UNDATEC FDL400 User manual

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
FDL400
Flow Data Logger

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
Contents
1. Declaration of Conformity
2. Safety instructions, intended use
3. Description
4. FDL400
4.1 Specifications
4.2 Dimensions
4.3 Pin Assignment
4.4 Connection Diagram
5. Programming
6. USB - port and SD – card
6.1 Reading data from the SD - card and processing in a spreadsheet program
7. Mounting instructions
8. Type Examination Certificate (ATEX)
Errors, mistranslation and technical improvement may change these data !

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
1. EC - Declaration of Conformity
Name of Manufacturer: UNDATEC Flow Measurement Solutions
Wiesenstraße 23
75248 Oelbronn-Duerrn - Germany
Product Name: FDL400 Flow Data Logger
Product Type: Flow measurement for water and wastewater
The designated product is in conformity with the following European Council Directives
for harmonization of the laws of Member States.
The conformity of the designated product with the provisions of the Directive
evidenced by full compliance with the following harmonized European standards.
2014/30 / EU EMC - Directive
DIN EN 61000-6-2: 2006 + Corrigendum 1: 2011-06
DIN EN 61000-6-4: 2011-09
2014/34 / EU ATEX - Directive II 2G Ex ib IIB T4
EN 60079-0: 2012 + A11: 2013
EN 60079-11: 2012
With the application of CE - marking we confirm the successful EMC – test
Report No.: 15024C20
Affixing CE marking: 2014
Location: Oelbronn-Duerrn Date: 29/01/2014 Signature / Technical Manager:
Frank Stark
This declaration certifies compliance with the indicated directives but contains no assurance of properties
in the sense of the product liability law.
The safety instructions of the supplied product documentation must be observed.

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
2. Safety instructions
Safe operation is only ensured if the notes and warnings will be noted in this manual.
Important! Seal, seals and labels
Opening or removing seals or labels, eg etc. with serial numbers, will result in immediate loss
of warranty claims result.
Determination and proper use
The manufacturer is not liable for damages resulting from improper use.
Compare and information before operating the conformity of the supply voltage match the
specifications on the nameplate.
If it becomes that a safe operation is no longer possible (eg visible damage), put the appliance
out of service immediately and secure it against unintentional starting.
Improper or unauthorized use, dangers may arise from the device, which is why we refer to
consistent observance of safety notes.
Mounting, commissioning and operating personnel
Mounting, electrical installation, commissioning and maintenance of the device must be carried out only
by trained specialist personnel authorized by the plant operator.
Technical personnel must have read and understood this manual, and follow their statements.
The device may only be operated by persons who are authorized and trained by the facility operator. The
instructions in this manual must be followed.
Make sure that the device is in accordance with the electrical terminals connected correctly.
Repairs
Repairs should only be performed by trained service personnel.
Please contact in this case to the company Dömges.Detectronic UG.
Technical progress
The manufacturer reserves the right to modify technical data without special announcements on the basis
of development of technical progress. About the activities and possible extensions of this manual can be
supplied by the company Dömges Detectronic UG.

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
3. Description
This manual is valid for the following device version
- Hardware from 2.0 and after
- Firmware from 2.1 and after
The Flow Data Logger FDL400 suitable for flow measurement in partially filled channels or
pipes.
The evaluation is performed according to the flow formula Q = VA.
A combined measuring sensor, mounted on the bottom or the wall of a channel detects the flow speed V
in m / s according to the Doppler principle. About a sound beam that shines against the flow direction in
the medium, the speed data is collected and further processed in the processor.
The integrated pressure sensor is used to detect the level in m;
above this value each perfused segment surface is again in the processor through a pre-programmed
channel cross section calculated in m2.
With the two multiplicands will calculated the sum Q in l / s.
The programming of all necessary parameters can made via 16 buttons on the front-panel keyboard.
The clear 4 lines display the individual values appear
- Velocity
- Level
- Flow
- Totalizer
The measuring sensor is powered from the FDL400 device with the necessary voltage. The two values for
flow rate and liquid level are directly on the sensor 4-20 mA signal is available. This allows the device to
be mounted further away from the location on your channel. Without any further installation work is
usually sufficient to cable extension of the probe.

