Sofraser 9000 User manual

Technical Manual
Viscosity and temperature
blind transducer
(To drive one MIVI sensor)
9000
Original version
REF.: 381/3

Technical Manual 9000
REF: 381/3 2
IMPORTANT
THE OFFSET ADJUSTMENT IN THE AIR
MUST BE THE FIRST TASK COMPLETED.
1. CLEAN AND DRY THE SENSOR ROD.
2. BE SURE THE PROCESS IS EMPTY. THE ROD MUST BE VIBRATING IN THE AIR.
3. INSTALL THE SENSOR ON THE PROCESS AND FIX IT WITH ITS 4 SCREWS.
4. POWER ON THE DEVICE, WAIT 15 MINUTES.
The MIVI+9000 is a complete digital system that can only communicate with other devices
through its serial RS-485 port. Hence, to make the zero adjustment in the air, only a digital
transmission through this port can be applied, it must be installed with its 9510 or 9710 ac-
cessory device, the zero adjustment in the air should be done on the 9510/9710 device (see
9510/9710 technical manual delivered with the unit).

Technical Manual 9000
REF: 381/3 3

Technical Manual 9000
REF: 381/3 4
TABLE OF CONTENT
1. TRANSDUCER PRINCIPLE ............................................................................4
2. TRANSDUCER TECHNICAL CHARACTERISTICS ......................................... 5
2.1 DIN-rail mounting IP20 enclosure .....................................................................5
2.1.1
Mechanical characteristics
............................................................................................5
2.1.2
Safety considerations
........................................................................................................5
2.1.3
System connection
.............................................................................................................6
2.2 IP66 Enclosure ...........................................................................................................7
2.2.1
Mechanical characteristics
............................................................................................7
2.2.2
Safety considerations
........................................................................................................7
2.2.3
System connection
.............................................................................................................8
2.3 ATEX Enclosure..........................................................................................................9
2.3.1
Mechanical characteristics
............................................................................................9
2.3.2
Safety considerations
........................................................................................................9
2.3.3
System connection
.............................................................................................................9
3. ACCESSORY DEVICES .................................................................................11
3.1 9510 ..............................................................................................................................11
3.2 9710 ..............................................................................................................................12

Technical Manual 9000
REF: 381/3 4
1. Transducer principle
The measuring chain is composed of three inseparable elements: the sensor, its cable and the
9000 transducer that controls it. The sensor cannot be used with another transducer or an-
other cable type or length because they are all a part of the resonant loop so they are
matched together as one vibrating system. The MIVI+9000 is specially designed to be used
with Sofraser HMI/PLC accessories (see chapter 3).
The active part of the sensor is composed of a vibrating rod held in oscillation at resonance fre-
quency by driving magnets. When the rod is immersed into a viscous material, the ampli-
tude of the vibration is dampened. The vibration amplitude varies according to the product
viscosity where the rod is immersed.
The sensor receiving coil detects the response and the signal is converted to a viscosity value
through the electronic device. The factory calibration is performed with standard oils.
The transducer acquires the coils’ amplitudes and frequency and generates various signals. These
signals represent the properties being measured. It is also in charge of powering the whole
system. It gives viscosity and temperature information through the serial communication.

Technical Manual 9000
REF: 381/3 5
2. Transducer technical characteristics
2.1 DIN-rail mounting IP20 enclosure
2.1.1
Mechanical characteristics
Hereunder are the different views of the electronics.
The processor must be connected to a stabilized and filtered 24 VDC (± 2.4 VDC) power supply. The
polarity of the power supply is highly important, in order to avoid any damage on the elec-
tronic board.
The operating temperature for this electronic device should be constant, the maximum is 50°C.
The electronic device delivers the viscosity and temperature information to the selected HMI
through the serial RS-485 communication port.
2.1.2
Safety considerations
Do not touch wires while the processor is ON.
Do not connect either the “Neutral” or “Line” signal of the 110/220VAC to the device’s 0V
pin.
In the event of voltage fluctuations or non-conformity to voltage power supply specifica-
tions, connect the device to a filtered and stabilized power supply.
Double-check all wiring before turning on the power supply.
Install at maximum distance from high-voltage cables and power equipment.

Technical Manual 9000
REF: 381/3 6
To maximize the system performances, avoid electromagnetic interferences by mounting
the processor on a metal panel and earthing the power supply.
2.1.3
System connection
Only the link between the sensor and the terminal block has to be operated by the user according
to the following description.
1
2
3
4
5
6
7
8
9
10
RS 485 (B)
RS 485 (A)
Earth
0 V
+24 VDC
Sensor (A)
Sensor (B)
Sensor (C)
Sensor (D)
Pt100 (E)
Pt100 (F)
Pt100 (G)
Earth
1
2
3
4
5
6
7
8
The wires at the end of the cable have the following allocations.
Wire
Color
Item
Transducer
terminal block
A
blue
receiving coil
1
B
brown
2
C
transparent
driving coil
3
D
black
4
E
red
Pt100
5
F
G
yellow
green
6
7
N/A
metal
Earth
8

