manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Landis+Gyr
  6. •
  7. Measuring Instrument
  8. •
  9. Landis+Gyr ULTRAHEAT T350 Series User manual

Landis+Gyr ULTRAHEAT T350 Series User manual

en
Operating Instructions Ultrasonic Heating and Cooling Meter
ULTRAHEAT
®
T350
(2WR6…)
ULTRACOLD
®
T350
(2WR6…)
26.05.2014
Translation
T350 (2WR6…)
Landis+Gyr GmbH Subject to change without prior notice 26.05.2014 1
Note:
These operating instructions remain with the end user following activation.
Note:
In the following text, the term meter refers to both
the heat
ing meter and the cooling meter, unless other-
wise specified.
1. General
1.1 Use
The T350
(2WR6...)
is used as a meter for heating or cooling
consumption measurement in systems with water.
The meter consists of two temperature sensors and an elec-
tronic unit which calculates the energy consumption from the
volume and temperature difference.
Note:
The meter cannot be opened without damaging
the
security seal.
1.2 General notes
The meter left the factory in a faultless condition where safety
is concerned. The manufacturer will provide additional tech-
nical support on request. Calibration relevant security seal on
the meter must not be damaged or removed. Otherwise the
guarantee and calibration validity of the meter will lapse.
•Keep the packaging so that you can transport the meter in
its original packaging following expiry of the calibration va-
lidity.
•Lay all cables at a minimum distance of 500 mm to high
voltage and high frequency cables.
•A relative humidity of < 93 % at 25 °C is permissible (with-
out condensation).
•Avoid cavitation in the whole system due to overpressure
i.e. at least 1 bar at qp and approx. 3 bar at qs (applies for
approx. 80 °C).
•
At a heating meter the mounting place of the flow sensor
cold side is equivalent to return. The mounting place of the
flow sensor hot side is equivalent to flow.
•At a cooling meter the mounting place of the flow sensor
hot side is equivalent to the return. The mounting place of
the flow sensor cold side is equivalent to flow.
2. Safety Information
The meters may only be used in building service
engineering systems and only for the applications
described.
The local regulations (installation etc.) must be ad-
hered to.
Adhere to the operating conditions according to the
dial plate during use. Non-adherence can cause
hazards and the guarantee will lapse.
Adhere to the AGFW requirements regarding circulat-
ing water (FW510).
The meter is only suitable for circulating water in
heating systems.
The meter is not suitable for drinking water.
Do not lift the meter by the electronic unit.
Be aware of sharp edges on the thread, flange and
measuring tube.
Only personnel, trained in the installation and opera-
tion of meters in heating and cooling systems, may
install and remove the meter.
Only install or remove the meter when the pipes are
pressure-less.
After installing the meter, check the leak-tightness of
the system.
Guarantee and calibration validity will lapse if the
calibration relevant security seals are broken.
Only clean the meter from outside with a soft, lightly
wetted cloth. Do not use any spirit or cleaning solvent.
As far as disposal is concerned, the meter is a waste
electronic appliance in the sense of European Di-
rective 2002/96/EC (WEEE) and it must not be dis-
posed of as domestic waste. The relevant national,
legal regulations must be observed as the appliance
must be disposed of via the channels provided for this
purpose. The local and currently valid legislation must
be observed.
The meter contains lithium batteries. Do not dispose
