TR 900 User manual

Hygiene Equipment
Designed for People
1
Registered at Medical Products Agency,
Sweden as an MDR Class 1 product
according to MDR 2017/745.
TR reserves the right to make design
changes and/or improvements without
notice.
Always study this manual carefully before you
attempt to operate your bathing system. The
operator/user must always have access to the
operating manual.
This bathing system is intended for professional use
only where the patient will be under supervision of
trained personnel.
Operating instructions and recommended
maintenance procedures are explained in the
following pages.
If you have any questions regarding the operation or
maintenance of your bathing system and cannot find
the answer in this manual, please contact us.
The system is designed for years of trouble-free
operation, with proper use and maintenance.
Bathtub and shower water must be run at least every
three days.
Operating manual
Version 202
1
-
04
-
26
TR
900
Height adjustable Bathtub
Warning
Mandatory to read the Operating manual before using the equipment.

Hygiene Equipment
Designed for People
2
Table of contents
Manufacturer
Page
4
Introduction
Page
5
Unpacking
Page
6
Installation
Page
7
Initialization/Calibration
Page 1
2
-
Cleaning injector/mixture
Page 1
2
-
Thermostatic mixer
Page 1
3
Intended use
Page 1
4
Safety Instructions
Page 1
5
-
General safety instructions
Page 1
5
-
In case of an accident
Page 1
7
Features and controls
Page 1
8
-
TR 900 /
-
SS,
-
C,
-
CW
Page 1
8
-
TR 900A
(Autofill) /
-
SS,
-
C,
-
CW
Page 1
9
Product description
Page
20
-
TR 900 /
-
SS,
-
C,
-
CW
Page
20
-
TR 900A (Autofill) /
-
SS,
-
C,
-
CW
Page 2
1
-
Possible combinations
Page 2
2
Technical specification
Page 2
3
User instructions TR 900 /
-
SS,
-
C,
-
CW
Page 2
4
-
How to
fill the tub with water
Page 2
4
-
Showering
Page
2
5
-
Height adjustment
Page 2
6
-
Whirlpool
Page 2
7
-
Cleaning
Page 3
2
User instructions TR 900A (Autofill) /
-
SS,
-
C,
-
CW
Page 2
8
-
How to fill the tub with water
Page 2
8
-
Showering
Page 2
9
-
Height
adjustment
Page
30
-
Whirlpool
Page 3
1
-
Cleaning
Page 3
2
-
Patient handling
–
Ambulatory patients
Page 3
3
-
Patient handling
–
Recumbent patients
Page 3
4
System components
Page 3
5
-
Lifting columns
Page 3
5
-
Thermostatic mixer
Page 3
5
Labels and
symbols
Page 3
6
Routine maintenance
Page 3
8
-
General information / Warranty
Page 38
-
Important information regarding the Thermostatic mixer
Page 39
-
Before every use
Page
40
-
Daily maintenance
Page
40
-
Monthly maintenance
Page
4
1
-
Annual
maintenance
Page 4
2
-
Maintenance every 3 years
Page 4
4
-
Maintenance every 5 years
Page 4
5
-
Maintenance after 10 years
Page 4
6

Hygiene Equipment
Designed for People
3
Table of contents
Cleaning instructions
Page 4
7
-
General cleaning instructions
Page 4
7
-
Cleaning the Whirlpool system
(Option
-
CW, CWA)
Page 4
8
Disposal / Recycling
Page
50
Maintenance record
Page
5
1
Training record
Page 5
2

Hygiene Equipment
Designed for People
4
Manufacturer
TR Equipment AB
Rundelgatan 2
SE-573 35 Tranås
Sweden
www.trequipment.com
Phone: +46 (0)140 38 50 60
Group subsidiaries:
USA
TR Group Inc. Phone: +1 (800) 7526900
www.trequipment.com
information@trequipment.com
United Kingdom:
TR Equipment Ltd Phone: +44 (0) 844 335 8386
www.trequipment.com
Germany:
TranspoRehamed GmbH Phone: +49 (0) 212-645850-0
www.transporehamed.de

Hygiene Equipment
Designed for People
5
Introduction
Dear Customer,
Thank you for choosing one of our quality products. We hope you enjoy the TR 900 bathtub.
This manual shall, in conjunction with staff training, ensure that the operator/user is always using and/or
operating the TR 900 correctly.
New colleagues must be trained by qualified and experienced staff.
TR Equipment AB requires that each user has read and understands the entire operating manual and that
they can apply the instructions as specified. This applies to anyone who was not present at the in-service.
This instruction manual should generally be kept easily accessible for every user. Copying the whole manual is
encouraged since its sole purpose is to improve the standard of training and safety.
Please order another copy from us if the instruction manual is lost or unusable, or if it has become illegible
over time.
If you need further information, do not hesitate to contact TR Equipment AB or one of our qualified partners.
Warning
Mandatory to read the Operating manual before using the equipment.
Warning
Make sure you have read, understood, and applied the information on the following pages
before using the bathtub.

