Systech GASPACE 6000 User manual

Illinois Instruments, Inc. (U.S.)
2401 Hiller Ridge Road
Johnsburg
Illinois IL
60051
E: sales.usa@systechillinois.com
Systech Instruments Ltd
17 Thame Park Business Centre
Wenman Road
Thame
Oxfordshire
OX9 3XA
E: sales.uk@systechillinois.com
E: support.usa@systechillinois.com
E: support.uk@systechillinois.com
www.systechillinois.com
GASPACE 6000
HEADSPACE/MAP
ANALYZER
OPERATIONS MANUAL
Version 1.3 5th July 2018

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Table of Contents
1.0 PREFACE ....................................................................................................................................1
1.1 Important..........................................................................................................................................2
2.0 INTRODUCTION..........................................................................................................................3
2.1 Identification and Check List ............................................................................................................4
2.2 Options.............................................................................................................................................4
2.3 Operating Power..............................................................................................................................4
3.0 USING THE INSTRUMENT .........................................................................................................5
3.1 Front Panel Controls ........................................................................................................................5
3.2 Powering up.....................................................................................................................................5
3.3 Navigating the Menus ......................................................................................................................6
3.3.1 GS6600 Menu System......................................................................................................................9
3.3.2 GS6500 Menu System....................................................................................................................10
3.3.3 GS6700 Menu System....................................................................................................................11
4.0 SETTING UP THE GS6000 SERIES ANALYZER ....................................................................13
4.1 Setting a New Password for Page 1 or Page 2 Access..................................................................13
4.2 Setting Time...................................................................................................................................13
4.3 Setting Date ...................................................................................................................................13
4.4 Setting Year ...................................................................................................................................13
5.0 SETTING UP A TEST................................................................................................................14
5.1 Enabling Alarms(s) for O2and/or CO2............................................................................................14
5.2 Setting Alarm Levels......................................................................................................................14
5.3 Setting Alarm Polarity Low / Hi.......................................................................................................14
5.4 Setting the Test Method.................................................................................................................15
5.4.1 Timed.......................................................................................................................................15
5.4.2 Peak/Valley.............................................................................................................................. 15
5.4.3 Autosense/Continuous.............................................................................................................16
5.5 Taking a Measurement ..................................................................................................................17
5.6 Additional Test Methods.................................................................................................................18
5.6.1 Direct Injection (syringe)...........................................................................................................18
5.6.2 Testing Rigid Packages............................................................................................................19
5.7 Filters .............................................................................................................................................19
6.0 CALIBRATING THE INSTRUMENT..........................................................................................21
6.1 Calibration with Ambient Air for Oxygen and/or CO2 (all models)...................................................21
6.2 Calibration with Certified Calibration Gas.......................................................................................21
6.2.1 Oxygen Calibration with Certified Calibration Gas (GS6500 & GS6600)...................................21
6.2.2 Carbon dioxide Calibration Using Certified Calibration Gas (GS6600 & GS6700).....................22
7.0 SERIAL COMMUNICATIONS VIA USB....................................................................................23
7.1 Installing the USB Driver Software.................................................................................................23
7.2 Installing the Systech Illinois Client Utility Software .......................................................................25
7.3 Determining the Installed COM Port...............................................................................................28
7.4 Running the Systech Illinois Client Utility Software........................................................................29
7.5 Communications Commands for the GS6000................................................................................31
8.0 SETTING UP THE OPTIONAL EXTERNAL PRINTER FOR THE GS6000 .............................32
8.1 Setting the Time and Date for Printouts .........................................................................................32
8.2 Setting the Product Code and Description for Printouts.................................................................32

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
8.2.1 Changing Header Text on Printouts .........................................................................................34
8.2.2 Entering Product Line Description............................................................................................34
8.2.3 Choosing the Product Code .....................................................................................................35
8.2.4 Entering Descriptions for Corresponding Product Codes.......................................................... 35
9.0 PRINTER REPORTS .................................................................................................................36
9.1 Test report......................................................................................................................................36
9.2 Calibration Report..........................................................................................................................37
9.3 Analyzer Status Report ..................................................................................................................37
10.0 ERROR / MESSAGE CODES....................................................................................................38
11.0 WARRANTY...............................................................................................................................39
12.0 TECHNICAL SPECIFICATIONS ...............................................................................................40
13.0 SPARES & ACCESSORIES......................................................................................................41
13.1 General Spare Parts List..............................................................................................................41
13.2 Options Spare Parts List ..............................................................................................................42
Revision History
Issue
Issue Date
Changes
By
1.0
8th Jul 2015
Initial Release Version
Victor Kofman
1.1
7th Jan 2016
Changes according to ECN1610i
Victor Kofman
1.2
29th Jun 2018
Changes according to CAR0308
Tim Keep
1.3
10th Aug 2018
Changed per ECN1819s
Tim Keep
1.4
21st Aug 2018
Changed per ECN1860s
Tim Keep

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
1
1.0 PREFACE
The Gaspace 6000 Series is designed for measuring the concentration of modified atmosphere gases
in the headspace of flexible and rigid packages. Primary applications are for testing Modified
Atmosphere Packaged food and drink products, and pharmaceutical packaging and products.
