introduction
specification
Symbol & key button location
Operation
Power on/ off
Memory capacity
Thermocouple connection
Temp. Unit choice
Hold reading value
Time features
Time counting
Relative reading function
Memory function
Time setting
Record Time interval
Max. Min. Reading
Auto Power off
Low battery indication
Digital output
Appendex.
Software installation

Instructions for use
DTM-320/DTM-321/ DTM-322 is a handheld instrument with temperature and humidity measurement functions. DTM-321 / DTM-322 tool
Temperature Coefficient:
The following inaccuracies need to be added:
Notice:
When the operating temperature is not within the range of 18°C~28°C, each
until the memory is exhausted.
Each time you press the "REC" button to start recording, until you press "REC" again to end the recording, it is called a set of data. You can record unlimited sets of data until
Reading x 0.01% ± 0.06 °F
Identify the total error.
Reading x 0.01% ± 0.03 °C
The above accuracy is the specification of the instrument itself, but in actual application, the specification error of the thermocouple temperature rod itself needs to be considered. Please refer to the relevant specification sheet to confirm
https://tecpel.com
2. Specifications:
1
1. Introduction:
Temperature: [ DTM-320] -20~60 ÿ -4~140°F
T2: -200~200 ÿ 0.1ÿ; 200~1370 ÿ 1ÿ
Range
-200°C ~ 200°C
200°C ~ 400°C
400°C~ 1370°C -328°F
~ -200°F -200°F ~
200°F 200°F ~ 400°F
400°F ~ 2498°F
"Note 1":
Storage environment: Relative humidity 0 ~ 80% RH, no condensation
Display mode: four-digit liquid crystal display (LCD), three groups of display values.
Resolution: Humidity: 0.1%RH
” Appendix (Ambient temperature 23 ± 5°C, humidity below 80%RH)
Accuracy
Signal output: RS-232 data output
T2: -200~1370
,
Relative humidity 0 ~ 90% RH, no condensation
Read out for analysis, printing or archiving purposes. Large amounts of data can also be recorded by a computer (depending on your computer's memory and hard drive space).
K
Response time: Humidity: 75 seconds at slow wind speed
-200~200°F 0.1°F; other 1°F
Accuracy: Humidity: ±2.5%RH at 25ÿ
It has "K" type thermocouple temperature rod measurement function. TECPEL DTM-322 also has a memory device (Datalogger), which can connect up to
,
Temperature -10 ~ 60°C (14 ~ 140°F) No condensation
Measuring range: Humidity: 0% ~100%RH
±(0.3% reading + 1°C) ±(0.5%
reading + 1°C) ±(0.3% reading
+ 1°C) ±(0.5% reading +
2°F)
±(0.3% reading + 2°F) ±(0.5%
reading + 2°F) ±(0.3% reading
+ 2°F)
Temperature: T1: 0.1 0.1° ÿ F
[ DTM-321 / 322 ] T1: -20~60 ÿ -4~140°F
"
Temperature 0 ~ 50°C (32 ~ 122°F) No condensation
,
ÿ -328~2498°F
Continuously memorizes 16,000 entries (Note 1). You can directly record the temperature and humidity values in the meter without connecting to the computer, and then transfer the memorized data via the computer.
Temperature: T1: ±0.7 ±1.4° ÿ F; T2:
K type, as shown in the following table:
Operating environment: Altitude 0-2000 meters
,
,
Temperature: T1 40 seconds at slow wind speed
Machine Translated by Google

HOLD
C
OPEN
TIMER
F
•9Maximum and minimum reading value buttons•3 LCD display
•4Power switch button •10 reading value retention button
•13 tripod connector
•11 digital output connection hole
•7Record keystrokes
•12 AC power adapter socket
•14Battery box cover
•1 Sensing head protective cover
•8 Temperature unit selection button•2 sensing rod
•5 time display button
•6
timer button
MANUAL FOR SAFE
PLEASE READ
140F
%RH100
POWER-UP OPTIONS
9V BATTERY
NEDA 1604 6F22 00
TEMP: 20C
HUMIDITY: 0
4F 60C
3. Symbol definition and key position
2
DC9V
RS-232
2
6
8
4
1
3
8
Instructions for use
https://tecpel.com
DTM-320
Batteries used: Battery 9 VDC, 006P or IEC1604
AC converter 9 VDC/10mA or above, 3.5ÿ×1.35mm plug
Battery life: approximately 100 hours (alkaline battery)
Appearance dimensions: Table body: 186 length x 64 width x 30 height mm
Sensing rod: 190 length x 15 mm diameter
Weight: about 320g
Accessories: instruction manual, 9V battery, leather case, "K" type wire temperature cable (DTM-321 / DTM-322 ),
Windows operating software (DTM-322 ), USB transmission cable (DTM-322 )
Optional accessories: AC/DC power converter, RS-232 computer cable
Machine Translated by Google

