User Guide 7100 Bubble Humidier Home Use
The 7100 Bubble Humidier is a disposable, single-patient
use device for use in health care facilities and home care. The
humidier is used to provide additional humidity to oxygen
therapy for ow rates at 2 to 5 LPM. The Bubble Humidier has
a 3 PSI safety Pop-O valve, which sounds and releases excess
pressure if the oxygen tubing becomes blocked.
Product Use
1. Wash your hands and remove the Bubble Humidier
from packaging.
2. Use sterile, distilled or boiled tap water. If using boiled tap
water, cool water to room temperature before adding to the
Bubble Humidier.
3. Unscrew the lid and add water to the maximum level marking
(290 mL) located on the side of the bottle.
4. Replace the lid. Ensure that the threads are properly aligned
and that the lid is screwed on tight.
5. Attach the Bubble Humidier to an oxygen source, e.g.
owmeter, oxygen concentrator or liquid tank.
6. Test for leaks. Turn the ow to 3 LPM and block the oxygen
outlet port on the Bubble Humidier (1). A sound should emit
within 5 seconds. If there is no sound, check the lid and other
connections for leaks. Repeat the leak test.
7. Attach the oxygen supply tubing to the oxygen outlet port (2)
and adjust oxygen ow to the prescribed setting.
8. Check for bubbles in the water. If there are no bubbles check
for leaks. If the device sounds, check for tubing obstruction.
9. Apply your nasal cannula or oxygen mask.
10. If water level drops below the minimum level marking on
the Bubble Humidier add more water.
11. Empty and replace the water in Bubble Humidier daily.
3 PSI Safety Pop-O- 2 LPM to 5 LPM
Please read and follow the Instruction for Use
provided with product before using the bubble humidier.
Maximum
Water Level
Pressure Relief
Valve
Oxygen Source
Oxygen
Outlet
Port
Diuser
Minimum
Water level
1
Oxygen Source
2
Oxygen Source
Oxygen
Tubing
To Patient