18 of 40 Operating Manual: SU780XLE 20050 | 2.10.17
5.3.2 PIN
A Password Identification Number (PIN) may be set, but is not displayed. Use the PIN
to restrict access to GUI controls. The default PIN is 1234 and may be changed by the
user to any four-digit numeric code or disabled.
5.3.3 – GUI Menu
A hierarchical menu system is used to display and adjust system values. Each screen
has a button to return to the default "Home" screen and a“Back” button to return to
the prior level of the menu. Screen names (where applicable) are within brackets.
Home – display of current freezer temperature.
Set-up [Data and Setup] – display Password Identification Number (PIN) status (enabled/
disabled), door defrost status, temperature set point, system date and time, and alarm status.
• Set-up [Setup] – display current temperature set point. PIN required if enabled.
• Change [Change Set Point] – display and adjust over temperature set point,
temperature set point, and under temperature set point
• Battery [Battery Replace Date] – display and adjust battery replacement date.
• Alarms [Alarm Set-up] – display and change alarm delay settings
• Temp. Delay [Temperature Delay] – duration of time in which the audible alarm
is delayed when temperature is out-of-specification.
• Aud. Delay [Silence Duration] – duration of time in which the audible alarm is
delayed after the 'Silence' button is tapped.
• Ext. Delay [External Contact Delay] – duration of time in which the alarm dry
contact activation is delayed after the alarm condition is detected.
• Door Delay [Door Delay] – duration of time that the door may remain open
before an audible alarm sounds.
• Date & Time [Date/Time Setup] – display and change system date and time.
• Door Defrost [Door Defrost] – adjust defrost heater between 0 to 100%. A warning
will be displayed on 'Home' screen if set to 0.
• PIN – disable/enable the system PIN and change PIN.
• Silence – temporarily silence audible alarm.
• Event Log [Event Log] – track the last 200 alarm conditions such as: door openings,
temperature set point changes, over temperature condition, and power outages.
• Chart History [Chart History] – display the cabinet interior air temperature over the last 12
hours, 24 hours or 7 days. Increase chart resolution by tapping the screen.
• About [About] – display freezer model number, serial number, firmware version, engine
number, run hours, temperature set point, and date.
• Alarm Test – test the system alarm.
• Service Contact [Service Contact] – display manufacturer contact information.
• Service [Service] – display cabinet, control, and cold-head RTD temperatures,
warm-head thermistor temperature, as well as engine voltage, current and power.
Service screen information is used for service by authorized personnel only.
Service PIN required.
• RTD Calibration [Calibration] – display RTD readings and calibration offsets.
Calibration is used to compensate for discrepancies between Display RTD and
user-supplied temperature measurement.