VELOCICALC 9565-P Operation manual

VELOCICALC®
AIR VELOCITY METER
MODEL 9565 SERIES
OPERATION AND SERVICE MANUAL
P/N 6004851, REVISION F
JUNE 2018
www. .com information@itm.com1.800.561.8187

www. .com information@itm.com1.800.561.8187

i
Copyright
TSI Incorporated / 2011-2018 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
Limitation Of Warranty And Liability (effective February 2015)
Seller warrants the goods, excluding software, sold hereunder, under normal use and
service as described in the operator's manual, to be free from defects in workmanship
and material for 24 months, or if less, the length of time specified in the operator's
manual, from the date of shipment to the customer. This warranty period is inclusive of
any statutory warranty. This limited warranty is subject to the following exclusions and
exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date
of shipment;
b. Carbon monoxide (CO) Electro-chemical sensors are warranted for 12 months from
the date of shipment.
c. Pumps are warranted for hours of operation as set forth in product or operator’s
manuals;
d. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of
shipment;
e. Seller does not provide any warranty on finished goods manufactured by others or on
any fuses, batteries or other consumable materials. Only the original manufacturer's
warranty applies;
f. This warranty does not cover calibration requirements, and seller warrants only that
the instrument or product is properly calibrated at the time of its manufacture.
Instruments returned for calibration are not covered by this warranty;
g. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in the
manual allow an operator to replace consumables or perform recommended
cleaning;
h. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or cleaned
according to the requirements of the manual. Unless specifically authorized in a
separate writing by Seller, Seller makes no warranty with respect to, and shall have no
liability in connection with, goods which are incorporated into other products or
equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS
stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT,
SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND
EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS.
BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE
PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S
OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER
OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
www. .com information@itm.com1.800.561.8187

ii
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS
BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR
OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE
REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE
REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE,
SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO
DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO
EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS
INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR
INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR
INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No
Action, regardless of form, may be brought against Seller more than 12 months after a
cause of action has accrued. The goods returned under warranty to Seller's factory
shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY
AND LIABILITY, which contains the complete and exclusive limited warranty of Seller.
This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or
its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are
to our customers, our service policy is designed to give prompt attention to any
problems. If any malfunction is discovered, please contact your nearest sales office or
representative, or call Customer Service department at (800) 874-2811 (USA) or (1)
651-490-2811 (International).
www. .com information@itm.com1.800.561.8187

iii
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............. 1
Bluetooth® Safety and Compliance................................................. 1
CHAPTER 2 SETTING-UP ................................................................. 5
Supplying Power to the Model 9565 Series .................................... 5
Installing the Batteries ............................................................... 5
DIP Switch Settings ................................................................... 5
Using the AC Adapter ................................................................ 6
Connecting Ventilation or IAQ Probes ............................................ 7
Using the Telescoping Thermoanemometer Probes ...................... 7
Extending the Probe .................................................................. 7
Retracting the Probe .................................................................. 7
Using the Pressure Ports (9565-P) ................................................. 8
Connecting the Static Pressure Probe ...................................... 8
Connecting an Optional Pitot Probe or Airflow (straight
pitot) Probe ............................................................................ 9
Thermocouple Ports ..................................................................... 10
Connecting the Thermocouples .............................................. 10
Connecting the Optional Bluetooth Portable Printer Device
(Models 9565-P and 9565-X only) ............................................. 11
Connecting to a Computer ............................................................ 11
CHAPTER 3 OPERATION ............................................................... 13
Keypad Functions ......................................................................... 13
Common Terms ............................................................................ 14
Menus ........................................................................................... 15
ZERO PRESSURE ............................................................................. 15
DISPLAY SETUP ............................................................................... 16
SETTINGS ....................................................................................... 17
FLOW SET UP .................................................................................. 17
VOC SETUP ..................................................................................... 19
ACTUAL/STANDARD SETUP .............................................................. 20
DATA LOGGING ............................................................................... 21
Measurements ......................................................................... 21
Log Mode/Log Settings ............................................................ 21
Choose Test ............................................................................ 27
Name Test ............................................................................... 27
View Data ................................................................................ 28
View Stats ................................................................................ 28
View Samples .......................................................................... 29
Print Test ................................................................................. 30
Delete Data .............................................................................. 31
% Memory................................................................................ 32
www. .com information@itm.com1.800.561.8187

