Teseq NSG 3040 User manual

1
NSG 3040
EMC TEST SYSTEM
USER MANUAL
601-279F

NSG 3040 EMC test system
NSG 3040
EMC TEST SYSTEM
USER MANUAL

1 Explanation of symbols 9
2 Introduction 10
2.1 General description 10
3 Standards and applications 12
3.1 Burst test 12
3.2 Combination wave test 13
3.3 Mains quality test 13
3.4 Magneticeldswithmainsfrequency 14
3.5 Pulsedmagneticelds 14
3.6 Slow Damped Oscillatory Wave (NSG 3040-SOW only) 14
4 Safety instructions 16
4.1 General 16
4.2 Installation 17
4.3 InstallationofanEUTpowerswitch 18
4.4 Applicablesafetystandards 18
4.5 Testexecution 19
4.6 Userwarnings-Generator 20
4.7 DangersconcerningtheEUT 21
5 First steps 22
5.1 InstallationoftheNSG3040system 23
5.2 Connectingthetestsystemtothegroundreferenceplane 24
5.3 Mountingina19”rack 24
5.4 Reargroundbrackets(INA165) 24
6 Mainframe description 25
6.1 Front panel 25
6.1.1 EUToutput 25
6.1.2 Highfrequencygroundterminal 26
6.1.3 Surge or SOW output 27
CONTENTS

6.1.4 Burst output 27
6.1.5 IndicatorLEDs 27
6.1.6 Touchscreenanduserinterface 28
6.2 Rearpanel 28
6.2.1 Instrumentsupply 28
6.2.2 EUTpowerinput 29
6.2.3 DCEUTinput 31
6.2.4 Ground connection point 31
6.2.5 Systeminterfaceconnector25pinDsub 31
6.2.6 Synchro-Bussystem 32
7 The standard user interface (SUI) 34
7.1 Main menu 35
7.2 System window 36
7.2.1 GENERALsettings 36
7.2.2 EQUIPMENTscreen 39
7.2.3 COMMUNICATIONscreen 40
7.2.4 MONITORINGscreen 42
7.3 UpdatingSUIsoftwareviatheSD-card 43
8 Parameter setting window 47
8.1 Theredmenubar 47
8.2 Thebottombar 52
8.2.1 Loadusertest 52
8.2.2 Loadstandardtest 53
8.2.3 Savetest 56
8.2.4 Keypad 56
8.2.5 Showgraphic 57
8.3 Burstgeneratorsetting 57
8.3.1 Testcongurationwithpowerlinecoupling 58
8.3.2 Testcongurationwithexternalcoupling 58
8.3.3 Burstparameterswindow 59
8.3.4 Voltage 59
8.3.5 Frequency 59
8.3.6 Phase 59
8.3.7 Coupling 60
8.3.8 Bursttime 61
8.3.9 Repetitiontime 61
8.3.10 Testduration 61

8.3.11 Burstgeneratortechnicaldata 62
8.3.12 Derating 62
8.4 Combinationwave(Surge)–parametersetting 67
8.4.1 Testcongurationforpowerlinecoupling 67
8.4.2 Testcongurationforexternalcoupling 67
8.4.3 Voltage 69
8.4.4 Impedance 69
8.4.5 Phase 69
8.4.6 Coupling 70
8.4.7 Repetitiontime 71
8.4.8 Testduration 71
8.4.9 Surgegeneratortechnicaldata 71
8.5 Dips,interruptsandvariations 72
8.5.1 Examplesofdips&interrupts 73
8.5.2 Dips&interruptsgenerator 73
8.5.3 VoltageUVar 74
8.5.4 Phase 74
8.5.5 Repetitiontime 74
8.5.6 T-Event 74
8.5.7 Testduration 74
8.5.8 Dipsandinterruptstechnicaldata 75
8.5.9 Variationtesttechnicaldata 75
8.6 Powermagneticeldtesting(-4-8)parametersetting 77
8.7 Pulsedmagneticeldtesting(-4-9)parametersetting 78
8.8 SlowDampedOscillatoryWavetest(-4-18)parametersetting 78
8.8.1 SlowDampedOscillatoryWaveparameterswindow 79
8.8.2 Coupling 81
8.8.3 Outputimpedance 82
8.8.4 Timingparameters 82
8.8.5 Testduration 83
8.8.6 Phasesynchronization 83
8.8.7 Derating 83
9 Description of the 25 pin D-Sub signals 86
9.1 Interlock 86
9.2 Triggertoscopeoutputsignal 87
9.3 Synchronization(Sync)signal:Outputsignal 87
9.4 Pulseenable/nextstepinput 88

