
Chapter 3 – Operation
Page 3-4 TestEquity 115R Temperature Chamber
Performance Considerations
The performance of all chambers is significantly affected by the characteristics of your test
sample. Factors include size, weight, material, shape, and power dissipation if energized.
The test sample should be placed in the chamber in a manner that allows for air circulation. The
air plenum is located on the top ceiling of the chamber, where air is sucked in from the left side
top and exits from the fan on the right-side top. You should not place the test sample directly on
the chamber floor. It should be placed on the shelf. Multiple test samples should be distributed
throughout the chamber to ensure even airflow and minimize temperature gradients. If necessary,
additional shelves should be used to evenly distribute the load. Verify that the temperature
gradients are within acceptable limits, by measuring the chamber temperature at strategic points
using a multipoint thermocouple meter or data logger.
You may find that the temperature throughout the chamber is even, but always different from
what the temperature controller indicates. The correct way to adjust what the temperature
controller “displays” compared to what is measured at some point other than the controller’s
sensor is with the “Calibration Offset” parameter, NOT by recalibrating the controller.
Avoiding Moisture
Any time the ambient air is subjected to temperatures below the dew point, moisture will
condense out of the air. The effect is ice or frost during low temperature operation, or water
when maintaining over 0°C and cooling is required.
To avoid moisture condensation, make sure the port plugs are inserted at all times. Also, avoid
opening the chamber door while the chamber is operating at temperatures below room ambient.
When a low temperature test is completed, warm the chamber to at least room ambient before
opening the chamber door and before removing your test sample.
Internal Test Fixtures
Some applications require internal fixtures to support test samples and provide a convenient
method of connecting wires and sensors. Fixtures must be designed to minimize their impact on
chamber functionality and performance.
Fixtures should be designed for easy removal to permit maintenance and cleaning of the
chamber. The chamber liner should never be drilled or screwed into. This will compromise the
integrity of the liner and permit moisture migration due to condensation into the insulation,
which will eventually impact performance and lead to premature rusting of the outer cabinet.
Fixtures should be constructed from stainless steel or aluminum. All screws and fasteners should
be made from stainless steel. If stainless steel is used, all welds should be passivated. To prevent
rust and corrosion, never use iron or mild steel even if it is painted or plated. Aluminum has
almost twice the specific heat of stainless steel but weighs less than one-half as much per unit
volume. Careful consideration in fixture design balances the test fixture’s mass times its
material specific heat to arrive at the lowest effective total.
Make sure that all connectors, wiring, pc boards, and auxiliary components can withstand the
temperature extremes that they will be subjected to. In some cases, these components may not be
able to last after repeated tests and should be considered expendable.