5Measuring Procedure Examples.................................................................................8
2-6Function......................................................................................................................9
2-6-1Internal memory....................................................................................................9
2-6-2LED .......................................................................................................................9
2-6-3Alert Beep.............................................................................................................9
2-6-4Keypad Sound ......................................................................................................9
2-6-5Language..............................................................................................................9
2-6-6Power....................................................................................................................9
2-6-7 LCD(Display)....................................................................................................9
3.Measurement...................................................................................................................10
3-1 Direct Measurement(Use base unit with sensor assy inserted.)............................10
3-2 Samplers Measurement(Use sensor as stand-alone passive sampler).................14
3-3 Continuous Measurement (Use base unit with sensor assy inserted) .......................19
4.Settings............................................................................................................................23
4-1Display Setting..........................................................................................................23
4-1-1Date and Time setting.........................................................................................23
4-1-2Power save mode ...............................................................................................23
4-1-3LCD Brightness...................................................................................................24
4-2Functions ..................................................................................................................24
4-2-1Units....................................................................................................................24
4-2-2Keypad Sound ....................................................................................................25
4-2-3Language............................................................................................................25
4-3Initialize .....................................................................................................................25
5.Viewing............................................................................................................................26
5-1Data Viewing .............................................................................................................26
5-2Battery Status............................................................................................................27
5-3Temperature and Humidity .......................................................................................27
6.Error.................................................................................................................................28
6-1No Memory Space.....................................................................................................28

Table of Contents
6-2Saturated Sensor ......................................................................................................28
6-3Data Error(Samplers)............................................................................................28
6-4Reading Error............................................................................................................29
6-5Malfunction ...............................................................................................................29
6-6Data Error..................................................................................................................29
7.Unit Dimension................................................................................................................30
8.Maintenance and Servicing.............................................................................................31
8-1Replacement of button battery.................................................................................31
8-2Troubleshooting........................................................................................................32
8-3How to clean the unit ................................................................................................32

Introduction
1
1.Introduction
1-1Supplied Components List
Formaldehyde Multimode Monitor FMM-MD kit has been packaged carefully and
includes all the components necessary for full operation.
Upon receipt, check the kit contents carefully to find you received the following items
in good condition.
The instrument kit contains;
Formaldehyde Multimode Monitor
FMM-MD Base unit 1
Formaldehyde Sensor Assy with diffusion filter
FMM-SA 5
“AA” Alkaline Battery 2
AC Adaptor 1
USB connection cable 1
CD-ROM(Data filing software)1
Operation Manual 1

Introduction
2
1-2Warning
WARNING
●Do not use any adaptors except supplied AC Adaptor.
●Do not use in explosive atmosphere
The meter is not designed for explosion proof. Do not use it where combustible
and/or explosive gases exist.
●Do not dismantle the base unit.

Introduction
3
1-3Cautions
CAUTIONS
●Do not install, use, or store the unit at a place where:
- Ambient temperature is not in the range of -10 to 40℃. (14-104°F)
- Ambient humidity exceeds 90% R.H.
- Condensation occurs due to rapid temperature change.
- Corrosive and/or explosive gases exists.
- Direct vibration and/or steam occurs.
- Mush dust, salty materials, and/or iron powders exist.
- Static electrons, magnetic, and/or electronic noise can occur.
- Cool or warm air from conditioners has direct influence.
- The unit is exposed to direct sunlight.
- Heating occurs due to heat radiation.
●Do not turn off the power during detection.
●Do not remove the
button battery besides exchanging the battery is
needed. It causes loss of the the internal memory data and initializes
Date and Time.
- Exchange new batteries if battery status displays “Low.”
- Remember to transfer the the internal memory data to your PC before taking
out the battery.
- Taking out the battery causes loss of the internal memory data
. Check if no
necessary data is in the memory.
●Store the sensor assy at cool and dark place (below 25°C/77°F) in
the
sealed state aluminum bag before and after opening.
●
Accustom the sensor assy at ambient temperature before opening the
aluminum bag.