NSG 3040 EMC test system
9.5 EUTfailinput 88
9.6 EUTpoweroff 88
9.7 Highvoltageactive 89
10 Coupling network CDN 3041 90
11 Various NSG 3040 versions 92
11.1 NSG3040IEC 92
11.2 NSG3040“alacarte”(customerspecic) 92
11.2.1 MainframeNSG3040-MF 93
11.2.2 MainframeforexclusiveremotecontrolNSG3040-MF-ERC 93
11.2.3 CombinedwavesurgemoduleCWM3450 94
11.2.4 DipsandinterruptsmodulePQM3403 94
11.2.5 Electricalfasttransient/burstmoduleFTM3425 94
11.3 NSG3040-xxx-ERCseries 94
11.4 NSG3040-xxx-EPOseries 96
11.5 NSG3040-DDV 97
11.6 NSG3040-SOW 98
12 Maintenance and function check 99
12.1 General 99
12.2 Cleaning 99
12.3 Functioncheck 100
12.4 Calibration 100
12.5 Warranty 101
13 Declaration of conformity (CE) 102
14 Accessories 103
14.1 PCsoftware 103
14.2 Couplingdecouplingnetworksformultiplephases,higher
currents and voltages 104
14.2.1 CDN 3043 - 32 A series 105
14.2.2 CDN 3063 series - 63 A and 100 A series 106
14.3 Variablevoltagesource 107
14.3.1 AutomaticVariacs 107
14.3.2 Manualsteptransformer:INA6501 112
14.3.3 Manualsteptransformer:INA6502 116
14.4 Magneticeldoptions 121
14.4.1 Manualsolution:MFO6501 125
14.4.2 Automaticsolution:MFO6502 129
14.5 PulsewaveshapeadapterINA752 134

14.6 Couplingdecouplingnetworksfordatalines 136
14.6.1 BurstEFTcouplingclampNSG3425-andsafetycoverINA3825 136
14.6.2 SurgeCDNforunsymeticdatalinesCDN117 139
14.6.3 SurgepulseCDNforsymmetricdatalinesCDN118 140
14.7 Measuring accessories 142
14.7.1 MD200andMD200Adifferentialhighvoltageprobes 142
14.7.2 MD 300 surge pulse current probe set 144
14.7.3 MD 310 SOW pulse current probe set 146
14.7.4 Burst/EFTpulsevericationkit 148
14.8 Cables,plugsandadapters 149
14.8.1 Calibrationadapters 149
14.8.2 Testadapters 150
14.8.3 Variouscablesandplugs 151
14.9 Mountingaccessories 153
15 System description 154
16 Addresses 156

8
NSG 3040 EMC test system
WARNING - Lethal danger from high voltages and the
risk of radiating illegal electromagnetic interference.
This system must be used only for EMC test purposes
as specied in these operating instructions.
The NSG 3040 must be installed and used only by autho-
rized and trained EMC specialists.
Personnel tted with a heart pacemaker may not
operate the instrument and must not be in the vicinity
of the test setup while it is in operation.
When the system is used in conjunction with options,
accessories or other equipment the safety instructions
concerning those devices must also be observed.

9
Pleasetakenoteofthefollowingexplanationsofthesymbolsusedinorder
toachievetheoptimumbenetfromthismanualandtoensuresafetyduring
operationoftheequipment.
The following symbol draws your attention to a circumstance where non-
observationofthewarningcouldleadtoinconvenienceorimpairmentinthe
performance.
Example:
1 EXPLANATION OF SYMBOLS
This connection must not be confused with the Equip-
ment under Test (EUT) power input.
The following symbol draws your attention to a circumstance where non-
observationofthewarningcouldleadtocomponentdamageordangertothe
operating personnel.
Example:
Never connect or disconnect the EUT while the test
system is performing a test.