iv
ZERO CO ......................................................................................... 33
APPLICATIONS ................................................................................. 34
CALIBRATION .................................................................................. 35
BLUETOOTH FUNCTIONS (Models 9565-P and 9565-X only) ..................... 35
Discover Devices ..................................................................... 35
Discoverability .......................................................................... 36
PINcode ................................................................................... 36
# AutoConnects ....................................................................... 36
Printing Data Using the Portable Printer (Models 9565-P and
9565-X only) ............................................................................... 36
TrakPro™ Data Analysis Software ............................................... 37
LogDat2™ Downloading Software ................................................ 37
CHAPTER 4 MAINTENANCE .......................................................... 39
Recalibration ................................................................................. 39
Cases ............................................................................................ 39
Storage .......................................................................................... 39
CHAPTER 5 TROUBLESHOOTING ................................................ 41
APPENDIX A SPECIFICATIONS ..................................................... 43
APPENDIX B OPTIONAL PLUG-IN PROBES ................................ 47
www. .com information@itm.com1.800.561.8187

1
Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping
container. Check the individual parts against the list of components
below. If anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
3. Static pressure tip (9565-P and 9565-P-NB models only)
4. Rubber tubing (9565-P and 9565-P-NB models only)
5. USB cable
6. TrakPro CD-ROM with data analysis software
7. LogDat2 CD-ROM with downloading software
8. AC adapter
This manual supports the following models:
Model
Description
9565-P
Multi-Function Ventilation Meter with Differential
Pressure Sensor
9565-P-NB
Multi-Function Ventilation Meter with Differential
Pressure Sensor, No Bluetooth
9565-X
Multi-Function Ventilation Meter, No Differential
Pressure Sensor
9565-X-NB
Multi-Function Ventilation Meter, No Differential
Pressure Sensor, No Bluetooth
Bluetooth® Safety and Compliance
This product uses Classic Bluetooth® v2.0 to communicate with
software applications.
Hereby, TSI Inc. declares that this Classic Bluetooth® test and
measurement instrument is in compliance with Directive 2014/53/EC.
Transmit Power Rating is +4 dBm, MAX @ 2.4 – 2.5 GHz.
www. .com information@itm.com1.800.561.8187

Chapter 1
2
NO TE
This device may not cause interference; this device must accept
any interference, including interference that may cause undesired
operation of the device.
l’appareil ne doit pas produire de brouillage; l’appareil doit accepter
tout brouillage radioélectrique subi, même si le brouillage est
susceptible d’en compromettre le fonctionnement.
NO TE
This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules
(FCC ID: PI4411B). These limits are designed to provide
reasonable protection against harmful interference in a residential
installation. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference
will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which
can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the
following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from
that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for
help.
NO TE
This device complies with Industry Canada licence-exempt RSS
standard(s). Operation is subject to the following two conditions: (1)
this device may not cause interference, and (2) this device must
accept any interference, including interference that may cause
undesired operation of the device.
www. .com information@itm.com1.800.561.8187

3
W A R N I N G
Changes or modifications not expressly approved by
the party responsible for compliance could void the
user's authority to operate the equipment.
www. .com information@itm.com1.800.561.8187