General Description
4
2.General Description
FMM-MD consists of portable base unit and detachable formaldehyde sensor assy.
The sensor assy can be used as stand-alone passive sampler and also can be used
in combination with base unit.
The sensor assy can be used repeatedly up to the usage limit.
Base unit reads absorbance change induced in the sensor before and after the each
measurement respectively and converted result into formaldehyde concentration.
Measurement can be done in multiple modes as follows;
1. Direct Measurement mode(Use base unit with sensor assy inserted.)
Base unit in combination with the sensor works for normal Direct Measurement for
duration of 30 min or 1 hour.
2. Samplers Measurement mode (Use sensor assy alone as passive sampler
and base unit reads out measurement result.)
Sensor itself can work as stand-alone passive sampler. No fan, nor pump necessary.
Use base unit to read the relative absorbance change induced in the sensor before
and after the sampling and covert that into formaldehyde concentration.
3. Continuous Measurement mode(Use base unit with sensor assy inserted.)
Base unit reads the absorbance change every 30 min and store the data sequentially.
All the measurement results can be viewed on LCD panel, and also filed into PC with
using Data filing software.

General Description
5
2-1Specification
2-2Instrument Features
2-2-1Base Unit(FMM-MD)
Operating
Temperature
Range -10 to 40℃(14-104°F)
Operating Humidity
Range 20-90%rh(No dew condensation)
Storage
Temperature
Range -20-50℃(-4-122°F)/below 90%rh (No dew condensation)
Detection Range 20ppb-1000ppb
Accuracy ±10% at 40, 80, 160ppb
Resolution 1ppb
Concentration Unit
ppb orμg/m3
Power Source Two “AA” size alkaline batteries or AC adaptor
Display Digital LCD display (128×64 LCD)
Humidity Sensor RHI-112D(Shinyei Technology Co., Ltd.)
Memory Up to 250 sensors and 4500 data
150
90
46.9
46.9
3
4
5
6
2
1
7

General Description
6
No.
Name Function
①LED Green light indicates detection in progress.
Flashing light indicates the error.
②Keypad The device function operated by pressing four buttons
firmly.
③Power Switch Turn on and off the device.
④Tripod Socket
Hole The hole is used for mounting the device on tripod stand.
⑤USB Port Use attached USB cable to connect the device to PC.
⑥AC Adaptor Port AC adaptor is connected to the port.
⑦Sensor Port A formaldehyde sensor is inserted to the port.
2-2-2Accessory
Name Function
AC adaptor This adaptor is exclusively used for FMM-MD. Do not use it for
other purposes.
CD-ROM Data filling software. It can be installed into your PC to transfer
data from FMM-MD.
USB Cable Data stored in the internal memory can be sent to PC with this
cable.
2-3Formaldehyde Sensor Assy(FMM-SA)
2-3-1Description
・Use the sensor assy till the date shown on the sealed bag.
・ The sensor assy is sealed with the diffusion filter capped.
・Store the sensor assy with the diffusion filter attached at cool and dark place(below
25℃/77°F) in the sealed state aluminum bag before and after opening.
・Accustom the sensor assy at ambient temperature before open the aluminum bag.
・For reliable measurement result, the sensor is designed to secure its primarily
performance by going through the prior exposure to measurement environment for
60 min, before the actual measurement.
・Be sure to attach diffusion filter to the sensor assy for Samplers Measurement,
unless reliable measurement cannot be detected.
・The sensor assy can be used repeatedly up to the usage limit.
・Use fresh (new) sensor assy to avoid unintended interruption of measurement,
when you plan for long-term measurement and/or expect high concentration
measurement such as ≧200ppb or so.

