
Operating instructions
-Please read the manual carefully before use. Failure to follow the
instructions in this manual may lead to incorrect results.
-Avoid any shock, vibration, electrical discharge or contact with chemicals
that could cause a malfunction of the unit or produce incorrect results.
-The Speed ReaderTM analyser can only be used for reading in vitro veterinary
diagnostic tests of the Speed Biomarker range. The use of other tests could
cause malfunctions or provide incorrect results. Bio Veto Test and its
distributors disclaim all liability for any unauthorised operation.
-Before being used, the Speed ReaderTM analyser must be placed on a clean,
flat and stable surface, away from moisture, heat, dust, direct sunlight or
any liquid.
-Do not use the device outdoors. Do not place anything on the device.
-The operating temperature is between +15°C and +35°C. For an optimal
use, it is advisable to put the reader in a room having a temperature higher
than 20 ° C. Recommended maximum humidity is 70%. If the unit is placed
in an environment that does not meet these conditions, it is advisable to
move the Speed ReaderTM analyser and wait at least 30 minutes before
using it again.
-Always operate the reader with the stylus provided
-You should always wear clean and properly maintained individual protection
equipment when operating the Speed tests.
-The Speed ReaderTM analyser and its reagents must only be used by the
veterinary practitioner. The interpretation of the result should be carried out
by the veterinarian who must always take into account the medical history
and the clinical examination of the animal and any other diagnostic test
result, given that no diagnostic method is 100% accurate. The final
diagnosis remains the prerogative and responsibility of the veterinarian.
-Avoid applying intense pressure during the insertion of the tests in the
loader or when handling the touchscreen.
-In case of a major malfunction, immediately switch off the ON/OFF button
located at the rear of the unit, unplug the Speed ReaderTM analyser and
contact technical support.
-When necessary, you should clean your Speed ReaderTM analyser with a soft
and dry cloth, after disconnecting its power supply.
-After each use and before turning off the unit, check that the test loader is
properly introduced in the unit. If necessary, re-insert the loader, using the
button for this purpose on the main screen.
-To power off the device, strictly follow the instructions in this manual.
-You should unplug the unit when not using it for several days.
-Never disassemble the unit. This operation must only be performed by an
authorised technician.
-Do not disassemble randomly.
-Since the laser module is inside the product, do not disassemble it during
use.
-Caution - Use of controls or adjustments or performance of procedures other
than those specified herein may result in hazardous radiation exposure.