5
Chapter 2
Setting-up
Supplying Power to the Model 9565 Series
The Model 9565 VelociCalc® Air Velocity Meter can be powered in
one of two ways: four size AA batteries or the AC adapter.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on
the inside of the battery compartment. The Model 9565 is
designed to operate with either alkaline or NiMH rechargeable
batteries. Battery life will be shorter if NiMH batteries are used. If
NiMH batteries are used the DIP switch will need to be changed.
Refer to the next section, DIP Switch Settings. Carbon-zinc
batteries are not recommended because of the danger of battery
acid leakage.
DIP Switch Settings
To access the DIP switch, remove the batteries from the battery
compartment. On the inside of the battery compartment, there is
a window with a single DIP switch (see figure below). The table
below shows the functions for the switch.
C a u t i o n
Make certain that power is turned off before changing
the DIP switch settings.
Switch
Function
Settings
1
NiMH
OFF: Alkaline Batteries
ON: NiMH Rechargeable Batteries
VelociCalc is a registered trademark of TSI Incorporated.
www. .com information@itm.com1.800.561.8187

6 Chapter 2
DO NOT attempt to charge alkaline batteries.
Using the AC Adapter
The AC adapter can be used to power the instrument or to
charge the NiMH batteries when the DIP switch in the battery
compartment is set to NiMH. If the DIP switch is set to
Alkaline, and the AC power adapter is connected, then the
batteries will be bypassed and the meter will be powered by
the AC adapter. Be sure to provide the correct voltage and
frequency, which is marked on the back of the AC adapter.
AC Adapter input
www. .com information@itm.com1.800.561.8187

Setting-up 7
Connecting Ventilation or IAQ Probes
The ventilation and IAQ probes have a “D” shape overmolding on the
mini-DIN connector which must align with the connector at the base of
the 9565 series meter. This will ensure the probe is properly
connected and remains so during use. Once connected and turned
on, refer to the DISPLAY SETUP for displaying the desired
measurements.
Using the Telescoping Thermoanemometer Probes
The telescoping probe contains the velocity, temperature, and
humidity sensors. When using the probe, make sure the sensor
window is fully exposed and the orientation dimple is facing upstream.
N O T E
For temperature and humidity measurements, make sure that at
least 3 inches (7.5 cm) of the probe is in the flow to allow the
temperature and humidity sensors to be in the air stream.
Extending the Probe
To extend the probe, hold the handle in one hand while pulling on
the probe tip with the other hand. Do not hold the cable while
extending the probe as this prevents the probe from extending.
Retracting the Probe
To retract the probe, hold the handle in one hand while gently
pulling the probe cable until the smallest antenna section is
retracted.
“D” Shaped
mini-DIN connector
www. .com information@itm.com1.800.561.8187

8 Chapter 2
DO NOT use the instrument or probes near hazardous
voltage sources since serious injury could result.
Using the Pressure Ports (9565-P)
The 9565-P includes pressure ports that can be used to measure
static and differential pressures in ductwork. For more information,
see Application Note TSI-107.
Connecting the Static Pressure Probe
The Static Pressure probe included with the 9565-P is connected
to the + port on the 9565-P using the included tubing. The Static
Pressure probe is used to measure the duct static pressure and
features a magnet which holds the probe to the ductwork.
Negative (-) Pressure Port
Positive (+) Pressure Port
www. .com information@itm.com1.800.561.8187

Setting-up 9
Connecting an Optional Pitot Probe or Airflow (straight
pitot) Probe
When connected to a pitot probe, air velocity or air volume can be
measured. A pitot probe can be connected to the “+” and “-”
pressure ports on the Model 9565-P using two pieces of tubing of
equal length. The total pressure port of the pitot probe connects
to the “+” port on the meter, and the static pressure port of the
pitot probe connects to the “-” port on the meter.
For information on how to perform a duct traverse, refer to
Application Note TSI-106.
N O T E
The pitot velocity needs a valid temperature to perform the
standard or actual velocity correction. This is accomplished
in the “Actual/Std Setup” menu. If no probe capable of
measuring temperature (plug in probe or thermocouple) is
connected, the“Temp Source” must be set to “Entered”. The
duct air temperature must then be manually inputted by the
user using the “Entered Temp” setting. If the “Temp Source”
is set to Probe or Thermocouple 1 or 2, and no probe is
connected, dashes (------) will appear on the display.
For more information on entering the temperature manually,
refer to the Actual/Standard Setup section of this manual.
www. .com information@itm.com1.800.561.8187