This symbol is known as the ‘Crossed-out Wheelie Bin Symbol’. When this symbol is marked
on a product it means that consideration should be given to the disposal of the product, parts,
or accessories. Only discard electrical/electronic items in separate collection schemes which
cater for the recovery and recycling of the materials contained within. Your co-operation is vital
to ensure the success of these schemes and for the protection of the environment.
GS6600
GS6500
GS6700

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
2
1.1 Important
Please read this manual before attempting to install or operate the equipment.
In some variations, the sample gas in this equipment is heated to very high temperatures and therefore
no flammable gases should be present in a combustible concentration.
No responsibility is accepted by Systech Illinois for accidents resulting
from improper use of this equipment.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
3
2.0 INTRODUCTION
Instruments are available and configured for the following applications:
Designation
Gases Measured and Ranges
Technology
GS6500
Oxygen 0.001% - 100%
Zirconia, solid state
ceramic, non-depleting
GS6600
Oxygen 0.001% - 100%, Carbon dioxide 0.1% - 100%
Zirconia, solid state
ceramic, non-depleting,
Dual detector Infra-Red
GS6700
Carbon dioxide 0.1% - 100%
Dual detector Infra-Red
The analyzer is designed to measure the headspace of modified atmosphere packaged products.
Great care has been taken to ensure that the instrument is simple to use for the operator, while at the
same time offering many features for the technical owner.
We are always trying to improve our product, of which this manual is part. We would greatly appreciate
any feedback or information that you can give us in relation to the analyzer or the manual.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
4
2.1 Identification and Check List
Unpack the Gaspace 6000 and inspect it visually.
The following items will be included:
•GS6000 Analyzer
•IEC standard power cable
•USB flash drive with GS6000 Operations Manual and Calibration Certificate
•Sampling Kit, GS6000, including the following items:
1. Pack of needles 21G X 1.5” (50)
2. Pack of needles 23G X 1.25” (1)
3. Pack of needles 25G X 0.7” (50)
4. Needle probe (Wand) assembly (1)
5. Adhesive septa (100)
6. Luer filters (50)
2.2 Options
•External printer (Part number OPT-6000-PRN), including the following items:
1. Roll of printer paper (1)
2. Printer cable
•Direct Injection (Part number 898 626)
1. Syringe valve
2. Syringe (30ml)
Sampling systems are available for accessing the headspace within rigid packages and cans. Please
contact Systech Illinois for more information.
2.3 Operating Power
The analyzer is operated from AC line power (universal) between 90 and 260 VAC, 50/60 Hz. Line
voltage is automatically sensed by the analyzer.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
5
3.0 USING THE INSTRUMENT
3.1 Front Panel Controls
3.2 Powering up
Connect the IEC power cable to the analyzer first, and then plug it into an outlet. The power switch is
mounted on the back of the analyzer just above the IEC receptacle. After switching the analyzer on, all
LED’s and displays will light up. The version of software will flash momentarily, followed by the model
number as shown below.