of the meter and the batteries with domestic waste.
Observe the local stipulations and laws on disposal.
You can return the lithium batteries to the manufac-
turer for appropriate disposal following use. When
shipping please observe legal regulations, in particu-
lar, those governing the labelling and packaging of
hazardous goods.
Do not open the batteries. Do not bring batteries into
contact with water or expose to temperatures above
80 °C.
The meter does not have any lightning protection.
Ensure lightning protection via the in-house installa-
tion.
32 14 101 003 a
Landis+Gyr GmbH Subject to change without prior notice 26.05.2014 2
3. Operating
3.1 Operating elements
Note
:Both display range and data displayed can differ
from this description depending on the meter param
e-
teri
sation. Certain button functions can also be
blocked.
Switching the display
Please proceed as follows to switch between the display val-
ues:
•Press the button briefly to show the next line of the current
loop.
The first display value will appear again after the final display
value.
•Press the button for a longer time (longer than 10 s) in
order to display the service loop.
•Press the button for 3 s in the service loop in order to leave
the service loop.
User loop ▼
User loop Display type
0054567 kWh Energy accumulated
00065.43 m³ Volume accumulated
888888 kWh Segment test In o
F - - - In case o error: message with error
code
In o
Service loop ▼
Service loop Display type
0.534 m³/h Current low
22.9 kW Current power
84 47 °C Current temperature hot side / cold
side
04.06.11 D Date
786 Bh Operating hours
56 Fh Missing hours
3792701 G Device number, 7 digit In o
PulSE
CH Remote read out mode (optional) In o
123 A Primary address or M-Bus option In o
2345678 K Property number, 7 digit In o
18.02.11 F0 Date stamp or F0 early warning In o
3- 01 FW Firmware version In o
31.12.10 V Storage day previous year Previous year
0034321 kWh Energy previous year on set day Previous year
00923.12 m³ Volumes: previous year on set day Previous year
12 Fh Missing hours previous year Previous year
- -.- -.- - C Code input or parameterisation
Service loop Display type
01.06.11 M Storage day month 1-15 Previous month
over 3s button press:
0034321 kWh Energy previous month on re er-
ence date
Previous month
00923.12 m³ Volume previous month on re er-
ence date
Previous month
12 Fh Missing hours previous month on
re erence date
Previous month
3.2 Monthly values
The meter stores the following values for 15 months on the
monthly set day
•Energy (meter status)
•Volume (meter status)
•Missing hours (meter status)
When the meter shows the monthly set day, press the button
for 3 s in order to display the previous month's values.
You can also read the monthly values via the optical interface.
4. Functional Details
If the respective operation thresholds are exceeded and flow
and temperature difference are positive, the meter summates
the energy and the volume. All segments of the display are
switched on for control purposes during the segment test.
The meter status for energy, volume and missing time is en-
tered into a log for the previous year on an annual basis on the
annual set day.
The flow, power and temperature difference are recorded with
the appropriate +/- signs. If the operation threshold has
dropped, a "u" is displayed in front. The current temperatures
are presented together as whole numbers in °C in a display
line.
You can set the 8 digit property number (secondary address
for M-Bus operation) in the parameterising mode. The highest
value point is suppressed in the display. The device number is
issued by the manufacturer.
