USER' S MANUAL
ED-400
COATING THICKNESS GAUGE
FOR NON-FERROUS METALS
TX TESTING INSTRUMENTS CO., LTD.

READ THE USER ’ S MANUAL BEFORE USING THE GAUGE.
Contents
1. APPLICATION
2. TECHNICAL PARAMETERS
3. STANDARD PACKAGE
4. OPERATION
5. CALIBRATION
6. ADDITIONAL FUNCTIONS-
7. NOTICE
8. ERROR FACTORS
9. TROUBLESHOOTING
10. DAILY MAINTENANCE•••••

ED400 COATING THICKNESS GAUGEFOR NON-FERROUS METALS
1. APPLICATION
The ED-400 coating thickness gauge is suitable for
measuring the thickness of the non-conductive coating
on non-ferrous metal. The typical application is to
measure the thickness of the anodic oxide coating of the
aluminium alloy. It can also be used to measure the
thickness of plastic foil and paper.
The ED-400 coating thickness gauge conforms to
international standard ISO 2360-2003.
2. TECHNICAL PARAMETERS
Measuring range:
Accuracy:
Resolution :
Operating temperature:
Power supply:
Power consumption:
Dimonsio 门 s:
0-500 m m ;
0-50 m
m:
±1|jm;
50-500mm:±2%;
0-50 |J
m:
0.1|Jm;
50-500mm:1mm;
0-500mm:1mm;(Optional)
5-45°C;
9Vbattery
80mw
150mmx80mmx30mm
Weight: 260g
ED400COATINGTHICKNESSGAUGEFORNON-FERROUSMETALS
3.STANDARDPACKAGE
• Gauge
• Probe
• Plastic cover
• Standard substrate (6063 aluminum alloy)
• Calibration foil with inspection report
• User,smanual
• Certificate of conformity
• Certificate of warranty
• Carrying case
Spare Parts
• Spare probe
• Standard substrate (6063 aluminum alloy)
• Calibration foil with inspection report
4. OPERATION
Press the "ON/OFF" button and wait until "0" or
"0.0" is shown on the display.
4.1 Measurement
Hold the plastic part of the probe and place it vertically
onto the clean and dry workpiece surface. Hold it steadily
and wait for the measurement to appear in the LCD and a
beep is heard (unless the beeping is disabled). Lift the
probe for at least 10 mm from the workpiece and for at
least two seconds.

ED400 COATING THICKNESS GAUGEFOR NON-FERROUS METALS
4.2 Statistics
Press the “ STAT" button after serveral
measurements, and the following data will show in turn:
MEAN - average value
Max 一 maximum value
MIN - minimum value
S - Standard deviation
N - Number of measurements
4.3 Delete
During measurement, if an obvious error value
appears, it can be deleted by pressing the “ DEL"
button.
5. CALIBRATION
The gauge should be calibrated under the following
circumstances: it has not been used or calibrated for a
long period of time; it is found obviously inaccurate; it is
reseted; or the probe is changed.
In calibration mode, each measurement should be
made in the same area. If an erroneous measurement
result appears, it can be deleted by pressing the "DEL"
button. Press once to delete the last result, and press
twice to delete all measuring results in the current
calibration sequence.
5.1 Single point calibration
ED400 COATING THICKNESS GAUGEFOR NON-FERROUS METALS
The standard substrate or an uncoated workpiece is
used in single point calibration, and the existing zero point
will be adjusted.
5.1.1 Press the "CAL" button, then the gauge enters
calibration mode. The LCD will display "ZERO" and
"0.0".
512 Measure on the uncoated substrate for at least 10
times. The LCD will display the average measuring
result.
5.1.3 Press the "CAL" button twice to complete the
calibration. The LCD will display “ ” and then
“0” or “0.0”,then the gauge is back to measuring
Mode.
5.2 Dual point calibration
The calibration foil and the standard substrate or an
uncoated workpiece are used in single point calibration,
and the full measuring range will be adjusted.
5.2.1 Press the "CAL" button, then the gauge enters
calibration mode. The LCD will display "ZERO"
and “0.0” .
5.2.2Measure on the uncoated substrate for at least 10
times. The LCD will display the average measuring
result.
5.2.3Press the "CAL" button, then the instrument will
come to the next calibration stage. The LCD will display
"STD1" and "0.0".