Technical Manual 9000
REF: 381/3 7
2.2 IP66 Enclosure
2.2.1
Mechanical characteristics
Hereunder are the different views of the IP66 enclosure.
The processor must be connected to a stabilized and filtered 24 VDC (± 2.4 VDC), 1 A power supply.
The polarity of the power supply is highly important, in order to avoid any damage on the
electronic board.
The transducer box is IP66 can be mounted on a panel or a plate through 4 screws.
The operating temperature for this electronic device should be constant, the maximum is 50°C.
This enclosure is equipped with 3 cable glands.
The electronic device delivers the viscosity and temperature information to the selected HMI
through the serial RS-485 communication port.
2.2.2
Safety considerations
Do not touch wires while the processor is ON.
Do not connect either the “Neutral” or “Line” signal of the 110/220VAC to the device’s 0V
pin.
In the event of voltage fluctuations or non-conformity to voltage power supply specifica-
tions, connect the device to a filtered and stabilized power supply.
Double-check all wiring before turning on the power supply.
Install at maximum distance from high-voltage cables and power equipment.
To maximize the system performances, avoid electromagnetic interferences by mounting
the processor on a metal panel and earthing the power supply.

Technical Manual 9000
REF: 381/3 8
2.2.3
System connection
All the connections to the electronic device are made on the terminal block located inside the blue
enclosure.
All the wiring between the electronic board and the terminal block are not to be touched.
Only the link between the sensor and the terminal block has to be operated by the user according
to the following description.
In case a cable gland is not used, this one has to be sealed with a cap.
1
2
3
4
5
6
7
8
9
10
RS 485 (B)
RS 485 (A)
Earth
0 V
+24 VDC
Sensor (A)
Sensor (B)
Sensor (C)
Sensor (D)
Pt100 (E)
Pt100 (F)
Pt100 (G)
Earth
1
2
3
4
5
6
7
8
The wires at the end of the cable have the following allocations.
Wire
Color
Item
Transducer
terminal block
A
blue
receiving coil
1
B
brown
2
C
transparent
driving coil
3
D
black
4
E
red
Pt100
5
F
G
yellow
green
6
7
N/A
metal
Earth
8

Technical Manual 9000
REF: 381/3 9
2.3 ATEX Enclosure
2.3.1
Mechanical characteristics
Hereunder are the different views of the ATEX enclosure.
2.3.2
Safety considerations
Do not touch wires while the processor is ON.
Do not connect either the “Neutral” or “Line” signal of the 110/220VAC to the device’s 0V
pin.
In the event of voltage fluctuations or non-conformity to voltage power supply specifica-
tions, connect the device to a filtered and stabilized power supply.
Double-check all wiring before turning on the power supply.
Install at maximum distance from high-voltage cables and power equipment.
To maximize the system performances, avoid electromagnetic interferences by mounting
the processor on a metal panel and earthing the power supply.
2.3.3
System connection
When installed in an ATEX enclosure, the 9000 transducer is in its standard IP20 enclo-
sure. The connections are therefore identical to the ones described in 2.1.3.
1
2
3
4
5
6
7
8
9
10

Technical Manual 9000
REF: 381/3 10
RS 485 (B)
RS 485 (A)
Earth
0 V
+24 VDC
Sensor (A)
Sensor (B)
Sensor (C)
Sensor (D)
Pt100 (E)
Pt100 (F)
Pt100 (G)
Earth
1
2
3
4
5
6
7
8
The wires at the end of the cable have the following allocations.
Wire
Color
Item
Transducer
terminal block
A
blue
receiving coil
1
B
brown
2
C
transparent
driving coil
3
D
black
4
E
red
Pt100
5
F
G
yellow
green
6
7
N/A
metal
Earth
8

Technical Manual 9000
REF: 381/3 11
3. Accessory devices
The 9510 and 9710 devices have been designed as accessories to the MIVI+9000 sensor.
3.1 9510
The 9510 is an accessory processor that is to be connected to the 9000 transducer of the MIVI sen-
sor in order to add numerous functions to the whole system.
The 9510 electronic device has been designed in order to control 1 MIVI sensor. This means the MI-
VI’s transducer 9000 and the 9510 are connected in a RS-485 serial loop as shown in the
scheme just above.
See user manual ref. 390 for detailed information about the 9510 processor.
Cable supplied by Sofraser
Cable not
supplied by
Sofraser
Transducer
9000
S/N
Connection
to PORT 2

Technical Manual 9000
REF: 381/3 12
3.2 9710
The 9710 is an Human Machine Interface (HMI) that is to be connected to the 9000 transducer of
the MIVI sensor in order to add numerous functions to the whole system.
The connection between the MIVI+9000 and the 9710 has to be done in a similar way as for the
9510 but connecting the RS-485 cable to PORT 1.
Cable supplied by
Sofraser
Cable not sup-
plied by Sofraser
Connection
to PORT 1
Transducer
9000
S/N
Other manuals for 9000
1
Table of contents
Other Sofraser Transducer manuals
Popular Transducer manuals by other brands

Balluff
Balluff MICROPULSE Series user guide

Spectris
Spectris HBM U2B Mounting instructions

Kongsberg
Kongsberg Simrad ES38-10 manual

Electro-Voice
Electro-Voice 920-8B Technical specifications

S+S Regeltechnik
S+S Regeltechnik THERMASGARD TM54 Series Operating Instructions, Mounting & Installation

Endress+Hauser
Endress+Hauser Nivotester FTL375P Safety instructions