Supco DAVM+ User manual

TABLE OF CONTENTS
1. FEATURES.............................................................................. 1
2. SPECIFICATIONS.................................................................... 2
3. FRONT PANEL DESCRIPTION.................................................. 5
3-1 ON/ESC Button................................................................ 5
3-2 Hold Button..................................................................... 5
3-3 REC/Ente Button............................................................. 5
3-4 Unit/Ze o/ (CFM/CMM) Butto...................................... 5
3-5 °C/°F/ Lux/Ft-cd, A ea-Set Button................................ 5
3-6 Function/ Button........................................................ 5
3-7 Anemomete Senso ......................................................... 5
3-8 The mocouple Input Socket.............................................. 5
3-9 Humidity Senso ............................................................... 5
3-10 Light Senso ................................................................... 5
3-11 LCD display.................................................................... 5
3-12 Batte y Compa tment/Cove ............................................ 5
3-13 W istlet......................................................................... 5
4. MEASURING PROCEDURES..................................................... 6
4-1 AIR VELOCITY MEASUREMENT.......................................... 6
4-2 TEMPERATURE MEASUREMENT
(The mocouple)..................................................................... 6
4-3 RELATIVE HUMIDITY MEASUREMENT................................ 7
4-4 LIGHT MEASUREMENT......................................................7
4-5 AIR FLOW (CFM. CMM) MEASUREMENT............................. 8
5. OTHER FUNCTIONS................................................................ 9
6. BATTERY REPLACEMENT.........................................................10

1. FEATU ES
* 5 in 1 p ofessional measu ing inst ument: Anemomete ,
Ai flow, Hyg omete , The momete , and Light mete .
* Multi channel display fo elative humidity, tempe atu e,
and ai velocity measu ed values at the same time.
* Ai velocity measu ing units selectable by p essing button on
the f ont panel fo five kinds of units.
* Ai flow (CFM. CMM) measu ement by setting the desi ed a ea
dimension.
* Low-f iction ball bea ing mounted wheel design p ovides
high accu acy at high and low ai velocity.
* High p ecision thin-film capacitance humidity senso with
fast esponse to the humidity changes.
* °F/°C selectable by p essing button on the f ont panel.
Lux/Feet-candle selectable by p essing button on the f ont panel.
* Standa d type K (NiC -NiAl) the mocouple input jack
suitable fo all kinds of type K p obe.
* Built-in mic op ocesso ci cuit assu es excellent
pe fo mance and accu acy.
* Button a angement allows easy ope ation.
Maximum and minimum value with ecall.
* Hold function to f eeze the cu ent eading value.
* Lightweight and small size case design
a e suitable fo handling with one hand.
* W istlet design p ovides ext a p otection to the inst ument
especially fo use one hand ope ation.
1

Display 8 mm LCD display
Measurement Anemometer, Humidity, Temperature,
Light, Air flow ( CFM/CMM .
Operating Max. 80% RH.
Humidity
Operating 32 to 122°F (0 to 50°C
Temperature
Over Input Indication of - - - -
Display
Power Supply 006P DC 9V battery (Heavy duty type
Power Approx. DC 6.2 mA
Consumption
Weight 5.6 oz (160g
Dimensio 6.14x2.36x1.29 inch
156 x 60 x 33 mm
Standard SCASE Carrying Case
Accessory
Optional TPFG Thermocouple Probe
Accessories TPFDA Approved Thermocouple
TPP Thermocouple Piercing Probe
TPS Thermocouple Surface Probe
2
2. SPECIFICATIONS
2-1 General Specifications

