SONBEST SM2110V-10 User manual

SM2110V-10 Us e r Ma nual http://www.sonbus.com/
Sha nghaiSonbest Industrial Co., Ltd Page1
SM2110V-10
Protective DC10V voltage type temperature and humidity sensor
User Manual
File Version: V21.11.15
SM2110V-10 using the standard DC0 ~ 10V voltage output signal,easy access to PLC,DCS and
other instruments or systems for monitoring temperature,humidity state quantities.The internal use of
high-precision sensing core and related devices to ensure high reliability and excellent long-term
stability,can be customized RS232,RS485,CAN,4-20mA,DC0~5V\10V,ZIGBEE,Lora,WIFI,GPRSand
other output methods.
Technical Parameters
Technical parameter
Parameter value
Brand
SONBEST
Temperature measuringrange
-30℃~80℃
Temperature measuringaccuracy
±0.5℃@25℃
Humidity measuring range
0~100%RH
Humidity accuracy
±3%RH @25℃
Output Interface
DC0~10V
Power
DC12~24V 1A
Running temperature
-40~80°C
Working humidity
5%RH~90%RH
Wiringinstructions
Any incorrect wiring can cause irreversible damage to the product. Please carefully wire the cable as
follows in the case of power failure, and then connect the cable to confirm the correctness and then use it
again.
ID
Core color
Identification
Note
1
Red
V+
Power V+
2
Green
V-
Power V-
3
Yellow
H+
Humidity output
4
Blue
T+
Temperature output
In the case of broken wires, wire the wires as shown in the figure. If the product itself has no leads, the
core color is for reference.
1. temperature and DC0-10Vvoltage computing relationship
For example, the range is -30~80℃, the analog output is 0~10VDC0-10Vvoltage signal, temperature
and DC0-10Vvoltage The calculation relationship is as shown in the formula: C= (A2-A1) * (X-B1) / (B2-B1)
+ A1, where A2 is temperature range upper limit, A1 is the lower limit of the range, B2 is DC0-10Vvoltage
output range upper limit, B1 is the lower limit, X is the currently read temperature value, and Cis the
SM2110V-10 Us e r Ma nual http://www.sonbus.com/
Sha nghaiSonbest Industrial Co., Ltd Page2
calculated DC0-10Vvoltage value. The list of commonlyused values is as follows:
DC0-10Vvoltage(V)
temperatureValue (℃)
Calculation Process
0
-30
(80-(-30))*(0-0)÷ (10-0)+-30
1
-19
(80-(-30))*(1-0)÷ (10-0)+-30
2
-8
(80-(-30))*(2-0)÷ (10-0)+-30
3
3
(80-(-30))*(3-0)÷ (10-0)+-30
4
14
(80-(-30))*(4-0)÷ (10-0)+-30
5
25
(80-(-30))*(5-0)÷ (10-0)+-30
6
36
(80-(-30))*(6-0)÷ (10-0)+-30
7
47
(80-(-30))*(7-0)÷ (10-0)+-30
8
58
(80-(-30))*(8-0)÷ (10-0)+-30
9
69
(80-(-30))*(9-0)÷ (10-0)+-30
10
80
(80-(-30))*(10-0)÷ (10-0)+-30
As shown in the above formula, when measuring 5V, current DC0-10Vvoltage is55℃。
2. humidity and DC0-10Vvoltage computing relationship
For example, the range is 0~100%RH, the analog output is 0~10VDC0-10Vvoltage signal, humidity
and DC0-10Vvoltage The calculation relationship is as shown in the formula: C= (A2-A1) * (X-B1) / (B2-B1)
+ A1, where A2 is humidityrange upper limit, A1 is the lower limit of the range, B2 is DC0-10Vvoltage
output range upper limit, B1 is the lower limit, X is the currently read humidity value, and C is the calculated
DC0-10Vvoltage value. The list of commonlyused valuesis as follows:
DC0-10Vvoltage(V)
humidityValue (%RH)
Calculation Process
0
0.0
(100-0)*(0-0)÷ (10-0)+0
1
10.0
(100-0)*(1-0)÷ (10-0)+0
2
20.0
(100-0)*(2-0)÷ (10-0)+0
3
30.0
(100-0)*(3-0)÷ (10-0)+0
4
40.0
(100-0)*(4-0)÷ (10-0)+0
5
50.0
(100-0)*(5-0)÷ (10-0)+0
6
60.0
(100-0)*(6-0)÷ (10-0)+0
7
70.0
(100-0)*(7-0)÷ (10-0)+0
8
80.0
(100-0)*(8-0)÷ (10-0)+0
9
90.0
(100-0)*(9-0)÷ (10-0)+0
10
100.0
(100-0)*(10-0)÷ (10-0)+0
As shown in the above formula, when measuring 5V, current DC0-10Vvoltage is50%RH。
Disclaimer
This document provides all information about the product, does not grant any license to intellectual
property, does not express or imply, and prohibits any other means of granting any intellectual property
rights, such as the statement of sales terms and conditions of this product, other issues. No liability is
assumed. Furthermore, our company makes no warranties, express or implied, regarding the sale and use
of this product, including the suitability for the specific use of the product, the marketabilityor the
infringement liability for any patent, copyright or other intellectual property rights, etc. Product
specifications and product descriptions may be modified at any time without notice.
Contact Us
Company: Shanghai Sonbest Industrial Co., Ltd
Address:Building 8,No.215 North east road,Baoshan District,Shanghai,China
SM2110V-10 Us e r Ma nual http://www.sonbus.com/
Sha nghaiSonbest Industrial Co., Ltd Page3
Web: http://www.sonbest.com
Web: http://www.sonbus.com
SKYPE: soobuu
Email: sale@sonbest.com
Tel: 86-021-51083595 / 66862055 / 66862075 /66861077