10
NSG 3040 EMC test system
2.1 General description
TheNSG3040testsystemisamultifunctiongeneratorthatsimulatescable-
borneelectromagneticinterferenceeffectsforimmunitytestingtointerna-
tional,national,andmanufacturers’standards.
Thesystemisdesignedtofulllconductedelectromagneticcompatibility(EMC)
testrequirementsforcompliancetestingofhousehold,ofce,lightindustrial
orcommercialequipment,includingcombinationwavesurge,ElectricalFast
Transient(EFT)pulses,andPowerQualityTesting(PQT).
TheNSG3040’smodulararchitectureandindustrystandardinterfacesallowit
tobeeasilyexpandedandcustomizedtomeetindividualtestingneeds.
Thesystemisdesignedasaseriesofinteroperablefunctionunitswithamaster
controllerthathandlestherealtimefunctionsandcommunicateswiththe
functionmodules.Eachfunctionunitcontainsaslavecontroller;allfunction
unitsareconnectedtogetherthroughtheirslavecontrollersandnetworked
withthecentralmastercontrollerviaaeldbus(Interbus).Informationconcern-
ingspecialfeaturesandtheiradjustableparametersarestoreddirectlyinthe
functionmodules.
Thismodularityenablesthefunctionunitstobecombinedintocustomized
testsystems,andlaterreconguredtoaddresschangingtestingrequirements.
Thefunctionunitscanbereadilymodiedtoaddresstherequirementsofnew
standards,andnewfunctionunitsfornewparametersmaybeincorporated
inexistingsystems.
TheNSG3040iscontrolledthroughitsstandarduserinterfaceviaatouch
paneldisplay.ThesystemcanalsobecontrolledbyaremotePCviaitsEthernet
interface.
2 INTRODUCTION

11
To ensure optimal user and equipment safety, only industry-standard and
correctlyspeciedplugsandsocketsareusedthroughout.

12
NSG 3040 EMC test system
TheNSG3040testsystemisdesignedprimarilyforcable-bornetransientinter-
ferencetestsasspeciedintheEuropeangenericstandardsIEC/EN61000-6-1
coveringequipmentforhousehold,ofceandlightindustrialuse,andIEC/EN
61000-6-2forapplicationsinindustrialenvironments.TheNSG3040generates
thesetestsinaccordancewithIEC/EN61000-4-4,-4-5,-4-11,-4-18and-4-29,
dependingofthemodel.Accessoriesareavailableforgeneratingoptionaltests
toIEC/EN61000-4-8and-9.
TheEUdirectiveNo.2004/108/EC(fortheassignmentoftheCEsymbol)refers
tothesestandardsandtothistypeofequipment.
3.1 Burst test
BursttestsincompliancewithIEC/EN61000-4-4simulatethehighvoltage/high
frequencyinterferencepulsestypicallyproducedwhenaninductivelyloaded
switchisoperated.Withoutcountermeasures,suchinterferencemayoccur
whenacurrentthroughanelectromagneticdevice,e.g.motor,circuitbreaker,
relay,uorescentlamp,etc.isswitchedoff.
Thistypeofinterferencecanaffectotherequipmentineitheroffollowingtwo
ways.Firstly,theinterferencecanbecoupleddirectlyintothetargetequipment
viathemainspowercable.Theinterferencecanbetransmittedfromthesource
alongthemainspowercableconnectedtothetarget.Interferencefromthe
mainscanreachanyotherpieceofequipmentconnectedtothesamepower
sourceinasimilarway,howeverthisdoesnotallhavetooccurinthesame
sectionofabuilding.
Alternatively,theinterferencecanbecapacitivelycoupledintoanytargetdevice
inthevicinity.
Thesystemenablesatesttobeperformedusingbothstandardizedcoupling
3 STANDARDS AND APPLICATIONS