10 Chapter 2
DO NOT use the instrument or probes near hazardous
voltage sources since serious injury could result.
Thermocouple Ports
The 9565 series includes two thermocouple ports at the base of the
meter. Any K-Alloy thermocouple with mini-connector can be
attached. See Display Setup for setting the thermocouple
temperature readings to be displayed as TC1, TC2, or TC1-TC2.
Connecting the Thermocouples
The K-Alloy thermocouple with mini-connector has one terminal
wider than the other. The wider terminal will be inserted into the
bottom of the TC1 or TC2 connector port.
Thermocouples from an alternate TSI supplier must have
the metal sheath electrically isolated from the wires
inside. Failure to meet this requirement may result in
false readings, electrical shock, or fire hazard.
DO NOT use the instrument or probes near hazardous
voltage sources since serious injury could result.
K-alloy Thermocouple
TC1
TC2
www. .com information@itm.com1.800.561.8187

Setting-up 11
Connecting the Optional Bluetooth® Portable Printer Device
(Models 9565-P and 9565-X only)
To connect the Bluetooth printer to the Model 9565, power on the unit
and the printer. Then press the MENU soft key. From the Menu use
the and keys to highlight Bluetooth Functions and press the
key. Highlight Discover Devices and press the key. If other TSI
Bluetooth-printers are in the area, turn them off before searching. The
Model 9565 will then search for and list all available Bluetooth
devices.
For more information on establishing Bluetooth connections, refer to
TSI Applications Note TSI-150.
Connecting to a Computer
Use the Computer Interface USB Cable provided with the Model 9565
to connect the instrument to a computer for downloading stored data.
C a u t i o n
This symbol is used to indicate that the data port of the
Model 9565 is not intended for connection to a public
telecommunications network. Connect the USB data
port only to another USB port on a safety certified
computing device.
Protection provided by the instrument could be impaired
if used in a manner other than specified in this user
manual.
Bluetooth is a registered trademark of Bluetooth SIG.
USB Communications Port
www. .com information@itm.com1.800.561.8187

13
Chapter 3
Operation
Keypad Functions
ON/OFF ( ) Key
Press the ON/OFF key to turn the Model 9565 on
and off. During the power up sequence the display
will show the following: Model Number, Serial
Number, and Software Revision. To turn the
instrument off, press and hold the ON/OFF Key for 3
seconds. The instrument will count down (OFF2,
OFF1, OFF). If AC Adapter is attached, the Battery
and ON/OFF Key is bypassed. If the ON/OFF Key is
pressed while the AC adapter is attached, the
instrument instructs you to “Unplug the instrument
to turn off unit”. To turn the instrument on again,
attach the AC adapter or press the ON/OFF Key.
Arrow () Keys
Press to scroll through choices while setting a
parameter. Pressing the keys simultaneously
will lock the keypad to prevent unauthorized
adjustments to the instruments. To unlock the
keypad, press the keys simultaneously.
Enter ( ) Key
Press to accept a menu selection, value or
condition. Press to Start or Stop datalogging when
in Continuous Key mode.
www. .com information@itm.com1.800.561.8187

