Vibra HJ-K Series User manual

 220003M01










******************************
**
*HJ-Kseries*
*Relaycontactoption*
*Userʼsmanual*
**
******************************
DocumentHJ-KseriesRelaycontactoutputoptionuserʼsmanual 頁1/5


Introduction
ThisoptionisanoutputofrelaycontactsignalcorrespondingtojudgmentresultsoftheHJ-K
balance,whichcanclassifyintomaximum5ranks,obtainedthroughDtype25pinconnector
Signalfor“Onmeasurement”anderrorareavailable,too.


DocumentHJ-KseriesRelaycontactoutputoptionuserʼsmanual 頁2/5


1.Specification
1-1pinassignmentatconnectorandthefunctions(connector:D-SUB25P)
1-1-1Pinsassignmentandfunctions
Pin# Signal Input・
Output Functions・remarks
1 Rank1(LO)NO Output ONwhenjudgedasRank1(LO)
2 Rank1(LO)NC Output OFFwhenjudgedasRank1(LO)
3 Rank1(LO)COM - Rank1(LO)contactcommon
4 Rank2(OK)NO Output ONwhenjudgedas2(OK)
5 Rank2(OK)NC Output OFFwhenjudgedas2(OK)
6 Rank2(OK)COM - Rank2(OK)contactcommon
7 Rank3(HI)NO Output ONwhenjudgedas3(HI)
8 Rank3(HI)NC Output OFFwhenjudgedas3(HI)
9 Rank3(HI)COM - Rank3(HI)contactcommon
10 Rank4NO Output ONwhenjudgedas4
11 Rank4NC Output OFFwhenjudgedas4
12 Rank4COM - Rank4contactcommon
13 Rank5NO Output ONwhenjudgedasRank5
14 Rank5NC Output OFFwhenjudgedasRank5
15 Rank5COM - Rank5contactcommon
16 ERRNO Output ONwhenERRisdisplayed
17 ERRNC Output OFFwhenERRisdisplayed
18 ERRCOM - ERRcontactcommon
19 “Onmeasurement”NO Output ONwhendisplayedvalueismorethan5timesas
readability
20 “Onmeasurement”NC Output OFFwhendisplayedvalueismorethan5timesas
readability
21 OnmeasurementCOM - “Onmeasurement”contactcommon
22 BuzzerNO Output ONwhenbuzzersounds
23 BuzzerCOM - Buzzercontactcommon
24 Externalsignal(+) Input Contactoutputcontrol(+)
25 Externalsignal(-) Input Contactoutputcontrol(-)
*NO:NormalOpen(Acontact)
NC:NormalClose(Bcontact)






 Caution  Always unplug theAC adaptor before connecting.
1
2345
14 15 16 17
D-SUB25P male connector : view from rear
6 7 8 9 10 11 12 13
18 19 20 21 22 23 24 25
DocumentHJ-KseriesRelaycontactoutputoptionuserʼsmanual 頁3/5


1-1-2Relaycontactcircuit





1-1-3Remarksoninductiveloadinthecontact






 Supplyspark-offcircuitwheninductiveload(relay,solenoidetc)isconnected

1-1-4EquivalentcircuittoInput






1-1-5exampleofrelayinput






 Besuretoapplyspecifiedcurrent,byavoidingmistakeofwrongconnection,andusing
controlresistorandsoon.Wrongconnectionorexcessivecurrentsupplymaydamagetheinternal
circuittakeatleast100msforONperiod.
V R
5V 270Ω
12V 1kΩ
24V 2.2kΩ
NO
NC
COM
Capacity at contac
t
(resistive load)
AC 125V / 0.5A
DC 30V / 1A
V
F=0.8〜1.4V
Photo coupler
V
F
+
-
100Ω
+
-
V
Current
Control resistor
Relay contact control
Terminal at
External <example of constant>
DocumentHJ-KseriesRelaycontactoutputoptionuserʼsmanual 頁4/5


2.SettingoftheHJ-Kbalance
Activate the limit function for outputting the result of comparison through relay contact
limit function should be activated even when ERR, on weighing signal only are used 

■Condition of the function setting
This is a standard setting generally working. If it is not acceptable, apply your setting by
referring to the operation manual of the balance.
 Additionalfunction [



