steute MFS Microscope SW2.4LE-MED User manual

Product name: MFS Microscope SW2.4LE-MED
Material-Nr.: 1380391
Created on: 23.11.2022 Revision of data sheet: - D045071
Page
1 / 3
steute Technologies
GmbH & Co. KG
Brückenstraße 91
32584 Löhne, Germany
Phone +49 (0) 5731 745-0
Fax +49 (0) 5731 745-200
www.steute.com
Specification
Footswitch with 14 functions
Steute-wireless SW2.4LE-MED
Top housing color NCS-S-2502-B (light grey)
Printing: steute logo left, numbers 1 to 6 and symbols in black
Lower housing color NCS-S-7000-N (dark grey)
With M16x1,5 closing screw
2 rocker switches with foot-rack, collar in NCS-S-7000-N (dark grey)
6 pushbuttons with sleeve-color NCS-S-7000-N (dark grey)
Joystick, switching diagonal, with rubber gaiter and mushroom-type knob, color NCS-S-7000-N (dark grey)
2 LED (Display: Connection / Battery status)
Indicator laser-marked in similar mouse gray RAL7005

Product name: MFS Microscope SW2.4LE-MED
Material-Nr.: 1380391
Created on: 23.11.2022 Revision of data sheet: - D045071
Page
2 / 3
steute Technologies
GmbH & Co. KG
Brückenstraße 91
32584 Löhne, Germany
Phone +49 (0) 5731 745-0
Fax +49 (0) 5731 745-200
www.steute.com
Technical Data
Applied standards
IEC 60601-1 (Basic Safety)
IEC 60529 (Degrees of protection)
Conformity assessment by customers!
Top housing PC/ABS, UL 94-V0, similar NCS-S-2502-B (light grey)
Lower housing PC/ABS, UL 94-V0, color NCS-S-7000-N (dark grey)
Rockerswitch 4 switching functions / > 250.000 lifetime per switch
Pushbutton > 250.000 lifetime per pushbutton
Joystick 4 switching function (diagonal switching) / > 250.000 lifetime per switch
Connection type Radio
Degree of protection IPX8 (1m / 35 Min.) according to DIN EN 60529
Dimensions 420x242x95mm (LxBxH)
Weight < 2,6 kg
Electrical Ratings
Supply voltage 3 batteries LR14 (Type C) = 4,5 VDC
Inrush current ~ 30 mA
Radio specifications
Performance
Distance 10 m -30 m (with internal antenna, depends on spatial conditions and the surrounding)
Output rating max. 3 dBm (max. 7 dBm optional)
Input sensitivity -93 dBm
Antenna Internal antenna with U.FL-plug
Radio technology
Frequency range 2402 to 2480 MHz
Chanel spacing 2 MHz
Modulation GFSK, adaptive frequency hopping of 40 channels
Applied wireless standards
Europe EN 300 328 V2.1.1, EN 301 489-1 V2.2.0, EN 301 489-17 V3.2.0
Transmitter and wireless module registration
USA , Part 15C , single Modular, Identifier: XK5-SW24LE
Canada RSS-247 Issue 1, Identifier: 5158A-SW24LE
Japan ARIB STD-T66, Identifier: 204-650001
Please note the instruction manual for the SW2.4LE system (D023336) (Document on request)

Product name: MFS Microscope SW2.4LE-MED
Material-Nr.: 1380391
Created on: 23.11.2022 Revision of data sheet: - D045071
Page
3 / 3
steute Technologies
GmbH & Co. KG
Brückenstraße 91
32584 Löhne, Germany
Phone +49 (0) 5731 745-0
Fax +49 (0) 5731 745-200
www.steute.com
Environmental conditions
Storage/Transport Operating
Temperature range -18 °C up to +55 °C 0 °C up to +45 °C
Relative humidity 10 % up to 100 % 10 % up to 100 %
Air pressure 500 hPa – 1060 hPa 700 hPa – 1060 hPa
General notes
The foot switch as a component of a medical device can be evaluated only together in the overall system of the customer. Therefore the conformity
evaluation including the classification of the overall system according to the Medial Device Directive 93/42/EEC respectively EU Medical Device
Regulation 2017/745 must be carried out at the customer. In addition, validation can also only be carried out by the customer in the overall system.
Furthermore, the intended use and the basic functions are verifiable only in interaction with the overall system. As a result, the customer must involve
the footswitch in his risk management, his usability analysis, the verification of the electromagnetic compatibility and, if applicable, the verification of the
biocompatibility.
Please refer to the technical description MFS Microscope SW2.4LE-MED (D038004).
Maintenance/cleaning
During cleaning, avoid wiping with a cloth under the pedal, because this may cause the springs to move out of position.
Safety instructions
Despite careful production and inspection as well as redundant design of important components, an entirely safe function cannot be guaranteed at all
times (in case of malfunction, for example, a switch contact can turn off incorrectly).
Risk of injury due to incorrect handling!
• The plug connections must not be subjected to mechanical loads!
•
Components, especially pedals or other actuating elements, may not be dissembled or dismembered under any circumstances! If t
his happens the
products need to be send back immediately.
Errors and omissions excepted.
This manual suits for next models
1