14 Chapter 3
Arrow (or ) and
Menu Soft Keys
Press arrow keys to change choices while setting a
parameter. Press the Menu soft key to select the
Menu selections, which are Pressure Zero, Display
Setup, Settings, Flow Setup, VOC Setup,
Actual/Std Setup, Data Logging, Zero CO,
Applications, Calibration, and Bluetooth Functions.
Common Terms
This manual has several terms used in various places. The following
is a brief explanation of the meanings of those terms.
Sample
Consists of all of the measurement parameters
stored at the same time.
Test ID
A group of samples. The statistics (average,
minimum, maximum, and count) are calculated for
each test ID. The maximum number of test IDs
is 100.
Time Constant
The time constant is an averaging period. It is used
to dampen the display. If you are experiencing
fluctuating flows, a longer time constant will slow
down those fluctuations. The display will update
every second, but the displayed reading will be the
average over the last time constant period. For
example, if the time constant is 10 seconds, the
display will update every second, but the displayed
reading will be the average from the last 10
seconds. This is also referred to as a “moving
average”.
Log Interval
The logging interval is the period over which the
instrument will average the logged sample. For
example, if the logging interval is set to 30 minutes,
each sample will be the average over the previous
30 minutes.
Test Length
This is the time over which the data will be logged
in the “Continuous-Time” mode of data logging.
www. .com information@itm.com1.800.561.8187

Operation 15
Menus
The menu structure is organized to allow easy navigation and
instrument setup utilizing the arrow keys and button. To exit a
menu or menu item, press the ESC key.
To access the Menu items, press the Menu soft key.
To select a parameter, use the Arrow keys to highlight the
selection and press the button.
ZERO PRESSURE
Turn the instrument on for a minimum of five minutes to let it warm up
before zeroing the pressure sensor. To zero the pressure reading,
leave the positive and negative ports open to atmosphere and select
the Zero Press. The pressure sensor should be zeroed after initial
warm-up and checked periodically. The instrument will indicate if the
pressure zero was successful.
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
www. .com information@itm.com1.800.561.8187

16 Chapter 3
DISPLAY SETUP
Display Setup menu is where you will setup the desired parameters to
be displayed on the instrument screen. With a parameter highlighted
you can then use the ON soft key to have it show up on the instrument
screen or select the OFF soft key to turn off the parameter. Use
PRIMARY soft key to have a parameter show up on the instrument
screen in a larger display. A total of 5 parameters can be shown on
the display, 1 primary (large font) and 4 secondary. Parameters
shown in the Display Setup screen are dependent on the type of
probe currently connected.
When set to PRIMARY, measurement will be the large font on the
display.
When set to ON, measurement will be displayed as a secondary
parameter (up to 4 can be displayed).
When set to OFF, measurement will not be displayed.
N O T E
Pitot Velocity and AFProbe Velocity cannot both be ON at the
same time, nor can one be set to PRIMARY and the other to ON.
Only one can be ON at any given time.
MENU
Zero Press
Display Setup
Settings
Flow Setup
VOC Setup
Actual/Std Setup
Data Logging
Zero CO
Applications
Calibration
Bluetooth Functions
DISPLAY SETUP
Velocity *ON
Flow OFF
Pitot Velocity OFF
Pressure ON
Temperature ON
Baro Press OFF
ON PRIMARY OFF
www. .com information@itm.com1.800.561.8187
This manual suits for next models
3
Table of contents
Other VELOCICALC Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Varian
Varian FRG-700 series instruction manual

Bedfont
Bedfont Micro Plus Smokerlyzer quick start guide

PureAire
PureAire 99088 instruction manual

FlowKinetics
FlowKinetics FKS 1DP-PBM-E user manual

Ludlum Measurements
Ludlum Measurements 177 manual

Bindicator
Bindicator MP Series Installation & operation manual

WIKA
WIKA 2XX.34 Operating & installation instructions

Elvaco
Elvaco CMeX20w manual

Phenix Technologies
Phenix Technologies MRM10E user guide

Hinds Instruments
Hinds Instruments EXICOR 150AT user manual

PCB Piezotronics
PCB Piezotronics 214B Installation and operating manual

DANOPLUS
DANOPLUS DP-101 user manual