3
2-2 Measurement Specifications
Measurement Range Resolution
Ai velocity ft/min 80 to 5,910 ft/min 1 ft/min
m/s 0.4 to 30.0 m/s 0.1 m/s
km/h 1.4 to 108.0 km/h 0.1 km/h
MPH 0.9 to 67.0 mile/h 0.1 MPH
knots 0.8 to 58.3 knots 0.1 knots
Tempe atu e 32 to 122 °F 0.1 °F
(the miste ) 0 to 50 °C 0.1 °C
Measurement Range Resolution
Ai flow CFM 1,908,400 CFM 0.001 to 100
cube feet/min CFM
CMM 54,000 CMM 0.001 to 1
cube meter/min CMM
Measurement Range Resolution
Humidity % RH 10 to 95 % RH 0.1 % RH
Tempe atu e 32 to 122 °F 0.1 °F
( the miste ) 0 to 50°C 0.1 °C
Measurement Range Resolution
Light Lux 0 to 2,200 Lux 1 Lux
1,800 to 20,000 Lux 10 Lux
* auto range Ft-cd 0 to 204.0 Fc 1 Ft-cd
170 to 2,000 Fc 1 Ft-cd
Measurement Range Resolution
Tempe atu e (Type K) -148 to 2,372 °F 0.1 °F
Optional -100 to 1,300 °C 0.1 °C

Measurement Range Accuracy
Air velocity 80 to 5,910 ft/min
0.4 to 30.0 m/s < 20 m/s : + 3% F.S.
1.4 to 108.0 km/h > 20 m/s : + 4% F.S.
0.9 to 67.0 mile/h
0.8 to 58.3 knots
32 to 122 °F + 2.5 °F
0 to 50 °C + 1.2 °C
Remark :
ft/min : feet per minute MPH : miles per hour
m/s : meters per second knots : nautical miles per hour
km/h : kilometers per hour Ft-cd : feet candle
Measurement Range Accuracy
Humidity 10 to 95% RH < 7 % RH :
+ 4% RH
> 7 % RH :
: + (4%rdg+1.2%RH
32 to 122 °F + 2.5 °F
0 to 50 °C + 1.2 °C
Measurement Range Accuracy
Light 0 to 20,000 Lux + 5% rdg + 8 dgt
0 to 2,000 Fc
Measurement Range Accuracy
Temperature -148 to 2,372 °F +(1% rdg + 2°F
(Type K -100 to 1,300 °C +(1% rdg + 1°C
Measurement Area setting
CFM 0.01 to 322.92 feet square
CMM 0.001 to 30.000 meter square
4

3. F ONT PANEL DESC IPTION
5
Fig. 1
3-1 ON/ESC Button 3-6 Function/ Button
3-2 Hold Button 3-7 Anemometer Sensor
3-3 REC/Enter Button 3-8 Thermocouple Input Socket
3-4 Unit/Zero/ 3-9 Humidity Sensor
(CFM/CMM Button. 3-10 Light Sensor
3-5 °F/°C/ 3-11 LCD display
Lux/Ft-cd 3-12 Battery Compartment /
Cover
Area-set Button 3-13 Wristlet

4. MEASU ING P OCEDU E
4-1 Air Velocity Measurement
1 Power on the instrument by pressing the ON Button
(3-1, Fig.1 .
2 Select the Anemometer function by pressing Function
Button (3-6, Fig.1 .
3 Press the Unit/Zero Button (3-4, Fig.1 to select unit
that you want and then face the Anemometer Sensor
(3-7, Fig.1 to the source of wind.
4 Allow time for the reading to become stable and note the
value indicated. From a practical point of view the velocity
may fluctuate.
4-2 Temperature Measurement
(Thermocouple)
1 Power on the instrument by pressing the ON Button
(3-1, Fig. 1 ..
2 Plug a type K thermocouple probe in the Thermocouple
Input Socket (3-8, Fig.1 .
3 Select the Temperature function by pressing Function
Button (3-6, Fig.1
4 Press the °F/°C Button (3-5, Fig.1 to select °F or °C
unit.
5 Contact the Thermocouple Sensor Head with measuring
object and the reading value will be displayed on the LCD
display.
Measuring Consi eration of Temperature
Measurement ( Thermocouple )
* Please make sure the polarity is correct when you plug a
thermocouple probe in the Temp. input socket.
6

