Antenna installation:
Supplied antenna is screwed on antenna connector. When installing antenna, hold antenna by knurled
end.
Power supply:
According to operation method the radon probe can be supplied:
1) By internal accumulator for portable use –Radon probe includes internal accumulator which is
able to ensure autonomous operation of probe for more than 12 months without charging.
Depends on frequency of wireless data downloading from probe and depends on climatic
condition of probe use. Accumulator is charged with USB port and provided USB cable. The USB
cable is possible to connect to PC or to delivered power adapter. Status of accumulator and
charging process are described in paragraph ´Basic Maintenance/Accumulator charging´
2) By mains power supply 230V/50Hz for stationary use –Radon probe is permanently supplied by
delivered power adapter. Power adapter is connected to probe via provided USB cable. In case of
blackout internal accumulator ensures UPS function.
Configuration:
Setting and configuration is different according to operation way of radon probe.
If the probe is used for autonomous measurement and data downloading via USB interface then the
setting and configuration are realized by usbRADONview application. UsbRADONview application, drivers
and user manual with detail configuration description are free downloaded on website of producer .
If the probe is used for wireless measurement and wireless data downloading then setting and
configuration of probe and whole measuring system is realized by Central Unit connected to PC and
TERAview application. Central Unit is not included with package of probe and it is sold and delivered extra.
TERAview application, drivers and user manual with detail configuration description are free downloaded
on website of producer. For successful probe configuration in measuring system is essential to know probe
radio channel number (communication wireless channel) and P2P address (identification in wireless net).
Both parameters are printed out on probe serial number plate. Probe radio channel number is possible to
change by TERAview application and it must be identical to central unit radio channel number. P2P address
is permanent and it can occur in one big wireless net only once. Central Unit P2P address can be identical to
P2P address of other elements in network.
Installation
The probe has hanging holder for installation into hole or in places where is risk of water flowage.
Generally probe is possible to place in random position but it is necessary not to cover of botom of probe
and bottom of probe prevent from drops of water. Therefore vertical position is the best.