3.1.3 Igniting the exposure lamp (Controller CIC1200):
The CIC1200 controller will be powered on automatically once system is powered on. Wait
until “STAND BY” show on the alphanumeric display and make sure all “Air”, “Nitrogen”, and
“Vacuum” showing “OK” on the MAIN MENU screen (see step 3). Then press the key
“ON” in the control panel of the CIC1200. Wait until “READY” show on the alphanumeric
display.
Press the key “CP”. Wait until “=>>START” shows up.
Press the key “START”. The display shows “LAMP COLD” and in the annunciator group
FAILURE the LED of LAMP LIFE/POWER is flashing until the lamp reaches the operating
condition. Wait until the completion of the warm up sequence and the real value of the
power of the exposure lamp shows up (350W ).
Note: The machine will warn users by beeping if it does not work properly.
In case it happens when the lamp is on, turn the main switch to “OFF”,
leave the system as it is and report to specialist immediately.
3.1.4 Loading the mask:
Caution! Before loading or unloading a mask, the microscope must be moved to the “up”
position. This is accomplished by pressing the MICROSCOPE IS DOWN button on the
computer screen. The tool will beep, the display will change to MICROSCOPE IS UP
and the microscope will move to the “up” position.
Loosen the knurled knobs on the side of the mask holder frame and slide out the mask
holder. Turn over the mask holder so that the vacuum ring faces up. (Blow any unwanted
particles off your mask with the nitrogen gun before loading if necessary.) Place the mask
completely over the vacuum ring with the emulsion/chrome side up.
Adjust the mask to the proper position. Press the button “Mask vacuum is off” once. The
mask will stick to the holder.
Turn over the mask holder so that the mask is facing down. Slide the holder back to the
frame with the mask and tight the knurled knobs with fingers.
3.1.5 Loading the sample:
Pulling out the transport slide gently until it stops. Put the sample on top of the chuck.
Press and hold the hand valve on the front right side of the transport slide to switch on
the transport vacuum and carefully insert the slide back until it reaches the end stop.
3.1.6 Setting the mask sample thickness:
In order to achieve best result, the WEC SETTING needs to be reset for a new sample or if the
thickness of masks/substrates changes.
On the MAIN MENU screen, press the button “Setting” to setting menu and then press the
button “WEC SETTING”.
Unlock the level adjustment knob by putting the small black switch on the knob to upper
position. Lower the sample by turning the knob counterclockwise to ‘10’mm position.