SMART 433-CMU Operating and maintenance manual

Pager
433-CMU
Wireless Central Monitor and
Nurse Call Paging System
Installation and Use Instructions
+_
A23-12V
433-PGD
Caegiver Pocket Pager with
LCD Display
433-PG
Caegiver Pocket Pager
433-NC
Wireless Nurse Call Button
Optional Accessories
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE / ADD
RECORD/ ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
Pager

2
This page intentionally left blank.

PAGER
DING DONGMUSIC OFF
PAGER
+
+
-
-
AA AA
+_
A23-12V
C C
C C
3
1 Insert Batteries into Each Component
433-CMU Economy Central Monitor
433-MS Motion Sensor (optional) 433-NCA Adaptor (optional)
433-EXT Window/Door Exit Sensor (optional) 433-PG Caregiver Pager (optional)
433-PGD Caregiver Pager w/ Display (optional)
433-NC Nurse Call Button (optional)
+_
A23-12V
+_
A23-12V
Front
Top
Inside
Insert a flathead screwdriver
into notch and gently twist
until the button pops open.
Gently pry open
using tab at top
of 30A-EXT.
Back of pager
Front of
pager
Batteries already
installed.
Back of CMU
Batteries not included
Batteries not included
Batteries not included
Batteries not included
Batteries included
Batteries included
+
+
-
-
AA AA
+
-
AA
Back of
433-EXT
Inside of
433-EXT
+
+
-
-
AA AA
+_
A23-12V

WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
433-CMU
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
433-CMU
4
b
3 Program Components to Work with Central Monitor
2 Adjust Language and Volume on Central Monitor and Pager
433-NC Nurse Call Button (optional) Programming Instructions
Adjust volume by pressing
“Volume” buttons on Central
Monitor and/or optional pager.
Choose from 3 volume options
by repeatedly pressing “Volume
Button” until you nd desired
volume. Monitor is shipped
with alarm volume set at 90
decibels (90 decibels is the
legally accepted occupational
noise allowance).
Press and release the Nurse Call
Button. The Central Monitor will
speak and display a number.
Number shown is the Nurse Call
Button’s identification number
on the Central Monitor. Wait for a
melodic tone before programming
additional buttons/components
or exiting.
Before programming the Central
Monitor and components, make sure
the switch on the side of the Central
Monitor is set to “Unlock.” When all
programming is complete, prevent
tampering by setting this switch to
“Lock.”
Press and release the next Nurse
Call Button or other component
that you would like to program,
repeating step b until all
buttons/components have been
programmed.
Call
Alarm
Unlock
Lock
Pager
VOL LANGUAGE REVIEW
Pager
VOL LANGUAGE REVIEW
To adjust language or
select a song, press the
“Language” buttons
on the Central Monitor
and/or optional pager
(ON/OFF switch on side
of pager must rst be
switched to “ON” or ).
Press “Language” button
repeatedly to scroll
through the available
options. When you nd
your desired selection,
stop pressing the button.
433-PG 433-PG
Before programming the Central Monitor
and components, be sure to read through
directions for all of the components you will be
programming to work with your Central Monitor.
Make sure you are prepared to program all of
the components before you begin. The Central
Monitor automatically exits programming mode
after 30 seconds of inactivity. If you are not
prepared with all of your components, you may
be forced to start again at the beginning.
433-NC
433-NC
+_
A23-12V
+_
A23-12V
WARNING: If you do not wait for the Central Monitor to make a melodic tone before pressing the next Nurse Call Button or other
component, each component will be programmed to the same number on the Central Monitor. In this situation, the Central Monitor will
speak one number for the component, but show a different number to represent the room. When multiple components are programmed to
the same number on the Central Monitor, THE NUMBER SPOKEN WILL NOT CORRESPOND WITH THE NUMBER DISPLAYED. This may be
useful when multiple devices are being used in the same room, but not for any other situation.
433-CMU
dPress the “Reset” button to
exit programming mode
when programming is
complete.
a c
Press and hold the
“Record/Add” button
until you hear 2 dings
and a melodic tone.
433-CMU

WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WARNING: If you do not wait for the Central Monitor to make a melodic tone before pressing the next Nurse Call Button or other
component, each component will be programmed to the same number on the Central Monitor. In this situation, the Central Monitor will
speak one number for the component, but show a different number to represent the room. When multiple components are programmed to
the same number on the Central Monitor, THE NUMBER SPOKEN WILL NOT CORRESPOND WITH THE NUMBER DISPLAYED. This may be
useful when multiple devices are being used in the same room, but not for any other situation.
a
d
d
c
c
b
433-NCA Monitor Adaptor (optional) Programming Instructions
Plug Monitor Adaptor into
Nurse Call Port located on the
side of desired the Fall Alarm
Monitor (TL-2100 series)
Plug the sensor pad, floor mat,
or seat belt into phone-style port
at bottom of desired Fall Alarm
Monitor (TL-2100 series)
Press sensor pad, step on floor mat,
or buckle and unbuckle seat belt.
Central Monitor will speak and
display a number. Number shown is
the Monitor Adaptor’s identification
number on the Central Monitor.
Repeat steps “a” and “b” for all Monitor Adaptors that you plan to
add into the Central Monitor system.
Voice
Reset
Safety Auto Reset
Reset
Status
Caregiver
Key
TM
Pad & Monitor
Active when Blinking
Pad & Monitor
Active when Blinking
Low Battery Pad Lost
Voice
Reset
Safety Auto Reset
Reset
Status
Caregiver
Key
TM
Pad & Monitor
Active when Blinking
Pad & Monitor
Active when Blinking
Low Battery Pad Lost
Press and hold “Record/
Add” button until you
hear 2 dings and a
melodic tone
Sensor
Pad
fWait for the Central Monitor to make
a melodic tone. Then, repeat step “d”
with the next Monitor Adaptor or
follow programming directions for
another component that you would
like to program.
e
a
b
433-MS Motion Sensor (optional) Programming Instructions
Press and hold the
“Record/Add” button until
you hear 2 dings and a
melodic tone.
Turn on the Motion Sensor and
wave a hand in front of the “eye.”
The Central Monitor will speak and
display a number. The Number
shown is the Motion Sensor’s
identification number on the Central
Monitor.
ON OFF
Press “Reset” button to exit
programming mode.
Wait for the Central Monitor to make
a melodic tone. Then, repeat step “b”
with the next motion sensor or follow
programming directions for another
component that you would like to
program.
Press “Reset” button to exit
programming mode.
433-NCA
30A-CMU
30A-CMU
433-CMU
433-CMU433-CMU
433-MS
5
Sensor Pad
Monitor

WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WARNING: If you do not wait for the Central Monitor to make a melodic tone before pressing the next Nurse Call Button or other
component, each component will be programmed to the same number on the Central Monitor. In this situation, the Central Monitor will
speak one number for the component, but show a different number to represent the room. When multiple components are programmed to
the same number on the Central Monitor, THE NUMBER SPOKEN WILL NOT CORRESPOND WITH THE NUMBER DISPLAYED. This may be
useful when multiple devices are being used in the same room, but not for any other situation.
433-CMU
433-CMU
433-CMU
433-CMU
433-CMU
433-CMU
a
b
a
e
c
c
CordLess® Transmitter (TM-03) (optional) Programming Instructions
Press on the sensor pad or step on floor
mat. The Central Monitor will speak and
display a number. The number shown is the
Cordless Transmitter’s identification number
on the Central Monitor.
Plug a CordLess® Transmitter into a CordLess®
Sensor Pad or Floor Mat. Repeat this step for
every CordLess Pad and Transmitter
you plan to use with the Central
Monitor system.
Press and hold “Record/
Add” button until you
hear 2 dings and a
melodic tone.
Press and hold “Record/
Add” button until you hear
2 dings and a melodic
tone.
Wait for the Central Monitor
to make a melodic tone. Then,
repeat steps “c” & “d” with the
next Monitor Adaptor or follow
programming directions for
another component that you
would like to program.
Wait for the Central Monitor to
make a melodic tone. Then, repeat
step “c” with the next Exit Alarm
or follow programming directions
for another component that you
would like to program.
Turn on the transmitter. The
ON/OFF switch located on the
side of the Transmitter.
Sensor
Pad
d
TM-03
CordLess®
Transmitter
ON/OFF Switch
b
433-EXT Door/Window Exit Alarm (optional) Programming Instructions
Touch the two parts of the Exit
Alarm together and then the Central
Monitor will display and speak the
number the Exit Alarm has been
programmed to.
f
d
Press the “Reset” button to
exit programming mode.
Press the “Reset” button to
exit programming mode.
433-EXT
TM-03
TM-02
6