Other SONBEST Measuring Instrument manuals

SONBEST SM1212B-12 User manual

SONBEST

SONBEST SM1212B-12 User manual

SONBEST SM1911B User manual

SONBEST

SONBEST SM1911B User manual

SONBEST SM5090T User manual

SONBEST

SONBEST SM5090T User manual

SONBEST SL5090B User manual

SONBEST

SONBEST SL5090B User manual

SONBEST SM2130B-C2H2 User manual

SONBEST

SONBEST SM2130B-C2H2 User manual

SONBEST SD2100B User manual

SONBEST

SONBEST SD2100B User manual

SONBEST SM3691B User manual

SONBEST

SONBEST SM3691B User manual

SONBEST SM2801B User manual

SONBEST

SONBEST SM2801B User manual

SONBEST KEC900+ User manual

SONBEST

SONBEST KEC900+ User manual

SONBEST SM2120B User manual

SONBEST

SONBEST SM2120B User manual

Popular Measuring Instrument manuals by other brands

Crown RTA-2 Brochure & specs

Crown

Crown RTA-2 Brochure & specs

TR-Electronic LLB-65 user manual

TR-Electronic

TR-Electronic LLB-65 user manual

Cirrus CR:262 Quick setup guide

Cirrus

Cirrus CR:262 Quick setup guide

Ranger 2150 Reference manual

Ranger

Ranger 2150 Reference manual

Keller dv-2 operating manual

Keller

Keller dv-2 operating manual

Brigade MDR-504 Series Installation and operation guide

Brigade

Brigade MDR-504 Series Installation and operation guide

Circutor DM45 F quick start guide

Circutor

Circutor DM45 F quick start guide

Mallinckrodt Nellcor NPB-70 Service manual

Mallinckrodt

Mallinckrodt Nellcor NPB-70 Service manual

Cygnus M5-C2P operating manual

Cygnus

Cygnus M5-C2P operating manual

SICK GMS815P-PS-3G Supplementary operating instructions

SICK

SICK GMS815P-PS-3G Supplementary operating instructions

Delmhorst Instrument Co C-2000 owner's manual

Delmhorst Instrument Co

Delmhorst Instrument Co C-2000 owner's manual

Kimo SDA Technical data sheet

Kimo

Kimo SDA Technical data sheet

Group3 DTM-133 user manual

Group3

Group3 DTM-133 user manual

LaserLiner LaserRange-Master Gi7 Pro operating instructions

LaserLiner

LaserLiner LaserRange-Master Gi7 Pro operating instructions

Airthings View Series user manual

Airthings

Airthings View Series user manual

Bosch DUS 20 plus operating instructions

Bosch

Bosch DUS 20 plus operating instructions

Mag-ic Probe BLE quick start guide

Mag-ic Probe

Mag-ic Probe BLE quick start guide

Solinst 101B instructions

Solinst

Solinst 101B instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.