Sample
port
Sample
probe
Green TEST
button
Pushbutton keys
GS6600
Sample
probe
holder

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
6
Menu
Digit
Select
Digit
Increment
GS6500 and GS6600: After flashing the model number, the oxygen display (right-hand display on the
GS6600) will flash “hEAt” (Heat) as illustrated below. Since the analyzer cannot produce meaningful
readings until the sensor is at the proper operating temperature, the display will continue to flash
“hEAt” (Heat) until the correct operating temperature is reached (usually about five minutes). In
critical, high-accuracy applications, or when measuring at very low concentrations of oxygen, it is
advisable to wait at least thirty minutes before taking a measurement. This will allow the analyzer to
reach optimal operating conditions.
GS6600 Shown
GS6700: After flashing the model number, the display will show “0.0”. Since the detector will drift
slightly until fully warm, it is suggested that the analyzer be allowed to warm up for thirty minutes before
critical measurements are made.
3.3 Navigating the Menus
The analyzer is configured using the keypad shown above.
The ‘Menu’button allows you to change the menu listing, the ‘Digit Select’button is used to choose
which digit of the display will be changing. The ‘Digit Increment’button allows for digit change, and the
‘Enter’button will store a new configuration in memory.
Enter

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
7
Changing the program codes to configure parameters is simple. There are three pages of information.
Page 0 contains all the parameters that a test operator may want to use frequently such as changing
Product Code.
Page 1 contains all test configuration parameters which can be (optionally) password protected. The
factory default is password protection disabled. The default password is 1234.
Page 2 parameters are those which are used to configure instrument options that do not need to
accessed frequently. Page 2 parameters are always password protected. The factory default password
is 1234.
To access the parameters in Page 0 press the ‘Menu’ button:
The LED above it will light and the display will
briefly read “Prod” then indicate the current Product
Code. Press the ‘Menu’ button to cycle through
options until the display will read “PA9E” (Page)
briefly then display “0” to indicate you are on Page
0. Available parameters depend on the instrument
model and its configuration.
Every time the ‘Menu’ button is pressed a new parameter will be displayed briefly followed by the value
for that parameter. Note: some parameters are read only.
Continual pressing of the ‘Menu’ button when on Page 0 will eventually show “– O2 –” (for GS6600
and GS6500) or “CO2”(for GS6700) and this signals the end of the Programming mode and the
instrument is returned to the measurement mode (shown below is GS6500 display).
To access the parameters in Page 1 or 2, follow the steps below:
Press the ‘Menu’ button 3
times
The display will briefly read “PA9E” (Page) then
“0” (to indicate you are on page 0).
Now press ‘Digit Select’button
The display will flash.
Now press the ‘Digit
Increment’ button
The display will change from “0”to “1”.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
8
Now press the ‘Digit
Increment’ button again for
menu 2 if required
The display will change from “1”to “2”.
Now press the ‘Enter’button
The display will briefly read “PAS?” (Password) if
menu 2 was selected (optionally on menu 1) and
then 4 digits. To access the page you must enter a
correct password. At this point the display will read
“0000” and the left-hand “0” will flash, prompting
you to enter the password. The factory set
password is 1234. Enter this password using the
‘Digit Increment’ and ‘Digit Select’ buttons. Press
the ‘Enter’ button. The display will show “yES”
(YES), for a correct password or “nO” (NO) if
incorrect. If you have the correct password, access
is allowed to the page parameters.
Note: to get out of Page 1 or 2 back to Page 0 and return to the measurement mode (“O2” screen for
GS6500 and GS6600 or “CO2” screen for GS6700), you must keep pressing the ‘Menu’ key until you
get to the “PA9E” (Page) parameter, then change the code to “0” using the method shown above (no
password will be needed). Anytime the ‘Enter’ button is pressed, “Stor” (Store) will flash once on the
display to indicate a change has been accepted.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
9
3.3.1 GS6600 Menu System
When navigating through the menus, the right-hand display will flash the parameter name followed by
its value.
