Tenmars TM-305U User manual

Page 1
Temperature/Humidity Datalogger
Users Manual V.1.0
All manuals and user guides at all-guides.com
all-guides.com

Page 2
TABLE OF CONTENTS
FEATURES & SPECIFICATIONS ..................................................................................... 4
START UP AND GENERAL OPERATION ........................................................................ 7
LCD Display................................................................................................................................................................... 8
Sampling Rate............................................................................................................................................................... 9
Recording and Datalogging .................................................................................................................................... 10
Calibration ................................................................................................................................................................... 11
33% Calibration........................................................................................................................................................ 11
75% Calibration........................................................................................................................................................ 13
25C Calibration......................................................................................................................................................... 15
PC SOFTWARE INSTALLATION.................................................................................... 16
USING PC SOFTWARE WITH METER DATALOGGER ................................................ 18
Connect Meter Datalogger to PC ........................................................................................................................... 19
Setup Screen............................................................................................................................................................... 22
Download Records .................................................................................................................................................... 23
Graphic & Chart Analysis ........................................................................................................................................ 26
Data in Listing Format .............................................................................................................................................. 28
Data Export.................................................................................................................................................................. 30
All manuals and user guides at all-guides.com

Page 3
Model Number
Temperature/Humidity Datalogger
All manuals and user guides at all-guides.com

Page 4
FEATURES & SPECIFICATIONS
Features
•Measure humidity and temperature.
•Datalogging- 50,000 records.
•Fast USB download .
•Easy- to-Use PC software for parameter settings and data analysis.
•Internal Humidity Temperature alarm.
All manuals and user guides at all-guides.com

Page 5
Specifications
General
Operation Condition -40-185℉(-40-85℃)
1-99% RH, non-condensing
Relative Humidity Sensor
Type Capacitor
Measurement Range 1%-99%
Accuracy ±3.0%RH(20~80%)
±5.0%RH(<20%,>80%)
Resolution 0.1%
Temperature Sensor
Type Thermistor
Measurement Range -40-185℉(-40-85℃)
Accuracy ±0.6℃(±1.2℉)from -20-50℃(-4 to 122℉)
±1.0℃(±1.8℉) all other ranges.
Resolution 0.1℃
Power Supply
Battery One 3.6V ½ AA lithium battery
Battery Operation 1 Year
External Power Supply 5 VDC with 0.5A.
All manuals and user guides at all-guides.com

Page 6
Applications
•Building HVAC monitoring.
•Indoor air quality survey.
•Locate the presence of combustion fumes from vehicles and appliances.
LCD Display
Meter LCD
1. Temperature Display 4. USB Communication
2. Humidity Display 5. Battery Level Indicator
3. Recording
All manuals and user guides at all-guides.com
all-guides.com

Page 7
WHAT’S IN THE BOX
START UP AND GENERAL OPERATION
•Press the Start button for 2 seconds to turn on Meter Datalogger
•The unit is ready to use.
1.Main Unit 2. USB Cable 3. Installation Disk
4. 3.6V ½ AA lithium
battery
All manuals and user guides at all-guides.com

Page 8
LCD Display
•Press Sel to jump to the next LCD display.
•Main Screens (1-3)
1. 2. 3.
Main Screen
Temperature(C)
Main Screen
Temperature.(F)
% of Memory Left
Total 24,000 records
Main Screens
All manuals and user guides at all-guides.com

Page 9
Sampling Rate
(5sec / 10sec / 1min / 5min / 10min / 30min / 1 hour/ 2 hour)
Can only be set from PC setup screen.
All manuals and user guides at all-guides.com

Page 10
Recording and Datalogging
•Press the Start button to start data recording.
•The LCD will display REC when Meter is in the datalogging mode.
•Press the Start button again to stop recording.
All manuals and user guides at all-guides.com

Page 11
Calibration
33% Calibration
Press S
t
art + SEL to enter 33% humidity calibration mode.
•Press Start to start 33% calibration.
All manuals and user guides at all-guides.com
all-guides.com

Page 12
•During the calibration, the LCD will display an antenna sign.
•Meter Datalogger will start 33% Rh sampling 45 minutes after the Start button is
pressed.
•The COM icon will show up when calibration is completed.
•Press SEL to select the next calibration page.
All manuals and user guides at all-guides.com

Page 13
75% Calibration
•Press and SEL button then press the Start button at the same time to enter the
calibration mode.
•Press Sel one more time to switch to the 75% Rh calibration page.
•Press Start to start 75% calibration.
All manuals and user guides at all-guides.com

Page 14
•Meter Datalogger will start 75% Rh sampling 45 minutes after the Start button is
pressed.
During the calibration, the LCD will display an antenna icon.
•The COM icon will show up when calibration is completed.
•Press SEL to select the next calibration page
All manuals and user guides at all-guides.com

Page 15
25C Calibration
•Press Sel one more time to jump to the 25C calibration page.
•Press Start to start 25C calibration.
•During the calibration, the LCD will display an antenna icon.
•When the calibration is completed, the COM will show up.
All manuals and user guides at all-guides.com

Page 16
PC SOFTWARE INSTALLATION
Installation Requirements:
Minimum system requirements for the software:
-Pentium III - 500 MHz processor or equivalent with 128 megabytes of RAM.
Optimal Performance:
-Pentium 4 - 2.8 gigahertz processor or equivalent with 256 megabytes or RAM.
Operating System Requirements:
-Windows 2000
-Windows XP
-Window 7
-Window 8
All manuals and user guides at all-guides.com
all-guides.com

Page 17
•Insert the CD into the disk drive.
•Click Setup to start installation.
All manuals and user guides at all-guides.com

Page 18
•Follow the steps on the Installation Shield Wizard and click Finish to complete the
installation.
USING PC SOFTWARE WITH Meter Datalogger
•Start Window PC software.
•Click on All programs, Click on “Meter Logger”
•Click on Logger to start the datalogger program.
All manuals and user guides at all-guides.com

Page 19
Connect Meter Datalogger to PC
No Connection Screen
All manuals and user guides at all-guides.com

Page 20
•Connect Meter Datalogger to the PC with a USB cable.
All manuals and user guides at all-guides.com
Other manuals for TM-305U
1