ED400 COATING THICKNESS GAUGEFOR NON-FERROUS METALS
5.2.4Put the foil onto the uncoated substrate then
measure it above ten times. The LCD will display
the average measuring result.
5.2.5 Press the “ •“ or “•” button to adjust thedisplay
value from 50.0 to the foil thickness marked.
5.2.6 Press the "CAL" button to finish the calibration.
After “一一”,LCD will display “0” or "0.0" . Then
you can make the measurement.
Press the "CAL" button to complete the calibration.
The LCD will display “ ” and then "0” or
"0.0",then the gauge is back to measuring mode.
6. ADDITIONAL FUNCTIONS
6.1 Button combinations
By pressing two buttons at the same time, the
additional functions shown in Table 1 can be used. The
LCD will display “一一一” and then “0” or “0.0”,then
the corresponding function is set.
Table 1
Combination button Function description
DEL+STAT Reset (see Section 6.2)
STAT+ "A" Activate beeping
STAT+ “ •“ Inactivate beeping
CAL+ "•” Show decimal (0-50 m m)
CAL+ "•“ Show integer (0-500 m m)
ED400 COATING THICKNESS GAUGEFOR NON-FERROUS METALS
6.2 Reset
The reset operation will reload the factory settings
and calibration data. It can be completed by pressing the
"DEL" and "STAT" buttons at the same time. The
LCD will display “一一一”,then “0” or “0.0” . After the
reset operation, a dual point calibration must be
completed. 6.3 Low Battery
When the battery is low, the LCD will display
"LOBAT" . When the gauge run out of battery, it will
shut down automatically.
6.4 Power saving
If no measurement is detected within one minute, the
gauge will turn off automatically.
7. NOTICE
7.1 Before using the gauge, it should be checked for
accuracy. Normally it can be checked by measuring on
the uncoated substrate. For further assurance, measure
with the calibration foil and check if the measuring result
matches the marked thickness.
7.2 When measurements are repeated on the same
workpiece, the results may vary. Take the average value
for better accuracy.

ED400 COATING THICKNESS GAUGEFOR NON-FERROUS METALS
8. ERROR FACTORS
According to ISO 2360-2003, the following factors
will influence the accuracy of measurement.
8.1 Thickness of coating
If the coating thickness is less than 5 m m,theaverage
valueshouldbetaken.Ifthecoatingthicknessisless
than3mm,itcannotbemeasuredaccuratelywiththis
gauge.
8.2Edgeeffect
Whenthemeasuringarealessthan150mm2orthe
widthislessthan12mm,thegaugeshouldbere-
calibratedonthesameuncoatedworkpiece.8.3
Curvature
Ifthediameteroftheworkpieceislessthan50mm,
thegaugeshouldbere-calibratedonthesameuncoated
workpiece.
8.4Surfaceroughness
Measureatdifferentspotsoftheworkpiecetoreduce
theinfluenceoftheroughsurface.
8.5Conductivityoftheworkpiece
8.6Thicknessoftheworkpiece
8.7Gapbetweenprobeandworkpiecesurface
8.8Pressingforceontheprobe
8.9Perpendicularityofprobe
8.10Temperatureofprobe
ED400COATINGTHICKNESSGAUGEFORNON-FERROUSMETALS
9.TROUBLESHOOTING
9.1Failures,reasonsandsolutions
Table2
Failure Reason Solution
PowersavingPressON/OFFbutton
Nodisplayor
Nobatteryorincorrect
batteryinstallationReinstallthebattery
immediateshutdown
EmptybatteryChangebattery
Brokenbattery
connector
Contactmanufacturer
NoresponseonLCD
ThicknessoverrangeCheckworkpiece
tomeasurement
ProbefailureChangeprobe
AllcharactersProbeorcircuitChangeprobeor
displayed boardfailureContactmanufacturer
CurvingsurfaceorCalibrateonuncoated
smallmearingareasameworkpiece
CalibrationwrongRe-Calibrate
Obviouslywrong
Incorrectprobe
handling Press"DEL"button
measuringresultCalibrationdata
disorder
Resetandcalibrate
Probefailure
Changeprobe
Probechanged
withoutcalibration
Re-Calibrate
UnabletoturnoffCircuitproblemContactmanufacturer

ED400 COATING THICKNESS GAUGEFOR NON-FERROUS METALS
9.2 Error codes
Table 3
Code Error Solution
E4 Probe failure,connection problem
or circuit board failure
Connect probe again or
Go门tact manufacturer
E5
Probe failure
Change probe
E6 Measuring range exceeded
Re-Calibrate
E7
Faulty calibration Re-Calibrate
10. Daily maintenance
10.1 The probe must be kept dry and clean. If the
instrument was used on the assembly line, the workpiece
should be wiped dry so that no water or other fluid runs
into the probe. Do not add lubricant to the probe.
10.2 When not in use, the gauge should be put into the
case to avoid impact and dirt.
10.3 Remove the battery if the gauge will be stored for a
long period of time.