SMART CRUZER Series User manual

E) SETTINGS
Press the M button and then switch on the scale. The scale has the following
user settings in this menu:
1. DUMMY ZERO: The user can enable or disable the DUMMY ZERO display
with this setting. Select YES or NO depending on whether you want to display
the dummy zero or not.
2. AUTO ZERO: After dummy zero selection the display will show“A-0 X”.
Here the user can select the range of the autozero. “A-0 0” means auto
zero function disabled and the display will show even the first count from the
zero level. “A-0 3” means the auto zero range is 3d, meaning that the scale
will auto zero any reading between 0d and 3d at the zero level, to maintain
the correct zero on the scale.
F) CALIBRATION
External Calibration Setting
Press the ZERO Button and switch on the scale. The display will show “PASS”
and then “000000”. Press the ZERO Button. The display will show “CAL”. Press
the ZERO the display will briefly show “no Load” and then the internal division at
no load condition. Make sure for the decimal point to stop blinking, and then
press the ZERO Button. The display will show “LOAD” briefly and then again the
internal counts. Keep the calibration weight and press the ZERO Button, the
display will show “000000”. Now enter the value of the calibration weight kept
using the UP and LEFT Arrow Keys. Finally press the ZERO Button, the display
will show the calibrated weight. If the difference between this external
calibration and the previous internal calibration is more than 5%, the new
calibration will not be stored and the scale will pick up the values / setting
according to the old internal calibration only. If the user feels that the new
external calibration is correct and still the scale is not accepting it, then the
correct way to do the calibration will be to do the internal calibration again.
G) BATTERIES
The scale is capable of running on rechargeable battery or 6 pcs of AA size
batteries, if the scale is not connected to the mains through the adaptor for any
reason. The following is recommended to maximize the life of the batteries
•If the scale is not used for long, it is highly recommend to remove the
batteries from the scale, as there are chances that batteries may leak and
damage the scale beyond repair.
•Please charge the battery for at least 20 hours before using for first time.
•The battery automatically gets charging once the scale adaptor is plugged in to
the mains; hence only make sure that the GREEN LED on the top right side of the
display is ON.
G) MAINTENANCE
# The user should never put weight more than the designed capacity of the scale
# When the scale is switched on, whenever the user will put more load than the
designed capacity of scale, then display will show “-OL-” meaning that the scale is
overloaded. Even if the scale is no in use and is switched off, if load above the
designed capacity of the scale is applied on the weighing pan, by keeping some
weight, or by hand, or any other means, the scale can be damaged beyond repairs.
# Never clean the scale with highly intensive solvents like petrol, thinner or acetone
etc. the best way to clean the scale is to use a slightly moist cloth, and then use a dry
cloth to wipe again.
OWNERS MANUAL
SHIPPING SCALES
Owner’s Manual Shipping Scales Cruzer Series
4 1
Version 2.01
Cruzer
A) TECHNICAL DETAILS
B) SETTING UP THE SCALE:
Take the scale out of the box and place on a stable surface. Place the Top pan
correctly on the scale. Insert the DC pin of the adaptor, and connect the adaptor
to the nearest mains point and / or install 6 pcs of AA batteries by removing the
battery cover.
For best performance, do not place the scale on a surface where:
The surface has effect of Vibrations
The air movement is more, like directly under a fan, or in open in windy condition
ON / OFF Switch: The on/off switch is located on the rear side of the scales. The
switch has 3 positions. At the centre position the scale is turned off and can be
pushed left or right to run the scale on rechargeable battery or dry batteries.
Note : The scale has 3 power inputs a) Rechargeable battery, Dry Battery and
power adaptor.
C) OPERATION
a) Buttons
o ZERO: This button is used to zero an unwanted reading on the display. This
button is also used for going in to the calibration mode.
o HOLD: This button is used for holding the weight on the display.
o M / UP Arrow: This button is used for Setting the LO and High Limit in the
scale. The button is also used as the up arrow key in the setting menu.
o Counting / Left Shift Key: This button is used for counting function. This
button is also used as the left shift key in the setting menu.
b) Indications: There are three visual indications in the scale which are as follows:
o Power LED: When you connect the adaptor to the scale, the GREEN LED on
the bottom right side of the display will glow. When the LED will glow, you
can be sure the scale is being powered from the main supply, and that the
battery is being charged.
o Stable weight indication: In the weighing or the counting mode, when you
keep the weight you will notice that the decimal point is blinking. The blinking
decimal point means that the weight is not stable. Once the weight is
stable, the decimal point stops blinking and the user can be sure that the
weight being shown is the correct weight.
o Hold function: In the weighing mode if you press the hold button after
keeping the weight, the displayed weight will start blinking this means that
the hold function is active. Once the display stops blinking the hold mode is
turned off.
D) FUNCTIONS
a) HOLD Function: The scale has a hold function which can be used to hold the
weight display. In weighing mode, keep a weight and wait for the decimal point to
stop blinking. Once the decimal point is stable, now user can press the HOLD
button. The display will start blinking, and the weight will be hold on the display
without any movement if you add some weight or remove some weight. Also if
you remove the total weight, the previous weight will still be displayed at which
the hold button was pressed. Now to come out of the hold mode, simply press
the hold button again, or keep a new weight. The scale will automatically
understand a new weight has been kept, and it will exit the HOLD mode, and
display the new weight correctly.
b) SET POINT Function: The function is mainly used for packing or check weighing
function. The user can pre define the level at which he needs the alarm to sound.
When the display is showing “0.00”, press the M button, the display will show
“SET LO”, now press the zero button, the display will show “000000” or the
previous entered value. Enter a value using the Up arrow and the left shift arrow
and then press the Zero Button. Now the display will show “SET HI” and the user
can similarly enter the HI value. The following logic is followed in this mode:
o SET LO value is entered only: If the user only enters the SET LO value and
keep the SET HI value as “0.00”, the scale will sound the beep sound below
the SET LO value.
o SET HI value is entered only: If the user only enters the SET HI value and
keep the SET LO value as “0.00”, the scale will sound the beep sound above
the SET HI value.
o SET LO & SET HI both are entered: If the user enter both SET LO and SET HI
value, the beep will not sound below between the SET Lo and SET HI value
range, and the beep will sound below the LO value and above the HI value.
Please note the following points:
o The beep sound will sound only when the stable light stops blinking, that is the
weight is stable.
o The BUZZER is not provided as standard, and it needs to be installed by the
dealer / reseller. Without the buzzer installed, the beep sound cannot be
heard.
c) Counting Function: The scale has counting mode which can be used for parts
counting. In the weighing mode, keep a weight on the scale. When the decimal point
stops blinking, press the Counting Button. The display will show the value according
to the previously entered data. Now press the M button, the Display will show
“SM0010” which is selection mode for the number of pcs. Press the M button again
display will show “SM0020”, so that the user can select the appropriate number as
per the number of samples kept on the scale in the beginning. After correct selection
press the ZERO Button, the display will show the pieces counting. To go back to the
weighing mode, press the Counting Button again. You will come back to the weighing
mode and the display will show the weight. Press the counting button again and
again, to toggle between the counting mode and weighing mode.
Owner’s Manual Shipping Scales Cruzer Series
2
Owner’s ManualShipping ScalesCruzer Series
3
Measurement Data Weighing Scales
Capacity 20kg - 75kg
Tare range subtractive 100% subtractive
Repeatability +- 1d
Display Resolution 6000 ~ 10,000 divisions
Application Modes Weighing, Counting
Product Features Pre-Set Tare, Glow Control, Set HI, LO
Keyboard 4 membrane keys (Hol d, Mode, Counting, and Zero)
Weighing Units Kg
Calibration Digital with external weight.
Weight 1.8kg approx
Front Display Type 0.56” Bright red LED display
Stabilization 2 seconds
Power 9V Adaptor / built in battery for backup to
25hrs. / 6 x AA dry battery
Operating Temp. 50° to 98° F / 10° to 35° C