7
* The temperature difference between thermocouple
probe and thermometer will cause an inaccurate
measuring result. Therefore, for the best measuring
and accuracy performance, whenever changing a probe
or plugging a new probe, thermal equivalent between
probe plug and meters input socket is a necessary
condition. Thermal equivalent procedure may take few
minutes and apply only when the probe has been exposed
to an ambient temperature different from the meter.
4-3 Humi ity & Ambient Temperature
Measurement
1 Power on the instrument by pressing the ON Button
(3-1, Fig. 1 .
2 Select the Relative Humidity function by pressing
Function Button ( 3-6, Fig. 1 .
3 At the mean time the reading value of relative humidity
and temperature will be displayed on the LCD display.
4 When the meter is applied in a new environment, a few
minutes are required to reach a stable condition.
4-4 Light Measurement
1 Power on the instrument by pressing the ON Button
(3-1, Fig.1 .
2 Select the Light Measurement function by pressing the
Function Button ( 3-6, Fig. 1 until the light value is
displayed. The light display digits are oriented 180° from
the other function displays for easy exposure and output
reading of the light sensor.
3 Press the Lux/Ft-cd Button (3-5, Fig.1 to
select measuring unit Lux or Ft-cd .

8
Zero Offset A justment of Light Function :
* For best results zero the light sensor prior to use in a
dark environment. Placing the light sensor end of the
meter under a desktop or flat surface so as to block
any light can accomplish this. Then press the
Unit/Zero Button (3-4, fig.1 to set the meter
indication to zero.
* Zero point can drift due to environment temperature
and battery power change as well as for other reasons.
It is recommended that the zero be checked
frequently using the above procedure.
* Zero adjustment only can be executed if the measurement
Light value < 20 Lux.
4-5 AIR FLOW (CFM. CMM) MEASUREMENT
1 Power on the instrument by pressing the ON
Button (3-1, Fig.1 .
2 Select the Air Flow function by pressing
Function Button (3-6, Fig.1 until the display show
unit of CMM or CMM.
* Air flow unit CFM or CMM can be select by
pressing the CFM/CMM Button (3-4, Fig.1
* Area size can be set by pressing the Area-Set
Button (3-5, Fig.1 , then use the
Button (3-5, Fig.1
Button (3-6, Fig.1
Button (3-4, Fig.1
to select the desired area size.

9
* For the CFM measurement, the area size is feet square ,
the lower display will show F-2 .
For the CMM, the area setting size is from 0.01 to 322.92 feet
square.
* For the CFM measurement, the area size is meter square ,
the lower display will show m-2 .
For the CMM, the area setting size is from 0.001 to 30.000
meter square.
4 Use the hand to hold the meter,
face the Anemometer Sensor (3-7, Fig.1 to the
measured wind. In the same time the air flow value
(CFM, CMM will show on the LCD display.
5. OTHE FUNCTIONS
5-1 Hol Function
Pressing the Hold Button (3-2, Fig.1 will freeze
the current reading value with a HOLD symbol on the
display.
5-2 Data Recor Function
1 The Data Record function records & displays the maximum
and minimum reading values. Start the Data Record function
by pressing the REC Button (3-3, Fig.1 once.
There will be a REC symbol on the display.

10
2 With the REC symbol on the display :
(a Press the REC Button (3-3, Fig.1 once and the
Max symbol along with the maximum value will appear
on the display.
(b Press the REC Button again, the Min symbol
along with the minimum value will appear on the display.
(c To exit the memory record function, press the REC
Button continuously for at least 2 seconds. The display
will revert to the current reading.
(d Clear the Max./Min. value recorded by pressing the
Hold Button (3-2, Fig.1 once. Previous recorded
Max./Min. value will be given up and then revert to the
REC. function keep on recording.
5-3 Auto Power Off Disable
In order to prolong the battery life, the instrument has Auto
Power Off function. The meter will switch off automatically if
no buttons are pressed for around 10 minutes.
6. BATTE Y EPLACEMENT
1) When the LCD display shows symbol,
it is necessary to replace the battery. However measure
ment may still be made for several hours after the low
battery indicator appears.
2) Open the Battery Compartment / Cover (3-12, Fig.1
and remove the battery.
3) Install a 9V battery (Alkaline or Heavy duty type and
then reinstate the cover.

11
Table of contents
Other Supco Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Haltian
Haltian Thingsee AIR TSAR01 installation instructions

VOLTCRAFT
VOLTCRAFT VC-10T operating instructions

Keysight Technologies
Keysight Technologies M9602A Configuration guide

Advanced Instruments
Advanced Instruments OsmoTECH PRO quick start guide

IFM Electronic
IFM Electronic Efector 300 SI5002 operating instructions

Conrad
Conrad 67 25 06 operating instructions