13methods.TheEUTisconnectedtothemainspowersocketonthefrontpanel
ofthetestsystemforthedirectmainsinjectiontest.Capacitivelycoupledtests
requiretheinterferencetobesuperimposedontothesignalordatalinecables
viaanexternalcouplingclampthatisconnectedtotheburstoutputonthefront
panelofthesystem.
3.2 Combination wave test
Thesurgetest,incompliancewithIEC/EN61000-4-5,duplicateshighvoltage/
highenergyinterferenceasexperiencedwithalightningstrike.Generallyspeak-
ingtheinterferencendsitswayintohouseholdequipmentviathemainspower
supply.
Thiskindofinterferencecanaffectequipmentineitheroftwoways.Firstly,the
interferencecanbecoupleddirectlyintotheequipmentviathemainssupply.
Theinterferenceisconveyeddirectlyfromthesource(e.g.lightningstriketo
externalpower cables).Every item of equipment connectedtothis power
sourcewillbeaffectedbytheinterferencepulses.
Alternatively,thepulsesfromthesourceoftheinterferenceoritsassociated
mainscablescanbecoupledintootherequipmentpositionednearby.
Surgepulseinterferencecanalsooccuronsignalanddatalinesthroughcou-
plingeffectsandelectricaldischarges.
Thesystemenablesteststobecarriedoutusingbothcouplingmethods.The
EUTisconnectedtothemainspowersocketonthefrontpanelofthetest
systemfordirectmainsinjectiontests.Externallycoupledtestsrequirethe
interferencetobesuperimposedontosignal/datalinecablesviaanexternal
couplingunitthatisconnectedtothesurgeoutputonthefrontpanelofthe
system.
3.3 Mains quality test
Themainsqualitytestincludesthesimulationofdipsandinterruptsofthemains
powersupplyinaccordancewithIEC/EN61000-4-11andforDCpowersupplies
inaccordancewithIEC/EN61000-4-29.
Avoltagedipoccurswhenthesupplyvoltagefallsconsiderablybelowthe

14
NSG 3040 EMC test system
nominallevelforarelativelyshorttime,e.g.forafewcycles,whereasaninter-
ruptmeansthatthevoltagefallstozeroforasimilarperiod.
3.4 Magnetic fields with mains frequency
Mainsfrequencymagneticeldtests,orPOWERMtests,involvethesimulation
ofthemagneticeldstypicallygeneratedbythecurrentowinpowersupply
cablesasspeciedinIEC/EN61000-4-8.Suchmagneticeldscanaffectthe
operationofitemsofequipmentthataresensitivetothem.TheNSG3040
performsthistestbycausingaheavycurrenttoowinamagneticeldcoil
suchthatthecurrentandfrequencyproduceaproportionaleldwithinthe
coil parameters.
Themagneticeldcoils,availableasaccessories,areconnectedtothemag-
neticeldoption(MFO)which,inturn,isconnectedtothesystem.
3.5 Pulsed magnetic fields
Testswithpulsedmagneticelds,orPULSEMtests,simulatethetypeofinter-
ferenceproducedbysurgepulsesasaresultoflightningstrikestobuildings
andothermetallicstructuressuchasfreestandingmasts,groundconductors,
groundingnetworks,etc.asspeciedinIEC/EN61000-4-9.Magneticeldsof
thistypecanupsettheoperationofinstallationsthatndthemselveswithin
suchelds.TheNSG3040performsthistestbycausingaheavycurrenttoow
inamagneticeldcoilsuchthattheamplitudeofthepulsecurrentproduces
aproportionaleldwithinthecoilparameters.
Themagneticeldcoils,availableasaccessories,areconnectedtothesurge
pulseoutputsocketviaanINA752pulseshapingnetwork.
3.6 Slow Damped Oscillatory Wave (NSG 3040-SOW only)
Thedampedoscillatorywavephenomenaaredividedintotwoparts.Therst
partisreferredtoastheslowdampedoscillatorywave(SOW)andincludes
oscillationfrequenciesof100kHzand1MHz.Thesecondpartisreferredtoas
thefastdampedoscillatorywave(FOW),anditincludesoscillationfrequencies
3MHz,10MHzand30MHz.
Theslowdampedoscillatorywaveimmunitytestsimulatesswitchingofdis-
connectorsinHV/MVopen-airsubstations,andisparticularlyrelatedtothe

