SIAM DDIM-2 User manual

Tomsk Scientific Industrial Introduction Company “SIAM” LLC
DYNAMOMETER «DDIM-2»
Operating manual
Russia

Dynamometer “DDIM-2”

Dynamometer “DDIM-2”
CONTENTS
1. GENERAL PRODUCT INFORMATION 5
2. SAFE OPERATION GUIDELINES 6
2.1. General Provisions (regulatory framework) 6
2.2. Requirements to Personnel 6
2.3. Requirements to Test Objects 7
3. BASIC SPECIFICATIONS 9
4. ENSURING EXPLOSION-PROOFNESS OF THE PRODUCT. 11
4.1. Explosion-proofness of the product 11
4.2. Measures to ensure and keep the device explosion-proofness
during the assembly, disassembly and repair of the device. 12
5. DELIVERY SET 14
6. PREPARATION OF TEST OBJECTS FOR DYNAMOMETRY 15
6.1. Dynamometry sensor installation 15
6.2. Dynamometry sensor removal 16
7. PRODUCT DESIGN AND OPERATION 16
7.1. General information 16
7.2. Dynamometry Sensor 16
7.3. Switching On and Switching Off 17
8. SENSOR OPERATION IN THE “DB SIAM” SOFTWARE 18
8.1. Connecting the Bluetooth sensor to a computer 18
8.2. Working with sensors in “DB SIAM” 20
8.3. Study procedure 25
9. WORK WITH SENSORS IN THE “SIAM SERVICE” PROGRAM 26
10. STORAGE AND TRANSPORTATION OF THE PRODUCT 28
11. CLAIMS 29
APPENDIX 1. 31

Dynamometer “DDIM-2”
The operating manual applies to “DDIM-2” dynamometer
(hereinafter referred to as the device) designed for operation at the
wellhead equipped with a sucker-rod pumping unit (SRPU) in oil and gas
fields. The device is explosion-proof designed (intrinsically safe circuit
type of protection) in accordance with the requirements of GOST 31610.0-
2014 (IEC 60079-0:2011), GOST 31610.11-2014 (IEC 60079-11.), GOST
31610:2011), and have 1Ex ib IIB T3 Gb Хexplosion proof mark.
“DDIM-2” is designed for indoor and outdoor installation in hazardous
areas where atmospheric explosives of IIA, IIB categories and T1, T2, T3
groups can form according to GOST R IEC 60079-20-1-2011 in the
operating temperature range from -40 ºC to +50 ºC. The device design is
comply with the assigned explosion proof mark, with the requirements of
GOST IEC 60079-14-2013 and other regulatory documents governing the
use of electrical equipment in hazardous areas.
The operating manual provide the details about technical
specifications of the device, delivery set, service life, storage and
manufacturer's guarantees, as well as other information about the device
used for the operation process.
Observe the requirements of the Federal norms and regulations in
the field of industrial safety "Safety rules in oil and gas industry"as well as
this manual, when operating, maintaining and repairing the device.
The device may only be operated by specially trained personnel
who have studied the operating documentation for the device, received the
relevant safety instructions and permission to work.

Dynamometer “DDIM-2”
1. GENERAL PRODUCT INFORMATION
Please read the operating instructions before using the
dynamometer “DDIM-2”.
The device is explosion-proof designed (intrinsically safe circuit
type of protection) in accordance with the requirements of GOST 31610.0-
2014 (IEC 60079-0:2011), GOST 31610.11-2014 (IEC 60079-11.), GOST
31610:2011), and have 1Ex ib IIB T3 Gb Хexplosion proof mark.
“DDIM-2” device is designed for indoor and outdoor installation in
hazardous areas where atmospheric explosives of IIA, IIB categories and
T1, T2, T3 groups can form according to GOST R IEC 60079-20-1-2011
in the operating temperature range from -40 ºC to +50 ºC. The device
design is comply with the assigned explosion proof mark, with the
requirements of GOST IEC 60079-14-2013 and other regulatory
documents governing the use of electrical equipment in hazardous areas.
The inter-gaging period recommended for the device is 1 year. To
check the device transmission characteristics for compliance with the TS
requirements and to bring them into compliance with the TS, calibrate the
device on the dynamometer control bench on the expiry of inter-gaging
period.

Dynamometer “DDIM-2”
2. SAFE OPERATION GUIDELINES
2.1. General Provisions (regulatory framework)
The design of the wellhead equipment is to be approved by the
Russian Technical Surveillance Agency (RTSA).
Preparation of the well for tests should be conducted in accordance
with the requirements of the current Manual and internal guidelines
for maintaining and testing wells approved by the head of the
company.
Test equipment is to be operated according to the operating
instructions supplied with the equipment.
Measures for ensuring safety are regulated by labor protection
instructions for respective types of work approved by the
Company’s Department of Labour Protection and the Safety
Regulations in Oil and Gas Industry approved by the RTSA.
2.2. Requirements to Personnel
Well tests are to be carried out by individuals above 18 who have
secondary education and who are medically cleared for work.
Personnel are to take professional training followed by knowledge
checking and have a relevant permit category in accordance with
qualification requirements: not below category 5 for oil and gas
production operators; not below category 4 for well test operators.
Personnel are to be trained to work with test equipment.