8
3
5
9
3
5
9
4
9
4
2 2
6
1
7
1
F
POWER-UP OPTIONS
C
F
9V BATTERY
C60
HUMIDITY:
140
(200
TEMP: 20C TEMP:
C 1370
) 200C
POWER-UP OPTIONS
C
60
PLEASE READ
F
C328 %RH100
0
NEDA 1604 6F22 00
F140
MANUAL FOR SAFE
44F 20
2498
F328 100%RH
F
(
2498
)
C
HUMIDITY: 0F
1370
CAL
RS-232
DC9V
DTM-321/322
DTM-322DTM-321
https://tecpel.com
Instructions for use
HOLD
FREC CF
REL HOLD TIME
C
OPEN
12
SETUP
CLOCK INTV
main display
Second display
Automatic shut-down
Sensor type
Low battery indicator
relative temperature
minimum value
Recording
The reading value retains the maximum relative value
years
) K-TYPE
T2 (T2 (K-TYPE)T1 T1
•13 digital output connection hole
•1 Sensing head protective cover
•8Record key presses •14 AC power adapter socket
•6 Relative value display button •12 K type 0°C calibration button
•7 time display button
•4 LCD display •10 Maximum and minimum reading button •16 Battery box cover
•11 Reading value retention button•5Power switch button
•2 sensing rod
•9 Temperature unit selection button
•3 Thermocouple connector socket •15 tripod connector
REC
Machine Translated by Google

4.3 Connect the temperature thermocouple: ( DTM-321/DTM-322 model )
Instructions for use
4.7 Timer function: ( DTM-320 model )
4.8 Relative reading function (return to zero): ( DTM-320 model )
4.2 Method of expressing remaining memory space: (DTM-322 model )
4.9 Memory function: ( DTM-322 model )
4.6 Time function: ( DTM-322 model )
4.5 Lock the reading value:
Press the "HOLD" button to lock the displayed reading value, and press it again to unlock the reading value; when the reading value is locked, the ÿREL, MAX/MIN and "°C/°F" functions
will not be executed.
After powering on, a number will appear below the LCD, which represents the remaining recordable size of the instrument's
memory. As shown in the picture on the right: it represents the remaining 16,000 pieces of data available for use in the memory.
When the "REL" key is pressed , the temperature and humidity meter will record the current reading and display the difference between the current reading and the memory reading. Press
again to deactivate this function.
Press the " TIME " key to switch to display time, which will display the year, month, day and time. Press any key to leave the time display mode and restore the reading value display mode.
Note: If the time is
2000/01/01 00:00 every time it is turned on, it means that the clock battery needs to be replaced. Remove the bottom cover and replace the 3V button battery.
Insert the connector plug of the thermocouple temperature measuring rod into the connector at the top of the thermometer and hygrometer. Please pay attention to the positive and negative polarity of the connector.
4.1 Power on and off: Press
the I key to power on or off the thermometer and hygrometer.
4.4 Temperature unit selection: Press
the "°C/°F" key to select the temperature unit displayed in °C or °F.
Press the "REC" button to start recording according to the set recording time interval. After starting recording, the "REC" symbol
will appear in the upper right corner of the LCD. Pressing the "REC" key again will stop recording, thus completing a set of data.
When the memory is full, the "REC" symbol will flash once a
second. To view the memory contents, you need to use the included RS232 cable and software to transfer the memory contents
to the computer. If the memory is full and you want to record again, you must clear the memory first. To clear
the memory, first
turn off the phone, then press and hold "REC", then press and hold the power button for two seconds, and then release all keys. ,
at this time "CLR" will appear on the LCD (as shown on the right), indicating that the memory is being cleared. After the clearing is
completed, "CLR" will disappear, and the "REC" symbol will stop flashing and disappear.
Press the "TIMER" button to start the timing function, which is displayed as minutes: seconds (m:s). Press it again to stop the timing action; press it again to continue the timing action, so
that the entire timing program can be repeated. When the timer exceeds 59 minutes and
59 seconds, the display unit will change to hours: minutes (h:m). When the "TIMER" button is pressed for 2
seconds, the timer will be reset.
"CLR" appears, indicating that the memory is
being cleared.
https://tecpel.com
4. Operation instructions:
4
Machine Translated by Google