Other SMART Scale manuals

SMART HAX Series User manual

SMART

SMART HAX Series User manual

SMART SHSAP3000 User manual

SMART

SMART SHSAP3000 User manual

Popular Scale manuals by other brands

Klarstein 10030612 manual

Klarstein

Klarstein 10030612 manual

Aurora AU 4302 manual

Aurora

Aurora AU 4302 manual

HOFFEN KS-8310-18M instruction manual

HOFFEN

HOFFEN KS-8310-18M instruction manual

OHAUS EXPLORER Service manual

OHAUS

OHAUS EXPLORER Service manual

Beurer BG55 Instructions for use

Beurer

Beurer BG55 Instructions for use

AWS VANILLA-5K user manual

AWS

AWS VANILLA-5K user manual

HiMirror HDBS6160X56 user manual

HiMirror

HiMirror HDBS6160X56 user manual

Scarlett SC-BSD33E898 instruction manual

Scarlett

Scarlett SC-BSD33E898 instruction manual

Weight Watchers WW67WC Instructions for use

Weight Watchers

Weight Watchers WW67WC Instructions for use

Beurer BF 66 Instructions for use

Beurer

Beurer BF 66 Instructions for use

ADE Vanessa instruction manual

ADE

ADE Vanessa instruction manual

Princess 493000 Instructions for use

Princess

Princess 493000 Instructions for use

Sanitas SBG 21 Instructions for use

Sanitas

Sanitas SBG 21 Instructions for use

Conair WW72C Instructions for use

Conair

Conair WW72C Instructions for use

Omron FULL BODY SENSOR BODY COMPOSITION MONITOR AND SCALE... instruction manual

Omron

Omron FULL BODY SENSOR BODY COMPOSITION MONITOR AND SCALE... instruction manual

OHAUS Catapult 1000 quick start guide

OHAUS

OHAUS Catapult 1000 quick start guide

HoMedics SC-322 Instruction manual and  warranty information

HoMedics

HoMedics SC-322 Instruction manual and warranty information

Soehnle Body Balance Shape F3 operating instructions

Soehnle

Soehnle Body Balance Shape F3 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.