General Description
7
2-3-2 Attach and Remove the sensor assy
Insert sensor assy without diffusion filter for detection.
2-4Default Settings
*Taking out the button battery causes initializing to the default settings.
Date and Time :Setting*
Power Save :ON
LCD Brightness :0
Units :ppb
Keypad Sound :ON
Language :English
Temperature Unit :°F
*Taking out the button battery causes Date and Time setting as below.
「2000/01/01 00:00:00」
Date and Time is set by two ways:
・Set by the base unit. (Go to pp.23)
・Set by PC. (See Data filing software manual)
Sensor Assy Insertion Slot
Take off diffusion filter.

General Description
8
2-5Measuring Procedure Examples
Mounting the device on tripod stand
(Direct/Continuous Measurement)Hanging the sensor assy
(Samplers Measurement)

General Description
9
2-6Function
2-6-1Internal memory
Detection data is stored in the internal memory of FMM-MD up to 4500 data.
You can see the data from the base unit or PC by using Data filing software.
(See Data filing software manual)
2-6-2LED
Green light indicates detection in progress.
Green light blinks when error occurs.
2-6-3Alert Beep
Alert works if error occurs or if the usage limit is exceeded.
*Alert does not work if power save mode is selected.
2-6-4Keypad Sound
Keypad sound can be turned on or off from Setting menu.
2-6-5Language
Japanese or English can be selected.
2-6-6Power
Use two “AA” alkaline batteries or AC adaptor.
*Battery Life…1 week if Continuous Measurement(at25℃/77°F)/Power Save
Mode ”ON”/”AA” Alkaline batteries are used.
(Battery life varies depending on usage environment.)
2-6-7 LCD(Display)
Full
Good
Direct 10:12
Sampling No. = 001
PREV
NEXT
Low
Measurement Mode
Time
Back to the previous screen. Go to the next.

Measurement – Direct Measurement
10
3.Measurement
3-1 Direct Measurement(Use base unit with sensor assy inserted.)
Base unit in combination with the sensor works for normal Direct Measurement
for duration of 30 min or 1 hour.
1.Select “Direct Measurement” from the menu screen.
2.Enter Sampling No.
3.Select Sampling Period.
●“Sampling No.” can be used to assign identifying number to the sensor so that
you may manage and locate the target measurement data within multiple
measurement data. (Assignable up to 250)
●All stored data are identified by sampling number.
●Writing down sampling number to the sensor assy is recommended.
Direct 10:12
SamplingNo. = 001
PREV NEXT
Move pointer to slide digit places.
Change number to set Sampling No.
Back to the previous screen.
Go to the Next.
Move pointer to select sampling period.
Go to the next.
Back to the previous screen.
SamplingPeriod
30min.
60min.
PREV
NEXT
MeasurementMode
Direct
Samplers
Continuous
Move pointer to select “Direct Measurement.”
Go to the next.

Measurement – Direct Measurement
11
4.Insert a formaldehyde sensor assy.
5.Retrieving sensor status.
6.Usage Expectancy information is displayed.
7.Prior exposure procedure.
Direct 10:12
Retrieving
Sensor Status…
Direct 10:12
Insert
SensorAssy
PREV
Direct 10:12
InitialCheck
to Start
PREV START
Direct10:12
PriorExposure
remains 58m30s
SKIP
Back to the previous screen.
Current sensor status shows it can be repeatedly used
for up to 50 times of 30m
in. Direct Measurement at
80ppb.
The sensor can be repeatedly used for up to 20 times of
30min. Direct Measurement at 200ppb.
Go to the next.
(The next screen appears automatically in 15 sec.)
*To skip prior exposure procedure, please go to pp.13.
Sens
or automatically enters prior exposure
procedure for 60 min.
Once
prior exposure is completed, baseline
value detection(next step) starts automatically.
UsageExpectancy
50times@80ppb
20times@200ppb
NEXT
Go to the next step.
Back to the previous screen.
*Expectancy information
can be fluctuated depending on temperature of the
environment and concentration of formaldehyde.
*
Do not eject sensor assy while sensor status is
retrieved.
*Erro
r screen appears when memory overflow
occurs. (pp.28)
*Insert the sensor assy without diffusion filter.