Dynamometer “DDIM-2”
2.3. Requirements to Test Objects
Pressure measurement and control of fluid level is carried out in oil
and gas producing wells of different operational modes (flowing, gas lift,
artificial lift, etc.), as well as in injection, intake, control and other wells
driven by a beam pumping unit of SKN series according to GOST 5866-
56, SKD according to GOST 5866-76, SKD according to OST 26-16-08-
87, including all the standard and similar foreign manufacture sizes.
Wells are to be tied and operated in accordance with the Safety
Regulations in Oil and Gas Industry.
When controlling the dynamogram, the rod string is to end with a
polished rod 16… 39 mm in diameter.
The polished rod is to be sealed with SUS1 or SUS2 gasket. The
rod and the beam pumping unit are to be connected with a clipping
of the wellhead rod of PSH type. The limits of the clipping
configuration are shown in Figure 1.

Dynamometer “DDIM-2”
НЕ МЕНЕЕ 45 мм
НЕ МЕНЕЕ 105 мм
Figure 1.
The pumping unit control station is to have a mode select switch
with the manual control option. The moving parts of the pumping
unit (the crank-conrod mechanism, the V-belt drive) are to be
properly fenced.
The service platform shall comply with the requirements of the
"Safety Rules in Oil and Gas Industry". Scheme of the platform is
shown in Figure 2.

Dynamometer “DDIM-2”
Figure 2. Wellhead equipment
If the height of the polished rod packing is low (up to 2 m above
ground level), portable platforms are to be used provided they
comply the above requirements.
3. BASIC SPECIFICATIONS
The device is designed for operation at the wellhead equipped with a
sucker rod pumping unit (SRPU), in oil and gas fields and provides:
Measurement of the load change on the polished rod of the SRPU
within (0 - 10000) kgf, movement of the polished rod within (500 -
9999) mm of at least one oscillation cycle of the SRPU horsehead.
Measured data record and saving in a non-volatile storage device.
Wireless transfer of stored parameters to an external device.
Оборудование
устья скважины

Dynamometer “DDIM-2”
The device maintains its operability at ambient temperatures from
minus 40 C to + 50 C.
The device operates autonomously and is powered by a special
internal battery with 3.6 V voltage. The minimum operating voltage of the
battery which does not lead to a loss of performance of the device is 1.8 V.
The basic specifications are given in Table 1.
Table 1
Parameter name
Norm on specifications
Explosion proof mark
1Ex ib IIB T3 Gb X
Upper limit of load change, kgf
10000
Load channel resolution, (kgf)
/displacement, (m)
10 / 0,010
Displacement measurement range, m
0,5 –9,999
Permissible SRPU horsehead pump
speed, SPM
1,5 –15
Permissible SPRU polished rod
diameter, mm
16 –37
Continuous operation time in
registration mode (once a day), at least,
years
2
Max. working power, mW
30
Weight, kg
1

Dynamometer “DDIM-2”
4. ENSURING EXPLOSION-PROOFNESS OF THE PRODUCT.
4.1. Explosion-proofness of the product
Explosion-proofness of the device is provided by the type of
protection "intrinsically safe electric circuit" “ib” level, according to the
requirements of GOST 31610.11-2014 (IEC 60079-11:2011), which is
achieved by the following:
Basic circuit diagram includes the protective component “Fib”. The
protective component “Fib” is an intact spark protection unit with the
short-circuit current limitation at 1.85 A (maximum), using current limiting
resistors and semiconductor fuses in series. Double redundancy is used in
the “Fib” protective component to ensure greater reliability. The “Fib”
component is integrated into the battery compartment of the device. The
design of the protective component “Fib” is met with the requirements of
GOST 31610.11-2014 (IEC 60079-11:2011), including leakages and
clearances. The minimum width of conductors on the PCB is 0.2 mm,
copper thickness is not less than 18 µm. Thus, the electrical circuit coming
out of the battery compartment of the device is intrinsically safe.
The electrical circuit principal and applied third party components
provide maximum power consumption not more than 0.030 W from the
internal battery with maximum possible voltage on it Uo=3.6 V. The total
maximum capacity of the electrical circuit is 10 μF, maximum inductance
is not more than 20 µGn. The maximum current in the circuit during
normal operation is not more than 7 mA.