DISPLAYED PARAMETER
FUNCTION
Page 0
Selects which product code and its corresponding description will be
printed. The user may enter up to 9 product codes (1 through 9).
Shows the current temperature of the internal zirconia probe furnace
(This parameter is ‘read-only’).
Selects the menu page. Page 1 or 2 parameters may be password
protected. The factory default password is 1234.
Page 1 (optionally password protected. Default password is 1234)
Sets the duration of a timed test in seconds (1 to 200).
Sets the delay time between tests for Autosense in seconds. 0=disable
Autosense, 1=Continuous mode, >1=delay time up to 200 seconds.
Sets the Peak/Valley measurement mode for CO2/O2 with the first digit
corresponding to CO2and the second digit corresponding to O2. 0=disable
Peak/Valley, 1=Peak, 2=Valley i.e. “12” will set the Peak for CO2and
Valley for O2.
Sets the alarm option.
0=alarm disabled, 1=alarm enabled.
Sets the value at which the oxygen alarm will be triggered (0 to 100). This
will not be shown if the alarms are disabled (“ALEn” or Alarm Enable set to
“0”).
Sets the value at which the carbon dioxide alarm will be triggered (0 to
100). This will not be shown if the alarms are disabled (“ALEn” or Alarm
Enable set to “0”).
Sets the oxygen alarm polarity. 0=low alarm, 1=high alarm. This will not
be shown if the alarms are disabled (“ALEn” or Alarm Enable set to “0”).
Sets the carbon dioxide alarm polarity. 0=low alarm, 1=high alarm. This
will not be shown if the alarms are disabled (“ALEn” or Alarm Enable set to
“0”).
Selects the menu page. Page 0 should be selected to return to the
measurement mode. Page 2 parameters are password protected. The
factory default password is 1234.
Page 2 (password protected. Default password is 1234)
Selects whether a password is requested to access menu page 1. 0 is
disabled, 1 is enabled.
Sets the password to access the Page 1 menu.
Sets the password to access the Page 2 menu.
Sets the percent concentration of oxygen for calibration using certified gas
(1% to 10% or 30% to 100%).
Sets the percent concentration of carbon dioxide for calibration using
certified gas (15% to 100%).

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
10
Sets the time in 24-hour format (HHMM).
Sets the date in format DDMM.
Sets the year four-digit format (YYYY).
Selects the menu page. Page 0 should be selected to return to the
measurement mode. Page 2 parameters are password protected. The
factory default password is 1234.
3.3.2 GS6500 Menu System
When navigating through the menus, the display will flash the parameter name followed by its value
DISPLAYED PARAMETER
FUNCTION
Page 0
Selects which product code and its corresponding description will be printed. The user
may enter up to 9 product codes (1 through 9).
Shows the current temperature of the internal zirconia probe furnace (This parameter is
‘read-only’).
Selects the menu page. Page 1 or 2 parameters may be password protected. The
factory default password is 1234.
Page 1 (optionally password protected. Default password is 1234)
Sets the duration of a timed test in seconds (1 to 200).
Sets the delay time between tests for Autosense in seconds. 0=disable Autosense,
1=Continuous mode, >1=delay time up to 200 seconds.
Sets the Peak/Valley measurement mode for CO2/O2 with the first digit corresponding to
CO2and the second digit corresponding to O2. 0=disable Peak/Valley, 1=Peak, 2=Valley
i.e. “12” will set the Peak for CO2and Valley for O2.
Sets the alarm option.
0=alarm disabled, 1=alarm enabled.
Sets the value at which the oxygen alarm will be triggered (0 to 100). This will not be
shown if the alarms are disabled on Page 1 (“ALEn” or Alarm Enable set to “0”).
Sets the oxygen alarm polarity. 0=low alarm, 1=high alarm. This will not be shown if the
alarms are disabled on Page 1 (“ALEn” or Alarm Enable set to “0”).
Selects the menu page. Page 0 should be selected to return to the measurement
mode. Page 2 parameters are password protected. The factory default password is
1234.