7
Functions of the 433-CMU Economy Central Monitor
Test system by activating each component. For example,
press a Nurse Call Button or wave a hand in front of a Motion
Sensor that has been programmed into the Central Monitor.
Pager (optional) and Central Monitor should both alarm.
The Central Monitor and optional Pager with Display
(433-PGD) should show correct identification number
of each device as it is tested. The optional Pager will only
alarm if within 300 feet from a Central Monitor. Any
433-PG or 433-PGD Pager will respond to any 30-CMU
or 433-CMU Central Monitor within 300 feet. The Central
Monitor and Pager should continue to alarm until they
are silenced.
To silence the Central Monitor, press either the “Call
Review” button or the “Reset” button. When the
Central Monitor is silenced, any Pager responding
to that Central Monitor will also be silenced. For a
better understanding of “Call Review” and “Reset”
features, refer to pages 7 and 8 of these instructions.
ON OFF
VOL LANGUAGE REVIEW
Pager
VOL LANGUAGE REVIEW
up to 300 feet
IMPORTANT: The end user is responsible for testing system at its furthest range of use.
1) Call Review Button
2) Volume Button
3) Language Button
4) Replace/Add Button
5) Record/Add Button
6) Programming Light
7) Reset Button
8) Antenna
9) Hook
10) Battery Compartment
11) Base
12) Call/Alarm Switch
13) External Speaker Outlet
14) Lock/Unlock Switch
15) DC Input
9
10
11
12
13
14
15
433-PG
433-PG
433-NC 433-MS
433-CMU Front
433-CMU Back 433-CMU Left Side
To silence the pager, switch the ON/
OFF switch on the side to “OFF.”
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
433-CMU
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
433-CMU
3 Test the system - Central Monitor and Optional Pager
+_
A23-12V
1
1
2
3
4
5
6
7
8
Call Review
Reset

Call Review Button
Press once to silence the Central Monitor without erasing
call memory.
Press to review recent calls - press repeatedly to scroll
through last 6 calls.
Volume Button
Controls the volume of the Central Monitor - press
repeatedly to scroll through 3 volume options
Language Button
Controls the language of the Central Monitor - press
repeatedly to scroll through 11 language options (English,
French, German, Dutch, Spanish, Italian, Portuguese, Russian,
Japanese, Cantonese and Mandarin) and 3 tone options.
Replace/Add Button
Use this button when you wish to:
1. Add an additional Nurse Call Button or other component
to your system.
2. Replace an existing Nurse Call Button or other component
with a new Nurse Call Button or other component.
Call Review
Volume
Language
Replace/Add
To add an additional Nurse Call Button or other component to your system:
1. Press and hold the “Replace/Add” Button until you hear two dings and a melodic tone.
2. Press the “Record/Add” Button once. Wait for two dings. After the second ding, press and release the Nurse Call
Button, apply pressure to the pad being programmed, turn a motion sensor on, or touch the parts of an Exit Alarm
together. Wait for a melodic tone before programming any additional Nurse Call Buttons or other components.
The Central Monitor should speak and display the channel that the new component has been programmed to.
Refer to previous pages for programming directions specic to each component.
3. Press the “Reset” button to exit programming mode.
4. Test the system with the new Nurse Call Button or other new component.
To replace an existing Nurse Call Button or other component with a new Nurse Call Button or other component:
1. Press and hold “Replace/Add” Button until you hear 2 dings and a melodic tone.
2. Repeatedly press the “Call Review” Button to scroll through the numbers on the Central Monitor until you get to
the number of the Nurse Call Button or other component that you would like to replace. Press the new Nurse Call
Button or other new component. Central Monitor will ding when new Nurse Call Button or other new component
has been accepted. Refer to previous pages for programming directions specic to each component.
3. Press “Reset” button to exit programming mode.
4. Test system with new Nurse Call Button or other new component.
WARNING: If you do not wait for the Central Monitor to make a melodic tone before pressing the next Nurse Call Button or other
component, each component will be programmed to the same number on the Central Monitor. In this situation, the Central Monitor will
speak one number for the component, but show a different number to represent the room. When multiple components are programmed
to the same number on the Central Monitor, THE NUMBER SPOKEN WILL NOT CORRESPOND WITH THE NUMBER DISPLAYED. This may
be useful when multiple devices are being used in the same room, but not for any other situation.
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30

WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
Programming Light
Indicates when the Central Monitor is in programming mode. If the
light is on, the Central Monitor is in programming mode; if the light is
o, it is not in programming mode.
1. Press once to exit programming mode.
2. Press once to silence the Central Monitor and erase/clear call memory.
Extend the antenna fully to ensure the best possible reception and range
between the Central Monitor and any programmed Nurse Call buttons or
other components or to optimize the range between the Central Monitor
and pagers.
1. Slide the base off of the Central
Monitor
2. Screw base into wall so
that hooks face out
3. Slide the Central Monitor onto the base (now
screwed onto the wall) so that the hooks from
the Central Monitor connect to the hooks on
the base.
Programming
Light
Reset
Button
Antenna
Screw Holes
Hooks
Hooks
Base
Reset Button
Antenna
Wall Mounting the 433-CMU
To program Nurse Call Buttons or other components into a new or empty system:
1. Refer to previous pages for instructions specic to each component.
To erase all Nurse Call Buttons and other components currently programmed into the Central Monitor System:
1. Press and hold “Record/Add” Button until you hear 2 dings and a melodic tone.
2. Press “Reset” button to exit programming mode.
WARNING: Pressing the “Record/Add” button will erase/clear all programming from the Central Monitor. After pressing the “Record/Add”
button, all Nurse Call Buttons or other components that were programmed into the Central Monitor will no longer be recognized by the
Central Monitor until they are reprogrammed. Numbers will flash across the Central Monitor to indicate that no Nurse Call Buttons or other
components are currently programmed to work with the Central Monitor.
WARNING: Pressing the “Reset” button will erase/clear the Central Monitor’s
memory of call history so that the “Call Review” button will be unable to recall any
calls made prior to pressing the “Reset” button.
433-CMU
433-CMU
433-CMU Back
433-CMU Back 433-CMU
Back
9
Record/Add Button
Use this button when you wish to
1. Program Nurse Call Buttons or other components into a new or
empty Central Monitor system
2. Erase all Nurse Call Buttons and other components currently
programmed into the Central Monitor system
Record/Add
WARNING: If you do not wait for the Central Monitor to make a melodic tone before pressing the next Nurse Call Button or other component,
each component will be programmed to the same number on the Central Monitor. In this situation, the Central Monitor will speak one
number for the component, but show a different number to represent the room. When multiple components are programmed to the same
number on the Central Monitor, THE NUMBER SPOKEN WILL NOT CORRESPOND WITH THE NUMBER DISPLAYED. This may be useful when
multiple devices are being used in the same room, but not for any other situation.
WIRELESS CALL SYSTEM
CALL REVIEW
VOLUME
LANGUAGE
REPLACE/ ADD
RECORD/ADD
RESET
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30

• Call: When switched
to this setting, the
Central Monitor will
alert the caregiver
by voicing the
programmed number
of the activated
device.
• Alarm: Call (above)
is preceded by a siren
sound.
• Unlock when
programming
• Lock to prevent
tampering or loss
of programming.
To plug the Central
Monitor into a wall outlet,
use a Smart AC-04 9-Volt
AC adaptor.
An external speaker can
be plugged into this
outlet if you wish for
louder volume.
Call/Alarm
Unlock/Lock
DC Input
Speaker
Call/Alarm Switch
Unlock/Lock Switch DC Input
External Speaker Outlet
Call
Alarm
Unlock
Lock
Call
Alarm
Unlock
Lock
Call
Alarm
Unlock
Lock
Call
Alarm
Unlock
Lock
•Notes
• Low power aects performance. Symbol indicates that battery is low. Replace batteries
when this symbol appears on the Central Monitor.
• There is no ON/OFF switch for the Central Monitor. If you choose not to use the AC-04 AC adaptor
to power the Central Monitor and depend on batteries, it is recommended that you remove
batteries when the Central Monitor is not in use. When batteries are replaced, all Nurse Call
Buttons and other components that were programmed to work with the Central Monitor prior
to removing the batteries will still be programmed to work when the batteries are replaced. They
will not have to be re-programmed.
• Avoid outside interference. When in programming mode, be sure that no Nurse Call Button or
other component that you are not currently programming is active within range of the Central
Monitor (for example: do not press the Nurse Call Button, do not separate Exit Alarm). The
symbol indicates that a component is active within range of the Central Monitor.
• To ensure optimal wireless performance and reduce electrical interference, do not place the
Central Monitor close to metal objects or other electronic devices.
• Do not expose the Central Monitor to heat or direct sunlight.
• Avoid dampness and water.
• Numbers ashing across the screen of the Central Monitor indicates that there are no Nurse Call
Buttons or other components currently programmed to work with the Central Monitor.
433-CMU
433-CMU
433-CMU
433-CMU
10

This page intentionally left blank.