Hygiene Equipment
Designed for People
6
Unpacking
When you receive a TR 900 bathtub from us it’s placed on a pallet surrounded by a wooden crate wrapped in
plastic. Start unpacking by carefully cut and remove the plastic. DO NOT use any sharp tools that may scratch
the bathtub.
Unscrew the wooden crate from the pallet.
Carefully lift the wooden crate away from the pallet. When lifted, do so briefly and use at least two people.
We advise to use the transport Wheel beam and ramp (Options) to get the bathtub off the pallet and to get the
bathtub in place.
Place the Wheel beam under the chassis. Position the Wheel beam just behind the adjustable feet.
Carefully
roll the tub off the pallet.

Hygiene Equipment
Designed for People
7
Installation instructions
The TR 900 bathtub is delivered fully assembled, ready to be connected to water and power supply at the
customer’s premises and should be installed according to local guidelines and regulations.
When the bathtub is in place adjust the adjustable feet to get the
bathtub chassis in level (Blue line).
The tub must be cleaned before the first use!
Bathroom layout
Recommended room size is 2400mm x 3000mm, or larger.
Plumbing
TR900 is delivered with an installation kit containing 2pcs 500mm connecting hoses R20, 2pcs ball valves
R20, and 2pcs check valves. DO NOT bend the connecting hoses.
If your TR900 is equipped with a cleaning unit, the installation kit contains 1pcs 500mm connecting hose R15
and 1pcs check valve. DO NOT bend the connecting hose.
Installation of the bathtub shall be performed by certified plumber according to national/local regulations.
To eliminate pipe debris, it is essential that supply pipes are thoroughly flushed before the bathtub is being
connected. The drainage hose shall be connected to floor- or wall drainage. To level the bathtub, adjust the
adjustable feet.
Autofill
Set the shower flowrate by adjusting the ball valve inside the panel. The ball valve is located on the pipes of
the shower unit inside the panel.
Installation kit
R20
Hot
-
and Cold Water
Installation kit
R15
Cleaning unit
Warning
The bathtub may only be used after it has been successfully and correctly installed.

Hygiene Equipment
Designed for People
8
Installation instructions
Electrical
The bathtub is designed for 100-120V – 50/60Hz, 10A or 220/240V – 50/60Hz, 10A.
The control box for the lifting actuator is specified for 110-230VAC, +/-10%.
Voltage outside these norms will damage the linear actuator and control box.
The bathtub must be connected to the mains via a cord that is connected to a GFCI, Ground Fault Circuit
Interrupter, including an easily accessible switch near the bathtub (Not supplied by TR Equipment).
Warning
Polarity and grounding must be correct. Improper connection may result in injury.
Warning
The bathtub may only be used after it has been successfully and correctly installed.
All electric conduit, fitting and wiring to bath supplied must be installed by a certified
electrician in accordance with national/local regulations. Main switch, main fuse and Ground
Fault Circuit Interrupter are NOT included at delivery and must be supplied by others.

Hygiene Equipment
Designed for People
9
Installation instructions

Hygiene Equipment
Designed for People
10
Installation instructions

Hygiene Equipment
Designed for People
11
Installation instructions
Floor drainage
Assemble the drainage pipe bracket to
the front of the panel. Cut the plastic
pipe to a suitable length.
Wall drainage
Assemble the drainage pipe bracket to
the back of the panel.