15switchingofHVbusbars,aswellastothebackgrounddisturbanceinindustrial
plantsasspeciedinIEC/EN61000-4-18andANSIC37.90.1.
TheNSG3040-SOWisnotonlyabletoperformtheslowdampedoscillatory
waveimmunitytestaccordingtobothofthesestandards,butincludesawide
rangeofover-testingcapabilitieslikehigherpulsevoltage,higherpulserepetition
ratesandaselectablesourceimpedancefortestsclosertorealityconditions.

16
NSG 3040 EMC test system
TheNSG3040systemanditsaccessoriesoperateathighvoltages.
4 SAFETY INSTRUCTIONS
WARNING - Improper or careless operation can be fatal!
Theseoperatinginstructionsformanessentialpartoftheequipmentandmust
beavailabletotheoperatoratalltimes.Theusermustobeyallsafetyinstruc-
tions and warnings.
NeitherTeseqAG,Luterbach,Switzerland,noranyofitssubsidiarysalesorga-
nizationscanacceptanyliabilityforpersonal,materialorconsequentialinjury,
lossordamagethatmayresultfromimproperuseofequipmentandacces-
sories.
4.1 General
TheNSG3040mustbeoperatedonlybyauthorizedandtrainedspecialists.
Thegeneratoristobeusedonlyforthepurposespeciedbythemanufacturer.
Theuserisdirectlyresponsibleforensuringthatthetestsetupdoesnotcause
excessiveradiatedinterferencewhichcouldaffectotherinstrumentation.The
testsystemitselfdoesnotproduceanyexcessiveEMradiation.However,the
injectionofinterferencepulsesintoaEUTcanresultinitand/oritsassociated
cablesradiatingelectromagneticradiation.Toavoidunwantedradiation,the
standardsorganizationsrecommendthatthetestsetupbeoperatedinsidea
Faraday cage.

17WARNING - NSG 3040 is not suitable for use in an explo-
sive atmosphere.
WARNING - Personnel tted with a heart pacemaker
must neither operate the instrument nor approach the
test setup while a test is being executed.
Onlyapprovedaccessories,connectors,adapters,etc.aretobeusedtoensure
safeoperation.
WARNING - Connect the EUT only after the initial system
self test has nished.
4.2 Installation
TheNSG3040testsystemconformstoprotectionclass1.Localinstallation
regulationsmustberespectedtoensurethesafeowofleakagecurrents.
WARNING - Operation without a ground connection is
forbidden!
Twoindependentgroundconnectionsarenecessary-oneforthetestsystem
andonefortheEUT.Thesemustbeconnectedbacktothelocalpermanent
installationortoaxed,permanentgroundconductor.
Operatetheequipmentonlyindrysurroundings.Anycondensationthatoccurs
mustbeallowedtoevaporatebeforeputtingtheequipmentintooperation.Do
notexceedthepermissibleambienttemperatureorhumiditylevels.Useonly
ofciallyapprovedconnectorsandaccessoryitems.
Ensure that a reliable return path for the interference current is provided
betweentheEUTandthegenerator.Thegroundreferenceplaneandtheground
connectionstotheinstruments,asdescribedintherelevantteststandards,
servethispurposewell.