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
4. FDL400 – Flow Data Logger
4.1 Technical Data
The transmitter are housed in an aluminum case and is suitable for wall mounting or DIN rail mounting.
The device with connected sensor required after turning a certain time to reach operating
temperature (warm-up time). This is about 10-15 minutes.
Power supply : 24V DC +/-20%
current consumption : 120mA – 220mA (depending of outputs - and sensor currents).
Fuse : The device is protected against reverse polarity.
To hedge serve soldered fuses.
(A blown fuse must not be repaired by the customer, since
this is special parts for ATEX-approved devices.)
Protection class : IP30
Dimension : 197 x 153 x 37,5mm (B x H x T)
Temperatur range : 0 – 50°C
Display : LCD 4 lines x 20 character for
- Velocity
- Level
- Flow
- Totalizer
Outputs : 3 x analog 0(4) – 20 mA (+/- 3%) / Burden 100Ω - 900Ω
freely configurable for flow standards, flow extra amount, velocity and level.
4 x digital , opto-coupled, p-type transistor outputs (PNP) -
12 - 30V / 50mA, short circuit protected.
1 x permanently programmed as an alarm .
3 x freely configurable for flow standards, flow extra amount, impulse,
velocity and the level.
All outputs are individually switchable (on or off).
Note : During the installation, make sure that there is enough space above and below the device for
cable laying and available operation. We recommend a minimum of 100mm.

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
4.2 Dimension

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
4.3 Pin Assignment
Attention !
In ATEX relevant applications never do operate the sensor without the safety
guard via plug !

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
4.4 Connection Diagram
The current outputs are not isolated from the power supply.
The switching outputs are electrically isolated by optocouplers from the device power supply and
therefore require a separate power supply. This can, for example, come from a connected controller (PLC)
or from another source.

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
5. FDL400 Programming
After switching on the device is displayed on the display device identification , hardware version, software
version, and copyright.
Example:
FDL400
H – Version 2.0
S – Version 2.3
copyright UNDATEC
After a few seconds the display will jump to the measurement menu.
The display shows velocity, level, flow and totalizer.
Example:
Velocity 0.00 m/s
Level 0.000 m
Flow 0000.0 l/s
Total 000000 x 1 m3
Function buttons
Button „B” = Break (Interrupting the measuring program for inputting parameters)
Button „E“ = Enter
Button „F“ = Correction
Button „A“ = Abbort
Button „C“ = no function
Button „D“ = special function
Functions (Button „B“ generated entry into the function menu. Here, the desired function will be
selected by entering the corresponding function number)
The functions are designed so that they are self-explanatory based on previous selection.
Function 1: Entering of channel shape and associated parameters. It can be a pre-programmed
channel shape
(Circle, rectangle, trapezoid, triangle or special) are selected.
If special channel shapes are available, they are read in via memory card (SD card).
Function 2: Enter the maximum values for V, H and Q.
(Maximum values refer to the 0/4 - 20mA current outputs. Maximum value is the value at
which for the selected output the current of 20mA has the full scale).
Function 3: Entering the limit values for V, H and Q.
(Limits refer to the switch outputs 1,2 and 3. Switch output 4 is for a different
Application allocates.
Value is the value at which the selected switching output switches).