Measurement – Direct Measurement
12
8.Retrieving baseline value of the sensor.
9.Once baseline value of the sensor is stored, data number (data ID) is displayed.
10.Measurement in progress
11.Upon measurement completes, concentration value is displayed.
12. Measured data successfully stored.
Direct 10:12
SamplingNo.:001
Result: 15ppb
Record
(Data number (Data ID) /Total number of storable data)
Dataretrieved
Sampling No.:001
( 1/4500) The next screen appears automatically in 5 sec.
Quick estimation in every 10 min is displayed.
(Display only. Not stored into memory)
Direct 10:12
Measurement・・・
Complete in 29m
Stop <10ppb
(Data number (Data ID) /Total number of storable data)
DataRecorded
SamplingNo.: 001
(1/4500)
EXIT
*Error screen appears when memory overflow occurs. ( pp.28)
Direct 10:12
Retrieving
Baselinevalue…
*Do not eject sensor assy while sensor s
tatus is
retrieved
*Error screen appears when memory overflow
occurs. (pp.28)
Press to shut down the measurement
Store the measurement result.
Finish the Direct Measurement. Go ba
ck to
the screen for another measurement.

Measurement – Direct Measurement
13
☆Skipping Prior Exposure procedure
●For reliable measurement result, the sensor is designed to secure its
primarily performance by going through the prior exposure to measuremen
t
environment for 60 min, before the actual measurement.
●In case this procedure is skipped, calculated correction
will apply to the
measurement result.
●
In case the sensor reaches the saturation point, measurement is
automatically stopped. LED light indicates when this occurs.
●Use fresh (new) sensor assy to avoid unintended interruption of
measurement, when you plan for long-
term measurement and/or expect
high concentration measurement such as ≧200ppb or so.
Direct10:12
PriorExposure
remains 58m30s
SKIP
When you can not afford to go through prior
exposure procedure, you may skip this.
Skip the procedure to go to baseline detection.
(to item 8 on pp.12)
*Going through Prior Exposure p
rocedure is recommended for more reliable
measurement result.

Measurement – Samplers Measurement
14
3-2 Samplers Measurement(Use sensor as stand-alone passive sampler)
Use base unit to read the relative absorbance change induced in the sensor
before and after the sampling and convert that into formaldehyde concentration.
1.Select “Samplers Measurement” from the menu screen.
2.Enter Sampling No.
3.Select “Baseline Value” of the sensor.
*If baseline value of the sensor is assigned to the same Sampling No. that laps over old
measurement data, you will be requested to select YES or NO whether you mean to
renew the baseline value or not.
Samplers 10:12
SamplingNo. = 001
PREV NEXT
Samplers10:12
BaselineValue
MeasuredResult
PREV NEXT
●”Sampling No.” can be used to assign identifying number to the sensor so
that you may manage multiple measurement data. (Assignable up to 250)
●All stored data are identified by sampling number.
●Writing down sampling number to the sensor assy is recommended.
MeasurementMode
Direct
Samplers
Continuous
Move pointer to select “Samplers Measurement.”
Go to the next.
*Error screen appears when memory overflow occurs. (pp.28)
*Before measurement
, take out sensor assy from the sealed bag and leave it in
measurement environment for approx 60 min as “Prior exposure procedure”
is
recommended for more reliable measurement result.
Go to the next.
Back to previous screen.
Move pointer to select “Baseline Value.”
Move pointer to slide digit places.
Change number to set Sampling No.
Back to the previous screen.
Go to the Next.

