GasSampleDrawPanel(5398-21)providesaneasyto
use method of drawing an air sample from a remote, or
hostile area to a more convenient location for analysis.
The Sample Draw Panel can use gas sensor modules
for Combustible or Toxic gases, either the Sentry style
modules or the Analog (4-20 mA output) modules. It
can use gas sensor modules for Combustible or Toxic
gases.
Theintegral sample pump to draw thesample at a high
flowratetoreducesample time. Thesampleis delivered
tothe sensorup-streamfromthepumptoavoid negative
pressure conditions. A bypass line exhausts excess
samplevolumetoavoidhighflowratesacrossthesensor.
Each panel is a self-contained system with the pump,
flow indication, valves to control sample draw and
calibration gas delivery, and the gas detection sensor
and transmitter (priced separately) all housed in an
enclosure. The flow systems design will deliver the
sample to the sensor with minimal pressure affecting
thesensorasthe excess flow and pressure are bledoff
viathebypassflowmeter.
Model 5398-21
Gas Sample Draw Panels
Specifications:
Power: 110 VAC @ 1 amp
Pump: OilFree,DiaphragmType
MaximumSampleDraw: 600feet with 1/4” tubing
Flow Rates: 0-500 cc/min.Adjustable sample to sensor
MaximumOperatingPressure: 50PSIG
Controls: 3-wayvalve forRunandCalibration
Outputs: Dry contact pump flow failure switch (contacts closed when
flowisdetected,open for alarm)
Filter: Coalescingfilterfor moisture knock-out for each sampleline
TubingConnections: 1/4" TubeFittingsfor connection of:
- sample lines - sample exhausts
- calibration gas - bypass exhaust
Enclosure type: NEMA4XFiberglass with window
Dimensions (H x W x D): 16.9 x 15.0 x 8.3 inches (42.9 x 38.1 x 21.1 cm)
Weight: 17.6 lbs (8.0 Kg) - without gas sensor module
Sensor/Transmitter: MountingprovisionsforallstandardSierraMonitorCorp.
SentryandAnaloggas sensormodules
Ordering Information:
Model5398-21 Single Channel Gas Sample Panel
Sierra Monitor Corporation 1991 Tarob Court Milpitas, CA 95035 USA (408) 262-6611 (800) 727-4377 Fax: (408) 262-9042
Visit our Web site: www.sierramonitor.com E-Mail: sierra@sierramonitor.com
© 2009 by Sierra Monitor Corporation
Model 5398-21 Dimensions
Model 5398-21
TheSampleDraw Panels areintendedforinstallationin
an area, which is accessible, by the user for periodic
routine inspection and calibration. The panel requires
120 VAC input and provides local visual as well as
remoteindication of pump and flow status.