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
Function 4: Input of V and H - offset. An offset can be positive or negative. Since there is no "-" button
is available,
is a selection of "pos" or "neg" a switchover using the "F" key is provided.
The function under the 4 (V or H - offset) value entered is in the calculation of V or H
taken into account.
Function 5: Input to determine the pulse output (selectable from 1 to 999 l / sec or 1-999 m3 / sec.
and the totalizer (potency 0-3 - means quantity counter
counts 0 = m3, 1 = 10m3, 2 = 100m3, 3 = 1000m3)
Note : The pulse output is limited to a maximum of 1 pulse per second. That is, a plurality of pulses per
second are not possible. The pulse output has to be, if necessary, one or more numbers, or a
potency be selected higher. The pulse length is about 250ms.
Function 6: Assignment of the analog outputs (current outputs). Here is the assignment of the issued
Parameters (V, H, Q) is set to the current outputs.
Function 7: Assignment of the switching outputs. Here, the assignment of the parameter to be output
(V, H, Q) to the Switching outputs set.
Function 8: Selection of the totalizer for the display and switching output. Possible are Totalizer and
totalizer per day. Totalizer means the constantly counter, totalizer per day (daily amount)
means a resettable counter, this accumulated the amount as long until it is reset.
Function 10: Calibration factor for velocity (V) - software-based amplification of the
V - value 0.001-9.999
Function 11: Calibration factor for level (H) - software-based amplification of the
H - value 0.001-9.999
Function 12: Set date and time (This function is important for the correct time - date allocation of
stored data)
If the device is de-energized longer time, is it neccessary to control the time and date via
the function 99 to checked and corrected if necessary. It is important for data storage.
Function 13: Language (german/english)
Function 14: Simulation.
In the simulation, the chosen "Function 1" channel is used as a basis for calculation.
With the input of a certain filling level and a specific velocity can be a corresponding flow
are simulated.
The flow is displayed according to the calculation. The pulse output turns real
according to the calculated flow.
A jumping out from the simulation requires a new selection button "B" then "Function 14"
with followed by selection on simulation "off"!
Function 20: Storage of measurement data - storage rate normal, adjustable 1-999 sec, 1-999 min..
Function 21: Storage of measurement data – storage rate quick, adjustment like storage rate normal.
Function 22: Level trigger for quick storage. Example: rain events can increase the level in channels.
From an adjustable level is to be saved faster. Level trigger is the adjustable fill height
for quick storage.
Function 40: Select the boot loader. Is needed for firmware updates.
Function 41: Setting the display brightness

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
Function 50: Saving the system configuration (system.cfg) on the SD card. Necessary before a
firmware update.
Function 51: Write back of the on SD card stored system configuration (system.cfg). Necessary after a
firmware update.
Function 52: This function allows the existing hardware and software version and thus the revision
level are displayed.
Function 53: This function is used to generate a user-defined password to protect a
unauthorized entry of parameters.
That is, when selecting the parameters input by "B" will prompt you to
Entering a four-digit password (in the delivery of 0000).
In the parameter input is now on the function: 53 the creation of a
customer password possible.
Here is a unique password generated, so this password is in the future
Selection of the parameters input by "B" valid.
Function 66: In this function, the values of the factory generated calibration value of the
analog current and / or voltage outputs can be displayed.
Function 80: Reset of Totalizer (select Totalizer or Totalizer per day).
Function 90: Automatic zeroing V and H (currently at the V and H – input pending value
is called as zero).
Function 91: Automatic zeroing V (currently on the V - input pending value is set as zero).
Function 92: Automatic zeroing H (currently on the H - input pending value is set as zero).
All three functions (90, 91 and 92) set for the selected parameter the currently measured input value as
zero (for display and calculation).
Function 93: Zeroing to 4mA, zero for V and H are set exactly 4mA.
Function 94: Zeroing to 4mA, zero for H are set exactly 4mA.
Function 95: Zeroing to 4mA, zero for V are set exactly 4mA.
All three functions (93, 94 and 95) for the selected parameter set 4 mA as zero (for display and
calculation).
This function can also be used to correct an erroneous zero point adjustment function with 90,91 or 92.
Function 98: Correction of the V - curve (depending on the filling level). Height-dependent V-correction
in 5 steps with entering the height cutoff point and the addition in %.
Function 99: Displaying Date and Time