Page 2 (password protected. Default password is 1234)
Selects whether a password is requested to access menu page 1. 0 is disabled, 1 is
enabled.
Sets the password to access the Page 1 menu.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
11
Sets the password to access the Page 2 menu.
Sets the percent concentration of oxygen for calibration using certified gas (1% to 10% or
30% to 100%).
Sets the time in 24-hour format (HHMM).
Sets the date in format DDMM.
Sets the year four-digit format (YYYY).
Selects the menu page. Page 0 should be selected to return to the measurement
mode. Page 2 parameters are password protected. The factory default password is
1234.
3.3.3 GS6700 Menu System
When navigating through the menus, the display will flash the parameter name followed by its value
.
DISPLAYED PARAMETER
FUNCTION
Page 0
Selects which product code and its corresponding description will be printed. The user may
enter up to 9 product codes (1 through 9).
Selects the menu page. Page 1 or 2 parameters may be password protected. The
factory default password is 1234.
Page 1 (optionally password protected. Default password is 1234)
Sets the duration of a timed test in seconds (1 to 200).
Sets the delay time between tests for Autosense in seconds. 0=disable Autosense,
1=Continuous mode, >1=delay time up to 200 seconds.
Sets the Peak/Valley measurement mode for CO2/O2 with the first digit corresponding to
CO2and the second digit corresponding to O2. 0=disable Peak/Valley, 1=Peak, 2=Valley i.e.
“12” will set the Peak for CO2and Valley for O2.
Sets the alarm option.
0=alarm disabled, 1=alarm enabled.
Sets the value at which the carbon dioxide alarm will be triggered (0 to 100). This will not be
shown if the alarms are disabled (“ALEn” or Alarm Enable set to “0”).
Sets the carbon dioxide alarm polarity. 0=low alarm, 1=high alarm. This will not be shown if
the alarms are disabled (“ALEn” or Alarm Enable set to “0”).
Selects page (Page 0, 1). Page 1 parameters are password protected. The factory
default password is 1234.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
12
Page 2 (password protected. Default password is 1234)
Selects whether a password is requested to access menu page 1. 0 is disabled, 1 is
enabled.
Sets the password to access the Page 1 menu.
Sets the password to access the Page 2 menu.
Sets the percent concentration of carbon dioxide for calibration using certified gas (15% to
100%).
Sets the time in 24-hour format (HHMM).
Sets the date in format DDMM.
Sets the year four-digit format (YYYY).
Selects page (Page 0 or Page 1). Page 1 parameters are password protected. The
factory default password is 1234.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
13
4.0 SETTING UP THE GS6000 SERIES ANALYZER
4.1 Setting a New Password for Page 1 or Page 2 Access
By default, the password to access Pages 1 & 2 is “1234”. To change the password, use the menu
system (refer to section 3.3) to get to the “PAS1” or “PAS2”(Password 1 or Password 2) parameters
on Page 2 as shown below:
Using the ‘Digit Select’ and ‘Digit Increment’ buttons, set the new password to any 4-digit number.
Press the ‘Enter’ button to save. If the password is lost, please contact Systech Illinois for assistance.
It is suggested that the new password is recorded and kept safe.
4.2 Setting Time
Use the menu system (refer to section 3.3) to get to the “tinE” (Time) parameter on Page 2 as shown
below:
The time format displayed on the analyzer is HHMM (hours and minutes) and cannot be changed.
Using ‘Digit Select’ and ‘Digit Increment’ buttons, set up hours and minutes. Press the ‘Enter’ button
to save.
4.3 Setting Date
Use the menu system (refer to section 3.3) to get to the “dAtE” (Date) parameter on Page 2 as shown
below:
The date format shown above will depend on the format you entered in section 8.1. The default
format is “0” – DD/MM. Using the ‘Digit Select’ and ‘Digit Increment’ buttons, set the day and month.
Press the ‘Enter’ button to save.
