Introduction
ThisoptionisanoutputofrelaycontactsignalcorrespondingtojudgmentresultsoftheHJ-K
balance,whichcanclassifyintomaximum5ranks,obtainedthroughDtype25pinconnector
Signalfor“Onmeasurement”anderrorareavailable,too.

DocumentHJ-KseriesRelaycontactoutputoptionuserʼsmanual 頁2/5
1.Specification
1-1pinassignmentatconnectorandthefunctions(connector:D-SUB25P)
1-1-1Pinsassignmentandfunctions
Pin# Signal Input・
Output Functions・remarks
1 Rank1(LO)NO Output ONwhenjudgedasRank1(LO)
2 Rank1(LO)NC Output OFFwhenjudgedasRank1(LO)
3 Rank1(LO)COM - Rank1(LO)contactcommon
4 Rank2(OK)NO Output ONwhenjudgedas2(OK)
5 Rank2(OK)NC Output OFFwhenjudgedas2(OK)
6 Rank2(OK)COM - Rank2(OK)contactcommon
7 Rank3(HI)NO Output ONwhenjudgedas3(HI)
8 Rank3(HI)NC Output OFFwhenjudgedas3(HI)
9 Rank3(HI)COM - Rank3(HI)contactcommon
10 Rank4NO Output ONwhenjudgedas4
11 Rank4NC Output OFFwhenjudgedas4
12 Rank4COM - Rank4contactcommon
13 Rank5NO Output ONwhenjudgedasRank5
14 Rank5NC Output OFFwhenjudgedasRank5
15 Rank5COM - Rank5contactcommon
16 ERRNO Output ONwhenERRisdisplayed
17 ERRNC Output OFFwhenERRisdisplayed
18 ERRCOM - ERRcontactcommon
19 “Onmeasurement”NO Output ONwhendisplayedvalueismorethan5timesas
readability
20 “Onmeasurement”NC Output OFFwhendisplayedvalueismorethan5timesas
readability
21 OnmeasurementCOM - “Onmeasurement”contactcommon
22 BuzzerNO Output ONwhenbuzzersounds
23 BuzzerCOM - Buzzercontactcommon
24 Externalsignal(+) Input Contactoutputcontrol(+)
25 Externalsignal(-) Input Contactoutputcontrol(-)
*NO:NormalOpen(Acontact)
NC:NormalClose(Bcontact)
Caution Always unplug theAC adaptor before connecting.
1
2345
14 15 16 17
D-SUB25P male connector : view from rear
6 7 8 9 10 11 12 13
18 19 20 21 22 23 24 25

DocumentHJ-KseriesRelaycontactoutputoptionuserʼsmanual 頁3/5
1-1-2Relaycontactcircuit
1-1-3Remarksoninductiveloadinthecontact
Supplyspark-offcircuitwheninductiveload(relay,solenoidetc)isconnected
1-1-4EquivalentcircuittoInput
1-1-5exampleofrelayinput
Besuretoapplyspecifiedcurrent,byavoidingmistakeofwrongconnection,andusing
controlresistorandsoon.Wrongconnectionorexcessivecurrentsupplymaydamagetheinternal
circuittakeatleast100msforONperiod.
V R
5V 270Ω
12V 1kΩ
24V 2.2kΩ
NO
NC
COM
Capacity at contac
t
(resistive load)
AC 125V / 0.5A
DC 30V / 1A
V
F=0.8〜1.4V
Photo coupler
V
F
+
-
100Ω
+
-
V
Current
Control resistor
Relay contact control
Terminal at
External <example of constant>

DocumentHJ-KseriesRelaycontactoutputoptionuserʼsmanual 頁4/5
2.SettingoftheHJ-Kbalance
Activate the limit function for outputting the result of comparison through relay contact
limit function should be activated even when ERR, on weighing signal only are used
■Condition of the function setting
This is a standard setting generally working. If it is not acceptable, apply your setting by
referring to the operation manual of the balance.
Additionalfunction [
] limitfunctioneffective
judgmentcondition [
] judgmentatalltimes(includingwhen
unstable)
judgment range [
] wholerangeabovezero
Number of set points [
] 2points(upper & lower) (HI/OK/Lo
judgment)
Judgment method[
] realvalue
Rank1Buzzer [
] NobuzzerdoesnotsoundinRank1(LO)
Rank2Buzzer [
] NobuzzerdoesnotinRank2(OK)
Rank3Buzzer [
] NobuzzerinRank3(HI)
Rank4Buzzer [
] NobuzzerinRank4
Rank5Buzzer [
] NobuzzerinRank5
Rank] indication[
] HI/OK/LO/rankindication
Relay output control[
] Outputatalltimes(notcontrolledbyexternal
signal)
■Settingupperandlowerlimits
referto“limitfunction“inoperationmanualofHJ-K
(example)

DocumentHJ-KseriesRelaycontactoutputoptionuserʼsmanual 頁5/5
3.Operations
1. Plug theAC adaptor and press the [ON/OFF] key
2. Enter the set point values ( upper & lower limits) to enable the comparator function
( Refer to in the previous section and the operation manual of HJ-K in detail )
3.TheRelaycontactworksasthecomparatorresultisdisplayed
4.Statusofthebalanceandthecontact
Statusofthebalance Display Contact
Doesnotmeetcondition Noresultisdisplayed AllOFF
Doesnotmeettherange Noresultisdisplayed AllOFF
Set points or values are
invalid Three
on left side
light AllOFF
o-Err/u-Err o-Err/u-Err *OFFinRank1〜5in“On
measurement”
*ONinERR
Duringsetting
(notonmeasurement) Settingdisplay AllOFF
Waitingforstabilitywith
tare 『M』flickers Thelatestresultishold
Standby standbyLEDlights AllOFF
Onsleep sleepLEDlights Output to the judgment
result
At start up until
measurementstarts 『
』
alllights AllOFF
5.Outputcontrolbyexternalsignal
1.Generallycontactsandbuzzersarecontrolledbyjudgmentresult,butcanbecontrolled
byexternalsignalbycontrolsetting.
2.Ifpowersourcesisconnectedtoexternalsignalinput,Whileoutputissettobecontrolled
byexternalsignal([
]control),judgmentresultisoutputforabout200ms
aftersignalstarts.
3.Ifyouapplyexternalsignalinpulsewave,takeONperiodforatleast100ms.
Other manuals for HJ-K Series
1
Popular Relay manuals by other brands

Siemens
Siemens SIRIUS 3UG5461-1AA40 Original operating instructions

Simons Voss Technologies
Simons Voss Technologies SmartRelay SREL Series manual

Elko
Elko HRN-43 manual

Zander Aachen
Zander Aachen SR3E operating instructions

Schweitzer Engineering Laboratories
Schweitzer Engineering Laboratories SEL-751 instruction manual

ABB
ABB Relion 615 series manual