800-433-2797 www.Sun-Med.com
User Guide 7900 Bubble Humidier Home Use
The 7900 High Flow Bubble Humidier is a disposable, single-patient use
device for use in health care facilities and home care. The humidier is used
to provide additional humidity to oxygen therapy for ow rates at 6 to 15
liters per minute (LPM). The Bubble Humidier has a 6 PSI safety pop-o
valve, which sounds and releases excess pressure if the oxygen tubing
becomes blocked.
Product Use
1. Wash your hands and remove the Bubble Humidier from packaging.
2. Use sterile, distilled or boiled tap water. If using boiled tap water, cool water
to room temperature before adding to the Bubble Humidier.
3. Unscrew the lid and add water to the maximum level marking (290 mL)
located on the side of the bottle.
4. Replace the lid. Ensure that the threads are properly aligned and
that the lid is screwed on tight.
5. Attach the Bubble Humidier to an oxygen source, e.g. owmeter, oxygen
concentrator or liquid tank.
6. Test for leaks. Turn the ow to 6 LPM and block the oxygen outlet port on the
Bubble Humidier (A). A sound should emit within 5 seconds. If there is no
sound, check the lid and other connections for leaks. Repeat the leak test.
7. Attach the oxygen supply tubing for your nasal cannula or oxygen mask and
adjust oxygen ow to the prescribed setting.
8. Check for bubbles in the water. If there are no bubbles check
for leaks. If the device sounds, check for tubing obstruction.
9. If water level drops below the minimum level marking on
the Bubble Humidier add more water.
10.Empty and replace the water in Bubble Humidier daily.
7900 High Flow Bubble Humidier – 6 PSI Safety Pop-o –
6 LPM to 15 LPM
Maximum
Water Level
Pressure Relief
Valve
Oxygen Source
Oxygen
Outlet
Port
Diuser
Minimum
Water level
A
Oxygen Source
B
Oxygen Source
Oxygen
Tubing
To Patient