4.4 Setting Year
Use the menu system (refer to section 3.3) to get to the “yEAr” (Year) parameter on Page 2 as shown
below:
Using the ‘Digit Select’ and ‘Digit Increment’ buttons, set the year. Press the ‘Enter’ button to save.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
14
5.0 SETTING UP A TEST
5.1 Enabling Alarms(s) for O2and/or CO2
Use the menu system (refer to section 3.3) to get to the “ALEn” (Alarm Enable) parameter on Page 1
as shown below:
Using the ‘Digit Select’ and ‘Digit Increment’ buttons, set “1” to enable alarms, or “0” to disable
alarms. Press the ‘Enter’ button to save.
5.2 Setting Alarm Levels
Alarms must be enabled to set levels (refer to section 5.1 above). Use the menu system (refer to
section 3.3) to get to “AL1” and/or “AL2” (Alarm 1 and/or Alarm 2) parameter(s) on Page 1 as shown
below:
a) The “AL1” (Alarm 1) parameter sets the alarm level for O2(GS6500 and GS6600) or for CO2
(GS6700).
Using the ‘Digit Select’ and ‘Digit Increment’ buttons, set alarm 1 to the desired level. Press ‘Enter’
button to save.
b) The “AL2” (Alarm 2) parameter (available only for GS6600) sets the alarm level for CO2.
Using the ‘Digit Select’ and ‘Digit Increment’ buttons, set alarm 2 to the desired level. Press ‘Enter’
button to save.
5.3 Setting Alarm Polarity Low / Hi
Alarm Polarity is the parameter that sets the alarm condition to trigger either low or high relative to the
alarm level selected (see section 5.2) for “AL1” (Alarm 1) and/or “AL2” (Alarm 2 for GS6600).
Alarms must be enabled to set polarity (refer to section 5.1 above). Use the menu system (refer to
section 3.3) to get to the “POL1” (Alarm Polarity) parameter on Page 1 as shown below:
a) “POL1” option sets alarm polarity for O2(GS6500 and GS6600) or for CO2(GS6700).
Using the ‘Digit Select’ and ‘Digit Increment’ buttons, enter a “0” to set the polarity to low or a “1” to
set the polarity too high. Press the ‘Enter’ button to save.
b) “POL2” option (available only for GS6600) sets the alarm polarity for CO2.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
15
Using the ‘Digit Select’ and ‘Digit Increment’ buttons, enter a “0” to set the polarity to low or a “1” to
set the polarity too high. Press the ‘Enter’ button to save.
NOTE: Alarms triggered on the display will also be shown triggered on the printout if the optional
printer is connected.
5.4 Setting the Test Method
There are several ways the analyzer can be used to measure residual gas in the headspace of MAP
packaged products.
The available test methods are: TIMED
PEAK/VALLEY
AUTOSENSE/CONTINUOUS
5.4.1 Timed
With the timed test method, the test time duration is user-selectable. When the test is started by
pressing the green ‘TEST’ button, the test will run for a fixed period. The reading is displayed during
the test and then remains frozen, so the values can be recorded. The internal sampling pump will stop
at the end of the test time duration. The test time duration can be defined by the user in the Page 1
menu.
Test time duration calculation:
The GS6000 analyzers have an approximate flow rate of 60cc/min (1cc/second). Estimate the
available headspace volume in cubic centimeters in the package or container to be tested. This number
is the approximate maximum test time duration in seconds. For example, if there is 30cc available
headspace in the package or container, the maximum test time duration is 30 seconds. It is important
not to exceed the maximum calculated test time to avoid the following:
1. Drawing product (particulates, liquids, etc.) into the sampling system.
2. Creating a vacuum within the package or container causing ambient air ingress.
3. Deadheading the pump (blocking the flow).
By default, the test time is set for 20 seconds. Use the menu system (refer to section 3.3) to get to
the “teSt” (Test) parameters on Page 1 as shown below:
Using the ‘Digit Select’ and ‘Digit Increment’ buttons, set a new test time in seconds. Press the ‘Enter’
button to save.
5.4.2 Peak/Valley
In the Peak/Valley test method, the final test result represents a peak or valley reading detected during
the test time.