Hygiene Equipment
Designed for People
12
Calibration
Calibrating the detergent mixture of the Cleaning System
Detergent is automatically mixed with the cold water,
in a pre-set mixture, at the factory, of 2 %.
If an alternatively calibration is needed follow the instructions
below to achieve a mixture of 2 % or what is recommended
by your detergent supplier.
To calibrate the mixing setting screw (17) placed on the injector
inside the panel, requires a thin screwdriver, a measuring cup
with a capacity of at least 20 cl, and a bucket with at least 10L
capacity.
- Remove the panel cover.
- Empty the detergent bottle by loosening the hose from the
injector. When empty, connect the hose to the injector again.
- Fill the measuring cup with 20 cl concentrated detergent, open
the detergent bottle lid (18) and pour it into the bottle.
- Place the empty 10L bucket in the tub and hold the cleaning spray
handle (15) with yellow hose over the bucket for filling.
- Open the cleaning safety valve (13) to a fully open position
and spray into the bucket.
- Observe level when bucket fills, while bottle empties. Bucket should
reach 10L mark at the same time as the bottle empties. If bucket
fills first, turn the detergent mixing setting screw (17) on injector
slightly counterclockwise, using the screwdriver. If bottle empties first,
turn the detergent mixing setting screw (17) slightly clockwise.
- Repeat the steps until the injector is calibrated to required mixture.
- Close cleaning safety valve (13) and return the cleaning
handle (15) to the bracket (16) on side of the panel.
- Attach the panel cover.
17

Hygiene Equipment
Designed for People
13
Calibration
Thermostatic mixer
See Product manual Rada 320 for instructions how to calibrate the thermostatic mixer. The manual is
supplied at delivery.

Hygiene Equipment
Designed for People
14
Intended use
- Height adjustable bathtub TR 900 is intended for bathing and showering patients.
- The patient is typically a person who is unable to take care of their own personal hygiene.
- The TR 900 is not intended to be used with patients that exceeds a weight of 150kg/330Ibs.
- The TR 900 is only intended for short term use (<60min) when in contact with breached or
compromised skin.
- The TR 900 is intended to be used indoors in all types of care facilities including hygiene areas.
- The user is any adult person who is capable of reading and understanding the Operating manual and
is qualified to care for dependent persons/patients.
Warning
Mandatory to read the Operating manual before using the equipment.
Warning
Make sure you have read, understood, and applied the information on the following pages
before using the bathtub.

Hygiene Equipment
Designed for People
15
Safety instructions
If you have questions regarding the operation or maintenance of your TR900 and cannot find the answer in
this manual, please contact us.
- The height adjustable, TR 900 bathtub may only be operated by qualified and trained staff.
- New colleagues must be trained by qualified and experienced staff.
- All safety and operating instructions must be read, understood, adhered to and applied when using
the bath.
- The TR 900 may only be installed according to the installation instructions in this Operating manual.
- All electrical conduit on site must be performed by a certified electrician and according to
national/local regulations.
- All plumbing installations must be performed by a certified plumber according to national/local
regulations.
- Even if you have chosen TR Equipment AB or a TR Equipment representative to install the bathtub, all
water supply and electrical wiring is under responsibility of the buyer.
- All repairs and maintenance must be performed by a certified service company or trained staff.
- Repairs shall only be carried out after consulting TR Equipment AB or a TR Equipment representative
and using original spare parts.
- Always physically check the water temperature before and during bathing and showering to ensure
that the thermostatic mixer is working properly.
- The thermostatic mixer is factory pre-set at maximum 43°under ideal installation conditions.
- Make sure that all parts that come in contact with the patient are intact and without sharp edges that
might hurt the patient.
- Check that hoses, tub controls, connections, thermostatic mixer and thermometer are free from
damage.
If the patient can enter the bath themselves but with assistance from care giver:
- Only step into the height adjustable TR 900 bathtub when the bathtub is in its lowest position.
- Only step out of the height adjustable TR900 bathtub when the bathtub is in its lowest position.

Hygiene Equipment
Designed for People
16
Safety instructions
If you have questions regarding the operation or maintenance of your TR900 and cannot find the answer in
this manual, please contact us.
- When using the TR 900 together with our Mobile bath lift TR 9650 the patient must be seated
comfortably on the chair and/or stretcher, with the safety belts securely attached. Safety belts must
be maintained in good condition.
- Raise the stretcher/chair to ensure clearance of all body parts over the tub. This to avoid pinching
injuries between the tub rim and chair/stretcher.
- Before raising or lowering the tub, make sure that the patient hands firmly grasps the stabilizing bar or
that the patient arms are crossed over the chest. This to avoid pinching injuries between tub rim and
chair/stretcher frame.
- Wipe beneath stretcher/chair to prevent water dripping on the floor.
- Always physically check the water temperature before and during bathing and showering to ensure
that the thermostatic mixer is working properly.
- The thermostatic mixer is factory pre-set at maximum 43°under ideal installation conditions.
Warning
Never leave the patient unattended
Warning
Always check the water temperature by hand before using the shower on the patient to
prevent patients from being scalded or shocked by cold/hot water.
Warning
Always disconnect the electricity before attempting any service or repair.
Warning
The TR 900 shall not be used in any way that varies from the safety precautions and operating
procedures described in this manual. The equipment shall not be operated by anyone until the
user has read and understands the Operating manual.

