manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Taylor Hobson
  6. •
  7. Measuring Instrument
  8. •
  9. Taylor Hobson Surtronic Duo User manual

Taylor Hobson Surtronic Duo User manual

Error codes 䫭䇃ҷⷕ
The following error codes may appear on the
display during the course of using the Duo
Instrument. Before returning an instrument,
ensure that the batteries are not low.
E=1. Infra Red communication error - Repeat
operation.㑶໪䖲᥹䫭䇃˗䞡ᮄ䖲᥹
E=2. Measurement attempted when not
calibrated. Calibrate the TU󰶖ޚҾ఼
E=3. Traverse Unit - Oscillator fault - If fault is
persistent, return to a Taylor Hobson Service
Centre. ᤃ㤵఼䫭䇃˗ᓎ䆂䖨ׂ
E=4. Traverse Unit - Motor Positioning fault - If
fault is persistent, return to a Taylor Hobson
Service Centre. 偀䖒ԡ㕂䫭䇃˗ᓎ䆂䖨ׂ
E=5. Traverse Unit - NVRAM Error - If fault is
persistent, return to a Taylor Hobson Service
Centre. NVRAM 䫭䇃˗ᓎ䆂䖨ׂ
E=6. Traverse Unit - Profile Over-range.
Re-attempt measurement.䕂ᒧ䍙䞣⿟˗䞡䆩
E=7. Traverse Unit - Parameter Over-range.
Re-attempt measurement.খ᭄䍙䰤˗䞡䆩
E=8. Traverse Unit - Motor Speed Setup error.
Re-attempt Setup procedur偀䖒䗳ᑺ䆒㕂䫭䇃
E=9. Traverse Unit - Gain Setup error.
Re-attempt Setup procedure.๲Ⲟ䫭䇃
E=10. Control Unit - Oscillator fault - If fault is
persistent, return to a Taylor Hobson Service
Centre.᥻ࠊܗ˖ᤃ㤵఼䫭䇃ˈᓎ䆂䖨ׂ
E=11. Traverse Unit - Calibration information
download error. Try again.󰶖ޚ䫭䇃ˈ䞡䆩
E=12. Traverse Unit - Not set up. Run the
Setup procedure.偅ࡼܗ˖᳾䆒㕂
E=13. <Reserved>
E=14. : Traverse Unit - Electronic Measure-
ment error - If fault is persistent, return to a
Taylor Hobson Service Centre.⬉䏃䫭䇃ˈ䖨ׂ
Specification
Traverse length. 5mm (0.2in)偅ࡼ䭓ᑺ
Measurement length. 4mm (0.16in)⌟䞣䭓ᑺ
Traverse speed. 2mm/s (0.08in/s)⌟䞣䗳ᑺ
Gauge range. 200μm (8000μin)䞣⿟
Gauge resolution. 0.05μm (2μin)ߚ䕼⥛
Display resolution:Ra. 0.01μm (0.4μin)
Display resolution: Rz, Rt*, Rp*, Rv*.
0.1μm (1μin)ᰒ⼎ߚ䕼⥛
Cut off. 0.8mm (0.03in) +/- 15%পḋ䭓ᑺ
Uncertainty. 0.1μm (4μin) or 5% of
result, to 95% confidence levelϡ⹂ᅮᑺ
Results range:Ra. 40μm (1600μin),
see ISO 4288-1996 ᳔໻5Dؐ
Results range: Rz, Rt*, Rp*, Rv*.
199.9μm (8000μin), see ISO 4288-1996
Stylus. Diamond, radius 5μm nominal , 90otip
angle, skid radius: 10mm nominal.⌟䩜
Filter type: 2CR, (200:1)Ⓒ⊶఼
Sample spacing: 1um䞛ḋ䯈䱨
Stylus Force. 10mN (1gf) max at mid range⌟࡯
Batteries. Control Unit: One x 3V Lithium ⬉∴
2450. Traverse Unit: Three x 3V Lithium 2450
Units. μm or μinԡࠊ
Operating conditions. 10 to 35oC,
80%RH non-condensing⦃๗㽕∖
Size. 125mm x 80mm x 38mm
(4.92in x 3.15in x 1.5in)ሎᇌ
Weight. 200g (7oz)䞡䞣
*5 parameter model only.
Surtronic Duo
112-2916 (2 parameter model) and 112-3115 (5 parameter model)
Operator’s Leaflet K505-54
Issue 1.1 January 2002
Leicester, England
www.taylor-hobson.com
Declaration of Conformity
Manufacturer’s Name: Taylor Hobson Limited
Manufacturer’s Address: 2 New Star Road, Leicester, England, LE4 9JQ
It is declared that the product: Surtronic Duo
Conforms to all applicable requirements of BS EN 50081-1:1992, BS EN 50082-1:1998 and
BS EN 61010:2001.
The above product complies with the requirements of the EMC Directive 89/336/EEC as amended.
The above product complies with the requirements of the EMC Directive 73/23/EEC as amended.
Traverse Unit偅ࡼܗ Control Unit᥻ࠊܗ
Parameter Selectorখ᭄䗝ᢽ
Mode selector῵ᓣ䗝ᢽ
Start buttonsᓔྟᣝ䪂
Stylus⌟䩜
Mode status῵ᓣ
Parameter statusখ᭄
Displayᰒ⼎ሣ
On buttonᓔᴎᣝ䬂
IrDA windows㑶໪にষ
Marketing CHINA:TEL:86 (010) 82951585 82946733 FAX:86 010 58859230
Getting Started
Storage. The Surtronic Duo is supplied in its
storage configuration, that isTraverse Unit
connected to Control Unit. Before use, it is
necessary to separate the Traverse Unit from
the Control Unit.
Note: It is recommended that the unit is stored
in the storage configuration when not in use,to
prevent damage to the stylus.
Positioning. Position the
Traverse Unit on the surface
to be measured as shown.
There are two modes for operating the
Surtronic Duo:
Remote. In remote mode the Con-
trol Unit is positioned within 1m of the
Traverse Unit and data is passed
between the two units over the
IrDA link.
Note: The alignment
between the Traverse
Unit and Control Unit
must be maintained
during the measurement
process.
Connected. In connected mode
the Control Unit is connected to
the Traverse Unit as shown
before the measurement is made.
To connect the Control Unit to the
Traverse Unit, insert the Control
Unit as shown and tilt until the
latch engages.
Inserting the batteries
Remove battery covers as shown.
Insert new batteries, one 3V Lithium 2450 in the
Control Unit and three 3V Lithium 2450 in the
Traverse Unit. Replace the battery covers.
CAUTION: ONLY USE THE RECOMMENDED
BATTERIES. 䄺˖᥼㤤Փ⫼ӣ䫖⬉
Battery Disposal
WARNING: DO NOT ATTEMPT TO RE-
CHARGE THE BATTERIES. USED AND OLD
BATTERIES MUST BE DISPOSED OF SAFE-
LY. DO NOT INCINERATE.
Switching on
If the Control Unit
is not used for five
minutes the unit
enters the power
saving mode, with
the display turned
off. Pressing any button will
reactivate the Contol Unit.
The Traverse Unit switches off completely after
five minutes without a new measurement. To
reactivate the Traverse Unit press the On
button. An LED flashes to confirm activation.
Low battery indicators: ⬉∴⬉Ԣ˖
Control Unit
battery low
Traverse Unit
battery low
Calibrate the gauge󰶖ޚ
Position the Traverse Unit to measure the
Calibration Standard supplied with the
instrument.偅ࡼܗᬒ㕂೼󰵼䜡ⱘ5DḋᵓϞˈ
ḋᵓ󰵼⿄ؐ5D ᖂ㉇
On the Control Unit press the Mode Select
buttons until is indicated on Mode status.
᥻ࠊܗ䗝ᢽ󰶖ޚ῵ᓣˈᣝᓔྟ䬂ˈҾ఼㞾ࡼ
ᅠ៤󰶖ޚDŽ⊼ᛣ5D XP˗5] XPDŽ
Press the Start button on the Control Unit. This
will initiate the gauge calibration procedure.
The Control Unit will display the results of the
calibration. The Ra value should be 5.81μm
and the Rz value 21.5μm.
Measuring⌟䞣Operation of the Control
identical,
whether the system is in Connected or
Remote modes,
Position the
Traverse Unit on
the surface to be
measured. On the
Control Unit press
the Select
parameter button o select the parameter that is
required to be displayed upon completion of
measurement. Using the Select mode button
select either metric or imperial units as required.
To begin the
measurement press
either of the Start
buttons. The
‘measurement in
progress’ display is
as shown.
The measured
parameter is
displayed as shown.
If there is an error
the appropriate
error code is
displayed, see Error Codes.1)将驱动单元放置在
工件上,使测针接触被测表面
2)按模式选择键指示至um;
3)按开始键,仪器将执行测量
4)按参数选择键依次查看各参数结果
All parameters (not just the selected
parameter) are transmitted to the Control Unit
at the same time and temporarily stored. The
stored results may then be stepped through
using the select buttons.
lk