In the Peak/Valley set up page, the test parameters for O2and CO2can be set. The Peak/Valley is
referenced to a normal air measurement, namely 20.9% for O2and 0.0% for CO2. If the package gas
is normally lower than 20.9% O2and higher than 0.0% CO2, then the test should be set up for ‘valley’
on O2and ‘peak’ on CO2. It should be set up on Page 1in “nodE” (mode) menu.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
16
By default, the Peak/Valley method is turned off. To change test method to Peak/Valley using menu
system, (refer to section 3.3) get to the “nodE” (mode) options on Page 0 as shown below.
Using the “Digit Select” and “Digit Increment” buttons, set mode to desired peak / valley setup (refer
to menu tables in section 3.3.1 for setup explanations). Press the ‘Enter’ button to save.
5.4.3 Autosense/Continuous
The Autosense test method may be used when measuring many packages. Autosense avoids having
to press the ‘TEST’ button to initiate each test.
Autosense is a user-programmable parameter which specifies the delay time between each test in
seconds. When Autosense mode is selected for the analyzer and the green ‘TEST’ button is pressed,
the first package is tested as described for a timed test. The next test starts after the delay time specified
in “Auto” setup on Page 1. Upon completion of the first test the user has the specified delay time to
withdraw the needle from the first package, prepare and insert the needle into the next package, and
so on until all the packages have been measured.
To pause Autosense, press the green ‘TEST’button or the ‘Enter’ button after the pump stops (before
next test starts). Press the green ‘TEST’ button again to re-start Autosense. To completely stop
Autosense, change “Auto” in Page 1 back to “0”.
To change Autosense to Continuous mode, change “Auto” in Page 1 to “1”. This will cause the pump
to run continuously until the green ‘TEST’button is pressed to stop/start measurements or Autosense
is turned off as described above.
By default, Autosense is turned off. To change Test method to Autosense, using menu system (refer
to section 3.3) get to the “Auto” options on Page 1 of menus as shown below:
Using the ‘Digit Select’ and ‘Digit Increment’ buttons, change “Auto” from “0” to the desired time in
seconds between consecutive tests. Press the ‘Enter’ button to save. To turn Autosense method off
change “Auto” time back to “0”.

Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
17
5.5 Taking a Measurement
Connect the sampling probe (Wand) to the blue Luer connector on the right side of the instrument and
screw hand tight. Connect the disposable filter and a needle to the sampling probe (Wand).
To take a measurement, follow the steps below:
1. Ensure the GS6000 is operational and ready to use, that it has a clean filter, a needle in good
condition and sampling probe (wand) with no leaks.
2. Situate the package to be measured in such a way as to maximize the available headspace.
3. Pullthe packaging away from itscontents to create an area which will be used to accept the needle.
4. Apply an adhesive-backed septum to the area of the package through which the needle will be
inserted –ensure the package is clean and dry in the area where the septum is applied (the septum
creates a seal around the needle to avoid atmospheric oxygen ingress).
5. Pierce the septum with the needle and keep the needle in the area created, awayfrom the product.
Do not allow the needle to exit the other side of the package (if this happens, discard the package
and start over with a new one).
6. Set the analyzer test time as mentioned in section 5.4.1. Initiate the measurement by pressing the
green ‘TEST’ button –do not move the needle during the measurement.
7. If the oxygen/carbon dioxide concentration reading is still changing after the test time expires,
consider extending the test time as mentioned in 5.4.1.
8. Once the measurement stabilizes, the test is complete. If connected, the external printer will
document the test results.
9. Withdraw the needle from the package and place back in the sample probe holder on the side of
the analyzer –dispose of tested packages as instructed.
10. Document test results in other ways if so instructed.
11. Visually inspect the upstream side of the filter after each test. If it begins to collect product,
exchange it for a new filter. Passing product through the filter may cause damage to the
analyzer which will not be covered under warranty.
12. Perform an Air Calibration at regular intervals, see section 6.1.
Table of contents
Other Systech Measuring Instrument manuals