
Typll! 59 Amplifier
3. Position a free-running trace (or spot, if you are not us-
ing a time-base module) so that it lies along one of the hori-
zontal lines of the graticule. This line will be used as a
zero reference line; its position in any given case will depend
upon the polarity and amplitude of the voltage to be mea-
sured. Do not adjust the POSITION control of the Type 59
module after this reference line has been established.
4. Set the AC-DC-GND. switch to DC and display the sig-
nal to be measured.
5. Measure the vertical distance, in graticule divisions, from
the reference line established in step 3 to the point whose
de level you wish to measure.
6. Multiply the distance obtained in step 5 by the deflection
factor established in step 1. This is the de level of the point
you wish to measure. If the point is above the zero refer-
ence line, its potential is positive with respect to zero; if the
point is below the zero reference line, its potential is nega-
tive with respect to zero.
Note that you can reestablish your zero reference line at
any time merely by placing the AC-DC-GND. switch to
GND.; you do not need to disconnect the probe from the
signal source. If desired, you can establish a reference other
than zero by placing the AC-DC-GND. switch to DC, con-
necting the signal probe to the desired reference voltage,
and positioning the trace as described in step 3 of the pro-
cedure.
CIRCUITDESCRIPTION
The Type 59 Amplifier module consists of an Input CF
stage, V434A, and an Output Amplifier stage, V434B-V444A.
The output stage is a cathode-coupled paraphase amplifier,
which converts the single-ended input signal to a push-pull
output signal. With the VARIABLEGAIN control set fully
clockwise (maximum gain position), the amplifier has a
gain of about 20.
The front-panel POSITION control effectively moves resis-
tance from the cathode circuit of one side of the paraphase
amplifier to the cathode circuit of the other. This varies the
quiescent currents through the tubes inversely to one another.
The quiescent current through V434B and \/444A are further
controlled by the setting of the POSITION RANGE adjust-
ment, thereby controlling the range through which the POSI-
TION control will position the signa1 on the screen. The
proper method of setting the POSITION RANGE adjustment
is described in the Calibration instructions.
The ferrite beads threaded on the plate leads of the two
Output Amplifier tubes are parasitic suppressors to control
high-frequency oscillations.
2
The Trigger Pickoff circuit samples the plate signal of
V444A and applies the sample through V444B to the trigger
input circuit of the time-base module in the other oscillo-
scope opening, if one is used. There is about 4 volts of
signal at the cathode of V444B for each division of deflec-
tion on the screen.
TROUBLESHOOTING
In the following discussion it is assumed that you have
read the general maintenance and troubleshooting informa-
tion contained in the oscilloscope manual and have definite-
ly isolated a trouble to the Type 59 module by the pro-
cedures described there.
Since there are only two tubes in the Type 59 module, the
first thing to do in case of trouble is to replace both tubes.
If this does not eliminate the trouble, check the rest of the
circuit by voltage and resistance measurements. Typical
voltages to be encountered at certain points throughout the
module are given on the schematic diagram at the rear of
this section. Resistance measurements will usually be point-
to-point checks for which the proper value can be approxi-
mated from the schematic diagram.
CALIBRATION
Calibration of the Type 59 module is performed with the
module inserted in either opening of a Type 560-Series
Oscilloscope. There need not be a module in the other
'oscilloscope opening, although either an amplifier or time-
base module may be there. The only test equipment re-
quired for the calibration is a de voltmeter.
POSITIONRANGEAdjustment
To properly set the POSITION RANGE adjustment (see
Fig. 2), set the POSITION control to midrange, and adjust
the POSITION RANGE adjustment to center a free-running
trace, or spot, on the screen.
INT.TRIG.DCLEVELAdjusiment
To properly set the INT. TRIG. DC LEVELadjustment (see
Fig. 2), proceed as follows:
1. Set the AC-DC-GND. switch to GND. and center the
trace or spot on the screen.
2. Set the INT. TRIG. DC LEVELadjustment so that the po-
tential at pin 11 of the interconnecting plug is exactly zero
with respect to ground.
@