ED400 COATING THICKNESS GAUGEFOR NON-FERROUS METALS
5.2.4Put the foil onto the uncoated substrate then
measure it above ten times. The LCD will display
the average measuring result.
5.2.5 Press the “ •“ or “•” button to adjust thedisplay
value from 50.0 to the foil thickness marked.
5.2.6 Press the "CAL" button to finish the calibration.
After “一一”,LCD will display “0” or "0.0" . Then
you can make the measurement.
Press the "CAL" button to complete the calibration.
The LCD will display “ ” and then "0” or
"0.0",then the gauge is back to measuring mode.
6. ADDITIONAL FUNCTIONS
6.1 Button combinations
By pressing two buttons at the same time, the
additional functions shown in Table 1 can be used. The
LCD will display “一一一” and then “0” or “0.0”,then
the corresponding function is set.
Table 1
Combination button Function description
DEL+STAT Reset (see Section 6.2)
STAT+ "A" Activate beeping
STAT+ “ •“ Inactivate beeping
CAL+ "•” Show decimal (0-50 m m)
CAL+ "•“ Show integer (0-500 m m)
ED400 COATING THICKNESS GAUGEFOR NON-FERROUS METALS
6.2 Reset
The reset operation will reload the factory settings
and calibration data. It can be completed by pressing the
"DEL" and "STAT" buttons at the same time. The
LCD will display “一一一”,then “0” or “0.0” . After the
reset operation, a dual point calibration must be
completed. 6.3 Low Battery
When the battery is low, the LCD will display
"LOBAT" . When the gauge run out of battery, it will
shut down automatically.
6.4 Power saving
If no measurement is detected within one minute, the
gauge will turn off automatically.
7. NOTICE
7.1 Before using the gauge, it should be checked for
accuracy. Normally it can be checked by measuring on
the uncoated substrate. For further assurance, measure
with the calibration foil and check if the measuring result
matches the marked thickness.
7.2 When measurements are repeated on the same
workpiece, the results may vary. Take the average value
for better accuracy.