Measurement – Samplers Measurement
15
4.Select whether or not the sensor went through prior exposure procedure.
5.Insert a formaldehyde sensor assy.
6.Retrieving sensor status.
7.Usage Expectancy information is displayed.
Expectancy information can be fluctuated depending on temperature of the environment and
concentration of formaldehyde.
Samplers10:12
Retrieving
SensorStatus…
Samplers10:12
Insert
SensorAssy
PREV
Samplers10:12
InitialCheck
to Start
PREV START Back to the previous screen.
Current sensor status shows it ca
n be repeatedly used for
up to 20
times of 30 min Samplers Measurement at 80
ppb.
The sensor can be repeatedly used for up to 24 hours at
less than 140 ppb.
Go to the next.
(The next screen appears automatically in 15 sec.)
UsageExpectancy
20times@ 80ppb
24hours@140ppb
NEXT
Go to the next step.
Back to the previous screen.
*
Do not eject sensor assy while sensor status is
retrieved.
*Error screen
appears when memory overflow
occurs. (pp.28)
*Insert the sensor assy without diffusion filter.
PriorExposure?
Yes
No
EXIT DONE
Go to the next.
Back to previous screen.
Move pointer to select.

Measurement – Samplers Measurement
16
8.Once baseline value of the sensor is stored, data number (data ID) is displayed.
9.Start measurement:
Eject a formaldehyde sensor assy and cap the diffusion filter, then place it to the
desirable measurement environment to start detection.
To handle multiple measurements with multiple sensors at the same time, repeat
Step. 2 – 8 and retrieve baseline value of each sensor with unique Sampling No.
respectively.
(Assignable up to 250)
10.Calculate Measurement result.
Recall the Sampling No. you set before the measurement.
Samplers 10:12
SamplingNo. = 001
PREV NEXT
(Data number (Data ID) / Total number of storable data)
Dataretrieved
Sampling No. : 001
(1/4500)
EXIT Back to Sampling No. screen.
*Error screen appears when memory overflow occurs. (pp.28)
*Be aware of sensor usage expectancy information.
*Be sure to cap diffusion filter to the sensor assy, otherwise reliable measurement
result cannot be expected.
*Sampling period and temperature information must be
noted which is necessary at
later stage to calculate measurement result.
Samplers 10:12
BaselineValue
MeasuredResult
PREV NEXT Go to the next.
Back to the previous screen.
Move pointer to select “Measured Result.”
Change number to set Sampling No.
(pp14. 3-2 2.)
Move pointer to slide digit places.
Back to the previous screen.
Go to the Next.

Measurement – Samplers Measurement
17
11.Insert a formaldehyde sensor assy.
12.Retrieving sensor status.
13.Enter temperature and sampling period.
14.Upon measurement completes, concentration value is displayed.
Samplers 10:12
Temp. Period
77°F 000h30m
PREV NEXT
Samplers 10:12
SamplingNo.: 001
Result: <15ppb
PREV RECORD
Move pointer to slide digit places.
Change number to set necessary information.
Go to the next.
Back to previous screen.
Samplers 10:12
Retrieving
SensorStatus…
Samplers 10:12
Insert
SensorAssy
PREV
Samplers 10:12
InitialCheck
toStart
PREV START Back to the previous screen.
Go to the next step.
Back to the previous screen.
*
Do not eject the sensor assy while sensor status
is retrieved.
*Error screen appears when memory overflow
occurs. (pp.28)
*Insert the assy with diffusion filter uncapped.
Store the measurement result.
Back to previous screen for re-entering
information.
Table of contents
Popular Monitor manuals by other brands

Ronan
Ronan X16PDM Series Hardware manual

Renkforce
Renkforce RF-SW-130 operating instructions

HANNspree
HANNspree HT161CGB2EX user manual

Panasonic
Panasonic BTLH2550E - LCD VIDEO MONITOR operating instructions

ViewSonic
ViewSonic VA502mb-1 Service manual

NEC
NEC LCD1990SX - MultiSync - 19" LCD Monitor user manual