
Operating
Instructions--~7A'18/7A18N
Panel Controls and Connectors), Turn the knob clockwise,
CH
2
POLARITY
Switch
then counterclockwise, until the GAIN control is engaged.
When the GAIN control is engaged, the vertical deflection
will change as the knob is turned. Turn the GAIN control
knob with the screwdriver until the deflection is set to
exactly four divisions, then remove the screwdriver.
Input Coupling
The Channel
1
and Channel 2 coupling (AC-GND-DC)
switches allow a choice of input coupling methods. The
type of display desiredand the applied signal will determine
the couplingto use.
The DC coupling position must be used to display the
DC component of the signal,
It
mustalso be used todisplay
AC signals below about
30
hertz (ten hertz with a
10X
probe) and square waves with low-frequency components as
these signals are attenuated inthe AC position.
In the AC coupling position, the DC component of the
signal
is
blocked byacapacitor inthe inputcircuit. IheAC
coupling position provides the best display of signals with a
DC component much larger than the AC components. The
precharge feature should be used with large DC inputs. 70
use this feature, first set the coupling toGND. Connect the
probe to the circuit and wait about two seconds for the
coupling capacitor to charge. Then set the coupling toAC.
The GND position provides a ground reference at the
input of the amplifier without externally grounding the
input connectors. However, the signals connected to the
inputs are notgrounded, and the same DC load is presented
to the signa! source.
VOLTSIDIV and VARIABLE Controls
The amount of vertical deflection produced bya signal is
determined by the signal amplitude, the attenuation factor
of the probe, the setting of the VOLTSlDlV switch, and
the setting of the VARlABLE control. Calibration deflec-
tion factors indicated by the settings of the VOLTS/DIV
witch apply only when the VARIABLE control is in the
calibrated (CAL IN)position.
The VARIABLE control provides variable, uncalibrated
settings between the calibrated steps of the VOLTS/DIV
switch. With the VARIABLE control fully counterclock-
wise and the VOtTSlDlV set to
5
voltsldiv the uncali-
brated vertical deflection factor
is
extended to at least 12.5
volts/division. By applying a calibrated voltage source to
the input connector, any specific deflection factor can be
set within the range ofthe VARIABLE control,
The CH
2
POLARITY switch may be used to invert the
displayed waveform ofthesignal applied totheCH
2
input.
This
is
particularly useful in added operation of the 7A18
when differential measurements are to be made. The CH
2
POLARITY switch has two positions, +UP and INVER1. In
the +UP position, the displayed waveform will have the
same polarity as theapplied signal and a positive DC voltage
will move the CRT trace up. In the INVERT position,
a
positive-goingwaveform at the CH
2
inputwill be displayed
on the CRT in inverted form and apositive DCvoltage will
move the trace down.
DISPLAY MODE Switch
For single-trace operation, apply tire signal either tothe
CH
1
inpttt or the CH
2
inputand set the DISPLAY MODE
switch tothecorresponding position: CH
f
or CH
2.
To display a signal inone channel independently when a
signal is also applied to the other channel, simply select the
desired channel by setting the DISPLAY MODE switch to
the appropriate CH
1
or CH
2
position.
Alternate Mode. The
ALT
position of the DISPLAY
MODE switch produces a display which alternates between
channel 1 and channel
2
with each sweep on the CRT.
Although the ALT mode can be used at all sweep rates, the
CHOP mode provides a more satisfactory display at sweep
rates below about
0.2
millisecond/division. At slow sweep
rates alternate mode switching becomes visually per-
ceptible,
Add Mode. "The ADD position of the DISPLAY MODE
switch can be used to display the sum or difference of two
signals, for common-mode rejectionto removean undesired
signal, or for DC offset (applying a DC voltage to one
channel to offset the DC component of a signal on the
other channel). The overall deflection factor in the ADD
mode with both VOLTSlDIV switches set to the same
position is the deflection factor indicated by either
VOLTS/DIV switch. However, if the CH
1
and CH
2
VOLTSlDIV switchesare set to different deflection factors,
the resultant amplitude is difficult to determine from the
CRT display. In this case, the voltage amplitude of the
resultant display can be determined accurately only if the
amplitude of the signal applied toone channel is known. In
the ADD mode, positioningof thetrace is contro:led by the
channel
1
POSITION controlonly.
Chop Mode. The CHOP position of the DISPLAY
MODE switch produces a display which
is
electronically
switched between channels at approximately a
500
kilo-
hertr rate (controlled by mainframe). Ingeneral the CHOP
mode provides the best display at sweep rates slower than
REV.
APR
1974
2-3
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com