18
NSG 3040 EMC test system
Thetestsystemmayonlybeopenedbyaqualiedspecialistuponspecic
instructiongivenbythemanufacturer.Sincetheinstrumentworks,onprin-
ciple,withtwoindependentpowersupplies(oneforthegeneratorandonefor
theEUT),theNSG3040mustbedisconnectedfrombothsourcesbeforeany
modicationstothetestsetupareundertaken.Besidesthemainsconnections
themselves,certaincomponentsalsooperateathighvoltages,andarenot
providedwithanyformofextraprotectionagainstaccidentalcontact.
4.3 Installation of an EUT power switch
TheEUTinputshouldbeconnectedthroughaproperlyratedpowerswitch
device,whichshouldbelocatedclosetothetestsetup.Inordertoensureeasy
andquickaccesstotheEUTpower,theswitchshouldbeclearlyandvisibly
labeledas“EUTpowerON/OFF”.
Thein-housepowerdistributionmustbeequippedwithapropercircuitbreaker
andanemergencyoffbuttonasperIEC61010-1:2001.
The test setup should only be accessible to trained per-
sonnel.
DimensioningofthemainssupplyandratingoffuseprotectionoftheACorDC
powersupplymustconformwithlocalelectricalcodesandEUTrequirements.
Inappropriatearrangement,mounting,cablingorhandlingoftheEUTorground
canhamperornegatetheeffectivenessoftheNSG3040’ssafetyfeatures.
4.4 Applicable safety standards
DevelopmentandmanufactureisincompliancewithISO9001.
ThesystemcomplieswiththesafetyrequirementsofIEC/EN61010-1(Safety
requirementsforelectricalequipmentformeasurement,controlandlaboratory
use).
Itistheuser’sresponsibilitytoensurethatthetestrigdoesnotemitexcessive
electromagneticinterference(EMI)thatmightaffectotherequipment.Thetest

19
WARNING - The test area must be organized so that
unauthorized persons do not have access during the
execution of a test. If a safety contact (Interlock) is used
as a means of access control to the test zone (e.g. a
Faraday cage), then an additional contact connected in
series is necessary to provide protection for parts of
the EUT that are likely to be touched accidentally.
During a test, the EUT together with its accessories and
cables are to be considered live at all times. The test
system must be stopped and the EUT supply disconnected
before any work can be carried out on the EUT. This can
be achieved simply by opening the interlock circuit.
The EUT is to be tested only in a protective cage or
under a hood which provides protection against electric
shock and all manner of other dangers pertaining to the
particular EUT (see: User warnings - Generator).
The user must observe safety instruction for all the
instruments and associated equipment involved in the
test setup.
Test setup conguration is to be strictly in compliance
with the methods described in the relevant standard to
ensure that the test is executed in a compliant manner.
systemitselfdoesnotproduceanyexcessiveradiation;however,theinjection
ofinterferencepulsesintotheEUTcanresultinthedeviceand/oritsassociated
cablesradiatingEMI.Toavoidradiatingunwantedinterferencethestandards
organzationsrecommendthatthetestsetupbelocatedinaFaradaycage.
Sincethepurposeofthetestsystemistoproduceinterferencesignalsfor
interferenceimmunitytesting,therequirementsintheIEC/EN61000series
concerninglimitingtheradiatedEMIcanonlybecompliedwithbyoperating
thetestsysteminsideaFaradaycage.
4.5 Test execution

20
NSG 3040 EMC test system
WARNING - Users must be aware of the following
dangers that can occur during testing:
Local burning, arcing, ignition of explosive gases.
EUT supply current surge caused by a ashover or
breakdown resulting from the superimposed high
voltage.
Disturbance of other, unrelated electronics, tele-
communications, navigational systems and hear
pacemakers through unnoticed radiation of high
frequency energy.
In the test system the interference voltage, corre-
sponding to the level called for in the relevant test
specication, is superimposed also on the EUT’s
protective earth conductor. Earth contacts or pins
(e.g. as in German and French mains plugs) as well as
the EUT earth itself can therefore be at an elevated
voltage level that would make touching dangerous. In
many power connectors even the screws are linked
to the protective earth.
4.6 User warnings - Generator
CAUTION! Warning of a danger spot
(refer to the documentation).
Warningsymbolsonthetestsystem:
Caution! Warning of electrial hazards!
Table of contents
Popular Inverter manuals by other brands

Sunforce
Sunforce 60 WATT SOLAR 12 VOLT POWER GENERATOR KIT user manual

INVT
INVT XG3-10KTL user manual

Axpert
Axpert 3KVA-110V Service manual

Huawei
Huawei SUN2000-33KTL-JP user manual

Champion Global Power Equipment
Champion Global Power Equipment 100515 Operator's manual

SOLINTEG
SOLINTEG M Series user manual