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
The unit has 4 switching outputs.
The switching output Q4 is switched on so-called security mode. This can be output as alarm conditions
such as power failure or sensor failure. The output is switched on correctly-working device, ie, it is in
tension. In the case of a not powered or not properly operating unit of output is switched off, ie, he is de-
energized (alarm state).
In the case of a disconnected or defective sensor, the output also turns off.
Sinks the analog input under about half of its nominal zero value of 4 mA (about 2mA), so turn off the
output and thus signals an alarm. The message on display shows !! Alarm !! - Cable break ".
Accouterment
1. In the first moment of the keyboard operation, the backlight of the display will automatically turn on.
During device operation, the display is then illuminated, so a pleasant operation is possible. About 10
minutes after the last operation of the keyboard, the display backlight will turn off automatically. This is
for a low power consumption, on the other hand to avoid unnecessary heating devices and thereby also
the increasing of life.
2. Measurement data are stored on an SD card (Secure Digital Memory Card). This is on top of the unit
under an openable flap. The presence and correct operation is indicated by lighting an LED to the side of
SD card. The saving of measuring data is visible by the flashing of this LED.
Measurement will stored on a daily basis. That is, for every day a new log file is created. Each log file has
the name of the log day, eg, 20140908 for the day 2014/09/08.
These are text files that can be opened, edited and archived with any standard spreadsheet software
(Excel, LibreOffice etc.)
3. The device can also be operated without an SD card. However, it can then not be stored measurement
data.
4. The unit has a mini-USB (AB). This is also on top of the device under the opening flap. An Update of
the operating software is with the bootloader on this interface possible.
5. In the device software under "Function 1" four types of channels selectable (circle, rectangle,
trapezoid and V-shape).
In cases other than those pre-programmed channel shapes it is necessary to provide a sketch of the
channel shape with dimensions available.
The customer then receives an SD card on which this channel is programmed.
The customer can then select "Function 1" its specialty channel with the additional choice of whether the
channel should be included in the instrument software or should only be read from the SD card.
After taking over in the device software, the SD card can be removed from the device when no data
needs to be stored.

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
Diagram and explanation of „Function 98“
The "Function 98" is useful for large channels or filling levels.
Since the flow rate distribution is non-linear in a channel, but is similar to a hyperbola, it may be
necessary to calculate a logarithmic curve at high flow levels from low flow levels relatively linear flow
rate curve.
This is necessary because of the velocity measuring sensor is able to measure up to a filling level of
about 400mm safely. An increasing velocity at higher filling levels can not be measured and then
calculated with the "Function 98".
Füllhöhe
(Level)
Einsatzpunkt 1
(Triggerpoint 1)
Fließgeschwindigkeit (Velocity)
Füllhöhe
(Level)
Fließgeschwindigkeit (Velocity)
Tatsächliche Fließgeschwindigkeit
(actual velocity)
Korrigierte Fließgeschwindigkeit
(corrected velocity)
Gemessene Fließgeschwindigkeit
(measured velocity)
Einsatzpunkt 2
(Triggerpoint 2)
Einsatzpunkt 3
(Triggerpoint 3)
usw.
etc.

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
6. Top panel (openable flap) with SD card and USB connection
Figure shows open flap on top of the device with SD - card and USB - connection.
The USB interface is neccessary for update the firmware to be made. This requires a connection to a USB
cable to a PC. The following figure shows the necessary cable.