This manual suits for next models

2

Popular Measuring Instrument manuals by other brands

BIRD Termaline 8920 Series Operation manual

BIRD

BIRD Termaline 8920 Series Operation manual

Omntec OEL8000II owner's manual

Omntec

Omntec OEL8000II owner's manual

FAURE HERMAN FH8400 user manual

FAURE HERMAN

FAURE HERMAN FH8400 user manual

Buchi Process Analyser PA2 Operation manual

Buchi

Buchi Process Analyser PA2 Operation manual

PCE Health and Fitness PCE-PHM 12 user manual

PCE Health and Fitness

PCE Health and Fitness PCE-PHM 12 user manual

Aeon Labs Z-Wave Home Energy Meter instructions

Aeon Labs

Aeon Labs Z-Wave Home Energy Meter instructions

MASHIN VAT-650 user guide

MASHIN

MASHIN VAT-650 user guide

Steren CCTV-090 user manual

Steren

Steren CCTV-090 user manual

Monroe Electronics 288 Operator's manual

Monroe Electronics

Monroe Electronics 288 Operator's manual

Shodex Asahipak GF-HQ Series Operation manual

Shodex

Shodex Asahipak GF-HQ Series Operation manual

Emerson Rosemount CT5100 Reference manual

Emerson

Emerson Rosemount CT5100 Reference manual

Konica Minolta CS-2000 brochure

Konica Minolta

Konica Minolta CS-2000 brochure

OEG PR 400 operating instructions

OEG

OEG PR 400 operating instructions

Micronics Ultraflo 2000 operating manual

Micronics

Micronics Ultraflo 2000 operating manual

UEi C165 instruction manual

UEi

UEi C165 instruction manual

VDO Viewline TU00-0752-8707102 installation instructions

VDO

VDO Viewline TU00-0752-8707102 installation instructions

Systec deltawaveVER2 user manual

Systec

Systec deltawaveVER2 user manual

SONIX 30D Instructions for assembly operation

SONIX

SONIX 30D Instructions for assembly operation

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.