Instructions for use
https://tecpel.com
Step 1: First press and hold the "MAX MIN" button, then press the power button, and then release these two
buttons (Picture 1)
4.10 Time setting: ( DTM-322 model )
Step 2: Press the "HOLD" (INTV) key to start entering the time interval setting (Figure 2)
4.14 Low battery indication:
(figure 1)
When it appears on the display, it means the battery is low and needs to be replaced with a new one.
(figure 2)
4.13 Automatic shutdown:
When turned on in the normal way, a symbol will appear on the upper left side of the LCD, indicating that it will automatically shut down if there is no key operation or RS232
connection and no data is recorded within 30 minutes. When you want to release the
automatic shutdown mode, you can press the "HOLD" button and then turn on the phone. The symbol on the LCD will not be displayed, indicating that the automatic shutdown has been released.
When RS-232 and REC are activated, the watch will automatically release the automatic shutdown function. .
4.11 Recording time interval setting: ( DTM-322 model )
Step 2: Press the "CLOCK" button ("TIME") to enter the time setting mode (Figure 2)
(figure 2)
4.12 Maximum and minimum reading function:
Step 3: Press "REC" ÿ or "°C/°F" ÿ to increase or decrease the number. If you want to quickly increase or
decrease the number, please hold this button and press "Time" to change the adjustment
items. The order is: year ÿ month ÿ Day ÿ Hour ÿ Minute ÿ Second, then press "TIME" to save
the setting value. During the setting process, if you want to leave the setting mode without saving
the setting value, you can press the power button to end it.
Step 3: Press "REC" ÿ or "°C/°F" ÿ to increase or decrease the number. To quickly increase or decrease the
number, please press and hold this button, and then press the "HOLD" (INTV) button to adjust
the next item, and finally press "HOLD" (INTV) to save the setting value. During the setting,
if you want to leave the setting mode without saving the setting value, you can press the power
button to end. [Note] The setting mode is divided
into :Second
(figure 1)
When the "MAX MIN" key is pressed, the temperature and humidity meter will record the maximum value and the
minimum value at the same time; when MAX appears, the displayed value is
the recorded maximum value. Press the "MAX MIN" key again , and the value displayed when MIN appears is the minimum
recorded value. Press the "MAX MIN" key again . When MAX MIN flashes at the same time, the value displayed is the current measurement value.
Press the "MAX MIN" key again to return to the MAX display. In this way, the user
can select the appropriate reading value display, and the reading value during the period will be updated at any time.
This operating state will end when the "MAX MIN" key is pressed for 2 seconds.
Step 1: Press and hold the "MAX MIN" button and then turn on the phone to enter the setting mode. (Picture 1)
5
Machine Translated by Google

Instructions for use
GND
RX
Tx
Bead probe -50ÿ to 200ÿ
TPK-01 (±3.6 or ±0.75 ÿ %)
Teflon insulation, maximum temperature resistance
260 ÿ
-58 to ÿ ÿ 392
https://tecpel.com
1. Do not touch the humidity sensor directly with your hands.
4.15 Digital output:
3. Do not immerse the sensor directly in water or solvent.
2. Do not expose the sensor directly to strong electromagnetic fields.
,
model
: (DTM-322/ DTM-321)
illustrate
:
4. If there is dust, please blow it away with clean air or wipe it gently with alcohol. Do not use other chemical solvents to wipe.
Output method: UART 9600bps N, 8, 1 normal High. Output
format: ASIC II character output and encoding status flag report
Accuracy
±2.2 or ÿ ±0.75%
scope
Signal output connection: please refer to the picture on the right
Appendix
This table includes a linear temperature rod with the following specifications:
Warning !
6
Machine Translated by Google

Instructions for use
Figure 5-1
Minimum hardware
5-1.Execute TestLink
_
Live data table window
Two-way control window
Install the TestLink (Humidity DataLogger) software; the attached CD is the
TestLink SE-310 program. Please install it on your computer and connect the computer cable to the temperature and humidity meter and the computer.
requirements: 486-100MHZ or above PC compatible, 16MB memory, 5MB or above available hard disk space.
Operating system requirements: Win98 /NT4.0 /NT2000 /XP /VISTA
Installation
method: Put the CD in the CD drive under Windows, and the installation screen will automatically appear. Select "Install TestLink SE-310" and the installation program
will be executed. If the installation screen does not appear automatically, select the CD in the file manager. The setup.exe in the SE-310 directory can execute the
installation program. Follow the instructions to complete the installation. After the installation is completed, the "TestLink" program item will appear in the "Program Set"
of "Start". Click on the "SE310" "Execute the program. For other detailed instructions, please refer to the online help.
Real-time data sampling rate editing grid
Real-time data graphics window
5-2. Real-time data graphics and real-time table windows
Click on it with the mouse, and TestLink will start collecting the data on the temperature and humidity meter into the graph window and table window according to the
time interval of the real-time data sampling rate. You can use the mouse to click on the editing box of the sampling rate, Then enter the new sampling rate.
Main menu and toolbar
https://tecpel.com
5. Software installation instructions: ( TECPEL DTM-322)
Program main screen
7
Machine Translated by Google

Instructions for use
8
https://tecpel.com
graphics window
Data Group Window – It will show how many data sets (when you press the “REC” button on the temperature and humidity meter to start recording and press
“REC” again to end the recording, it is called a data set) that has been loaded into the computer. To view the data of any set of data sets in detail, just click on
the data set in the data group window with the mouse, and TestLink will display the contents of this set of data sets in the graphics window and table on the
right in the form of graphics and text. on the window.
Data group window
Figure 5-2
Select "DataLogger" on the main menu, and TestLink will load the memory data on the temperature and humidity meter into the computer, and will display the
loading progress. After the loading is completed, the screen as shown in Figure 5-2 will appear:
table window
5-3. DataLogger ( only applicable to DTM-322)
5-4. For other operating instructions, please refer to the online instructions. ( Help on the main menu)