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
6.1 Reading data from the SD - card and processing in a spreadsheet program
To read the data must be taken the SD - card from the device and plugged into the appropriate card slot
of a PC.
In the period in which the SD - card is not present in the device, no data is stored.
These memory idle time can be circumvented by inserting of another SD - card.
The measurement data are stored according to a preselected time step in a so-called log file (eg
20150320.log = data from 03.20.2015).
For every day a new log file is created automatically. The log file of the day begins at midnight and ends
at midnight course again.
The saved data format is a text file that corresponds to the Excel format and can be practically applied
and processed in any spreadsheet program.
As separator is a space.
Example:
With the most spreadsheet programs, it is possible to design curves, graphs, and statistical analysis.
FDL400 - Logfile
Date Time V(m/s) H(m) Q(l/s) M(m3)
2015.03.15 14:42:13 0.02 m/s 12 mm 0000.0 l/s 30 m3
2015.03.15 14:42:43 1.44 m/s 50 mm 0030.1 l/s 30 m3
2015.03.15 14:43:13 0.57 m/s 67 mm 0018.8 l/s 31 m3
2015.03.15 14:43:43 0.53 m/s 69 mm 0017.5 l/s 32 m3
2015.03.15 14:44:13 0.51 m/s 66 mm 0016.8 l/s 32 m3
2015.03.15 14:44:43 0.49 m/s 65 mm 0015.4 l/s 33 m3
2015.03.15 14:45:13 0.49 m/s 65 mm 0015.4 l/s 33 m3
2015.03.15 14:45:43 0.50 m/s 65 mm 0015.4 l/s 33 m3
2015.03.15 14:46:13 0.52 m/s 64 mm 0016.0 l/s 34 m3
2015.03.15 14:46:43 0.48 m/s 64 mm 0015.1 l/s 34 m3
2015.03.15 14:47:13 0.49 m/s 63 mm 0014.8 l/s 35 m3
2015.03.15 14:47:43 0.50 m/s 64 mm 0015.4 l/s 35 m3
2015.03.15 14:48:13 0.48 m/s 62 mm 0014.4 l/s 36 m3
2015.03.15 14:48:43 0.48 m/s 65 mm 0014.8 l/s 36 m3
2015.03.15 14:49:13 0.46 m/s 63 mm 0013.5 l/s 37 m3
2015.03.15 14:49:43 0.51 m/s 62 mm 0014.9 l/s 37 m3
2015.03.15 14:50:13 0.51 m/s 62 mm 0014.7 l/s 37 m3
2015.03.15 14:50:43 0.50 m/s 62 mm 0014.0 l/s 38 m3

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
7. Mounting instructions
Conversion of the unit of wall mounting on DIN rail mounting
On delivery the unit for wall mounting is provided.
Using the supplied bracket, the unit can be mounted in the control cabinet using a DIN rail.
For this purpose, the 3 pcs. Screws with the rosettes on the back of the unit must be loosened.
With the screws the bracket must be mounted on the rear panel, the rosettes are not necessary for this
(see photo).
Auslieferungszustand
für Wandmontage
(state at the surrender
for wall mounting)
Zustand umgebaut für Montage
auf Normschiene
(state rebuilt for installation
on DIN rail)

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
Mount the unit on a DIN rail
The bracket for mounting on a standard rail has a spring. The insertion of the device onto the DIN rail is
done as shown in the photo. To engage the rail, the device is so compressed the spring down. This can
now be folded down in the DIN rail in the unit.
To disassemble the device, it is also compressed down and allow the spring, then it can be taken out
again.
Feder
(compression
spring)
Einrasten
(lock)
Gerät in
Normschiene
eingerastet
(device in
DIN rail
locked)
Druck
(pressure)
Druckrichtung
zum Ausrasten
(pressure direction
for unlocking)
Richtung
zum Ausrasten
(direction for
unlocking)

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
8. Type Examination Certificate

UNDATEC Flow Measurement Solutions
Frank Stark - Wiesenstrasse 23 - D-75248 Oelbronn-Duerrn / Germany - Telefon +49 (0) 7237-7804
Email: undatec@web.de
Table of contents
Popular Data Logger manuals by other brands

Measurement Computing
Measurement Computing USB-1608GX user guide

Reed Instruments
Reed Instruments R6020 instruction manual

Lascar Electronics
Lascar Electronics EasyLog EL-EnviroPad-TC quick start guide

Lumel
Lumel HT22IoT user manual

Siemens
Siemens SITRANS RD500 Application guide

Hobo
Hobo Pendant Event manual