Dynamometer “DDIM-2”
The device is powered by a special internal battery type ER18505
according to GOST 31610.11-2014 (IEC 60079-11:2011).
The maximum temperature of overheating of components and
connections in the circuit diagram during normal operation is 15 °C max.
Thus, the surface temperature of conductors and elements during operation
and at maximum operating temperature plus 50 ° C is not more than 65 °
C. Device enclosures are made with the degree of protection against
external influences not lower than IP54 according to GOST 14254-2015.
4.2. Measures to ensure and keep the device explosion-proofness
during the assembly, disassembly and repair of the device.
4.2.1 Measures to ensure explosion-proofness before the production
process:
- The elements used in the protective component of “Fib” are
subjected to an acceptance test:
- resistors are tested for their nominal resistance;
- the fuses are checked against the short-circuit current limit;
- Materials used for casting the protective component “Fib” are
subjected to an acceptance test according to the certificates presented.
4.2.2 Measures to ensure explosion-proofness during production:
- The protective component “Fib” with the installed battery is
subjected to an outgoing inspection:

Dynamometer “DDIM-2”
- the short-circuit current and open circuit voltage at the output of
the protective component are checked;
- visual inspection of the filling location is carried out in order to
confirm the absence of any foreign inclusions, bubbles, cracks or
stratification.
4.2.3 Measures to ensure explosion-proofness during operation:
- Check that the cover of the battery compartment and the
compartment itself (including the threaded connection) are free from
mechanical damage and corrosion. In case of corrosion or mechanical
damage, do not operate the device;
- meet the requirements specified by the “X”sign in the explosion
proof mark.
4.2.4 Measures to ensure explosion-proofness during repair:
- the device must be repaired only by the employees of the
manufacturer.
Repair by third parties or individuals is prohibited.
- the protective component “Fib” is not to be repaired. It is made
intact and must be replaced in case of failure. Repair of the “Fib”
component is not permitted.
- after repair, the device must be tested according to the test
procedure, with the “Fib” component tested for short-circuit current and
open-circuit voltage (together with the battery installed).

Dynamometer “DDIM-2”
5. DELIVERY SET
The device is supplied according to the set shown in Table 2.
Table 2
Title
Q-ty
1. Dynamometer «DDIM-2»
1
2. Certificate
1
3. Flash drive with the software «Siam Service 2»
1
4. «Siam Service 2» software user’s guide
1
5. Mounting plate IZM 8.683.235
2
6. Package
1

Dynamometer “DDIM-2”
6. PREPARATION OF TEST OBJECTS FOR
DYNAMOMETRY
Check the availability and serviceability of the handbrake, balancer
horsehead lock, protective fences, service platform.
If the beam pumping unit is operated in remote control mode or in
automatic mode, switch the control to manual according to the
instructions.
Hang up a sign on the starter: “Do not switch on, people work”.
Make sure that the wellhead packing has no fluid passes at any rod
position.
6.1. Dynamometer installation
Stop the beam pumping unit in its lower position, not reaching the
extreme 20 -25 cm and secure it with the handbrake.
Unload the clipping and spread the cross-arms according to the
approved technology.
Install the device into the space between the cross-arms, install the
safety mechanism. Lower supports should rest on the lower cross-
arm in the whole plane. If the crosshead has an uneven surface, a
steel gasket of appropriate thickness and configuration is to be
used.
Load the clipping and smoothly align the cross-arms according to
the approved technology.

Dynamometer “DDIM-2”
6.2. Dynamometer removal
Stop the SRPU in the lower clipping position and secure with the
brake.
Reload the clipping and smoothly extend the cross-arms according
to the approved technology.
Remove the device from the space between the cross-arms.
Remove any dirt from the device and put it in its original bag.
Load the suspension and smoothly align the cross-arms according
to the approved technology.
Start the SRPU.
7. PRODUCT DESIGN AND OPERATION
7.1. Dynamometer
The dynamometer is designed for measuring and recording
displacement and value of the load on the polished rod while the sucker-
rod pumping unit is in operation.

Dynamometer “DDIM-2”
Figure 3: DDIM-2 appearance
7.3. Switching On and Switching Off
The dynamometer designed without on/off button is triggered via a
smartphone via Bluetooth.
For the correct operation, the device must be held at a temperature
of at least 10 min before measuring.

Dynamometer “DDIM-2”
8. DEVICE OPERATION IN THE “DB SIAM” SOFTWARE
8.1. Connecting the Bluetooth device to a computer
The Bluetooth module must be activated on the computer/tablet
from which the wireless devices will be operated. The first time you use
each device with this computer, you must connect the device using
Windows - "Add Bluetooth device". The procedure for connecting may
vary depending on the OS version.
The device must be turned on and ready for connection. The sign of
readiness is the green color of the LED "Power". As a result of the
connection, a COM port is created through which the SIAM Database
software establishes a connection to the device.
To see which COM port has been created for a specific device,
proceed as follows:
point the mouse cursor over the Bluetooth icon in the right
part of the taskbar, click the right mouse button;
in the appeared context menu select item "Open parameters"
(see Figure 4);
in the appeared Bluetooth Options window on the COM
Ports tab, see which outgoing port is assigned to the connected device (see
Figure 5).

Dynamometer “DDIM-2”
Figure 4: Bluetooth Menu in Windows 7-10

Dynamometer “DDIM-2”
Figure 5. Bluetooth settings. Assigned Bluetooth COM ports to
devices.
In figure 5, the outgoing port COM20 is assigned to device DDIN
No. 356.
8.2. Working with devices in “DB SIAM”
To work with devices with the “DB SIAM” software:
start the program;
switch to the “Settings” tab;
press the “Devices” button (see Fig.6);
the “Devices” dialog box will appear (see Fig.7):
Figure 6. The “Settings” tab in the “DB SIAM”