] limitfunctioneffective
 judgmentcondition [



] judgmentatalltimes(includingwhen
unstable)
judgment range [



] wholerangeabovezero
Number of set points [



] 2points(upper & lower) (HI/OK/Lo
judgment)
Judgment method[





] realvalue
 Rank1Buzzer [





] NobuzzerdoesnotsoundinRank1(LO)
 Rank2Buzzer [





] NobuzzerdoesnotinRank2(OK)
 Rank3Buzzer [





] NobuzzerinRank3(HI)
 Rank4Buzzer [





] NobuzzerinRank4
 Rank5Buzzer [





] NobuzzerinRank5
Rank] indication[



] HI/OK/LO/rankindication
Relay output control[





] Outputatalltimes(notcontrolledbyexternal
signal)

■Settingupperandlowerlimits
 referto“limitfunction“inoperationmanualofHJ-K
 (example)


DocumentHJ-KseriesRelaycontactoutputoptionuserʼsmanual 頁5/5


3.Operations
1. Plug theAC adaptor and press the [ON/OFF] key
2. Enter the set point values ( upper & lower limits) to enable the comparator function
( Refer to in the previous section and the operation manual of HJ-K in detail )
 3.TheRelaycontactworksasthecomparatorresultisdisplayed
 

4.Statusofthebalanceandthecontact
Statusofthebalance Display Contact
Doesnotmeetcondition Noresultisdisplayed AllOFF
Doesnotmeettherange Noresultisdisplayed AllOFF
Set points or values are
invalid Three

on left side
light AllOFF
o-Err/u-Err o-Err/u-Err *OFFinRank1〜5in“On
measurement”
*ONinERR
Duringsetting
(notonmeasurement) Settingdisplay AllOFF
Waitingforstabilitywith
tare 『M』flickers Thelatestresultishold
Standby standbyLEDlights AllOFF
Onsleep sleepLEDlights Output to the judgment
result
At start up until
measurementstarts 『

』
alllights AllOFF

5.Outputcontrolbyexternalsignal
1.Generallycontactsandbuzzersarecontrolledbyjudgmentresult,butcanbecontrolled
byexternalsignalbycontrolsetting.
2.Ifpowersourcesisconnectedtoexternalsignalinput,Whileoutputissettobecontrolled
byexternalsignal([





]control),judgmentresultisoutputforabout200ms
aftersignalstarts.
3.Ifyouapplyexternalsignalinpulsewave,takeONperiodforatleast100ms.

Other manuals for HJ-K Series

1

Popular Relay manuals by other brands

Pilz PZA Series operating manual

Pilz

Pilz PZA Series operating manual

Compact Electric CDMR-61 user manual

Compact Electric

Compact Electric CDMR-61 user manual

Minebea MIS-304-P Mounting instruction

Minebea

Minebea MIS-304-P Mounting instruction

Pickering SYSTEM 10 user manual

Pickering

Pickering SYSTEM 10 user manual

Pilz PZE X5P operating instructions

Pilz

Pilz PZE X5P operating instructions

ABB CWC Instruction leaflet

ABB

ABB CWC Instruction leaflet

WURM ZCB-ADP/PAT Product information

WURM

WURM ZCB-ADP/PAT Product information

CD Automation REVO SSR user manual

CD Automation

CD Automation REVO SSR user manual

GE 750 instruction manual

GE

GE 750 instruction manual

Toro TEMPUS AIR PR user guide

Toro

Toro TEMPUS AIR PR user guide

Pilz PNOZ X2.1VP operating manual

Pilz

Pilz PNOZ X2.1VP operating manual

Deif MVR-200 Series Operator's manual

Deif

Deif MVR-200 Series Operator's manual

Reer SR ONE Installation, use and maintenance

Reer

Reer SR ONE Installation, use and maintenance

Panasonic TX Series SOLDERING AND CLEANING GUIDELINES

Panasonic

Panasonic TX Series SOLDERING AND CLEANING GUIDELINES

socomec ATyS 3S operating instructions

socomec

socomec ATyS 3S operating instructions

Pilz S1EN operating instructions

Pilz

Pilz S1EN operating instructions

ABB SPAS 348 C User manual and technical description

ABB

ABB SPAS 348 C User manual and technical description

Kyongbo Electronics GDR-C01 manual

Kyongbo Electronics

Kyongbo Electronics GDR-C01 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.