Hygiene Equipment
Designed for People
17
Safety instructions
Note:
In case of a serious incident when using the TR 900, this should be reported to the
manufacturer and to the local authorities where you are established.
TR Equipment AB
Rundelgatan 2
SE-573 35 Tranås
Sweden
www.trequipment.com
Phone: +46 (0)140 38 50 60
Group subsidiaries:
USA
TR Group Inc. Phone: +1 (800) 7526900
www.trequipment.com
information@trequipment.com
United Kingdom:
TR Equipment Ltd Phone: +44 (0) 844 335 8386
www.trequipment.com
info@trequipment.co.uk
Germany:
TranspoRehamed GmbH Phone: +49 (0) 212-645850-0
www.transporehamed.de
info@transporehamed.de

Hygiene Equipment
Designed for People
18
Features and controls
Features and controls for TR900 & TR900SS, -C, CW
1
Control button, height adjustment
UP/DOWN
2
Thermostatic mixer
3
Digital
T
hermometer
C°
Digital
T
hermometer F° (USA)
4
Tub filling valve, open/close
5
Bathtub filling spouts
6
Overflow drainage
6
Drainage control knob open/close
8
Adjustable footrest
9
Shower valve, open/close
10
Patient shower hose, White
11
Patient shower handle
12
Bracket for shower handle
13
Cleaning Safety valve open/close
14
Cleaning hose, Yellow
15
Cleaning handle
16
Bracket for Cleaning handle
17
Setting screw Cleaning injector
(inside the panel)
18
Cleaning detergent
filling
19
Cleaning adaptor for Whirlpool
20
Whirlpool ON/OFF button
21
Whirlpool jet (adjustable)
22
Whirlpool air control
23
Whirlpool inlet safety screen
1
2
3
4
5
6
9
20
8
22
21
23
8
6

Hygiene Equipment
Designed for People
19
Features and controls
Features and controls for TR 900A & TR 900A-SS, -C, CW
1
Control buttons, height
adjustment UP/DOWN
2
Thermostatic mixer
3
Digital
T
hermometer
C°
Digital
T
hermometer F° (USA)
4
Tub filling button ON/OFF
5
Bathtub filling spout
6
Overflow drainage
6
Drainage control knob open/close
8
Adjustable footrest
9
Shower button ON/OFF
10
Patient shower hose, White
11
Patient shower handle
12
Bracket for shower handle
13
Cleaning Safety valve
open/close
14
Cleaning hose, Yellow
15
Cleaning handle
16
Bracket for Cleaning handle
17
Setting screw Cleaning injector
(inside the panel)
18
Cleaning detergent filling
19
Cleaning adaptor for Whirlpool
20
Whirlpool ON/OFF button
21
Whirlpool jet (adjustable)
22
Whirlpool air control
23
Whirlpool inlet safety screen
1
2
3
4
9
5
8
22
21
23
8
6
20
1

Hygiene Equipment
Designed for People
20
Product description
TR 900 & TR 900SS, -C, CW
An height adjustable bathtub as standard delivered with a fiberglass tub, foot support, manual tub filling
valve, a manual shower valve, control button for height adjustment, thermostatic mixer, digital thermometer,
filling spouts, overflow drainage, shower hose, shower handle, drainage, drainage hose, adjustable feet and
lifting actuator.
TR900 SS is delivered with a stainless-steel tub.
TR 900 & TR900SS are available with following options:
Cleaning unit (-C)
Cleaning and Whirlpool (-CW)
Local treatment
Emergency lowering
UK-TMV3
UK-Risk 5 Gantry
All options are factory mounted
Accessories:
PVC Pillow
Transport wheel beam
This manual suits for next models
3
Table of contents
Other TR Bathtub manuals
Popular Bathtub manuals by other brands

Whirlpool
Whirlpool Ariel ARL-702 installation manual

Kohler
Kohler K-1151 Homeowner's guide

American Standard
American Standard 3052OD.X0X Installation instructions and owner's manual

Kohler
Kohler K-1111-H2 Roughing-In Guide

Kohler
Kohler K-1158 Roughing-In Guide

Kohler
Kohler Hourglass K-1209-LA Roughing-In Guide