The operating hours are counted from the first connection of
the power supply. Missing hours are summated if there is a
fault and the meter is thus unable to take measurements. The
date is incremented on a daily basis.
The number for the firmware version is issued by the manufac-
turer.
5. Error Messages
The meter continuously runs a self-diagnosis and can thus
recognise and display various installation or meter error mes-
sages.
The error message F0 means that no flow measurement is
possible, due to air in the volume measurement unit for exam-
ple; the system must be vented carefully.
The error message F4 means the battery must be replaced.
When any of the error messages F1, F2 or F5, F6, F8 are
displayed, the temperature sensors are defective. The mes-
sages F3, F7, F9 mean a defect in the electronics.
Contact the service department in all these cases.
Landis+Gyr GmbH Subject to change without prior notice 26.05.2014 3
6. Technical data
Note: The information on the meter must be observed!
General
Measuring accuracy Class 2 or 3 (EN 1434)
Environment class A (EN 1434) or indoor installation
Mechanical class M1 *)
Electromagnetic class E1 *)
*) according to 2004/22/EC Directive on Measuring Instruments
Ambient humidity < 93 % rel. humidity at 25°C without condensation
Max. height 2000 m above sea level
Storage temperature - 20 … 60°C
Electronic unit
Ambient temperature 5… 55 °C
Housing protection rating IP 54 according to EN 60529
Power supply Battery or 6 or 11 years
or 24 V AC/DC external (optional)
Operation threshold . ∆T 0.2 K
Temperature di erence ∆ T 3 K … 80 K
Temperature measurement
range
2 ... 180 °C
National type approvals may be di erent.
LCD indication 7 digit
Optical inter ace Standard, EN 1434-3
Communication Optional: M-Bus or pulse output
Protocol IEC 870, 300 baud in Nb
Separability Always detachable, cable length 1 m
Sensor
Type Pt 500 according to EN 60751
Connection type 2-wire technology, ixed connection
Cable length 1.5 m (optional 5 m)
Construction type DS direct short, M10 × 27.5 mm or PS pocket short
ø 5.2 x 45 mm according to EN 1434
Temperature range 0 ... 105 °C
National type approvals may be di erent.
Max. temperature 105 °C
Volume measurement unit
Protection class IP 54 / IP 65 according to EN 60529
Mounting place Hot side / cold side
Installation position Any, horizontal or vertical
Smoothing section None
Measuring range Approved 1:100, calibrated 1:50
Temperature range 5… 105 °C
National type approvals may be di erent!
recommended or...
...heating applications 15… 105 °C
...cooling applications 5… 50 °C
Maximum overload qs = 2 × qp, permanent
Nominal pressure PN16 (1.5 MPa; PS16)
qp Overall length and connection
0,6 110mm (3/4'') 190mm (1'')
1.5 110mm (3/4'') 190mm (1'')
2.5 130 mm (1'') 190mm (1'')
7. EU-Directives Declaration of Conformity
Landis+Gyr GmbH, Humboldtstr. 64, D-90459 Nürnberg here-
by declares that the type Ultraheat T350 / 2WR6 meter meets
the requirements of the following directives:
−2004/108/EC Electromagnetic Compatibility of Electrical
and Electronic Devices
−2006/95/EC Low Voltage Directive
−2004/22/EC Directive on Measuring Instruments
−2011/65/EU Directive on the Restriction of the Use of Cer-
tain Hazardous Substances in Electrical and Electronic
Equipment (RoHS2)
This declaration and the associated documents are deposited
with Mr. Fuchs c/o Landis+Gyr with the number CE 2WR6
007/03.14.
EC Type Approval Certificate
DE-06-MI004-PTB007
Certificate recognising the quality management system
DE-12-AQ-PTB006MID
Notified Body:
PTB Braunschweig and Berlin, Germany; identification number
0102 0102
Landis+Gyr GmbH
Humboldtstrasse 64
90459 Nuremberg
Germany

