
 
- 7 - 
5.3.1.3 In the turning on state, press key“
”for about 2 seconds, this instrument is turned 
off and saved current setting information. 
5.3.2 Measurement 
5.3.2.1 Examples and conditions 
There are two ways of measurements, normal measurement(flat surface or camber ) and 
continuity measurement, users can select it on the measurement mode, the selection ways and 
conditions are as follows: 
Example1: normal measurement 1”shortcut operation mode” (Factory setting mode) 
This  is  a  pseudo  temperature  compensation  mode,  it  doesn’t  need to consider the temperature 
value and coefficient, as long as the following two conditions are met. 
①Temperature of the material to be measured is almost the same as the standard temperature 
block, use the value of the standard block at 20℃to calibrate, then measure the conductivity for 
the testing material.. 
②The material temperature coefficient of the material to be measure is the same as the 
temperature of the standard block. If users want to measure the conductivity of the unknown 
material of copper and aluminium sheet, it is right to calibrate as the same temperature coefficient 
in the standard block (temperature coefficient of copper and aluminium is -0.004). 
2.Operation methods 
①When the material to be measured is of flat surface material, refer to “measurement mode” of 
the  menu  5.3.4.2  to  select  “close  camber  compensation”,  and  set  “temperature  compensation 
mode”  of  the  menu  5.3.4.3  into  pseudo  temperature  compensation.  Now  both  the  temperature 
value TAof measurement interface and standard interface and temperature coefficient value a0are 
display invalid “None”, that is to say it doesn’t have to consider the current temperature value and 
select the temperature coefficient when measuring and calibrating. As long as calibrate the 
higher/lower value before the measurement. Then return to measurement interface to measure the 
conductivity of the material. (refer to 5.3.3.4~5.3.3.6 and 5.3.2.8) 
②When testing pieces is of“convex”or“concave”material. Go to the “measurement interface” of 
menu  5.3.4.2  select  “opening  curvature  compensation”.  For  “convex”  camber,  first  change  the 
hand grip φ14mm probe into “ ”style, then calibrate higher and lower value on the calibration 
interface, return to the measurement interface, select the value η(φ) of curvature compensation 
coefficient  (refer  to  5.3.2.6),  adjust“ ”grip  to  the  suitable  position(refer  to  5.3.2.2-b), then 
measure conductivity of the camber material. 
③When measuring the thin plates or small pieces(area between φ8mm~φ14mm), should 
change to φ8mm(500KHz)style probe, select φ8mm-500KHz probe in the “frequency and probe” 
of menu 5.3.4.6, the measuring method is the same as ①. 
Example2: normal measurement2 “Different temperature operating mode” 
This mode is used that calibrations and measurements can’t be operated under the 
same temperature circumstance(or calibration temperature isn’t the same as that of 
testing material), such as tracking the conductivity change of the material under 
different temperatures, and fasting measure the conductivity under non-normal 
temperature conditions. This operation mode is relatively cumbersome, but it is 
suitable for any occasion, the accuracy is much higher. (refer to 6.1.2.2 “automatic