433-CMU Contents:
• 433-CMU Central Monitor
• 2 Screws
• 1 Mounting Bracket
433-CMU Optional Accessories
(sold separately):
• Caregiver Pager without LCD Display (433-PG)
• Caregiver Pager with LCD Display (433-PGD)
• Nurse Call Button (433-NC)
• Wireless Motion Sensor (433-MS)
• Wireless Bed & Chair Fall Alarms (Call for details)
• Exit Contact Switch for Door or Window (433-EXT)
• AC Adapter (AC-04)
Important Information:
30-Channel Receiver, Nurse Call Buttons and Pagers rely on wireless technology which is subject to physical and environmental considerations. These products
do not have an out of range function and as such should be tested periodically in the setting in which they are to be used to understand their area of eective
operation. Please be aware that the transmitter will not be 100% accurate if it is out of range at anytime. Wireless systems and monitors are intended as an adjunct
to good caregiving practices and are not a substitute for proper stang and patient management practices. We recommend that all personnel receive periodic
training in the operation of these systems and that the systems are tested before each use.
Warning: Failure to comply with warning may result in injury or death. This device is not suitable for all individuals. Other devices may be required. This device is
not a substitute for visual monitoring by caregiver. The manufacturer does not claim that this device will stop elopement and or stop falls. This device is to help
augment caregivers comprehensive resident mobility management program. Test this device before each use. Read the instructions and Legal Disclaimer.
Seller warrants that the goods sold will be free from defects of workmanship and manufacturing, for a period of one year from the date of sale. Other than this
limited warranty, seller makes no express or implied warranties. Except for this limited warranty, NO EXPRESS OR IMPLIED WARRANTIES ARE GIVEN AND NO
AFFIRMATION OF SELLER, BY WORDS OR ACTION, WILL CONSTITUTE A WARRANTY. THE GOODS BEING PURCHASED ARE BEING SOLD ON A “WITH ALL FAULTS”
BASIS. THE ENTIRE RISK AS TO THE QUALITY AND PERFORMANCE OF THE GOODS IS WITH THE BUYER. SHOULD THE GOODS PROVE DEFECTIVE FOLLOWING THEIR
PURCHASE, THE BUYER, AND NOT THE MANUFACTURER, DISTRIBUTOR OR RETAILER, ASSUMES THE ENTIRE COST OF ALL NECESSARY INSTALLATION, SERVICING OR
REPAIRS.
The Smart Caregiver Corporation 30-Channel Wireless Call System & Nurse Call Paging System are designed to be installed by the end-user. As such, it is the entire
responsibility of the buyer to insure that the system is properly installed and tested. Further, the system is not designed to replace good care giving practices
including, but not limited to: a) Direct patient supervision b) Adequate training for sta personnel for elopement c) Testing of the system before
each use
WHERE PERMISSIBLE, IT IS THE SELLER’S INTENT TO LIMIT ANY LIABILITY FOR SPECIAL, CONSEQUENTIAL OR INCIDENTAL ECONOMIC DAMAGES TO REFUND OF THE
PURCHASE PRICE OR REPLACEMENT OF THE GOODS. SELLER DISCLAIMS ANY LIABILITY FOR COMPENSATORY OR OTHER DAMAGES ARISING OUT OF THE USE OF
THE GOODS.
WARNING: The manufacturer does not claim that this device will stop elopement and or stop falls. This device is to help augment
caregivers comprehensive resident mobility management program. This product is to be used by a licensed caregiver only.
+_
A23-12V
433-PGD
Caegiver Pocket Pager with
LCD Display
433-PG
Caegiver Pocket Pager
433-NC
Wireless Nurse Call Button
Pager
Optional Accessories
WIRELESS CALL SYSTEM
CALL REVIEW