Other manuals for ULTRAHEAT T350 Series

1

This manual suits for next models

1

Other Landis+Gyr Measuring Instrument manuals

Landis+Gyr E230 User manual

Landis+Gyr

Landis+Gyr E230 User manual

Landis+Gyr ULTRAHEAT 50 Series User manual

Landis+Gyr

Landis+Gyr ULTRAHEAT 50 Series User manual

Landis+Gyr ULTRACOLD 2WR6 Series User manual

Landis+Gyr

Landis+Gyr ULTRACOLD 2WR6 Series User manual

Landis+Gyr ULTRAHEAT XS User manual

Landis+Gyr

Landis+Gyr ULTRAHEAT XS User manual

Landis+Gyr ULTRAHEAT T350 Series User manual

Landis+Gyr

Landis+Gyr ULTRAHEAT T350 Series User manual

Landis+Gyr 750 Guide

Landis+Gyr

Landis+Gyr 750 Guide

Landis+Gyr E650 Series 4 User manual

Landis+Gyr

Landis+Gyr E650 Series 4 User manual

Landis+Gyr Cashpower Gem Lite User manual

Landis+Gyr

Landis+Gyr Cashpower Gem Lite User manual

Landis+Gyr E450 User manual

Landis+Gyr

Landis+Gyr E450 User manual

Landis+Gyr ULTRAHEAT T230 User manual

Landis+Gyr

Landis+Gyr ULTRAHEAT T230 User manual

Landis+Gyr ZMQ202 User manual

Landis+Gyr

Landis+Gyr ZMQ202 User manual

Landis+Gyr E470 Series User manual

Landis+Gyr

Landis+Gyr E470 Series User manual

Landis+Gyr ULTRAHEAT T230 Assembly instructions

Landis+Gyr

Landis+Gyr ULTRAHEAT T230 Assembly instructions

Landis+Gyr UH50 Series User manual

Landis+Gyr

Landis+Gyr UH50 Series User manual

Landis+Gyr G130 User manual

Landis+Gyr

Landis+Gyr G130 User manual

Landis+Gyr T450 User manual

Landis+Gyr

Landis+Gyr T450 User manual

Landis+Gyr CASHPOWER SABRE MK 5 User manual

Landis+Gyr

Landis+Gyr CASHPOWER SABRE MK 5 User manual

Landis+Gyr E650 Series User manual

Landis+Gyr

Landis+Gyr E650 Series User manual

Landis+Gyr T450 User manual

Landis+Gyr

Landis+Gyr T450 User manual

Landis+Gyr FOCUS kWh User manual

Landis+Gyr

Landis+Gyr FOCUS kWh User manual

Landis+Gyr Ultracold T550 User manual

Landis+Gyr

Landis+Gyr Ultracold T550 User manual

Landis+Gyr T450 Parts list manual

Landis+Gyr

Landis+Gyr T450 Parts list manual

Landis+Gyr T450 User manual

Landis+Gyr

Landis+Gyr T450 User manual

Landis+Gyr E650 Series User manual

Landis+Gyr

Landis+Gyr E650 Series User manual

Popular Measuring Instrument manuals by other brands

Klein Tools ET110 instruction manual

Klein Tools

Klein Tools ET110 instruction manual

dacell PDN-10 user manual

dacell

dacell PDN-10 user manual

LaMotte WaterLink SpinTouch 3580-AU manual

LaMotte

LaMotte WaterLink SpinTouch 3580-AU manual

SCALTEC SMO 01 Installation and operating instructions

SCALTEC

SCALTEC SMO 01 Installation and operating instructions

DynAmp LKP Series Installation, operation and service manual

DynAmp

DynAmp LKP Series Installation, operation and service manual

Elster Instromet RABO G65 instruction manual

Elster Instromet

Elster Instromet RABO G65 instruction manual

Biuged Laboratory Instruments BGD 160/1 instruction manual

Biuged Laboratory Instruments

Biuged Laboratory Instruments BGD 160/1 instruction manual

Anritsu Spectrum Master MS2721B Maintenance manual

Anritsu

Anritsu Spectrum Master MS2721B Maintenance manual

OTT PLS 500 operating instructions

OTT

OTT PLS 500 operating instructions

HEDÜ D104 quick start guide

HEDÜ

HEDÜ D104 quick start guide

Chauvin Arnoux C.A 1310 user manual

Chauvin Arnoux

Chauvin Arnoux C.A 1310 user manual

ADE WS 2322 operating instructions

ADE

ADE WS 2322 operating instructions

Endress+Hauser Dosimag operating instructions

Endress+Hauser

Endress+Hauser Dosimag operating instructions

Fema Electronica S40-R user manual

Fema Electronica

Fema Electronica S40-R user manual

Sparex S. 24540 quick start guide

Sparex

Sparex S. 24540 quick start guide

Automation Dr. Nix CarCheck System PLUS user manual

Automation Dr. Nix

Automation Dr. Nix CarCheck System PLUS user manual

Haas VCQPS troubleshooting guide

Haas

Haas VCQPS troubleshooting guide

Agilent Technologies Acqiris DC110 user manual

Agilent Technologies

Agilent Technologies Acqiris DC110 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.